Does hydroxyapatite coating have no advantage over porous coating in primary total hip arthroplasty? A meta-analysis

  • Yun-Lin Chen1,
  • Tiao Lin1,
  • An Liu1,
  • Ming-Min Shi1,
  • Bin Hu1,
  • Zhong-li Shi1 and
  • Shi-Gui Yan1Email author
Journal of Orthopaedic Surgery and Research201510:21

https://doi.org/10.1186/s13018-015-0161-4

Received: 7 July 2014

Accepted: 8 January 2015

Published: 28 January 2015

Abstract

There are some arguments between the use of hydroxyapatite and porous coating. Some studies have shown that there is no difference between these two coatings in total hip arthroplasty (THA), while several other studies have shown that hydroxyapatite has advantages over the porous one. We have collected the studies in Pubmed, MEDLINE, EMBASE, and the Cochrane library from the earliest possible years to present, with the search strategy of “(HA OR hydroxyapatite) AND ((total hip arthroplasty) OR (total hip replacement)) AND (RCT* OR randomiz* OR control* OR compar* OR trial*)”. The randomized controlled trials and comparative observation trials that evaluated the clinical and radiographic effects between hydroxyapatite coating and porous coating were included. Our main outcome measurements were Harris hip score (HHS) and survival, while the secondary outcome measurements were osteolysis, radiolucent lines, and polyethylene wear. Twelve RCTs and 9 comparative observation trials were included. Hydroxyapatite coating could improve the HHS (p < 0.01), reduce the incidence of thigh pain (p = 0.01), and reduce the incidence of femoral osteolysis (p = 0.01), but hydroxyapatite coating had no advantages on survival (p = 0.32), polyethylene wear (p = 0.08), and radiolucent lines (p = 0.78). Hydroxyapatite coating has shown to have an advantage over porous coating. The HHS and survival was duration-dependent—if given the sufficient duration of follow-up, hydroxyapatite coating would be better than porous coating for the survival. The properties of hydroxyapatite and the implant design had influence on thigh pain incidence, femoral osteolysis, and polyethylene wear. Thickness of 50 to 80 μm and purity larger than 90% increased the thigh pain incidence. Anatomic design had less polyethylene wear.

Keywords

Hydroxyapatite Porous Harris hip score Survival Total hip arthroplasty

Introduction

Total hip arthroplasty (THA) is one of the most wide operations in orthopedic practice [1]. Cement was widely used, but high rates of failure of cemented femoral components in active patients have been reported. Cemented primary THAs showed excellent results in the short-term but deteriorated with time, while uncemented primary THAs were not only satisfactory in the short-term but also tended to improve with time. Moreover, some studies showed that uncemented and cemented THAs had the comparable clinical results during the follow-up of 6 years, which stimulated the development of implant with uncemented fixation. The uncemented THAs with porous coating allowed bone ingrowth to achieve a rigid fixation. But the problem of osteolysis and stress shielding meant that the long-term stability of uncemented THA was still in question [2,3]. Meanwhile, uncemented stem fixation had the shortcoming of thigh pain. To address such problem, bioactive coating has been added to uncemented component to enhance the fixation by osseointegration of implant, of which hydroxyapatite (HA) was the most popular one.

HA coating accelerates bone healing and enhances the biologic fixation of implant due to its biocompatibility and osteoconductive potential. Several studies have shown that it could reduce the migration of HA-coated prosthetic components and have better results and higher survival rate than identical press-fit components [4]. However, other studies have shown that HA particles delaminated from the stem surface may induce osteolysis either by stimulating bone loss or by migration to the joint space producing third-body wear [5]. Concerns have been raised regarding the technique and parameters that were used in applying the coating to the stem as well [6]. Most reviews of clinical results, X-ray findings, and revision rates were unable to find the significant difference between hydroxyapatite and non-hydroxyapatite-coated stems [7-12]. Some studies showed that hydroxyapatite-coated components could ensure earlier return to activity, reduction in thigh pain, and fewer radiolucent lines [13-16].

A previous meta-analysis by Gandhi et al. [17] based on nine studies including 1,764 samples showed that survival from aseptic loosening had no difference between the two groups at a mean follow-up of 6.5 years, and the mean Harris hip score (HHS) between these groups demonstrated the same. Another meta-analysis by Goosen et al. [18] with eight RCTs including 857 samples reported the clinical and radiographic results, and there was no difference in HHS, endosteal bone ingrowth, and radioactive lines in the surface area of the prosthesis.

The former analysis only included four RCTs of nine studies, and one of which did not show clear HHS data, and the HA stem was grit-blasted, while porous stem was not. The latter analysis did not target new studies published in the later years. It is essential to update the previous results based on the following reasons. (1) The up-to-date 12 RCTs and comparative studies enlarged the sample size to 9,860 and expanded the population distribution. (2) Compared to the average follow-up of 5.4 years in the previous analysis, the longer 7.5 years of the new analysis may lighten on the long-term efficacy of HA coating for HHS, survival, or radiolucent lines. (3) Insufficient data in the previous meta-analysis led to the incomprehensive evaluation of the potential influential factors on HA coating effects, including the thickness and purity of HA coating, the implant design, and the duration of follow-up. The present analysis updated the meta-analysis on the effects of HA-coated stems on clinical and radiographic results, such as HHS, survival, and thigh pain incidence.

Methods

Electronic databases were searched with the limited language of English. The result was last updated on Dec. 17, 2012. The search used the following term and Boolean operators: “(HA OR hydroxyapatite) AND ((total hip arthroplasty) OR (total hip replacement)) AND (RCT* OR randomiz* OR control* OR compar* OR trial*)”. The reference lists of all the selected articles were hand-searched for any additional trials. If necessary, we contacted authors to collect additional information.

The trials was included if (1) the patients had trauma, arthritis, or other diseases requiring total hip arthroplasty with age not less than 18, (2) the inclusion of the comparison between a proximally HA/porous coated femoral and a proximal porous coated stem who underwent primary uncemented total hip arthroplasty, (3) the measurement of outcome was at least by one of the following clinical and radiographic results, including Harris hip score, survival, thing pain incidence, radiolucent lines, femoral osteolysis, and polyethylene wear, and (4) they were published randomized controlled trials or comparison observational studies. Trials were excluded if (1) the patients underwent cemented THAs or revision uncemented THAs, (2) the trials were phase I or case report or review or animal models, or only the abstract was available, (3) the coating was only on the cup not the stem, (4) the follow-up was less than 1 year, and (5) they were sub-analysis of previously published meta-analyses.

Two of us (CYL and SMM) independently assessed each trial with a 12-item scale [19], assessing factors such as randomization, allocation concealment, and blindness. We resolved disagreements through discussion.

For each eligible trial, we extracted relevant data and checked the accuracy. In instances of unreported standard error for a mean difference in HHS, we calculated the standard error by converting the p value to a z-score and solving for the standard error with the formula: z = mean difference/standard error [20]. If the article did not have a certain p value and only a range, we just excluded the data [13,21,22]. For the trials [13], which had more than one intervention group with different ranges of HA coating, we combined group B and group C into one intervention group. For the trials [23,24], in which some identical patients were included but with different durations of follow-up, we included both of them to evaluate HHS.

Our main outcome measurements were mean postoperative HHS and the survival of prosthesis from aseptic loosening. We also assessed the incidence of thigh pain and radiographic results. To improve the clinical relevance, we used the weighed mean difference (WMD) and then estimated the relative difference in the change from baseline as the absolute benefit divided by the mean of all the baseline means of the control groups. With the fixed effect model, WMD and 95% confidence intervals (95% CI) were calculated and the data pooling was done using Review Manager 5.1.7.

We calculated the statistical heterogeneity using a χ 2 test on N − 1 degrees of freedom (N = sample size), with significance at 0.05. We also assessed the inconsistency I 2 using the formula [(Q − df)/Q] × 100% (Q = the χ 2 statistic, df = degree of freedom) to describe the percentage of the variability in effect estimates due to the heterogeneity [19]. We considered I 2 value of 25%, 50%, and 75% as low, medium, and high heterogeneity, respectively. A fixed effects model was used if there was no statistical heterogeneity among the studies; otherwise, we used the random effects model.

We developed several subgroup analyses not only to explain the heterogeneity but also to identify the factors potentially influencing the clinical and radiographic results, such as the design of study, thickness, and purity of hydroxyapatite, follow-up duration, and implant type. Because there was revalent data showing that 50 to 80 μm was the acceptable standard [25,26], 50 to 80 μm was chosen to be the interval of HA thickness. Six-year duration was chosen to be the cutoff follow-up duration as there was sufficient data available allowing for the subgroup analysis. Sensitivity analysis was performed through omitting trials to assess the changes in overall effect. Funnel plots were used to assess publication bias among the included trials graphically. Bias can be seen if the plots were widely skewed versus a plot resembling an inverted triangle which represents no bias [27].

Results

The literature search initially yielded 878 relevant trials. Two of us (CYL and SMM) reviewed the titles and abstracts of all reviews including two of hand-searched. Twelve RCTs [4,22-24,28-35] and nine comparative observation trials [9,10,12,14,16,36-39] were included after applying our eligibility criteria (Figure 1). Funnel plots indicated no publication bias (Figure 2). We recorded the characteristics of 21 trials that were included (Table 1) and details of co-factors and measurement (Table 2). All studies reported a minimum 1 year (median 7.5 years, range 1 to 17.7). We assessed the quality of included trials with the 12-item scale (Table 3). However the intention-to-treat analysis was rarely reported, and no outcome was selectively reported in all studies.
Figure 1

A flowchart illustrated the selection process of eligible trials in our meta-analysis.

Figure 2

Funnel plot for HHS shows no publication bias.

Table 1

Study characteristics

Study

Study design

Sample size (HA/control)

Mean ages of patients (years)

Sex distribution (female/male)

Inclusion criteria

Camazzola et al. [28]

RCT

61 (34/27)

48.2 ± 9.0/50.4 ± 8.7

22/39

Men younger than 60 years and women younger than 65 years having elective primary THA

Dorr et al. [16]

Retrospective, matched pair

30 (15/15)

55 (38–71)

10/5

Patients who underwent bilateral primary THA

Hamadouche et al. [34]

RCT

45 (22/23)

65/64

41/39

Patients with osteoarthritis of the hip requiring THA

Incavo et al. [32]

RCT

50 (24/26)

55

NR

Patients who underwent THA

Kim et al. [23]

RCT

100 (50/50)

45.3 (27–61)

14/36

Patients who underwent sequential bilateral primary THA

Kim et al. [24]

RCT

110 (55/55)

46.3 (27–63)

39/16

Patients who underwent bilateral primary THA

Lee and Lee 2007 [33]

RCT

40 (20/20)

44 (25–72)

2/18

Patients who had late-stage bilateral osteonecrosis were randomly treated with bilateral THA

Lombardi Jr. et al. [12]

Retrospective, observational study

131 (46/85)

52 (29–72)/51 (22–78)

67/97

Patients who underwent THA using a MHP

Mcpherson et al. [36]

Retrospective, matched pair

84 (42/42)

55 ± 11.4/56.5 ± 11.7

36/48

Patients of the same gender, bone type, activity level, and diagnosis, ages within 5 years, weight within 25 pounds, Charnley activity class

Parvizi et al. [9]

Prospective, matched-pair

86 (43/43)

66.8 ± 6.2/65.7 ± 5.9

NR

The patients matched for age, sex, weight, diagnosis, Charnley class, operative approach, bone quality, femoral head size, type of acetabular component, and duration of follow-up

Paulsen et al. [38]

Retrospective comparison

3,158/4,749

NR

3,834/4,073

Patients underwent primary uncemented THA, who were younger than 70 years of age at surgery

Ranawat et al. [35]

RCT

174 (92/82)

54.9 (29.4–67.5)/55.5 (28.6–71.8)

38/114

Patients received cementless THA with the Ranawat-Burstein metaphyseal-diaphyseal fit hip system

Rothman et al. [10]

Retrospective, matched pair

104 (52/52)

64 (31.2–86.1)

49/49

Consecutive THA with use of Taperloc stem, matched for age, sex, weight, diagnosis, Charnley class, operative approach, and duration of follow-up

Sanchez-Sotelo et al. [37]

Retrospective, matched-pair

136 (68/68)

54 (23–66)/56 (22–67)

56/80

Patients who had a primary hip replacement with insertion of either a porous-coated or HA-coated Omniflex femoral componene

Sano et al. [39]

Retrospective, observational study

55 (24/31)

64.0 (51–83)/62.7 (41–80)

49/3

Patients in whom surgery was performed at least 2 years before the present study

Santori et al. [14]

Retrospective, observational study

227 (158/69)

NR

NR

Patients underwent THA with the anatomic prosthesis

Søballe et al. [4]

RCT

26 (14/12)

56.8 (48–63)/58.6 (50–68)

NR

Patients who underwent THR to receive prosthetic with either Ti-alloy coating or HA coating

Tanzer et al. [22]

RCT

39 (17/22)

66 (54–80)/64 (43–78)

13/26

Patients undergoing a cementless THA

Tanzer et al. [31]

RCT

318 (164/154)

64.5 ± 9.9/63.1 ± 10.5

153/165

All patients who underwent cementless THA

Yee et al. [30]

RCT

62 (35/27)

48.2 ± 9.0/50.4 ± 8.7

11/22

Men younger than 60 years of age and women younger than 65 years of age undergoing primary THA

Yoon et al. [29]

RCT

75 (37/38)

45.3 (20–69)/46.0 (23–71)

14/49

Patients who underwent THA use a multilock femoral stem with or without HA/TCP coating

RCT randomized controlled trials, HA hydroxyapatite, NR not reported, THA total hip arthroplasty, TCP tricalcium phosphate.

Table 2

Details of co-factors and measurement of studies

Study

Purity and thickness of HA

Surgery approach

Co-factors

Follow-up (years)

Missing information

Hip implant

Outcome measurement

Camazzola et al. [28]

NR

Hardinge approach

Routine antibiotic prophylaxis; anticoagulation with dicumarin was used preoperatively and for a total of 3 months postoperatively, full weight-bearing for 6 weeks

13 years and 5 months (12–15 years and 3 months)

4 patients were lost to follow-up, 8 died. 1 refuced to participate in the questionnaire or clinical follow-up

Mallory-Head porous stem (Biomet)

HHS, radiographic outcome, thigh pain, survival

Dorr et al. [16]

94% purity, 50–60 μm

NR

NR

6.5 (5–7.9)

No

Anatomic porous replacement—I hip stem (Intermedics Ortho)

HHS, radiographic outcome

Hamadouche et al. [34]

100 ± 30 μm

Posterolateral approach and a Hardinge lateral approach

Postoperative management include administration of systemic antibiotics for 48 h, preventative anticoagulation therapy until full weight-bearing, and NSAID for 5 days to prevent heterotopic ossification, partial weight-bearing was allowed for 6 weeks followed by full weight-bearing

9.18 (3.93–10.28)

One patient from each group died from an unrelated cause at three months and at 27 months after surgery, three patients were lost to follow-up at a mean of three years, of these, one belonged to the HA and two to the GB group

Profile (DePuy)

HHS, radiographic outcome

Incavo et al. [32]

NR

NR

NR

4

no

ProWle (DePuy)

HHS, radiographic outcome

Kim et al. [23]

30 μm

Posterolateral approach

Stand on the second postoperative day. Partial weight-bearing with crutches as tolerated, full weight-bearing was allowed at 6 weeks after surgery

6.6 (5–7)

No

Cementless IPS femoral component (DePuy)

HHS, functional outcome, thigh pain, radiographic outcome, survivorship, complications

Kim et al. [24]

30 μm

Posterolateral approach

NR

15.6 (15–16)

3 were lost to follow-up, 2 died

IPS femoral stem (DePuy)

HHS, WOMAC, thigh pain, Los Angeles activity score, survivorship

Lee and Lee 2007 [33]

98% purity, 150–250 μm

Direct lateral approach (transgluteal approach)

Hip joint motion and ambulation using a wheelchair were allowed from the first postoperative week, crutch walking with partial weight-bearing began 3 to 4 weeks after the second operation. In addition, the patients used a cane for additional 2 to 4 months until they could walk well without any support

143 (123–168)

4 patients died, 5 patients were lost

Spotorno (Zimmer)

Merle d’Aubigne and Postel score, radiographic outcome

Lombardi, Jr. et al. [12]

95% purity, 50–75 μm

NR

Routine clinical evaluation was performed under the supervision of the operative surgeons

14.5 (10.2–16.6)/16.9 (11.4–18.5)

27 patients in the MHP group and 5 patients in the MHP HA group were decreased from causes unrelated to the index surgery. In addition, 14 patients (14 hips, 10.8%) in the MHP group and nine patients (10 hips, 16.4%) in the MHP HA group had not returned for minimum 10-year follow-up and were lost to contact

Mallory-Head porous stem (Biomet)

HHS, radiographic findings, survival

Mcpherson et al. [36]

94% purity, 50–60 μm

NR

NR

3

No

Anatomic porous replacement—I hip stem (intermedics orthopedics)

HHS, radiographic fixation scores

Parvizi et al. [9]

NR

NR

NR

9.2 ± 4.8/10.1 ± 4.6

Each one of the members of nine pairs of patients is dead

Taperloc, Biomet, Warsaw, Indiana

HHS, radiographic findings

Paulsen et al. [38]

95–97% purity, 50–75 μm

NR

NR

3.2

No

Biometric (Biomet)

Time to implant failure

Ranawat et al. [35]

95% purity, 50–75 μm

Posterolateral approach

Standard postoperative rehabilitation protocol

5 (3–8) for thigh pain, 17.7 + −0.8 (16.3–20)

53 patients were deceased, 28 patients were lost to follow-up

Ranawat-Burstein metaphyseal-diaphyseal fit femoral stem (Biomet)

HSS, functional outcome(patient assessment questionnaire), incidence of pain, radiographic outcome, stem subsidence Kaplan-Meier survivorship

Rothman et al. [10]

95% purity, 50–75 μm

NR

Prophylactic antibiotics were given intravenously at the time of the operation and were continued for 48 h. Ten milligrams of low-dose warfarin was given on the night of the operation, instructed to bear only 10% of the body weight on the affected limb for 6 weeks, at which time, they progressed to use of a cane

2.2 (2–3.4)

No

Taperloc stem (Biomet)

Charnley scores, radiographic outcome

Sanchez-Sotelo et al. [37]

NR

NR

NR

6.7 (2.4–9.1)/9.3 (2.2–11.4)

No

Omniflex stem (Osteonics Corporation)

HHS, radiographs, survival

Sano et al. [39]

NR

Posterior approach

Partial weight-bearing was allowed 1 week after the operation, with full weight-bearing after 3 weeks

34/52 m

No

Biomet (Warsaw); Stryker(Fairfield Rd)

HHS, BMD, radiographic outcome

Santori et al. [14]

70% purity, 80–130 μm

NR

Partial weight bearing with two canes was allowed on the fifth postoperative day and progressed to one cane on day 30

70 m (60–84)

No

Anatomic prosthesis (Zimmer)

HHS, thigh pain, radiographic evaluation

Søballe et al. [4]

50–75 μm

Posterolateral approach

Prophylactic antibiotics and anti-thromboembolic drugs, mobilized on the third postoperative day and instructed to walk with protected weight-bearing for the first six postoperative weeks

1

11 patients were excluded from RSA because of technical errors, 1 patient with bilateral THR died from unrelated disease

Biometric (Biomet)

HHS, the visual analog scale score, radiographs data, RSA

Tanzer et al. [22]

80% HA, 20% TCP, 80 μm

Posterolateral approach

All patients remained non-weight-bearing for 6 weeks, followed by progressive weight-bearing as tolerated

2

No

Cementless multilock stem (Zimmer)

HHS, periprosthetic BMD measurement

Tanzer et al. [31]

80% HA, 20% TCP, 80 μm

Posterolateral approach for 64% and 69% in groups uncoated and coated, lateral approach was used in the remainder

All patients were non-weight-bearing for 6 weeks postoperatively, followed by progressive weight-bearing as tolerated

37 m (2–5 years)

16 patients in the group with uncoated and 11 patients in the group with coated components withdraw or were lost to follow-up; 4 in uncoated and 3 in coated died

Cementless multilock stem (Zimmer)

HHS, WOMAC, radiographic data

Yee et al. [30]

95% purity, 50–70 μm

A modified lateral Hardinge approach

Routine prophylactic antibiotic(cefazolin sodium) was administered before surgery and 48 h after surgery; anticoagulation with dicumarin was given: 5 mg orally the night before surgery and daily for a duration of 3 months after surgery. Physical therapy was commenced on the first or second day after surgery. Tough weight-bearing with crutches for 6 weeks was allowed for uncomplicated cases. Progression to full weight-bearing as tolerated was allowed after 6 weeks

4.6 (3–7)

6 patients were lost to follow-up, 1 died of cardiac causes.1 patient declined additional participation in the study after surgery, 1 with bilateral THA was involved in a motor vehicle accident that resulted in a periprosthetic fracture of one hip

Mallory-Head porous femoral stem (Biomet)

HHS, routine radiographs, survivorship

Yoon et al. [29]

70% HA, 30% TCP, 70 μm

Hardinge’s lateral approach

Instruted to walk with partial weight-bearing with the aid of 2 crutches for 4 weeks after surgery

127.4 m (96–144)/127 (108–144)

2 patients in the coated group died of myocardial infarction and cerebral infarction, 2 patients in coated and 2 in uncoated were lost to follow-up

Multilock femoral stem (Zimmer)

HHS, radiographic evaluation, thigh pain

HA hydroxyapatite, TCP tricalcium phosphate, NR not reported, HHS Harris hip score, WOMAC Western Ontario and McMaster Universities Osteoarthritis Index scores, RSA Roentgen stereophotogrammetric analysis, HSS hospital for special surgery hip score.

Table 3

Methodologic quality of included studies

Study

Randomized adequately a

Allocation concealed

Similar baseline

Patient blinded

Care provider blinded

Outcome assessor blinded

Avoid selective reporting

Similar or avoided cofactors

Patients’ compliance b

Acceptable drop-out rate c

Similar timing

ITT analysis d

Quality e

Hamadouche et al. [34]

Yes

Unclear

Yes

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

Yes

No

High

Incavo et al. [32]

Yes

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

High

Kim et al. [23]

Yes

Yes

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

High

Kim et al. [24]

Yes

Unclear

Yes

Yes

Unclear

Yes

Yes

Yes

Yes

Yes

Yes

No

High

Lee and Lee [33]

Yes

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

High

Søballe et al. [4]

Yes

Yes

Yes

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

Yes

No

High

Tanzer et al. [22]

Yes

Yes

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

High

Tanzer et al. [31]

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

No

High

Yee et al. [30]

Yes

Unclear

Yes

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

Yes

No

High

Yoon et al. [29]

Yes

Yes

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

High

Camazzola et al. [28]

Yes

Unclear

Yes

Unclear

Unclear

Yes

Yes

Yes

Yes

No

Yes

No

Moderate

Dorr et al. [16]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Unclear

Yes

Unclear

Moderate

Lombardi, Jr. et al. [12]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

No

Yes

No

Moderate

Mcpherson et al. [36]

Unclear

Unclear

Yes

Yes

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

Moderate

Parvizi et al. [9]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

Moderate

Paulsen et al. [38]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

Moderate

Ranawat et al. [35]

Yes

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

No

Yes

No

Moderate

Rothman et al. [10]

No

Unclear

Yes

No

No

Yes

Yes

Yes

Yes

Yes

Yes

No

Moderate

Sanchez-Sotelo et al. [37]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yesc

Yes

Yes

Yes

No

Moderate

Sano et al. [39]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

Moderate

Santori et al. [14]

Unclear

Unclear

Yes

Unclear

Unclear

Unclear

Yes

Yes

Yes

Yes

Yes

No

Moderate

aOnly if the method of sequence generated was explicitly described could get a “Yes”; sequence generated by “Dates of Admission” or “Patients Number” received a “No”.

bIntermittent treatment or therapy duration less than 6 months means “Yes”, otherwise “No”.

cDrop-out rate ≥ 20% means “No”, otherwise “Yes”.

d ITT intention-to-treat, only if all randomized patients are analyzed in the group they were allocated to could receive a “Yes”.

eThe frequency of “Yes” as 7 or greater means “High”, greater than 4 but less means “Moderate”, 4 or less means “Low”.

The results showed that the HA presented higher HHS than the porous group (15 trials, N = 1,353, WMD = 1.66, 95% CI 0.71 to 2.60, p = 0.0006) (Figure 3), could decrease the thigh pain incidence (6 trials, N = 724, OR = 0.53, 95% CI 0.32 to 0.87, p = 0.01) (Figure 4), and had less femoral osteolysis (5 trials, N = 386, OR = 0.52, 95% CI 0.31 to 0.86, p = 0.01) (Figure 5), while there was no difference in the survivorship from aseptic loosening (16 trials, N = 9,472, RR = 1.00, 95% CI 1.00 to 1.00, p = 0.32) (Figure 6), polyethylene wear (4 trials, N = 347, WMD = −0.02, 95% CI −0.04 to 0.00, p = 0.08) (Figure 7) and radiolucent lines (6 trials, N = 566, OR = 0.95, 95% CI 0.67 to 1.35, p = 0.78) ( Figure 8) between the two groups.
Figure 3

The forest plot for Harris hip score shows HA coating can improve the post-operative HHS compared with porous coating. IV inverse variance, HA hydroxyapatite.

Figure 4

The forest plot for thigh pain incidence shows that HA coating can reduce it compared with porous coating. HA hydroxyapatite.

Figure 5

The forest plot for femoral osteolysis shows that HA coating has less osteolysis compared with porous coating. HA hydroxyapatite.

Figure 6

The forest plot for survival from aseptic loosening shows no difference between HA coating and porous coating. HA hydroxyapatite.

Figure 7

The forest plot for polyethylene wear shows HA coating has less wear compared with porous coating. IV inverse variance, HA hydroxyapatite.

Figure 8

The forest plot for radiolucent lines shows no difference between HA coating and porous coating. HA hydroxyapatite.

There were no heterogeneities for HHS (I 2 = 0%), survivorship from the aseptic loosening (I 2 = 0%), radiolucent lines (I 2 = 0%), or femoral osteolysis (I 2 = 46%). The results of thigh pain incidence and polyethylene wear was heterogeneous in some degree (I 2 = 56%; I 2 = 96%). The heterogeneity of thigh pain incidence can not be explained by the study design, quality of study, thickness of HA, implant design or follow-up duration maybe because of the potential co-factors, such as weight-bearing after the operation. As to the limitation of this analysis, we could not set this subgroup analysis. The polyethylene wear can not be explained by purity of HA and duration of follow-up. But when we classified the polyethylene wear into subgroups by the thickness of HA (50–80 μm and >80 μm or < 50 μm) and implant design (anatomic and non-anatomic), the heterogeneity could no longer be observed (I 2 = 0%).

Subgroup analysis for HHS, survival of implant from aseptic loosening, and radiolucent lines indicated that the included non-RCTs did not affect the RCTs (p = 0.42; p = 0.27; p = 0.98; p = 0.52), while as to incidence of thigh pain and femoral osteolysis, there was a slight difference (p = 0.02; p = 0.05). Our results showed that the longer duration of follow-up tends to have higher HHS (p = 0.11) than the shorter one (WMD = 2.21, 95% CI 1.05 to 3.37 for duration of follow-up > 6 years; WMD = 0.58, 95% CI −1.04 to 2.20 for duration < 6 years) and likewise for the survival (p = 0.11, RR = 1.01 95% CI 0.99 to 1.03 for duration of follow-up > 6 years, RR = 1.00, 95% CI 0.99 to 1.00 for duration of follow-up < 6 years). The thickness of HA was larger than 80 μm or less than 50 μm, the purity less than 90%, and the anatomic implant reduced incidence of thigh pain and the duration of follow-up did not affect it. It is doubtful that the RCTs and high-quality study had the high incidence of thigh pain and femoral osteolysis (Table 4). The overall result was not significantly altered by omitting trials with a sample sizes less than 70 or those with imputed data.
Table 4

Subgroup analysis of the included studies by different influential factors

Factors

Harris hip score

Survival

Incidence of thigh pain

Radiolucent lines

Femoral osteolysis

Polyethylene wear

 

Subgroups (numbers)

WMD (95% CI)

Subgroups (numbers)

RR (95% CI)

Subgroups (numbers)

OR (95% CI)

Subgroups (numbers)

OR (95% CI)

Subgroups (numbers)

OR (95% CI)

Subgroups (numbers)

WMD (95% CI)

Study design

RCT (10)

1.41 (0.30, 2.52)

RCT (10)

1.01 (0.99, 1.02)

RCT (4)

0.73 (0.42, 1.28)

RCT (4)

0.95 (0.65, 1.39)

RCT (2)

1.37 (0.45, 4.17)

RCT (4)

N.A.

Non-RCT (5)

2.29 (0.50, 4.07)

Non-RCT (6)

1.00 (0.99, 1.00)

Non-RCT (2)

0.14 (0.04, 0.49)

Non-RCT (2)

0.96 (0.41, 2.24)

Non-RCT (3)

0.39 (0.22, 0.70)

Non-RCT (0)

 

p = 0.42

 

p = 0.27

 

p = 0.02

 

p = 0.98

 

p = 0.05

 

Study quality

High (9)

1.41 (0.30, 2.52)

High (8)

1.01 (0.99, 1.02)

High (3)

1.29 (0.57, 2.88)

High (4)

0.95 (0.65, 1.39)

High (2)

1.24 (0.38, 4.13)

High (4)

N.A.

Moderate (6)

2.28 (0.50, 4.07)

Moderate (8)

1.00 (0.99, 1.00)

Moderate (3)

0.30 (0.15, 0.57)

Moderate (2)

0.96 (0.41, 2.24)

Moderate (2)

0.42 (0.24, 0.74)

Moderate (0)

 

p = 0.42

 

p = 0.27

 

p = 0.006

 

p = 0.98

 

p = 0.11

 

Thickness of HA

50–80 μm (8)

0.76 (−0.63, 2.15)

50–80 μm (10)

1.00 (0.99, 1.00)

50–80 μm (3)

0.71 (0.40, 1.27)

50–80 μm (5)

N.A.

50–80 μm (3)

N.A.

50–80 μm (2)

0.00 (−0.01, 0.01)

<50 μm or >80 μm (3)

2.03 (0.41, 3.64)

<50 μm or >80 μm (4)

1.02 (0.98, 1.06)

<50 μm or >80 μm (2)

0.19 (0.06, 0.58)

<50 μm or >80 μm (0)

<50 μm or >80 μm (0)

<50 μm or >80 μm (2)

−0.03 (−0.03, −0.03)

 

p = 0.24

 

p = 0.23

 

p = 0.04

   

p < 0.00001

Purity of HA

>90% (4)

−0.09 (−3.04, 2.86)

>90% (7)

1.00 (0.99, 1.00)

>90% (2)

0.68 (0.36, 1.26)

>90% (3)

0.73 (0.31, 1.71)

>90% (2)

0.66 (0.27, 1.62)

>90% (1)

0.01 (−0.06, 0.08)

<90% (3)

0.82 (−0.77, 2.42)

<90% (3)

1.00 (0.98, 1.02)

<90% (3)

0.23 (0.08, 0.65)

<90% (2)

0.93 (0.61, 1.40)

<90% (1)

1.24 (0.38, 4.13)

<90% (1)

0.00 (−0.01, 0.01)

 

p = 0.59

 

p = 0.72

 

p = 0.08

 

p = 0.63

 

p = 0.4

 

p = 0.79

Implant design

Anatomic (5)

1.65 (0.15, 3.16)

Anatomic (4)

1.02 (0.99, 1.05)

Anatomic (2)

0.19 (0.06, 0.58)

Anatomic (1)

0.32 (0.07, 1.47

Anatomic (2)

0.66 (0.27, 1.62)

Anatomic (2)

−0.03 (−0.03, −0.03)

Non-anatomic (10)

1.66 (0.44, 2.87)

Non-anatomic (12)

1.00 (1.00, 1.00)

Non-anatomic (4)

0.72 (0.41, 1.28)

Non-anatomic (5)

1.01 (0.71, 1.45)

Non-anatomic (3)

0.46 (0.25, 0.85

Non-anatomic (2)

0.00 (−0.01, 0.01)

 

p = 1.00

 

p = 0.24

 

p = 0.04

 

p = 0.15

 

p = 0.52

 

p < 0.00001

Follow-up duration

>6 years (9)

2.21 (1.05, 3.37)

>6 years (10)

1.01 (0.99, 1.03)

>6 years (4)

0.56 (0.29, 1.08)

>6 years (2)

0.58 (0.24, 1.43)

>6 years (4)

0.44 (0.24, 0.77)

>6 years (3)

−0.02 (−0.02, −0.01)

<6 years (6)

0.58 (−1.04, 2.20)

<6 years (6)

1.00 (0.99, 1.00)

<6 years (2)

0.50 (0.24, 1.05)

<6 years (4)

1.04 (0.71, 1.52)

<6 years (1)

1.00 (0.32, 3.15)

<6 years (1)

0.01 (−0.06, 0.08)

 

p = 0.11

 

p = 0.11

 

p = 0.82

 

p = 0.25

 

p = 0.2

 

p = 0.46

Study design and study quality would affect the incidence of thigh pain and study design has influence on femoral osteolysis. When the thickness of HA is <50 or >80 μm, it has less thigh pain incidence and polyethylene wear. The anatomic implant has less incidence of thigh pain and polyethylene wear.

WMD weighted mean difference, N.A. not available. HA hydroxyapatite.

Discussion

The primary finding is that HA coating could improve the postoperative HHS, reduce the incidence of thigh pain, and reduce the incidence of femoral osteolysis while there was no statistical difference of femoral stem survivorship from aseptic loosening, polyethylene wear, and radiolucent lines between the two groups. In addition, the subgroup analyses found that HHS tends to improve in the longer duration of follow-up and so was the prosthetic survival. The longer duration of follow-up, the better advantage of HA coating over porous coating for the HHS and survivorship from aseptic loosening.

In this meta-analysis, we asked: (1) which coating is better with regard to the clinical and radiologic measurements and (2) which modifying factors affect the comparative effect between both coatings.

To the best of our knowledge, the present meta-analysis is the first to comprise all the available comparative observational evidence and to comprehensively investigate the difference in HHS and survivorship and radiographic outcomes between HA and porous coating for THA. As the previous systematic review only included four RCTs of nine studies, one of the included studies did not show clear HHS data and the HA stem was grit-blasted, while porous stem was not. Another previous systematic review did not include new studies published in the later years. We included 12 RCTs and 9 comparison observation studies and developed explicit inclusion and exclusion criteria. Our analysis quintupled the sample size compared with previous meta-analysis (9,860 versus 1,764) and had longer duration of follow-up (7.5 years versus 6.5 years). We performed a comprehensive set of subgroup analyses and a sensitivity analysis not only to explain the heterogeneity but also to provide additional insights into the potential influential factors of HHS, survival, thigh pain, and other radiographic outcome.

Our meta-analysis has some limitations. (1) There was variability in the selection criteria of individual trials, including the primary disease, gender, ages of patients, and the type of prosthesis. (2) due to the limited number of included trials, we could not analyze the influence of other clinically relevant factors, such as complications of THA, BMD, and WOMAC osteoarthritis index. (3) Missing information such as declining participation and crossover led to incomplete data and potentially bias. (4) The small sample size in the subgroup analysis reduced the precision of the pooled estimates and the ability to detect the statistical significance of some variables, that is, polyethylene wear. More RCTs would be warranted to clarify them. (5) With the limitation of included studies, we can not analyze the effect of implant design, which needs more study to assess.

One of the most significant results of our analysis is that HA coating had higher HHS and less incidence of thigh pain. The advantage of an HA coating includes superior proximal femoral osteointegration and better preservation of periprosthetic bone quality. The patients with HA-coated stems demonstrated significantly lower incidence of activity-related trochanteric and thigh pain [35]. The thickness and purity of HA and implant design could affect the incidence of thigh pain. The incidence decreased abruptly after the first postoperative year [40]. Thigh pain following uncemented hip arthroplasty was generally transient and would disappear over time.

Femoral osteolysis and polyethylene wear could result to gradual subsidence or loosening of implant. Some studies showed that HA coating had less osteolysis and polyethylene wear [37,16], but in Almeida’s study, with the use of HA coated stem, they still found 38% of osteolysis and 41% of polyethylene wear in the hips. Their stems were mostly used in young patients, who had greater activity compared to the older patients which might have been the affecting factor. From the retrieved specimen, some studies found that HA coating could increase the amount of ingrowth and attachment of bone leading to the enhanced biological fixation [41]. Moreover, HA-coated Ti implants can achieve a much higher degree of bone apposition and mechanical stability compared to the implants without such a coating [26]. The heterogeneity of polyethylene wear can not be explained by purity of HA and duration of follow-up. But when we classified the polyethylene wear into subgroups by the thickness of HA (50–80 μm and >80 μm or < 50 μm) and implant design (anatomic and non-anatomic), the heterogeneity could no longer be observed.

The geometry design of the implant has a large impact on the clinical outcome. Joshi’s study explored the hypothesis that through redesign, a total hip prosthesis could be developed to substantially reduce stress shielding, then reduce the loosening of the prosthesis [42]. Dopico-González assessed effects of implant design geometry by probabilistic finite element tool, she thought the geometry of the implant design clearly affected the sensitivities of maximum nodal micromotion [43]. As the limitation of included studies, we just divided the implant into anatomic and non-anatomic. It needs more studies to assess the effect of implant design. Our study showed that anatomic implant had less incidence of thigh pain and reduced polyethylene wear. Ando et al. found that FMS-anatomic stem reduced the proximal stem-bone relative motion and transferred more load to the proximal femur compared to conventional symmetric stems and the FMS, which resulted in better biomechanical stability at least in the early postoperative period [44]. Another study showed an excellent clinical outcome and 98.3% survival of ABG II implant with HA coating [45]. In Cao’s study, they also thought that an anatomically designed prosthesis can provide good clinical results, with low incidence of thigh pain and loosening of the component [46].

HA is biocompatible and osteoconductive and in contact with bone often develops a mechanically tight bond. Human retrieval studies have shown that HA-coated stem observed significantly more ingrowth and attachment of the bone [47].

Conclusion

In conclusion, HA is better than porous coating. HA coating could improve the postoperative HHS, reduce the incidence of thigh pain, and reduce the incidence of femoral osteolysis, while there was no statistical difference of femoral stem survivorship from aseptic loosening, polyethylene wear, and radiolucent lines between the two groups.

Abbreviations

HA: 

hydroxyapatite

HHS: 

Harris hip score

THA: 

total hip arthroplasty

RCT: 

randomized controlled trial

WMD: 

weighed mean difference

CI: 

confidence intervals

RR: 

relative risk

BMD: 

bone mineral density

WOMAC: 

Western Ontario and McMaster Universities Arthritis Index

Declarations

Acknowledgements

This work is supported by the National Natural Science Foundation of China (81101377, 81101345, 81171687, 81371954, 81201414, 81201416, 81401785), the Key Project of Zhejiang Provincial Department of Science and Technology (2011C13033), the Zhejiang Provincial Natural Science Foundation of China (Y2100161,Y2090283), the Scientific Research Fund of Zhejiang Provincial Education Department (Y201018936), and the Medical Scientific Research Foundation of Zhejiang Provience of China (2012RCA032).

Authors’ Affiliations

(1)
Department of Orthopaedic Surgery, Second Affiliated Hospital, School of Medicine, Zhejiang University

References

  1. Prieto-Alhambra D, Javaid MK, Judge A, Maskell J, Kiran A, de Vries F, et al. Fracture risk before and after total hip replacement in patients with osteoarthritis: potential benefits of bisphosphonate use. Arthritis Rheum. 2011;63:992–1001.PubMedView ArticleGoogle Scholar
  2. Manley MT, D’Antonio JA, Capello WN, Edidin AA. Osteolysis: a disease of access to fixation interfaces. Clin Orthop. 2002;405:129–37.PubMedView ArticleGoogle Scholar
  3. Kim YH, Park JW, Patel C, Kim DY. Polyethylene wear and osteolysis after cementless total hip arthroplasty with alumina-on-highly cross-linked polyethylene bearings in patients younger than thirty years of age. J Bone Joint Surg Am. 2013;95(12):1088–93.PubMedView ArticleGoogle Scholar
  4. Søballe K, Toksvig-Larsen S, Gelineck J, Fruensgaard S, Hansen ES, Ryd L, et al. Migration of hydroxyapatite coated femoral prostheses. A Roentgen Stereophotogrammetric study. J Bone Joint Surg Br. 1993;75(5):681–7.PubMedGoogle Scholar
  5. D’Angelo F, Molina M, Riva G, Zatti G, Cherubino P. Failure of dual radius hydroxyapatite-coated acetabular cups. J Orthop Surg Res. 2008;3:35.PubMed CentralPubMedView ArticleGoogle Scholar
  6. Bloebaum RD, Beeks D, Dorr LD, Savory CG, DuPont JA, Hofmann AA. Complications with hydroxyapatite particulate separation in total hip arthroplasty. Clin Orthop. 1994;298:19–26.PubMedGoogle Scholar
  7. Geesink RGT, Hoefnagels NHM. Six year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg Br. 1995;77(4):534–47.PubMedGoogle Scholar
  8. Park YS, Lee JY, Yun SH, Jung MW, Oh I. Comparison of hydroxyapatite- and porous-coated stems in total hip replacement. Acta Orthop Scand. 2003;74(3):259–63.PubMedView ArticleGoogle Scholar
  9. Parvizi J, Sharkey PF, Hozack WJ, Orzoco F, Bissett GA, Rothman RH. Prospective matched-pair analysis of hydroxyapatite-coated and uncoated femoral stems in total hip arthroplasty. A concise follow-up of a previous report. J Bone Joint Surg Am. 2004;86-A(4):783–6.PubMedGoogle Scholar
  10. Rothman RH, Hozack WJ, Ranawat A, Moriarty L. Hydroxyapatite-coated femoral stems. A matched-pair analysis of coated and uncoated implants. J Bone Joint Surg Am. 1996;78(3):319–24.PubMedGoogle Scholar
  11. Sharp RJ, O’Leary ST, Falworth M, Cole A, Jones J, Marshall RW. Analysis of the results of the C-Fit uncemented total hip arthroplasty in young patients with hydroxyapatite or porous coating of components. J Arthroplasty. 2000;15(5):627–34.PubMedView ArticleGoogle Scholar
  12. Lombardi Jr AV, Berend KR, Mallory TH. Hydroxyapatite-coated titanium porous plasma spray tapered stem: experience at 15 to 18 years. Clin Orthop Relat Res. 2006;453:81.PubMedView ArticleGoogle Scholar
  13. Rasquinha VJ, Ranawat CS, Mauriello Jr AJ. Hydroxyapatite: catalyst or conjuror? J Arthroplasty. 2002;17 Suppl 1:113–7.PubMedView ArticleGoogle Scholar
  14. Santori FS, Ghera S, Moriconi A, Montemurro G. Results of the anatomic cementless prosthesis with different types of hydroxyapatite coating. Orthopedics. 2001;24(12):1147–50.PubMedGoogle Scholar
  15. Ciccotti MG, Rothman RH, Hozack WJ, Moriarty L. Clinical and roentgenographic evaluation of hydroxyapatite-augmented and nonaugmented porous total hip arthroplasty. J Arthroplasty. 1994;9(6):631–9.PubMedView ArticleGoogle Scholar
  16. Dorr LD, Wan Z, Song M, Ranawat A. Bilateral total hip arthroplasty comparing hydroxyapatite coating to porous-coated fixation. J Arthroplasty. 1998;13(7):729–36.PubMedView ArticleGoogle Scholar
  17. Gandhi R, Davey JR, Mahomed NN. Hydroxyapatite coated femoral stems in primary total hip arthroplasty: a meta-analysis. J Arthroplasty. 2009;24(1):38–42.PubMedView ArticleGoogle Scholar
  18. Goosen JH, Kums AJ, Kollen BJ, Verheyen CC. Porous-coated femoral components with or without hydroxyapatite in primary uncemented total hip arthroplasty: a systematic review of randomized controlled trials. Arch Orthop Trauma Surg. 2009;129(9):1165–9.PubMedView ArticleGoogle Scholar
  19. Furlan AD, Pennick V, Bombardier C, van Tulder M. Updated method guidelines for systematic reviews in the Cochrane Back Review Group. Spine. 2009;34:1929–41.PubMedView ArticleGoogle Scholar
  20. Higgins JPT, Green S. Cochrane Collaboration. Cochrane handbook for systematic reviews of interventions. Chichester, England and Hoboken NJ: Wiley-Blackwell; 2008.View ArticleGoogle Scholar
  21. Rorabeck C. Tapered hydroxyapatite-coated press-fit stems: any added value? J Arthroplasty. 2006;21(4 Suppl 1):85–8.PubMedView ArticleGoogle Scholar
  22. Tanzer M, Kantor S, Rosenthall L, Bobyn JD. Femoral remodeling after porous-coated total hip arthroplasty with and without hydroxyapatite-tricalcium phosphate coating: a prospective randomized trial. J Arthroplasty. 2001;16(5):552–8.PubMedView ArticleGoogle Scholar
  23. Kim YH, Kim JS, Oh SH, Kim JM. Comparison of porous-coated titanium femoral stems with and without hydroxyapatite coating. J Bone Joint Surg Am. 2003;85-A(9):1682–8.PubMedGoogle Scholar
  24. Kim YH, Kim JS, Joo JH, Park JW. Is hydroxyapatite coating necessary to improve survivorship of porous-coated titanium femoral stem? J Arthroplasty. 2012;27(4):559–63.PubMedView ArticleGoogle Scholar
  25. Jaffe WL, Scott DF. Rationale and clinical application of hydroxyapatite coatings in pressfit total hip arthroplasty. Semin Arthroplasty. 1993;4(3):159–66.PubMedGoogle Scholar
  26. Geesink RG. Osteoconductive coatings for total joint arthroplasty. Clin Orthop Relat Res. 2002;395:53–65.PubMedView ArticleGoogle Scholar
  27. Egger M, Davey Smith G, Schneider M, Minder C. Bias in meta-analysis detected by a simple, graphical test. BMJ. 1997;315(7109):629–34.PubMed CentralPubMedView ArticleGoogle Scholar
  28. Camazzola D, Hammond T, Gandhi R, Davey JR. A randomized trial of hydroxyapatite-coated femoral stems in total hip arthroplasty: a 13-year follow-up. J Arthroplasty. 2009;24(1):33–7.PubMedView ArticleGoogle Scholar
  29. Yoon KS, Kim HJ, Lee JH, Kang SB, Seong NH, Koo KH. A randomized clinical trial of cementless femoral stems with and without hydroxyapatite tricalcium-phosphate coating. An 8- to 12-year follow-up study. J Arthroplasty. 2007;22(4):504–8.PubMedView ArticleGoogle Scholar
  30. Yee AJ, Kreder HK, Bookman I, Davey JR. A randomized trial of hydroxyapatite coated prosthesis in total hip arthroplasty. Clin Orthop Relat Res. 1999;366:120–32.PubMedView ArticleGoogle Scholar
  31. Tanzer M, Gollish J, Leighton R, Orrell K, Giacchino A, Welsh P, et al. The effect of adjuvant calcium phosphate coating on a porous-coated femoral stem. Clin Orthop Relat Res. 2004;424:153–60.PubMedView ArticleGoogle Scholar
  32. Incavo SJ, Schnelder R, Elting J. The effect of surface coating of femoral prostheses implanted without cement: a 2- to 4-year follow-up study. Am J Orthop (Belle Mead, NJ). 1998;27(5):355–61.Google Scholar
  33. Lee JM, Lee CW. Comparison of hydroxyapatite-coated and non-hydroxyapatite-coated noncemented total hip arthroplasty in same patients. J Arthroplasty. 2007;22(7):1019–23.PubMedView ArticleGoogle Scholar
  34. Hamadouche M, Witvoet J, Porcher R, Meunier A, Sedel L, Nizard R. Hydroxyapatite-coated versus grit-blasted femoral stems, a prospective, randomised study using EBRA-FCA. J Bone Joint Surg Br. 2001;83(7):979–87.PubMedView ArticleGoogle Scholar
  35. Ranawat CS, Rasquinha VJ, Meftah M, Ranawat AS. The hydroxyapatite-tapered stem: any added value? Semin Arthroplasty. 2011;22:75–8.View ArticleGoogle Scholar
  36. McPherson EJ, Dorr LD, Gruen TA, Saberi MT. Hydroxyapatite-coated proximal ingrowth femoral stems. A matched pair control study. Clin Orthop Relat Res. 1995;315:223–30.PubMedGoogle Scholar
  37. Sanchez-Sotelo J, Lewallen DG, Harmsen WS, Harrington J, Cabanela ME. Comparison of wear and osteolysis in hip replacement using two different coatings of the femoral stem. Int Orthop. 2004;28(4):206–10.PubMed CentralPubMedGoogle Scholar
  38. Paulsen A, Pedersen AB, Johnsen SP, Riis A, Lucht U, Overgaard S. Effect of hydroxyapatite coating on risk of revision after primary total hip arthroplasty in younger patients: findings from the Danish Hip Arthroplasty Registry. Acta orthop. 2007;78(5):622–8.PubMedView ArticleGoogle Scholar
  39. Sano K, Ito K, Yamamoto K. Changes of bone mineral density after cementless total hip arthroplasty with two different stems. Int Orthop. 2008;32(2):167–72.PubMed CentralPubMedView ArticleGoogle Scholar
  40. Brown TE, Larson B, Shen F, Moskal JT. Thigh pain after cementless total hip arthroplasty: evaluation and management. J Am Acad Orthop Surg. 2002;10(6):385–92.PubMedGoogle Scholar
  41. Hernández CP, Nájera SO, Mesa RF, Pajares LM, Hernández HM. Hydroxyapatite-coated stems with metaphyseal and diaphyseal press-fit. Eleven-year follow-up results. Acta Orthop Belg. 2002;68(1):24–32.Google Scholar
  42. Joshi MG, Advani SG, Miller F, Santare MH. Analysis of a femoral hip prosthesis designed to reduce stress shielding. J Biomech. 2000;33(12):1655–62.PubMedView ArticleGoogle Scholar
  43. Dopico-González C, New AM, Browne M. Probabilistic finite element analysis of the uncemented hip replacement–effect of femur characteristics and implant design geometry. J Biomech. 2010;43(3):512–20.PubMedView ArticleGoogle Scholar
  44. Ando M, Imura S, Omori H, Okumura Y, Bo A, Baba H. Nonlinear three-dimensional finite element analysis of newly designed cementless total hip stems. Artif Organs. 1999;23(4):339–46.PubMedView ArticleGoogle Scholar
  45. Herrera A, Mateo J, Lobo-Escolar A, Panisello JJ, Ibarz E, Gracia L. Long-term outcomes of a new model of anatomical hydroxyapatite-coated hip prosthesis. J Arthroplasty. 2013;28(7):1160–6.PubMedView ArticleGoogle Scholar
  46. Cao CF, Zhou JJ, Pang JH, Chen XQ. A five-year clinical and radiographic follow-up of bipolar hip arthroplasty with insertion of a porous-coated anatomic femoral component without cement. Orthop Surg. 2011;3(2):88–94.PubMedView ArticleGoogle Scholar
  47. Coathup MJ, Blunn GW, Flynn N, Williams C, Thomas NP. A comparison of bone remodelling around hydroxyapatite-coated, porous-coated and grit-blasted hip replacements retrieved at post-mortem. J Bone Joint Surg Br. 2001;83(1):118–23.PubMedView ArticleGoogle Scholar

Copyright

© Chen et al.; licensee BioMed Central. 2015

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.