Biomechanical and clinical outcomes of 3D-printed versus modular hemipelvic prostheses for limb-salvage reconstruction following periacetabular tumor resection: a mid-term retrospective cohort study

Abstract

Background

Debates persist over optimal pelvic girdle reconstruction after acetabular tumor resection, with surgeons grappling between modular and 3D-printed hemipelvic endoprostheses. We hypothesize superior outcomes with 3D-printed versions, yet scarce comparative research exists. This study fills the gap, examining biomechanics and clinical results retrospectively.

Methods

From February 2017 to June 2021, we retrospectively assessed 32 patients undergoing en bloc resection for malignant periacetabular tumors at a single institution. Primary outcome: limb function. Secondary outcomes: implant precision, hip joint rotation center restoration, prosthesis-bone osteointegration, and complications. Biomechanical characteristics were evaluated through finite element analysis on pelvic defect models.

Results

In the 3D-printed group, stress distribution mirrored a normal pelvis, contrasting the modular group with elevated overall stress, unstable transitions, and higher stress peaks. The 3D-printed group exhibited superior functional scores (MSTS: 24.3 ± 1.8 vs. 21.8 ± 2.0, p < 0.05; HHS: 79.8 ± 5.2 vs. 75.3 ± 3.5, p < 0.05). Prosthetic-bone interface osteointegration, measured by T-SMART, favored 3D-printed prostheses, but surgery time (426.2 ± 67.0 vs. 301.7 ± 48.6 min, p < 0.05) and blood loss (2121.1 ± 686.8 vs. 1600.0 ± 505.0 ml, p < 0.05) were higher.

Conclusions

The 3D-printed hemipelvic endoprosthesis offers precise pelvic ring defect matching, superior stress transmission, and function compared to modular endoprostheses. However, complexity, fabrication expertise, and challenging surgical implantation result in prolonged operation times and increased blood loss. A nuanced consideration of functional outcomes, complexity, and patient conditions is crucial for informed treatment decisions.

Level of evidence

Level III, therapeutic study (Retrospective comparative study).

Introduction

Primary malignant bone tumors involving the pelvic girdle constitute around 10–15% of cases [11, 20]. Currently, limb-salvage surgeries are the primary surgical treatment for malignant pelvic bone tumors, owing to advancements in adjuvant chemotherapy, imaging, and surgical techniques [6, 8]. Surgical reconstruction options following tumor resection include arthrodesis [29], hip transposition [41], allograft/autograft reconstruction [2, 4, 34], and endoprosthetic reconstruction [1, 5, 11, 19, 20, 26, 27]. Among them, endoprosthetic reconstructions are preferred for their stability, aesthetics, early mobility, and the absence of risks of bone grafts-related complications.

In clinical practice, various hemipelvic endoprostheses are utilized, including saddle prostheses, ice-cream cone prostheses, modular prostheses, and 3D-printed hemipelvic prostheses [5, 7, 11, 18, 27]. However, saddle and ice-cream cone prostheses demand a significant volume of retained ilium for fixation, limiting their clinical application [30]. In contrast, modular hemipelvic endoprostheses offer the advantage of being assembled flexibly during pelvic tumor resection, even in cases involving extensive iliac bone resection [12, 21]. The smaller dimensions of these endoprostheses facilitate generous soft tissue coverage during surgical procedures, effectively minimizing residual dead space and ensuring robust muscle reconstruction. These advantages not only enhance the efficacy of hip joint function rehabilitation but also hold promise, in theory, for reducing the incidence of deep postoperative infections in patients [11, 22]. However, modular endoprostheses exhibit inadequate interface matching and their fixation method does not conform to optimal mechanical transmission [10]. Furthermore, the absence of osteointegration at the interface results in mechanical failures during later stages. [1, 27, 39]. Despite the incorporation of porous structures on the surface of modular prostheses by certain scholars, the fundamental issue of inadequate stress transmission persists [9, 23]. In recent years, 3D-printed hemipelvic endoprostheses have gained increasing attention [15]. The core advantage of 3D printing technology resides in its aptitude for precise customization. These innovative prostheses can be tailored to conform to any irregular pelvic bone deficiency. Moreover, the incorporation of a porous structure on the prosthesis surface promotes osteointegration [38]. As such, this technology has the potential to address the challenges of prosthesis integration with native bone and ensure its enduring stability over time. Over the recent decades, 3D-printed hemipelvic endoprostheses emerge as a promising alternative reconstruction option for patients with malignant tumors around the acetabulum [24, 37, 38, 40, 42].

Nowadays, pelvic girdle reconstruction after acetabular tumor resection lacks a consensus on the optimal approach. Surgeons face a dilemma between modular endoprostheses and emerging 3D printing technology, each with its own pros and cons. Regarding these two essential pelvic ring reconstruction methods, it appears that 3D-printed hemipelvic endoprostheses, when it comes to anatomical pelvic ring reconstruction, mimic stress transmission patterns akin to those seen in native human pelvises. However, no studies have yet been reported to ascertain whether they exhibit superior biomechanical performance compared to non-anatomical reconstruction using modular endoprostheses or if this leads to improved mid-term clinical outcomes. To the best of our knowledge, prior research on these two critical reconstruction methods has been lacking in controlled comparative studies or relevant biomechanical analysis. Herein two surgical approaches for pelvic girdle reconstruction after periacetabular tumor resection in 32 patients are compared both clinically and biomechanically in this study aimed at the identification of a better operative strategy.

Methods

Clinical study and patients

This single-center retrospective study was performed in accordance with the 1964 Helsinki Declaration and was authorized by the Ethics Committee of our hospital. Written informed consent was obtained from adult participants or parents of minors (below 16 years of age). The work has been reported in line with the STROCSS criteria [25].

We retrospectively analyzed the results of patients who underwent either 3D-printed or modular endoprosthestic reconstruction for the treatment of pelvic bone tumor between February 2017 and June 2021. Given the absence of significant differences in the indications for modular endoprosthesis and 3D-printed endoprosthetic reconstruction, both approaches are viable for reconstruction within the same patient. The decision on which reconstruction method to employ hinges on detailed communication with the patient, taking into account differences in postoperative reconstruction, prosthetic costs, and the time required for design and production. Ultimately, the choice is made based on the patient's preferences. In this retrospective analysis, we included patients with comparable numbers, tumor locations, and tumor sizes who participated in the study. Thus, the inclusion criteria were: (1) Pathological confirmation of a primary or metastatic pelvic malignant tumor. (2) Absence of contraindications for en bloc resection. (3) Life expectancy exceeding 6 months. (4) Comparable planned resection margin to that achieved in hemipelvectomy. (5) Expected preservation of reasonable function post-resection. (6) Utilization of 3D-printed custom-made hemipelvic endoprostheses or modular hemipelvic prostheses for reconstruction. (7) Availability of comprehensive follow-up data. The exclusion criteria were: (1) Inability to achieve a satisfactory surgical margin while pursuing limb-salvage procedures; (2) Incapacity to preserve a functional limb due to tumor involvement of the sacral or sciatic nerve; (3) Patients presenting with unresectable and/or extensively metastatic disease; (4) Active infection in the proximity of the prosthesis implantation site; (5) Allergy to metal implants; (6) Severe reduction in muscular strength of the affected limb or significant impairment of other joints impacting functional assessment; (7) Profound osteoporosis.

After careful assessment of eligibility criteria and obtaining informed consent from all participants, a total of 32 patients were included in this study. Patients with similar preoperative tumor locations (Type I + II, Type II + III, Type I + II + III) and comparable tumor sizes but differing in the surgical reconstruction approach were divided into two groups: the modular group and the 3D-printed group for a comparative study. Prior to their surgical procedures, all patients underwent a thorough pathological examination and were staged according to the Enneking classification system for tumor categorization. Furthermore, a comprehensive set of evaluations was conducted, encompassing physical examinations, biochemical analyses, and a range of imaging techniques, such as X-ray, 3D-CT, MRI, and SPECT. Thin-layer chest CT scans were specifically carried out to detect any potential lung metastases. To gauge pain levels, Visual Analog Scale (VAS) scores were documented, and the functional outcomes were assessed using the MSTS-93 scale. Following preoperative consultations between patients and medical teams, the patients were divided into two groups: the 3D-printed group (n = 19), and the modular group (n = 13). No statistically significant differences were observed between the two groups with respect to age, gender, BMI values, tumor volume, tumor location, preoperative MSTS and VAS scores, or the duration of follow-up. The baseline characteristics of the patients are summarized in Table 1.

Table 1 Characteristics of patients undergoing hemipelvic replacement surgery

Custom-made and modular hemipelvic prostheses

  • Custom-made hemipelvic endoprostheses: Designed by our clinical team and manufactured by Chunli Co., Ltd. (Tongzhou, Beijing, China), these prostheses are individually tailored to fit each patient’s pelvic defect shape with a bone-mimicking porous structure to facilitate bone ingrowth. Detailed information about the prostheses and specific design procedures can be found in previous report [38].

  • Modular hemipelvic prostheses: Provided by Chunli Co., Ltd. (Tongzhou, Beijing, China). These prostheses comprise four components: Chunli System fixation device (CS fixator), pubic plate, acetabular cup, and polyethylene acetabular liner [43].

Surgical techniques

All surgeries were performed by the same senior orthopedic surgeon. Patients in lateral decubitus position. Kocher-Langenbeck and Smith-Petersen approaches were combined, with optional inguinal extension [14]. Preserving hip joint muscles and their attachment points requires emphasis. For the 3D-printed group, precise osteotomy was assisted by cutting guides and validated with plastic prostheses. The prosthesis was implanted and fixed with multiple screws based on the preoperative plan. After pelvic reconstruction, the polyethylene liner angle within the acetabulum was appropriately adjusted (5°–10°) and fixed with bone cement. For the modular group, osteotomy was performed based on ensuring sufficient tumor margins while considering the actual situation. A suitable CS fixator was selected, and screws were fixed to the remaining sacroiliac bones. The acetabulum position was located using C-arm fluoroscopy, with an abduction angle of 45° and an anteversion angle of 15°–25°. If needed, the polyethylene cup angle was adjusted. In cases of tumor involvement in Zone III of the pelvis, an appropriate length of pubic plate was used for reconstruction. Finally, the installation of femoral head and neck prostheses and the reduction of the hip joint were carried out. The remaining muscles around the hip joint and their attachment points were subsequently sutured onto the prosthesis.

Postoperative management

Within the first week after surgery, patients performed non-weight-bearing early rehabilitation exercises to enhance hip muscle strength and balance. Two rehabilitation exercise methods were used: (1) Active maintenance of the affected limb with 15°–25° hip abduction, 60°–80° hip flexion, and 90° knee flexion. (2) Active knee extension and maintenance with 15°–25° hip abduction, 20°–30° hip flexion, and 30°–45° knee flexion. The postoperative rehabilitation plan is as follows:

  • 1–2 weeks post-surgery: Transitioned to non-weight-bearing standing and hip flexion training.

  • 2–4 weeks post-surgery: Gradually increased weight-bearing training (starting from 10 kg) to match the healthy limb’s force.

  • After 4 weeks: Encouraged hip abduction and extension exercises, assisted walking with aids.

  • First 3 months post-surgery: Used T-shaped pillows and anti-rotational shoes for sleeping.

  • After 3 months: Attempted walking without crutches, followed a one-week leg-crossing and squatting training. Achieved walking without crutches and hip flexion beyond 90°.

Follow-up routine

The follow-up routine includes systematic clinical and radiological evaluations at 1, 2, and 3 months, every 3 months for the first 2 years, and then every 6 months. These evaluations are independently assessed by an unbiased surgeon. Specific indicators assessed include:

  • Surgical indicators: Operation duration and blood loss. The bleeding is calculated by the anesthesiologist and surgical nurse.

  • Function: Lower-limb function is assessed using the Musculoskeletal Tumor Society 93 (MSTS-93) and Harris hip score (HHS) at each follow-up.

  • Complications: This includes monitoring for infection, local recurrence, dislocation, aseptic loosening, endoprosthetic breakage, and delayed wound healing.

  • Radiological outcome: Osteointegration is assessed using Tomosynthesis Shimadzu Metal Artefact Reduction Technology (T-SMART). The implant-host bone interfaces are analyzed to quantify bone integration efficacy in all patients. Furthermore, the accuracy of acetabular component reconstruction in terms of the hip center of rotation within the hemipelvic prosthesis was assessed postoperatively using pelvic X-rays (Anteroposterior view). The acetabular eccentricity (medial–lateral) was defined as the distance between two lines: one drawn vertically from the center of the femoral head to the horizontal line and the other drawn vertically from the pubic symphysis to the same horizontal line. The acetabular eccentricity (superior-inferior) was defined as the vertical distance between two lines drawn perpendicular to the central axis of the hemipelvic prosthesis/pelvis within the acetabulum.

Statistical analysis

Independent-samples Student’s t-test was used for normally distributed data, including operating time, intraoperative blood loss, HHS score, MSTS93 functional score, acetabular lever arm, and acetabular height. For non-normally distributed data, the Mann–Whitney U test was applied. Analysis was performed using SPSS 21.0 (IBM Corp., Armonk, NY). A p-value of less than 0.05 was considered statistically significant.

Biomechanical study

Creation of 3D finite element models

We established 3D finite element models for normal pelvis and prosthetic reconstruction by first selecting CT data from a healthy adult volunteer (height = 170 cm, weight = 71 kg, The pelvic anatomical parameters closely approximate the median values of the corresponding parameters in the clinical study patients). After importing the data into Mimics V20.0 (Materialise Corp., Leuven, Belgium), we constructed a normal pelvic model, distinguishing between cortical and trabecular bone. Next, under the guidance of experienced surgeons, preoperative simulations (bone resection, prosthetic implantation, and screw fixation) were performed using 3D CT/MRI data from actual pelvic tumor patients, resulting in finite element models for prosthetic reconstruction. The 3D-printed prosthetic models followed previously established methods [38], while the 3D finite element models for modular prosthetic reconstruction were based on manufacturers’ specifications (Fig. 1).

Fig. 1
figure 1

Three-Dimensional finite element models of pelvic defects post tumor resection and reconstruction: 3D FE models illustrating typical pelvic defects after acetabular tumor resection and reconstruction using 3D printing or modular hemipelvic prostheses. Key components (ilium, sacrum, prosthesis, and screws) are assembled in the models. The transformation from cortical bone to cancellous bone models is demonstrated through a global offset of "-2.0 mm." Internal cancellous bone regions are marked with asterisks (*)

Material assignment and mesh

In Abaqus 6.17 (Dassault Systèmes, Paris, France), all materials were assumed to be homogeneous, isotropic, and linearly elastic. Material properties, including elastic modulus and Poisson’s ratio, were assigned based on values from previous studies [17, 21, 44] (Table 2). For accurate geometry representation, tetrahedral elements with quadratic shape functions (C3D10) and displacement degrees of freedom were used to discretize the pelvis and implant. After meshing and optimization, the number of nodes and elements for each model is shown in Table 3.

Table 2 Material properties of the bone and implants
Table 3 Mesh division of each three-dimensional finite element model

Loads and constraints

This study simulated physiological pelvic loading by applying 2/3 of the volunteer’s body weight vertically above the sacrum, specifically in alignment with the fifth lumbar vertebra [3]. By applying constraints to various anatomical regions of the pelvis, this study simulated four physiological postures, and constraints were set as follows:

Normal pelvic model:

  • Bipedal stance: Both acetabula were constrained.

  • Unipedal stance: One acetabulum was constrained.

  • Sitting: Both ischial tuberosities were constrained.

Prosthetic-reconstructed pelvic model:

  • Bipedal stance: Both acetabula were constrained.

  • Standing on the affected leg: The acetabulum on the affected side was constrained.

  • Standing on the healthy leg: The acetabulum on the healthy side was constrained.

  • Sitting: Both ischial tuberosities were constrained, or in cases where the ischium was not reconstructed, the lower edge of the acetabulum was constrained. (Fig. 2a, b).

Fig. 2
figure 2

Loads and Constraints: a Simulated pelvic ring stress in three positions: a1 bilateral standing, a2 single-leg standing, and a3 sitting after acetabular tumor resection and reconstruction. b Arrows represent the vertical stress applied to the sacrum, simulating 2/3 of the body weight, depicting different mechanical characteristics of the pelvic ring under different conditions. Green circles (purple circles) indicate the constrained regions at the acetabulum (ischial tuberosity). Yellow lines represent spring elements simulating the interaction between the pelvic region and lower limb muscles (accounting for muscle attachment points). c Muscle load illustration and mesh display in the 3D printing and modular groups: Demonstrating muscle load distribution in three types of pelvic defect reconstructions. Mesh representation shows the 3D printing and modular prosthetic components in place. d FEA results of stress and displacement distribution in a normal pelvic ring

Furthermore, linear spring elements were used to simulate muscle forces with tension-only loading [28]. To create more realistic postoperative 3D finite element models with varying muscle strengths, muscle origin and insertion points were identified based on each patient’s postoperative 3D CT results (Soft tissue window setting) [17]. Muscle reconstruction ratios were calculated as the affected side’s muscle volume percentage compared to the healthy side (Additional file 1: Fig. S1). The 3D finite element model of the pelvis with prosthesis reconstruction was built using normal pelvic muscle loads (Table 4), and the stiffness of linear springs was adjusted to simulate muscle forces (Fig. 2c).

Table 4 Relevant muscles and stiffness values in normal pelvic model

Finite element analysis

The study used finite element analysis to assess stress distribution and displacement in the pelvis under different positions and muscle attachment conditions for (1) The normal pelvic model; (2) The prosthetic-reconstructed pelvic model as a whole; (3) The prosthetic components and screws.

Results

FEM results

Stress and displacement distribution in normal pelvis

The normal pelvic model exhibits a stable, continuous, and evenly distributed stress pattern in different positions, with no significant stress concentration areas observed (Fig. 2d). Specifically, during bipedal stance, stress was symmetrically transmitted through the sacroiliac joints and then transitions along the arcuate lines and greater sciatic notches to the acetabula, peaking at 12.11 MPa. The left unipedal stance showed stress transmission to the left acetabulum, peaking at approximately 20 MPa, while the right unipedal stance exhibited a symmetrically opposite stress pattern. In sitting, stress was symmetrically transmitted through the sacroiliac joints to the ischial tuberosities, with a peak of 13.67 MPa. The overall displacement is symmetric, with the maximum displacement being 0.8433 mm during bipedal stance, 3.434 mm during the left unipedal stance, and 1.052 mm during sitting.

Stress and displacement distribution in pelvic reconstruction with corresponding implants

Overall, the 3D-printed group exhibited a more uniform, continuous, and gentle stress distribution pattern in various bone defect reconstructions, resembling the physiological stress distribution observed in the normal pelvic model under muscle loading in the finite element analysis (FEA) results. Conversely, the modular group displayed a stiffer and discontinuous stress distribution, with noticeable stress concentrations, particularly in the CS fixator and certain screws, at times surpassing the yield stress of titanium alloy, markedly deviating from the FEA results of the normal pelvic model (Figs. 3, 4, 5). Detailed information on the highest stress, displacement, and their respective locations in different regions of the normal pelvis, 3D-printed monolithic, and group-assembly prosthetic-reconstructed pelvic models for tumor-related bone defects can be found in Table 5.

Fig. 3
figure 3

Biomechanical differences in pelvic ring reconstruction after Type I + II resection: Comparing different reconstruction methods using 3D-printed hemipelvic endoprostheses (a) and modular hemipelvic endoprostheses (b) in four common physiological postures, the stress distribution is more uniform and gradual in the 3D-printed prosthetic reconstruction. However, in the modular prosthetic reconstruction, significant stress concentration is evident

Fig. 4
figure 4

Biomechanical differences in pelvic ring reconstruction after Type II + III resection: Comparing different reconstruction methods using 3D-printed hemipelvic prostheses (a) and modular hemipelvic prostheses (b) in four common physiological postures, the 3D-printed prosthetic reconstruction exhibits a more uniform and gradual stress and displacement distribution. Conversely, the modular prosthetic reconstruction shows significant stress concentration at the pubic symphysis region

Fig. 5
figure 5

Biomechanical differences in pelvic ring reconstruction after Type I + II + III resection: Comparing different reconstruction methods using 3D-printed hemipelvic prostheses (a) and modular hemipelvic prostheses (b) in four common physiological postures, the 3D-printed prosthetic reconstruction exhibits a more uniform and natural stress and displacement distribution, closely resembling a normal pelvis. However, the modular hemipelvic prosthesis shows evident stress concentration at the pubic symphysis and CS fixator regions

Table 5 Peak stress and displacement distribution in 3D finite element models of reconstructed pelvises

Clinical outcomes

Demographics

This study involved 32 patients with primary malignant acetabulum tumors (18 males, 14 females). The average follow-up time for all patients was 41.6 ± 10.5 months. 19 patients received 3D-printed custom hemipelvic prostheses, with an average follow-up time of 42.3 ± 12.0 months, while 13 patients received modular prostheses, with an average follow-up time of 40.5 ± 8.0 months. Osteosarcoma and Ewing sarcoma patients had neoadjuvant chemotherapy, followed by post-surgery chemotherapy after incision healing. Radiotherapy was avoided to support wound healing.

Surgical outcomes

The 3D group exhibited significantly higher blood loss [2121.1 ± 686.8 ml (range, 1000.0–3500.0 ml) vs 1600.0 ± 505.0 ml (range, 1000.0–2500.0 ml), p < 0.05] and longer surgical time [426.2 ± 67.0 min (range, 300.0–592.0 min) vs 301.7 ± 48.6 min (range, 243.0–410.0 min), p < 0.05] compared to the modular group.

Functional assessment

The 3D group exhibited significantly higher mean Musculoskeletal Tumor Society (MSTS) scores compared to the modular group [24.3 ± 1.8 (range, 21.0–27.0) vs 21.8 ± 2.0 (range, 15.0–26.0), p < 0.05] and mean Harris Hip Score (HHS) [79.8 ± 5.2 (range, 72.0–85.0) vs 75.3 ± 3.5 (range, 70.0–82.0), p < 0.05] during the last follow-up, indicating improved hip joint function in the 3D group (Additional file 2: Fig. S2).

Complications

Intraoperative complications were absent. Postoperative complications occurred in 11 patients: 3 cases (26.3%) in the 3D group and 6 cases (46.2%) in the modular group. Poor wound healing was the most common postoperative complication among all patients.

In the 3D group:

  • One had an upper sacroiliac joint screw fracture (5.3%) one year postoperatively but remained asymptomatic without affecting function. (Fig. 3Sa).

  • One exhibited distal-bone interface loosening (5.3%) and screw fractures 3 years postoperatively, which improved with revision surgery (Fig. 3Sb).

  • Two cases of poor wound healing (10.5%) resolved with intensive wound dressing.

  • One patient (5.3%) experienced postoperative hip dislocation two days after surgery, which was promptly addressed through closed reduction. Stability was maintained using a T-shaped pillow and anti-rotation shoes (Additional file 3: Fig. S3c).

In the modular group:

  • Two cases had poor wound healing (15.4%). One case healed with enhanced wound dressing, while another required debridements and negative pressure Vacuum Assisted Closure (VAC) system therapy to achieve healing.

  • A deep prosthesis infection impacted one case (7.7%) and required extensive treatment one month after the surgery.

  • Three cases (23.1%) encountered screw loosening, with one being managed conservatively two years after surgery, and the other two undergoing successful 3D-printed hemipelvic revision surgery, one at 1 year post-surgery and the other at 2.5 years post-surgery. (Additional file 3: Fig. S3d).

Radiological outcome

In the 3D printing group, all patients underwent precise osteotomy, accurate prosthesis implantation, and planned screw fixation based on preoperative simulated surgery. The acetabular eccentricity (medial–lateral) on the healthy side measured 94.7 ± 4.3 mm (range, 91.5.0–97.1 mm), and on the affected side, it was 93.7 ± 5.3 mm (range, 91.2.0–97.5 mm) (Fig. 6). Similarly, the acetabular eccentricity (superior-inferior) on the healthy side was 54.3 ± 4.7 mm (range, 50.8–57.2 mm), and on the affected side, it was 55.6 ± 4.8 mm (range, 52.2.0–58.3 mm). The comparison between healthy and affected sides showed no statistically significant differences (p > 0.05), indicating accurate hip joint reconstruction. During the final follow-up, 97.1% of patients (66/68) exhibited successful osseointegration of their implants (Fig. 7), with only one case experiencing distal prosthesis loosening (Additional file 3: Fig. S3b).

Fig. 6
figure 6

Illustrates postoperative multi-directional pelvic X-ray assessment, demonstrating precise execution of the following: Precise osteotomy, prosthesis placement, and multi-level screw fixation. Anteroposterior views of the pelvis reveal an accurate hip joint rotation center (indicated by red circle) and precise femoral head eccentricity in horizontal (L1 = L2) and vertical (L3 = L4) directions. The inlet view displays accurately positioned sacroiliac joint screws in the anterior–posterior direction, carefully avoiding the posterior vertebral canal (highlighted by the yellow circle, representing the safe vertebral region). Complete small pelvic ring reconstruction is indicated by the red dashed circle. The outlet view shows accurate sacroiliac joint screw placement, avoiding S1/S2 sacral foramina (indicated by the red dashed box). Ischium and pubic bone screws are precisely positioned for enhanced stability and support

Fig. 7
figure 7

Successful prosthesis-bone integration using 3D-printed hemipelvic endoprosthetic reconstruction: In the last follow-up, T-SMART imaging confirmed effective osseointegration at the bone-implant interface

In the modular group, there was no effective bone integration at the prosthesis-bone contact. The healthy side acetabular eccentricity (medial–lateral) was 91.5 ± 5.8 (range, 89.0–94.0 mm), and on the affected side, it was 86.2 ± 9.0 mm (range, 64.0–99.0 mm). The healthy side acetabular eccentricity (superior-inferior) was 54.5 ± 5.8 mm (range, 50.2–57.2 mm), and on the affected side, it was 60.6 ± 6.0 mm (range, 55.3–65.0 mm). When comparing the healthy and affected sides, the differences in both acetabular eccentricity (medial–lateral) and acetabular eccentricity (superior-inferior) are statistically significant (p < 0.01). Moreover, postoperative X-rays revealed complications in some cases: 1 case showed inward bending of the pubic bone plate (Additional file 3: Fig. S3e), 1 case exhibited inward displacement of the acetabular cup (Additional file 3: Fig. S3f), and 2 cases had loose screws in the CS fixator area.

Discussion

In recent years, pelvic prostheses have gradually emerged as the primary method for reconstructing the pelvic region following the removal of malignant or invasive tumors due to their initial stability, high acceptance of their appearance, and relatively swift functional recovery. Modular and 3D-printed integrated prostheses, in particular, have gained significant attention. One significant difference between these two reconstruction methods lies in the fact that modular hemipelvic prostheses often do not encompass a complete reconstruction of the pelvic ring, while 3D-printed integrated prostheses have the capability to fully restore the anatomical shape of the pelvic ring during the recovery period. This dissimilarity results in markedly different stress distribution patterns within the pelvic ring, which theoretically could have varying effects on the mid-term stability and lifespan of the prostheses. Understanding the distribution of stress within the pelvic ring and the principles of its transmission under normal physiological conditions is of paramount importance. It provides essential guidance for surgeons tasked with reconstructing the pelvic ring after the resection of periacetabular tumors.

Based on the finite element analysis (FEA) results of our study, the normal pelvis, under physiological stress, follows a stress transmission pathway from the spine to the sacroiliac joint, the pelvic ring, and ultimately to the hip joint. It’s worth noting that the posterior pelvic ring plays a crucial role in carrying the load. However, it’s important not to underestimate the contribution of the anterior pelvic ring in enhancing stability and preventing lateral spreading. These findings are consistent with previous biomechanical analyses of the pelvis, such as those by Tile and others [31, 32], who suggested that the anterior and posterior structures contribute 40% and 60%, respectively, to the overall stability of the pelvic girdle. According to their “suspension bridge concept”, the posterior pelvic ring primarily bears the main load transfer, while the anterior ring acts like a pull bar (strut), similar to a suspension bridge, enhancing stability and preventing lateral spreading [33]. Furthermore, regarding the stress distribution in pelvic ring prosthesis reconstruction, for all types of resections (type I + II/II + III/I + II + III resections), the 3D-printed group consistently exhibits stress distribution patterns closer to those of a normal pelvis. Stress peaks and displacement amplitudes are consistently lower in the 3D-printed group compared to the modular group. These findings suggest that this reconstruction method carries a lower risk of mechanical failure, such as screw fracture or loosening, and provides improved post-reconstruction stability [13, 16].

Modular prostheses, despite fully reconstructing the pelvic ring, have a major drawback of relatively high postoperative loosening rates [36]. This issue can, in significant part, be attributed to the unnatural stress transmission pattern that emerges after hemipelvic reconstruction using modular prostheses. The stress distribution, exerting both pushing and pulling forces on the implant from two sides, is likely one of the key factors contributing to the heightened risk of loosening. Moreover, during reconstruction, achieving a satisfactory acetabular location and orientation often requires sacrificing the fit at the anchor part [10]. Our FEA results showed that this poor prosthesis-pelvic match leads to multiple stress concentrations in modular hemipelvic prostheses. For example, the pubic plate part experiences significant stress concentration, coinciding with the observed bending deformity of the pubic plate in one patient during clinical follow-up. Another significant factor contributing to the loosening of modular endoprostheses may be incomplete osseointegration. In this study, the high-stress concentration in the CS fixation area of the modular prosthesis, combined with the lack of a porous surface structure, hindered osteointegration, making this area a common site for loosening. Even with the incorporation of porous structures onto the surface of modular implants to augment their integration, the fundamental design incongruity with the stress propagation dynamics of the pelvic ring persists [9, 23]. Consequently, the precise manipulation of the hip joint's rotational center during surgical intervention remains challenging, which in turn may engender subsequent displacement under the influence of external forces during the extended postoperative period.

On the contrary, the 3D printing group achieved accurate bone resection with custom cutting guides, ensuring precise matching between the custom prosthesis and bone defect. FEA results demonstrated that the stress transmission in the reconstructed pelvic ring of this group closely resembled that of a normal pelvis. The stable initial reconstruction, along with the porous surface structure facilitating bone integration, further enhanced its mid-term stability. Consequently, the 3D printing group achieved satisfactory functional outcomes (average MSTS score of 76.7%), surpassing other investigations [5, 11, 22, 23]. However, achieving better postoperative functionality with 3D-printed hemipelvic prostheses comes with some trade-offs, such as longer design and production time, and higher costs. These may gradually decrease with wider adoption and market availability of the technology. The main concern lies in the ability to achieve precise preoperative planning and implantation. This demands utmost accuracy throughout the entire process, from preoperative cutting guides and prosthesis design to intraoperative bone resection, prosthesis implantation, and screw fixation. When contrasted with the flexible, convenient, and rapid implantation process of modular prostheses, it becomes understandable why the follow-up results reveal a somewhat longer operation time and slightly increased blood loss during surgery in the 3D printing group. Nevertheless, it's reassuring that these differences remain within acceptable ranges [23, 35, 45].

Conclusion

Modular and 3D-printed hemipelvic endoprostheses are commonly used for limb-salvage reconstruction after periacetabular tumor resection. Modular ones offer easy implantation and flexibility but have a higher loosening rate due to poor prosthesis-host bone matching. 3D-printed hemipelvic endoprostheses, customized for individuals, precisely restore pelvic ring anatomy, improving fit and alignment. Compared to modular prostheses, it offers a closer approximation of normal pelvic physiological stress transmission. Mid-term clinical follow-ups have shown improved functional outcomes and bone integration. However, this technique demands high expertise, time, and effort for prosthesis fabrication. Surgical implantation is more challenging and may lead to longer operating times and increased intraoperative blood loss as potential drawbacks. Surgeons must weigh postoperative functional outcomes, complexity, and patient condition to decide the best treatment.

Availability of data and materials

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

References

  1. Abudu A, Grimer RJ, Cannon SR, Carter SR, Sneath RS. Reconstruction of the hemipelvis after the excision of malignant tumours. Complications and functional outcome of prostheses. J Bone Joint Surg Br. 1997;79(5):773–9.

    Article  CAS  PubMed  Google Scholar 

  2. Ayvaz M, Bekmez S, Mermerkaya MU, Caglar O, Acaroglu E, Tokgozoglu AM. Long-term results of reconstruction with pelvic allografts after wide resection of pelvic sarcomas. ScientificWorldJournal. 2014;2014:605019.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Bergmann G, Graichen F, Rohlmann A, Linke H. Hip joint forces during load carrying. Clin Orthop Relat Res. 1997;335:190–201.

    Article  Google Scholar 

  4. Biau DJ, Thévenin F, Dumaine V, Babinet A, Tomeno B, Anract P. Ipsilateral femoral autograft reconstruction after resection of a pelvic tumor. J Bone Joint Surg Am. 2009;91(1):142–51.

    Article  PubMed  Google Scholar 

  5. Bus MP, Szafranski A, Sellevold S, Goryn T, Jutte PC, Bramer JA, Fiocco M, Streitbürger A, Kotrych D, van de Sande MA, et al. LUMiC(®) endoprosthetic reconstruction after periacetabular tumor resection: short-term results. Clin Orthop Relat Res. 2017;475(3):686–95.

    Article  PubMed  Google Scholar 

  6. Campanacci M, Capanna R. Pelvic resections: The Rizzoli Institute experience. Orthop Clin North Am. 1991;22(1):65–86.

    Article  CAS  PubMed  Google Scholar 

  7. Danışman M, Mermerkaya MU, Bekmez Ş, Ayvaz M, Atilla B, Tokgözoğlu AM. Reconstruction of periacetabular tumours with saddle prosthesis or custom-made prosthesis, functional results and complications. Hip Int. 2016;26(2):e14-18.

    Article  PubMed  Google Scholar 

  8. Enneking W. Resection and reconstruction for primary neoplasms involving the innominate bone. J Bone and Joint Surg. 1978;60:731–46.

    Article  CAS  Google Scholar 

  9. Fang C, Cai H, Kuong E, Chui E, Siu YC, Ji T, Drstvenšek I. Surgical applications of three-dimensional printing in the pelvis and acetabulum: from models and tools to implants. Unfallchirurg. 2019;122(4):278–85.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Gradinger R, Rechl H, Hipp E. Pelvic osteosarcoma. Resection, reconstruction, local control, and survival statistics. Clin Orthop Relat Res. 1991;270:149–58.

    Article  Google Scholar 

  11. Guo W, Li D, Tang X, Yang Y, Ji T. Reconstruction with modular hemipelvic prostheses for periacetabular tumor. Clin Orthop Relat Res. 2007;461:180–8.

    Article  PubMed  Google Scholar 

  12. Guo Z, Li J, Pei GX, Li XD, Wang Z. Pelvic reconstruction with a combined hemipelvic prostheses after resection of primary malignant tumor. Surg Oncol. 2010;19(2):95–105.

    Article  PubMed  Google Scholar 

  13. Hu X, Lu M, He X, Li L, Lin J, Zhou Y, Luo Y, Min L, Tu C. Hip reconstruction using a customized intercalary prosthesis with the rhino horn-designed uncemented stem for ultrashort proximal femur segments following tumor resection: a combined biomechanical and clinical study. BMC Musculoskelet Disord. 2022;23(1):852.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Hu X, Lu M, Wang J, Li L, Min L, Tu C. Combined and modified gibson and ilioinguinal approaches in type II + III internal hemipelvectomy for periacetabular tumors. Front Oncol. 2022;12:934812.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Hu X, Lu M, Zhang Y, Li Z, Wang J, Wang Y, Xing Z, Yang X, Tu C, Min L. Pelvic-girdle reconstruction with three-dimensional-printed endoprostheses after limb-salvage surgery for pelvic sarcomas: current landscape. Br J Surg. 2023;110:1712–22.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Hu X, Lu M, Zhang Y, Wang Y, Min L, Tu C. A biomechanical comparison between cement packing combined with extra fixation and three-dimensional printed strut-type prosthetic reconstruction for giant cell tumor of bone in distal femur. J Orthop Surg Res. 2022;17(1):151.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Hu X, Lu M, Zhang Y, Wang Y, Min L, Tu C. A biomechanical comparison between cement packing combined with extra fixation and three-dimensional printed strut-type prosthetic reconstruction for giant cell tumor of bone in distal femur. J Orthop Surg Res. 2022;17(1):1–13.

    Article  CAS  Google Scholar 

  18. Issa SP, Biau D, Babinet A, Dumaine V, Le Hanneur M, Anract P. Pelvic reconstructions following peri-acetabular bone tumour resections using a cementless ice-cream cone prosthesis with dual mobility cup. Int Orthop. 2018;42(8):1987–97.

    Article  PubMed  Google Scholar 

  19. Jaiswal PK, Aston WJ, Grimer RJ, Abudu A, Carter S, Blunn G, Briggs TW, Cannon S. Peri-acetabular resection and endoprosthetic reconstruction for tumours of the acetabulum. J Bone Joint Surg Br. 2008;90(9):1222–7.

    Article  CAS  PubMed  Google Scholar 

  20. Jansen JA, van de Sande MA, Dijkstra PD. Poor long-term clinical results of saddle prosthesis after resection of periacetabular tumors. Clin Orthop Relat Res. 2013;471(1):324–31.

    Article  CAS  PubMed  Google Scholar 

  21. Ji T, Guo W, Tang XD, Yang Y. Reconstruction of type II+III pelvic resection with a modular hemipelvic endoprosthesis: a finite element analysis study. Orthop Surg. 2010;2(4):272–7.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Ji T, Guo W, Yang RL, Tang XD, Wang YF. Modular hemipelvic endoprosthesis reconstruction–experience in 100 patients with mid-term follow-up results. Eur J Surg Oncol. 2013;39(1):53–60.

    Article  CAS  PubMed  Google Scholar 

  23. Ji T, Yang Y, Tang X, Liang H, Yan T, Yang R, Guo W. 3D-Printed modular hemipelvic endoprosthetic reconstruction following periacetabular tumor resection: early results of 80 consecutive cases. J Bone Joint Surg Am. 2020;102(17):1530–41.

    Article  PubMed  Google Scholar 

  24. Liang H, Ji T, Zhang Y, Wang Y, Guo W. Reconstruction with 3D-printed pelvic endoprostheses after resection of a pelvic tumour. Bone Joint J. 2017;99(2):267–75.

    Article  PubMed  Google Scholar 

  25. Mathew G, Agha R, Albrecht J, Goel P, Mukherjee I, Pai P, D’Cruz AK, Nixon IJ, Roberto K, Enam SA, et al. STROCSS 2021: Strengthening the reporting of cohort, cross-sectional and case-control studies in surgery. Int J Surg. 2021;96:106165.

    Article  PubMed  Google Scholar 

  26. Ogura K, Susa M, Morioka H, Matsumine A, Ishii T, Hamada K, Ueda T, Kawai A. Reconstruction using a constrained-type hip tumor prosthesis after resection of malignant periacetabular tumors: a study by the Japanese Musculoskeletal Oncology Group (JMOG). J Surg Oncol. 2018;117(7):1455–63.

    Article  PubMed  Google Scholar 

  27. Ozaki T, Hoffmann C, Hillmann A, Gosheger G, Lindner N, Winkelmann W. Implantation of hemipelvic prosthesis after resection of sarcoma. Clin Orthop Relat Res. 2002;396:197–205.

    Article  Google Scholar 

  28. Phillips AT, Pankaj P, Howie CR, Usmani AS, Simpson AH. Finite element modelling of the pelvis: inclusion of muscular and ligamentous boundary conditions. Med Eng Phys. 2007;29(7):739–48.

    Article  CAS  PubMed  Google Scholar 

  29. Puri A, Gulia A, Jambhekar NA, Laskar S. Results of surgical resection in pelvic Ewing’s sarcoma. J Surg Oncol. 2012;106(4):417–22.

    Article  PubMed  Google Scholar 

  30. Renard AJ, Veth RP, Schreuder HW, Pruszczynski M, Keller A, van Hoesel Q, Bökkerink JP. The saddle prosthesis in pelvic primary and secondary musculoskeletal tumors: functional results at several postoperative intervals. Arch Orthop Trauma Surg. 2000;120(3–4):188–94.

    Article  CAS  PubMed  Google Scholar 

  31. Shi D, Wang F, Wang D, Li X, Wang Q. 3-D finite element analysis of the influence of synovial condition in sacroiliac joint on the load transmission in human pelvic system. Med Eng Phys. 2014;36(6):745–53.

    Article  PubMed  Google Scholar 

  32. Tile M. Pelvic ring fractures: should they be fixed? J Bone Joint Surg Br. 1988;70(1):1–12.

    Article  CAS  PubMed  Google Scholar 

  33. Tile M, Helfet DL, Kellam JF, Vrahas M. Fractures of the pelvis and acetabulum. Baltimore: Williams & Wilkins; 1995.

    Google Scholar 

  34. Wafa H, Grimer RJ, Jeys L, Abudu AT, Carter SR, Tillman RM. The use of extracorporeally irradiated autografts in pelvic reconstruction following tumour resection. Bone Joint J. 2014;96(10):1404–10.

    Article  PubMed  Google Scholar 

  35. Wang B, Hao Y, Pu F, Jiang W, Shao Z. Computer-aided designed, three dimensional-printed hemipelvic prosthesis for peri-acetabular malignant bone tumour. Int Orthop. 2018;42(3):687–94.

    Article  PubMed  Google Scholar 

  36. Wang B, Sun P, Yao H, Tu J, Xie X, Ouyang J, Shen J. Modular hemipelvic endoprosthesis with a sacral hook: a finite element study. J Orthop Surg Res. 2019;14(1):309.

    Article  PubMed  PubMed Central  Google Scholar 

  37. Wang J, Min L, Lu M, Zhang Y, Lin J, Luo Y, Zhou Y, Tu C. Three-dimensional-printed custom-made hemipelvic endoprosthesis for the revision of the aseptic loosening and fracture of modular hemipelvic endoprosthesis: a pilot study. BMC Surg. 2021;21(1):262.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Wang J, Min L, Lu M, Zhang Y, Wang Y, Luo Y, Zhou Y, Duan H, Tu C. What are the complications of three-dimensionally printed, custom-made, integrative hemipelvic endoprostheses in patients with primary malignancies involving the acetabulum, and what is the function of these patients? Clin Orthop Relat Res. 2020;478(11):2487–501.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Windhager R, Karner J, Kutschera HP, Polterauer P, Salzer-Kuntschik M, Kotz R. Limb salvage in periacetabular sarcomas: review of 21 consecutive cases. Clin Orthop Relat Res. 1996;331:265–76.

    Article  Google Scholar 

  40. Wu J, Xie K, Luo D, Wang L, Wu W, Yan M, Ai S, Dai K, Hao Y. Three-dimensional printing-based personalized limb salvage and reconstruction treatment of pelvic tumors. J Surg Oncol. 2021;124(3):420–30.

    Article  PubMed  Google Scholar 

  41. Xu H, Li Y, Zhang Q, Hao L, Yu F, Niu X. Does adding femoral lengthening at the time of rotation hip transposition after periacetabular tumor resection allow for restoration of limb length and function? Interim results of a modified hip transposition procedure. Clin Orthop Relat Res. 2021;479(7):1521–30.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Zhang Y, Min L, Lu M, Wang J, Wang Y, Luo Y, Zhou Y, Duan H, Tu C. Three-dimensional-printed customized prosthesis for pubic defect: clinical outcomes in 5 cases at a mean follow-up of 24 months. BMC Musculoskelet Disord. 2021;22(1):405.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Zhou Y, Duan H, Liu Y, Min L, Kong Q, Tu C. Outcome after pelvic sarcoma resection and reconstruction with a modular hemipelvic prostheses. Int Orthop. 2011;35(12):1839–46.

    Article  PubMed  PubMed Central  Google Scholar 

  44. Zhou Y, Min L, Liu Y, Shi R, Zhang W, Zhang H, Duan H, Tu C. Finite element analysis of the pelvis after modular hemipelvic endoprosthesis reconstruction. Int Orthop. 2013;37(4):653–8.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Zhu D, Wang L, Fu J, Guo Z, Wang Z, Fan H. Comparison of customized 3D-printed prosthesis and screw-rod-cage system reconstruction following resection of periacetabular tumors. Front Oncol. 2022;12:953266.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

We thank Home for Researchers editorial team (www.home-for-researchers.com) for language editing service. We thank Engineer Fengda Guo from Beijing Chunlizhengda Medical Instruments Co., Ltd (Tongzhou, Beijing, China) for his assistance in prosthesis design.

Funding

The institution of one or more of the authors has received, during the study period, funding from Sichuan Science and Technology Agency (2022NSFSC0845) and National Key Research and Development Program of China (2023YFB3813002, 2023YFE0126900).

Author information

Authors and Affiliations

Authors

Contributions

XH: Conceptualization, data curation, methodology, visualization, writing—original draft, writing—review and editing), YW: Methodology, writing—review and editing, ML: Data curation, visualization, YZ: Data curation, methodology, YL: Data curation, visualization, XY: Conceptualization, data curation, methodology, visualization, writing—original draft, writing—review and editing, supervision, and funding acquisition, CT: Conceptualization, data curation, methodology, writing—review and editing, supervision, and LM: Conceptualization, data curation, methodology, visualization, writing—original draft, writing—review and editing, supervision, funding acquisition).

Corresponding authors

Correspondence to Xiao Yang, Chongqi Tu or Li Min.

Ethics declarations

Ethics approval and consent to participate

This single-center retrospective study was performed in accordance with the 1964 Helsinki Declaration and was authorized by the Ethics Committee of our hospital. Written informed consent was obtained from adult participants or parents of minors (below 16 years of age).

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests or personal relationships that could have appeared to influence the work reported in this paper.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. Fig. S1. Muscle Reconstruction Proportion Assessment Method: During surgery procedures, the extent of muscle resections varies among different muscle groups due to tumor involvement in specific regions. Postoperatively, the 3D CT measurement of the pelvic region includes the complete origin and insertion points of major hip joint functional muscle groups (adductor, abductor, and flexor muscles), represented by the longest dimensions (length, width, height) of each muscle group in the pelvic 3D CT. The product of these dimensions reflects the muscle content, and the ratio of affected side muscle volume to the healthy side is calculated, defining the muscle reconstruction rate. Based on the muscle reconstruction parameters, the stiffness of different regions in the pelvis with bone defects and different prosthetic reconstructions is proportionally adjusted to approximate the model to the real physiological state after hemipelvic prosthesis reconstruction.

Additional file 2

. Fig. S2 Functional Follow-up Photographs: The supplement figure illustrates functional follow-up photographs of patients who underwent pelvic ring reconstruction using 3D-printed and modular hemipelvic prostheses following tumor resection. The photographs depict the hip flexion function at 42 months post-surgery for three representative patients. In each patient's image, the left photograph captures the affected side during hip flexion in a standing position, while the right photograph portrays the healthy side during hip flexion in a standing position. These functional follow-up images provide valuable insights into the postoperative outcomes and the effectiveness of the two different hemipelvic prosthesis reconstruction techniques in restoring hip joint functionality after tumor resection.

Additional file 3

. Fig. S3 Typical postoperative complications in 3D-printed prosthetic hip reconstruction surgery: a Screw Fracture: Day 2 Postoperative X-ray (a1). One Year Postoperative X-ray (a2) displays a screw fracture (a2) at the uppermost part of the sacroiliac joint, evident by the red mark. Notably, the patient remained asymptomatic, and conservative observation was chosen. b Aseptic loosening: Postoperative 1-year Pelvic X-ray (b1) reveals loosening and fracture of the ischial screw. Postoperative 2-year X-ray (b2) shows loosening at the prosthesis-bone interface and multiple screw failures. c Hip Dislocation: Three Days Postoperative X-ray (c1) reveals hip dislocation. Successful closed reduction under general anesthesia was performed (c2). d Design and Application of 3D-Printed Hemipelvic Endoprosthesis for Revision of Aseptic Loosening: Preoperative Simulation d1, d2 Depicted endoprosthesis migration and fractured screws. Illustrations d3, d4 Demonstrated design of 3D-printed custom hemipelvic endoprosthesis and screw fixation. Pre-revision Radiographs d5, d6 Displayed aseptic loosening, screw fracture, and endoprosthesis migration. Post-implantation Radiographs d7, d8 Revealed successful reconstruction with custom hemipelvic endoprosthesis. e Three Years Post-Surgery: Pelvic X-ray showed suboptimal integration at modular hemipelvic endoprosthesis interface and inward acetabular cup movement. f Five Years Post-Surgery: Pelvic X-ray indicated inadequate integration at modular hemipelvic endoprosthesis interface, with pubic plate deformation and bending. (Reprinted with permission from Ref [68] ©2021 BMC Surgery).

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hu, X., Wen, Y., Lu, M. et al. Biomechanical and clinical outcomes of 3D-printed versus modular hemipelvic prostheses for limb-salvage reconstruction following periacetabular tumor resection: a mid-term retrospective cohort study. J Orthop Surg Res 19, 258 (2024). https://doi.org/10.1186/s13018-024-04697-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-024-04697-w

Keywords