Biomechanical analysis of the door-shaped titanium plate in single-level anterior cervical discectomy and fusion

Abstract

Background

Analyse and discuss the immediate stability of the cervical spine after anterior cervical discectomy and fusion using a door-shaped titanium plate and compare it with the traditional titanium plate, to provide biomechanical evidence for the rationality and effectiveness of the door-shaped titanium plate in clinical applications.

Methods

Ten adult goat C4/5 vertebral bodies were obtained, and models were prepared using denture base resin. Biomechanical experiments were performed on the specimens before internal fixation. MTS was used to conduct non-destructive biomechanical loading tests in six directions, including flexion, extension, left–right bending, and left–right torsion, recording the range of motion (ROM) and neutral zone (NZ) of each specimen. The specimens were then randomly divided into two groups: the study group was fixed with a door-shaped titanium plate, and the control group was fixed with a traditional titanium plate. ROM and NZ in each direction were measured again. After measurements, both groups were subjected to 0.5 Hz torsion loading with a torque of 2 N m for a total of 3000 cycles, followed by measuring ROM and NZ in six directions once more.

Results

Compared to before fixation, ROM and NZ in both groups significantly decreased in all six directions after fixation, with statistical significance (P < 0.05); after fixation, the study group showed slightly lower values for various mechanical reference parameters compared to the control group, with no statistical significance (P > 0.05); after 3000 torsional loads, both internal fixation groups showed increased ROM and NZ compared to after fixation but to a lower extent, and no screw or titanium plate loosening was observed. Compared to before fixation, the differences were still statistically significant (P < 0.05), with the study group having slightly lower ROM and NZ values in all directions compared to the control group, with no statistical significance (P > 0.05).

Conclusion

The door-shaped titanium plate exhibits mechanical properties similar to the traditional titanium plate in all directions, and its smaller size and simpler surgical operation can be used for anterior cervical endoscopic surgery, reducing surgical trauma. It is clinically feasible and deserves further research and promotion.

Background

Cervical spondylosis is a common degenerative disease often accompanied by symptoms such as neck pain, lower limb weakness, dizziness, and other discomforts. For patients with mild symptoms, conservative treatment is initially considered in clinical practice [1,2,3]. However, for those who do not respond well to conservative measures for at least 6 months, surgical treatment is recommended [4]. Anterior cervical discectomy and fusion (ACDF) is one of the classical surgical approaches for treating cervical spondylosis, exhibiting excellent efficacy in managing cervical spondylotic myelopathy (CSM) and cervical spondylotic radiculopathy (CSR) [5,6,7,8]. This surgical method was originally proposed by Smith, involving the removal of the diseased intervertebral disc and the use of autologous iliac bone for interbody fusion to restore cervical spine stability. Nevertheless, this initial method has its limitations, as a series of postoperative complications, including bone block dislocation, subsidence, and pseudoarthrosis formation, adversely affect surgical outcomes [9, 10]. The introduction of the anterior cervical plate system (ACPS) further enhances the immediate stability of the cervical spine after ACDF, reducing the duration of postoperative neck bracing and improving fusion results, with a bone fusion rate of approximately 90%. Moreover, the rate of subsidence of the fusion material is significantly reduced [11,12,13]. Therefore, ACPS has become a commonly used internal fixation system in clinical practice and a classic reference standard for testing the stability of newly developed internal fixation devices. However, traditional titanium plates, while offering many advantages, also have numerous shortcomings. Plates such as the Katia Anterior Cervical Internal Fixation System, AO Locking Plate, Caspar Plate, and Orion Plate are relatively thick and bulky. Consequently, they often lead to postoperative swallowing difficulties and complications such as adjacent segment degeneration (ASD), with an overall complication rate ranging from 13.2 to 19.3%. High rates of swallowing difficulties significantly impact patients’ quality of life [14,15,16,17,18]. Therefore, reducing the thickness of ACPS while ensuring bone fusion rate has become a focal point of research for new internal fixation devices. To address these limitations, our research team developed a door-shaped titanium plate that can be used in anterior cervical endoscopic systems. This plate is thinner, more compact, and combines the nail and plate, which has been applied in clinical practice with favourable clinical outcomes. However, there is limited biomechanical research on this device. In this study, we used goat cervical vertebrae specimens to evaluate the biomechanical performance of the door-shaped titanium plate.

Materials and methods

Internal Fixation Device: The traditional titanium plate group consists of the Katia Anterior Cervical Internal Fixation System and iliac bone blocks for fusion. The door-shaped titanium plate group is composed of the door-shaped titanium plate (Figs. 1 and 2) and iliac bone blocks for fusion. The door-shaped titanium plate was developed and designed by Director Zhang Changjiang of our institution and has been granted national utility model patent ZL202120000201.7 and design patent ZL202130000119.X. Both types of plates were provided by Shanghai Sanyou Medical Devices Co., Ltd.

Fig. 1
figure 1

3D view and physical drawing of the door-type titanium plate

Fig. 2
figure 2

A Clinical Application Example of the Door-Shaped Titanium Plate: a, b Preoperative cervical X-ray images in anteroposterior and lateral views; c, d Preoperative cervical MRI images, ef preoperative cervical CT images, and g, h Postoperative cervical X-ray images taken one week after the procedure

Acquisition and selection of specimens

The average weight of 10 mountain goats was 53.15 ± 3.83 kg, and specimens of the C4/5 vertebral segment were selected. All specimens underwent X-ray and bone density examinations to exclude abnormal specimens, such as tumours, bone destruction, trauma, degenerative changes, and osteoporosis. X-rays showed no obvious abnormalities in the specimens (Fig. 3), and bone density was within the normal range (1.651 ± 0.034 g/cm3). All specimens were carefully dissected to remove paravertebral muscles, anterior longitudinal ligaments, and fat tissues, while retaining appendages and major ligaments (anterior longitudinal ligament, posterior longitudinal ligament, ligamentum flavum, interspinous ligament, supraspinous ligament), joint capsules, and facet joints. Double-layer dressings were used to wrap and store the specimens at − 20 °C for later use in specimen model preparation.

Fig. 3
figure 3

a, b, c Radiographs of specimens before internal fixation and after performing different internal fixations

Model preparation

The upper 1/3 of the C4 vertebra and the lower 1/3 of the C5 vertebra were embedded in a mould using denture base resin, with the C4/5 vertebrae positioned at the centre of the mould, and the C4/5 intervertebral space parallel to the upper and lower modules formed by the denture base resin. After biomechanical experiments for all specimens before fusion fixation, 5 specimens were randomly selected for intervertebral disc removal, assisted by anterior titanium plate fixation and bone fusion, and designated as the research group. The remaining 5 specimens underwent intervertebral disc removal, assisted by the Katia cervical anterior fixation system and bone fusion, and were designated as the control group. Both groups of patients underwent X-ray examinations (Fig. 3), and X-rays showed that the implanted bone blocks were positioned in the intervertebral space, without entering the spinal canal, and the screws did not penetrate the endplate into the intervertebral space.

Mechanical experiment and loading sequence

MTS was used to simulate the physiological motion of the cervical spine in six directions: flexion–extension, lateral bending (left and right), and axial rotation for 10 specimens. A maximum torque of 2.5 N·m and a vertical physiological load of 20 N were applied. To minimize experimental errors caused by creep, a preloading of three cycles was applied to the specimens before data acquisition. The specimens were then allowed to rest for 60 s until they returned to their neutral position before formal testing. MTS collected the spatial coordinates of the end effector of the mechanical arm and obtained the torque values at various data collection points on the data acquisition board. Comprehensive curve graphs were generated through computer processing, providing the range of motion (ROM) and neutral zone (NZ) in each direction. ROM and NZ values were obtained for all samples before fixation using the above-mentioned method. The two groups of samples were then fixed, and ROM and NZ were measured for the research group and the control group. Subsequently, both groups of samples were subjected to a torsional load with a frequency of 0.5 Hz, 3000 cycles, and a torque of 2 N m. ROM and NZ of the specimens were measured after torsion (Fig. 4). Throughout the experiment, intermittent spraying of physiological saline was applied to keep the specimens hydrated.

Fig. 4
figure 4

a, b, c Three-Dimensional Motion Stability Test was performed on specimens from the pre-fixation, post-fixation, and post-torsion groups; d, e After completion of post-fixation biomechanical testing 3000 torsions were performed on both groups of specimens

Statistical analyses

All measurement results were statistically analysed using SPSS 26.0 software. Independent samples t tests were employed to compare ROM and NZ between the two groups after fixation, followed by paired t tests to compare the post-fixation data with the pre-fixation data. After 3000 cycles of torsion, independent samples t tests were performed again to compare ROM and NZ between the two groups, followed by paired t tests to compare the post-fixation and pre-fixation data, respectively. A significance level of P < 0.05 indicated statistical significance.

Results

All specimens exhibited normal morphology without bone damage, and intervertebral height was well preserved. Internal fixation in both groups effectively restored intervertebral height, reestablished the physiological curvature, and ensured proper bone placement, with secure fixation of adjacent vertebrae. There was no penetration of fixation screws into the spinal canal or endplates. Comparisons of ROM and NZ in six directions before fixation, after fixation, and after 3000 torsional cycles can be found in Tables 1 and 2, Figs. 5 and 6. The p values for all statistical analyses are provided in Tables 3 and 4. In both internal fixation groups, ROM and NZ in all directions significantly decreased compared to pre-fixation values, with statistical significance (P < 0.05). However, the research group exhibited slightly lower ROM and NZ in all directions compared to the control group, but the differences were not statistically significant (P > 0.05). After 3000 torsional cycles, ROM and NZ in all directions for both groups remained significantly lower compared to pre-fixation values (P < 0.05), with a slight increase compared to pre-torsion values (P < 0.05), but the magnitude of increase was minimal. In comparison, the research group still demonstrated slightly lower ROM and NZ in all directions compared to the control group, without statistical significance (P > 0.05).

Table 1 The ROM of C4/5 in different directions of two groups
Table 2 The NZ of C4/5 in different directions of two groups
Fig. 5
figure 5

ROM graphs of two sample groups before and after internal fixation, as well as after torsion

Fig. 6
figure 6

NZ graphs of two sample groups before and after internal fixation, as well as after torsion

Table 3 P values after statistical analysis of ROM
Table 4 P values after statistical analysis of NZ

Discussion

Anterior cervical discectomy and fusion (ACDF) is a widely accepted and safe surgical approach for treating degenerative cervical spine diseases. Initially, stand-alone cages were used, which not only restored intervertebral space height but also promoted bone fusion [19]. However, challenges such as cage migration, subsidence, and segmental kyphosis have been encountered [20, 21]. In order to achieve better clinical outcomes, various anterior cervical fixation plates have been developed, including the AO cervical spine locking plate, zero-profile structure, dynamic plate, and cervical disc replacement. Stephen A. Smith and colleagues conducted non-destructive testing on the cervical specimens of 5 fresh cadavers implanted with the AO cervical spine locking plate. The results indicated significant stability during flexion but insufficient stability during rotation [22]. A 4-year follow-up study by Barbagallo et al. reported that the zero-profile structure is safe and effective for both single-level and multi-level ACDF [23]. Chang BQ, through clinical research, found that the incidence of swallowing difficulties in patients using the Zero-profile device and traditional nail plate system was 9.52% and 29.17%, respectively, with a significant difference between the two groups [24]. However, Lee et al., through tracking and analysing clinical data of patients who underwent single-level interbody fusion, found that the zero-profile structure was inferior to the cage and plate structure in maintaining spinal structure and stability, preventing subsidence, and achieving bone fusion [12]. Pitzen et al. [25], through a clinical study of 132 patients, found that the risk of complications such as plate fracture and screw loosening with dynamic plate was lower than with traditional plates, and bone fusion occurred faster. However, there are also reports indicating that patients using dynamic plates have a higher rate of graft subsidence, accompanied by the loss of cervical physiological curvature [26]. Cervical disc replacement differs significantly from the anterior cervical plate system; it almost does not restrict cervical spine movement, avoiding the risk of postoperative loss of cervical range of motion and also reducing the probability of swallowing difficulties. However, it is also associated with potential risks such as accelerated degeneration of the facet joints, nonphysiological motion, and particulate wear [27]. In summary, although these plates have unique advantages, their shortcomings are also evident, such as the complexity of operations with too many screws, significant loss of normal cervical physiological curvature, and a decrease in intervertebral space height.

In addition, the choice of screws is crucial for the stability of the surgically fixed site. Unicortical screws and bicortical screws are frequently used in orthopaedic surgery, and many scholars have compared and analysed their fixation effects. Pater TJ et al. randomly allocated six pairs of matched fresh-frozen cadaver forearms into a unicortical locking group and a bicortical unlocking group, conducting bending and twisting tests to obtain structural stiffness and fixation strength. The results showed equivalent fixation strength in the two groups under bending, with the unicortical locking group having higher bending stiffness. However, the stiffness and strength of this group were significantly lower during twisting [28]. Little et al., through biomechanical studies on a model of externally fixed clavicular fractures in 15 cases, aimed to determine the effectiveness of three internal fixation methods: unicortical + locking plate, bicortical + locking plate, and bicortical + non-locking plate. The conclusion drawn was that the quantity of cortical bone has the greatest impact on fixation effectiveness. Unicortical locking plates may fail at significantly lower loads and stress levels. Bicortical fixation is significantly better than unicortical fixation, and whether the fixation device is a locking device does not significantly affect fixation effectiveness [29]. Sun et al. conducted a retrospective analysis of 12 cases of occipitocervical revision surgery using a bicortical screw and plate system from 2010 to 2018. These surgeries employed depth-limited drilling, progressively deepening the channel to reach the internal bone plate. Simultaneously, a probe was used to explore all walls of the channel, and finally, bicortical screws were inserted into the occipital bone plate. All patients achieved good fixation and bone fusion postoperatively, without major complications. Clinical symptoms were significantly relieved, leading to the conclusion that the posterior occipitocervical screw-plate system using bicortical screws has advantages such as safety, simplicity, and good efficacy, making it valuable for improving occipitocervical revision surgery [30]. These studies collectively indicate that bicortical screws have better fixation effects. In ACDF surgery, due to the thicker vertebral bodies and the presence of the spinal cord within the vertebral canal, it may be challenging to control the length of screws when using bicortical screws. Therefore, the current use of unicortical screws in the door-shaped titanium plate is understandable. However, in future development, the application of depth-limited drilling and probe can be employed to gradually deepen the channel and promptly investigate the internal conditions of the channel. Subsequently, the screws in the door-shaped titanium plate can be modified to bicortical screws, thereby further enhancing fixation stability under safe conditions.

In this experiment, a non-destructive approach was employed to evaluate the immediate stability performance of the internal fixation model. This design offers several advantages: (1) the application of axial load can compress the specimen, better simulating the loading conditions of the cervical spine in the human body. (2) Precise adjustment of loading in various directions was achieved through MTS, avoiding load loss caused by coupled movements and enabling synchronized data acquisition with torque loading, reducing experimental errors. (3) In contrast to most cervical spine mechanical experiments using 1.5 N·m as the motion load, this experiment, aimed at further verifying the mechanical performance of the door-shaped titanium plate, used 20 N as the axial load and 2.5 N·m as the motion load. (4) In addition to using the classic range of motion (ROM) as an indicator for stability assessment, this study also employed neutral zone (NZ). NZ represents the region where the spinal motion segment moves essentially free of applied loading, requiring minimal postural muscular effort [31,32,33]. In quantifying stability, it is often superior to ROM because it has higher sensitivity and can be applied to any setting of maximum loading bounds, even if the load–deflection curve is asymmetrical, it does not affect its use [33, 34]. (5) This experiment introduced short-frequency activities within the physiological load range, which could detect the immediate stability performance of both groups of cervical spines after performing a certain frequency of torsion, making the experimental results more convincing.

The results of this experiment show that, compared to pre-fixation, the research group post-fixation exhibited a significant decrease in ROM and NZ in all six directions, and they followed a consistent trend: flexion and extension ROM decreased by 88.07% and 81.83% respectively, while NZ decreased by 84.37%. Lateral bending ROM decreased by 91.59% on the left and 91.27% on the right, with NZ decreasing by 94.46%. Torsional ROM decreased by 45.61% on the left and 45.40% on the right, with NZ decreasing by 60.00%. In contrast, the control group, relative to pre-fixation, showed a decrease of 87.42% in flexion and 81.56% in extension ROM, with NZ decreasing by 84.89%. Lateral bending ROM decreased by 90.84% on the left and 90.59% on the right, with NZ decreasing by 93.76%. Torsional ROM decreased by 43.93% on the left and 44.07% on the right, with NZ decreasing by 50.00%. We can observe that both groups were able to effectively restore the stability of unstable cervical spine specimens, and the performance of the door-shaped titanium plate was slightly superior to traditional titanium plates. Crawford et al.’s [35] biomechanical experimental results suggest that two fixation screws combined with a titanium alloy plate can achieve similar biomechanical performance as a four-screw combined titanium plate, which is consistent with the results obtained in this experiment.

Short-frequency activities within the physiological load range can also serve as one of the methods for evaluating the immediate stability performance of cervical ACDF surgery [36]. Until the implanted bone blocks fully fuse, the stability of the operative segment of the cervical spine still relies on the internal fixation device to maintain it. Postoperatively, patients cannot avoid some degree of torsional movement, even with the use of a neck brace. In this experiment, after subjecting the internal fixation group to a torsional load of 2 N m, 0.5 Hz, and 2000 cycles, we again measured and recorded the ROM and NZ of each group. We found that the mechanical parameters of both groups had increased slightly, but still significantly decreased compared to the physiological group, with statistical significance. Furthermore, there were no statistically significant differences between the groups. There were no signs of internal fixation device failure during the experimental loading process. This indicates that both groups possess reliable immediate stability performance, and a certain frequency of torsional activity does not affect their stability performance.

Compared to the traditional titanium plate, the door-shaped titanium plate significantly reduces the incidence of complications. Complications associated with the traditional titanium plate primarily focus on difficulties in swallowing, loosening or fracture of screws, adjacent segment diseases, adjacent segment ossification, and occurrence rates [37,38,39]. According to previous literature, difficulty in swallowing is the most common and direct postoperative complication, with an incidence ranging from 1.7 to 67% [40, 41]. Factors such as the number of involved segments, surgical duration, plate size, and the extent of tissue dissection and retraction are considered potential contributors to swallowing difficulties, but the primary factor is the thickness of the plate [42, 43]. Lee and others, by comparing titanium plates of different thicknesses, believed that the thickness and texture of the titanium plate are directly related to swallowing discomfort. Thinner and smoother titanium plates exert less irritation to soft tissues and the oesophagus, resulting in a lower incidence of postoperative swallowing discomfort [44]. The door-shaped titanium plate, with a thickness of only 2 mm and a smooth, flat surface, reduces postoperative swallowing discomfort in patients. The door-shaped titanium plate’s plate and screws are made from a single piece of titanium alloy without gaps at the joints, thereby preventing screw loosening. The extremely high strength and hardness of titanium alloy make screw fracture unlikely. This theoretically allows the door-shaped titanium plate to ensure cervical stability for a longer duration, although this requires further destructive testing to confirm. Furthermore, since it does not require frequent screw installation, there is an increasing amount of clinical and radiological evidence indicating the accelerated degeneration of adjacent fusion segments of the spinal column [45, 46]. Changes in adjacent segment degeneration (ASD) may be related to the biomechanical effects of cervical fusion and/or the biological effects of cervical degeneration [47,48,49]. After fusion, changes in the biomechanics of adjacent segments can lead to increased mobility [50], increased load [51], or increased intervertebral disc pressure [52], ultimately accelerating disc degeneration [53, 54]. Clinical risk factor analysis of ASD following ACDF suggests that the plate-to-disc distance (PDD) may be an independent triggering factor for ASD [46, 55,56,57]. Chung and colleagues conducted follow-ups of over 10 years on 177 ACDF patients and found that most clinical cases of ASD occurred in patients with a PDD less than 5 mm. Therefore, they believe that a PDD greater than or equal to 5 mm is favourable for preventing ASD [46]. In addition, Yu et al. performed logistic regression analysis on 138 patients and also concluded that PDD less than 5 mm is a risk factor for ASD [58]. Similarly, adjacent segment ossification development (ALOD), another common complication of ACDF, can occur as early as 3 months postoperatively, and it is also associated with PDD [59]. Park and colleagues investigated the incidence of ALOD after ACDF, and their findings indicated that ALOD is more likely to occur when the distance between the ends of the plate and adjacent intervertebral discs is less than 5 mm [57, 60]. This suggests that controlling PDD to be greater than or equal to 5 mm is of significant importance in reducing the incidence of both ASD and ALOD. Compared to the larger and longer traditional titanium plates, the door-shaped titanium plate has a length of only 14–18 mm. After adapting to different sizes of cervical vertebrae, it meets the requirement of a PDD greater than 5 mm between the ends of the plate and adjacent intervertebral discs, significantly reducing the incidence of ASD and ALOD.

With the development of minimally invasive techniques for the posterior approach to the cervical spine, there is growing interest in endoscopic techniques for the anterior cervical spine. The primary focus is on percutaneous endoscopic anterior cervical discectomy (PEACD) [61, 62]. This procedure involves the endoscopic removal of cervical intervertebral discs from the anterior route, achieving complete spinal nerve decompression by percutaneously removing the responsible disc segments. However, because PEACD does not involve intervertebral fusion during the procedure, it has certain limitations. These include the inability to provide stability for the cervical surgical segment, the inability to restore the physiological curvature of the cervical spine, a higher risk of recurrence compared to fusion surgery, and a certain degree of height loss in the intervertebral space of the responsible segment. Therefore, it has a limited range of indications and is only suitable for soft intervertebral disc protrusions resulting from cervical injuries, CSM, CSR, and similar conditions. Patients with cervical diseases such as intervertebral disc displacement, calcified disc protrusions, ossification of the posterior longitudinal ligament, and narrow intervertebral disc spaces (less than 3 mm), as well as patients with neurological or vascular disorders similar to intervertebral disc protrusions, are not suitable candidates for this procedure [63, 64]. In comparison to PEACD, the development of mini open (air/water medium) endoscopy-assisted anterior cervical discectomy and fusion (MOEA-ACDF) has been relatively slow. This is because traditional titanium plates are large and not suitable for clamping, and they require frequent screwing, making them unsuitable for MOEA-ACDF. The door-shaped titanium plate is thinner and smaller, making it convenient for endoscopic clamping. It combines the plate and screws, reducing the intraoperative screw locking process, and can be easily hammered in with a bone hammer in just 10 s. Additionally, the screw foot has a spiky shape, allowing it to securely fasten to the vertebrae and apply continuous pressure to the vertebrae, promoting bone fusion. Therefore, it can fully function in MOEA-ACDF. MOEA-ACDF, based on intervertebral disc removal, enhances cervical stability through fusion, improving surgical effectiveness. Compared to PEACD, it broadens the range of indications for minimally invasive techniques, making them applicable to a wider range of cervical conditions.

Limitations

This study has experimentally demonstrated that the door-shaped titanium plate has many advantages. However, there are still the following shortcomings that need to be considered and addressed: (1) This experiment was only conducted on goat cadaver specimens and did not involve human cadaver experiments, making it unable to fully simulate the actual spinal motion in the human body, requiring further experimental research; (2) this experiment may have situations that were not observed due to the limited sample size; and (3) destructive experiments were not conducted, and the experimental results could not verify the maximum resistance to device failure.

Conclusions

In summary, this study confirms that the door-shaped titanium plate has mechanical properties similar to traditional titanium plates in all directions. Moreover, its smaller size and simpler surgical procedure make it suitable for cervical anterior endoscopic surgery, reducing surgical trauma and demonstrating clinical feasibility, worthy of further research and promotion.

Data availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

References

  1. Liu C, Liu KX, Chu L, et al. Posterior percutaneous endoscopic cervical discectomy through lamina-hole approach for cervical intervertebral disc herniation. Int J Neurosci. 2019;129(7):627–34.

    Article  PubMed  Google Scholar 

  2. Balasa A, Bielecki M, Prokopienko M, et al. Lateral approach for recurrent unilateral cervical radiculopathy after anterior discectomy with fusion. Report of two cases. Wideochir Inne Tech Maloinwazyjne. 2019;14(2):348–52.

    PubMed  Google Scholar 

  3. Eco LC, Brayton A, Whitehead WE, et al. Reconstruction of the anterior craniocervical junction using an expandable cage after resection of a C1 chordoma in a 5-year-old child: case report. J Neurosurg Pediatr. 2019;24(1):62–5.

    Article  PubMed  Google Scholar 

  4. Theodore N. Degenerative cervical spondylosis. N Engl J Med. 2020;383(2):159–68.

    Article  CAS  PubMed  Google Scholar 

  5. Burkhardt BW, Brielmaier M, Schwerdtfeger K, et al. Smith-Robinson procedure with and without Caspar plating as a treatment for cervical spondylotic myelopathy: a 26-year follow-up of 23 patients. Eur Spine J. 2017;26(4):1246–53.

    Article  PubMed  Google Scholar 

  6. Shangguan L, Ning G-Z, Tang Y, et al. Discover cervical disc arthroplasty versus anterior cervical discectomy and fusion in symptomatic cervical disc diseases: a meta-analysis. PLoS One. 2017;12(3):1–15.

    Article  Google Scholar 

  7. Zhao Y, Yang S, Huo Y, et al. Locking stand-alone cage versus anterior plate construct in anterior cervical discectomy and fusion: a systematic review and meta-analysis based on randomized controlled trials. Eur Spine J. 2020;29(11):2734–44.

    Article  PubMed  Google Scholar 

  8. Lu T, Liang H, Liu C, et al. Effects of titanium mesh cage end structures on the compressive load at the endplate interface: a cadaveric biomechanical study. Med Sci Monit. 2017;23:2863–70.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Wang J, Qian Z, Ren L, et al. A dynamic finite element model of human cervical spine with in vivo kinematic validation. Chin Sci Bull. 2014;59(33):4578–88.

    Article  Google Scholar 

  10. Li Z, et al. A biomechanical analysis of four anterior cervical techniques to treating multilevel cervical spondylotic myelopathy: a finite element study. BMC Musculoskel Disord. 2021;22(1):278.

    Article  Google Scholar 

  11. Smith MW, Romano DR, McEntire BJ, et al. A single center retrospective clinical evaluation of anterior cervical discectomy and fusion comparing allograft spacers to silicon nitride cages. J Spine Surg (Hong Kong). 2018;4(2):349–60.

    Article  Google Scholar 

  12. Lee YS, Kim YB, Park SW. Does a zero-profile anchored cage offer additional stabilization as anterior cervical plate? Spine. 2015;40(10):E563–70.

    Article  PubMed  Google Scholar 

  13. Lee CH, Hyun SJ, Kim MJ, et al. Comparative analysis of 3 different construct systems for single-level anterior cervical discectomy and fusion: stand-alone cage, iliac graft plus plate augmentation, and cage plus plating. J Spinal Disord Tech. 2013;26(2):112–8.

    Article  PubMed  Google Scholar 

  14. Oliver JD, Goncalves S, Kerezoudis P, et al. Comparison of outcomes for anterior cervical discectomy and fusion with and without anterior plate fixation: a systematic review and meta-analysis. Spine. 2018;43(7):E413–22.

    Article  PubMed  Google Scholar 

  15. Epstein NE. A review of complication rates for anterior cervical diskectomy and fusion (ACDF). Surg Neurol Int. 2019;10:100.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Yee TJ, Swong K, Park P. Complications of anterior cervical spine surgery: a systematic review of the literature. J Spine Surg. 2020;6(1):302–22.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Butler JS, Morrissey PB, Wagner SC, et al. Surgical strategies to prevent adjacent segment disease in the cervical spine. Clin Spine Surg. 2019;32(3):91–7.

    Article  PubMed  Google Scholar 

  18. Li Y, Shen H, Khan KZ, et al. Comparison of multilevel cervical disc replacement and multilevel anterior discectomy and fusion: a systematic review of biomechanical and clinical evidence. World Neurosurg. 2018;116:94–104.

    Article  PubMed  Google Scholar 

  19. Celik SE, Kara A, Celik S. A comparison of changes over time in cervical foraminal height after tricortical iliac graft or polyetheretherketone cage placement following anterior discectomy. J Neurosurg Spine. 2007;6(1):10–6. https://doi.org/10.3171/spi.2007.6.1.3.

    Article  PubMed  Google Scholar 

  20. Schmieder K, Wolzik-Grossmann M, Pechlivanis I, Engelhardt M, Scholz M, Harders A. Subsidence of the wing titanium cage after anterior cervical interbody fusion: 2-year follow-up study. J Neurosurg Spine. 2006;4(6):447–53. https://doi.org/10.3171/spi.2006.4.6.447.

    Article  PubMed  Google Scholar 

  21. Gercek E, Arlet V, Delisle J, Marchesi D. Subsidence of stand-alone cervical cages in anterior interbody fusion: warning. Eur Spine J. 2003;12(5):513–6. https://doi.org/10.1007/s00586-003-0539-6.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Smith SA, Lindsey RW, Doherty BJ, et al. Cervical spine locking plate: in vitro biomechanical testing. Eur Spine J. 1993;1(4):222–5.

    Article  CAS  PubMed  Google Scholar 

  23. Barbagallo GM, Romano D, Certo F, Milone P, Albanese V. Zero-P: a new zero-profile cage-plate device for single and multilevel ACDF. A single institution series with four years maximum follow-up and review of the literature on zero-profile devices. Eur Spine J. 2013;22(Suppl 6):S868-78. https://doi.org/10.1007/s00586-013-3005-0.

    Article  PubMed  Google Scholar 

  24. Chang BQ, Feng H, Yu CJ, Huang K, Gao X, Tang H, Jiang YC. Comparison of clinical effects between anterior cervical Zero-incision fusion system and traditional nail plate system in the treatment of cervical spondylotic myelopathy. Zhongguo Gu Shang. 2017;30(5):411–6. https://doi.org/10.3969/j.issn.1003-0034.2017.05.005.

    Article  PubMed  Google Scholar 

  25. Pitzen TR, Chrobok J, Stulik J, Ruffing S, Drumm J, Sova L, Kucera R, Vyskocil T, Steudel WI. Implant complications, fusion, loss of lordosis, and outcome after anterior cervical plating with dynamic or rigid plates: two-year results of a multi-centric, randomized, controlled study. Spine (Phila Pa 1976). 2009;34(7):641–6. https://doi.org/10.1097/BRS.0b013e318198ce10.

    Article  PubMed  Google Scholar 

  26. Ghahreman A, Rao PJ, Ferch RD. Dynamic plates in anterior cervical fusion surgery: graft settling and cervical alignment. Spine (Phila Pa 1976). 2009;34(15):1567–71. https://doi.org/10.1097/BRS.0b013e3181a99346.

    Article  PubMed  Google Scholar 

  27. Matgé G, Berthold C, Gunness VR, Hana A, Hertel F. Stabilization with the dynamic cervical implant: a novel treatment approach following cervical discectomy and decompression. J Neurosurg Spine. 2015;22(3):237–45. https://doi.org/10.3171/2014.10.SPINE131089.

    Article  PubMed  Google Scholar 

  28. Pater TJ, Grindel SI, Schmeling GJ, Wang M. Stability of unicortical locked fixation versus bicortical non-locked fixation for forearm fractures. Bone Res. 2014;1(2):14014. https://doi.org/10.1038/boneres.2014.14.

    Article  CAS  Google Scholar 

  29. Little KJ, Riches PE, Fazzi UG. Biomechanical analysis of locked and non-locked plate fixation of the clavicle. Injury. 2012;43(6):921–5. https://doi.org/10.1016/j.injury.2012.02.007.

    Article  CAS  PubMed  Google Scholar 

  30. Sun Y, Yang F, Ma H-N, Gong L, Wang Y-L, Song J-P, Hao Q-Y, Tan M-S. Occipitocervical revision surgery using the bicortical screw and plate system for failed craniovertebral junction stabilization. Orthop Surg. 2022;14:238–45. https://doi.org/10.1111/os.13086.

    Article  PubMed  Google Scholar 

  31. Wilke HJ, Wenger K, Claes L. Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J. 1998;7:148–54.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  32. Panjabi MM. The stabilizing system of the spine. Part II. Neutral zone and instability hypothesis. Clin Spine Surg. 1992;5(4):390–7.

    CAS  Google Scholar 

  33. Voinier SD, Agnew MJ, Carmouche JJ. Passive stiffness characteristics and neutral zone quality of the scoliotic lumbar torso in the principle anatomical planes of motion. Clin Biomech. 2020;80:105162.

    Article  Google Scholar 

  34. Costi JJ, Ledet EH, O’Connell GD. Spine biomechanical testing methodologies: the controversy of consensus vs scientific evidence. JOR Spine. 2021;4(1):e1138.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Crawford NR, Scholz M, Reyes PM, et al. Biomechanics of one-level anterior cervical discectomy and plating using two screws versus four screws. Spine J. 2011;11(3):234–40.

    Article  PubMed  Google Scholar 

  36. Wilke HJ, Kettler A, Claes L. Primary stabilizing effect of interbody fusion devices for the cervical spine: an in vitro comparison between three different cage types and bone cement. Eur Spine J. 2000;9(5):410–6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  37. Yue WM, Brodner W, Highland TR. Persistent swallowing and voice problems after anterior cervical discectomy and fusion with allograft and plating: a 5- to 11-year follow-up study. Eur Spine J Eur Sect Cerv Spine Res Soc. 2005;14(7):677–82.

    Article  Google Scholar 

  38. Sahjpaul RL. Esophageal perforation from anterior cervical screw migration. Surg Neurol. 2007;68(2):205–9.

    Article  PubMed  Google Scholar 

  39. Rhyne III AL, Spector LR, Schmidt GL, et al. Anatomic mapping and evaluation of the esophagus in relation to the cervical vertebral body. Eur Spine J Off Publ Eur Spine Soc Eur Spinal Deform Soc Eur Sect Cerv Spine Res Soc. 2007;16(8):1267–72.

    Article  Google Scholar 

  40. Tasiou A, Giannis T, Brotis AG, et al. Anterior cervical spine surgery-associated complications in a retrospective case-control study. J Spine Surg. 2017;3(3):444–59.

    Article  PubMed  PubMed Central  Google Scholar 

  41. Vaishnav AS, Saville P, McAnany S, et al. Predictive factors of postoperative dysphagia in single-level anterior cervical discectomy and fusion. Spine. 2019;44(7):E400-e407.

    Article  PubMed  Google Scholar 

  42. Wang T, Ma L, Yang DL, et al. Factors predicting dysphagia after anterior cervical surgery: a multicenter retrospective study for 2 years of follow-up. Medicine. 2017;96(34):e7916.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Riley LH 3rd, Skolasky RL, Albert TJ, et al. Dysphagia after anterior cervical decompression and fusion: prevalence and risk factors from a longitudinal cohort study. Spine. 2005;30(22):2564–9.

    Article  PubMed  Google Scholar 

  44. Lee MJ, Bazaz R, Furey CG, et al. Influence of anterior cervical plate design on dysphagia. J Spinal Disord Tech. 2005;18(5):406–9.

    Article  PubMed  Google Scholar 

  45. Lee JH, Park WM, Kim YH, et al. A biomechanical analysis of an artificial disc with a shock-absorbing core property by using whole-cervical spine finite element analysis. Spine. 2016;41(15):E893-e901.

    Article  PubMed  Google Scholar 

  46. Chung JY, Kim SK, Jung ST, et al. Clinical adjacent-segment pathology after anterior cervical discectomy and fusion: results after a minimum of 10-year follow-up. Spine J. 2014;14(10):2290–8.

    Article  PubMed  Google Scholar 

  47. Lawrence BD, Hilibrand AS, Brodt ED, et al. Predicting the risk of adjacent segment pathology in the cervical spine: a systematic review. Spine (Phila Pa 1976). 2012;37(22 Suppl):S52–64.

    Article  PubMed  Google Scholar 

  48. Asghar FA, et al. Radiographic Changes in the Cervical Spine Following Arthrodesis: Causation or Correlation?: Commentary on an article by Raj D. Rao, MD, et al.:“Radiographic Changes in the Cervical Spine Following Anterior Arthrodesis: A Long-Term Analysis of 166 Patients.” JBJS. 2016;98(19):85.

    Article  Google Scholar 

  49. Van Eck CF, Regan C, Donaldson WF, et al. The revision rate and occurrence of adjacent segment disease after anterior cervical discectomy and fusion: a study of 672 consecutive patients. Spine (Phila Pa 1976). 2014;39(26):2143–7.

    Article  PubMed  Google Scholar 

  50. Dong L, Xu Z, Chen X, et al. The change of adjacent segment after cervical disc arthroplasty compared with anterior cervical discectomy and fusion: a meta-analysis of randomized controlled trials. Spine J. 2017;17(10):1549–58.

    Article  PubMed  Google Scholar 

  51. Matsunaga S, Kabayama S, Yamamoto T, et al. Strain on intervertebral discs after anterior cervical decompression and fusion. Spine (Phila Pa 1976). 1999;24(7):670–5.

    Article  CAS  PubMed  Google Scholar 

  52. Eck JC, Humphreys SC, Lim TH, Jeong ST, Kim JG, Hodges SD, An HS. Biomechanical study on the effect of cervical spine fusion on adjacent-level intradiscal pressure and segmental motion. Spine (Phila Pa 1976). 2002;27(22):2431–4.

    Article  PubMed  Google Scholar 

  53. Schwab JS, Diangelo DJ, Foley KT. Motion compensation associated with single-level cervical fusion: where does the lost motion go? Spine (Phila Pa 1976). 2006;31(21):2439–48.

    Article  PubMed  Google Scholar 

  54. Ragab AA, Escarcega AJ, Zdeblick TA. A quantitative analysis of strain at adjacent segments after segmental immobilization of the cervical spine. J Spinal Disord Tech. 2006;19(6):407–10.

    Article  PubMed  Google Scholar 

  55. Chung JY, Park JB, Seo HY, et al. Adjacent segment pathology after anterior cervical fusion. Asian Spine J. 2016;10(3):582–92.

    Article  PubMed  PubMed Central  Google Scholar 

  56. Yang H, Lu X, He H, et al. Longer plate-to-disc distance prevents adjacent-level ossification development but does not influence adjacent-segment degeneration. Spine (Phila Pa 1976). 2015;40(7):E388–93.

    Article  PubMed  Google Scholar 

  57. Lee DH, Lee JS, Yi JS, et al. Anterior cervical plating technique to prevent adjacent-level ossification development. Spine J. 2013;13(7):823–9.

    Article  PubMed  Google Scholar 

  58. Yu C, Mu X, Wei J, et al. In-depth analysis on influencing factors of adjacent segment degeneration after cervical fusion. Med Sci Monit. 2016;22:4902–10.

    Article  PubMed  PubMed Central  Google Scholar 

  59. Park J-B, Watthanaaphisit T, Riew KD. Timing of development of adjacent-level ossification after anterior cervical arthrodesis with plates. Spine J. 2007;7(6):633–6.

    Article  PubMed  Google Scholar 

  60. Park J-B, Cho Y-S, Riew KD. Development of adjacent-level ossification in patients with an anterior cervical plate. J Bone Jt Surg. 2005;87(3):558–63.

    Article  Google Scholar 

  61. Ahn Y. Anterior endoscopic cervical discectomy: surgical technique and literature review. Neurospine. 2023;20(1):11–8.

    Article  PubMed  PubMed Central  Google Scholar 

  62. Chen X, Gao JA, Du Q, et al. Percutaneous full-endoscopic anterior transcorporeal cervical discectomy for the treatment of cervical disc herniation: surgical design and results. Pain Physician. 2021;24:E811–9.

    Article  PubMed  Google Scholar 

  63. Ej T. Cervical spine locking plate fixation for treatment of cervical spondylotic myelopathy in large breed dogs. Vet Surg. 2009;38(6):705–18.

    Article  Google Scholar 

  64. Kocis J, Wendsche P, Vesely R, et al. Complications during and after surgery of the lower cervical spine by isolated anterior approach with CSLP implant. Acta Neurochir (Wien). 2008;150(10):1067–71.

    Article  CAS  PubMed  Google Scholar 

Download references

Funding

This work was supported by Science and Technology Tackling and Key R&D Projects of Henan Provincial Science and Technology Department (232102310411).

Author information

Authors and Affiliations

Authors

Contributions

SL contributed to the conception of the study; SL and YF contributed significantly to literature search, data analyses, and manuscript preparation; RW contributed to improving the article for language and style; PY helped perform the analysis with constructive discussions; and CZ and PY revised the manuscript and approved the final version.

Corresponding author

Correspondence to Changjiang Zhang.

Ethics declarations

Ethics approval and consent for participate

The protocol for this study was approved by the Scientific Ethics Committee of the Second Hospital of Lanzhou University.

Informed consent

All data generated or analysed during this study are included in this published article.

Competing interests

The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, S., Yan, P., Fan, Y. et al. Biomechanical analysis of the door-shaped titanium plate in single-level anterior cervical discectomy and fusion. J Orthop Surg Res 18, 982 (2023). https://doi.org/10.1186/s13018-023-04474-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-023-04474-1

Keywords