Long-term outcomes of one single-design varus valgus constrained versus one single-design rotating hinge in revision knee arthroplasty after over 10-year follow-up

Abstract

Background

The appropriate degree of constraint in knee prosthetic revision is unknown, necessitating the use of the lowest possible constraint. This study aimed to compare the long-term clinical and survival results of revision with rotation hinge knee (RHK) VS constrained condylar constrained knee (CCK) implants.

Methods

Overall, 117 revision case were prospectively reviewed and dividing into two groups based on the degree of constraint used, using only one prosthetic model in each group (61 CCK vs 56 RHK). All implants were evaluated for a minimum of 10 years. Survival of both implants at the end of follow-up, free from revision for any cause, aseptic loosening, and septic cause was compared.

Results

Better results were seen with use of the RHK in joint ranges of (p = 0.023), KSCS (p = 0.015), KSFS (p = 0.043), and KOOS (p = 0.031). About 22.2% of the cases required repeat surgery (11.7% RHK vs 29.6% CCK, p = 0.023). Constrained condylar implants had a significantly lower survival rates than rotating hinge implants (p = 0.005), due to a higher aseptic loosening rate (p = 0.031).

Conclusion

Using a specific RHK design with less rotational constraint has better clinical and survival outcomes than implants with greater rotational constraint, such as one specific CCK.

Background

The increasing number of patients undergoing primary total knee arthroplasty (TKA) has been accompanied by a similar increase in the number of revision knee arthroplasty procedures [1], despite advances in instrumentation [2] and the more widespread use of computer assisted surgery [3], which have optimised length of stay and rehabilitation following primary TKA [4]. Aseptic loosening, infection, and instability are the three most common reasons for revision arthroplasty [5].

A minimally constrained conventional prosthesis can be used during revision knee arthroplasty in most situations [6,7,8]. However, the presence of severe bone loss and instability may necessitate the use of a more constrained prosthesis, such as condylar constrained knee (CCK) or a rotating hinged prosthesis (RHK). Adequate implant selection is indispensable to restore the function of the knee joint; however, inadequate constraint will lead to instability failure (third most common cause of revision), but excessive constraint may increase the risk of aseptic loosening (first cause of revision) [9]. This has popularized the idea of using the least constraint necessary, which is not always easy to determine [1].

The definition of constraint is determined by the degree of freedom an implant allows [10]. The greater constraint of the RHK over the CCK implant is classically accepted [1] due to the lower number of movements allowed by them. However, due to the existence of an internal/external rotation movement associated with each flexion–extension movement of the knee joint [11,12,13] and because most CCK implants almost completely restrict this rotational movement (due to the high congruence and trapezoidal shape of the post that limits this rotational movement), more surgeons tend to use hinged implants instead of constrained condylar implants, as they question the lesser constraint of these constrained condylar implants [14,15,16].

Our objectives are: (1) To compare the survival of two highly constrained implants in a group of patients with long-term follow-up (minimum 10 years of follow-up), (2) to assess and compare the differences, if any, in long-term clinical outcomes between these two types of implants, and (3) to determine if there is a higher rate of aseptic loosening with the use of the more constrained hinged implant. To the best of our knowledge, this is the first article that compares the clinical and functional outcomes of two highly constrained implants, a constrained condylar and a rotating hinge implant (CCK and RHK), with different degrees of constraint in over a 10 years of follow-up period.

Methods

All patients gave their informed consent before being included in this study. This was retroactive study that reviewed a prospectively collected database and was performed in accordance with the principles of the 1964 Declaration of Helsinki as revised in 2013 and was approved by the research ethics committee of our centre.

Between January 2004 and December 2009, 245 consecutive revision total knee arthroplasties were performed in 231 patients by 4 senior training orthopaedic surgeons in a single institution. Patients undergoing surgery after tumour surgery, periprosthetic fracture, or with prior alterations of the extensor apparatus were excluded. Two groups were defined based on the type of implant used. In Group 1, a constrained condylar implant was used (NexGen LCCK, Zimmer Biomet, Warsaw, IN, USA), while in Group 2, an intracondylar rotating hinge implant (Endomodel, Waldemar Link, Hamburg, Germany) was used. All surgeons had previous experience and implanted both implants. Patients that received an implant other than the two implants in this study (even if it was a CCK or rotating hinge implant) or patients using a LCCK implant with another insert other than the CCK (posterior stabilized) were also excluded.

A total of 131 knees were included, of which 117 completed a minimum follow-up of 10 years. The indication for the type of implant was at the discretion of the primary surgeon based on the bone defect, ligament status, and their own experience. In cases with absence of collateral ligament, difficulties to equalize gaps or posterior capsule dysfunction a RHK implant was selected. A total of 61 knees were operated with the CCK implant and 56 with the rotating hinge implant (Fig. 1). If severe ligament instability was observed, a rotating hinge implant was selected.

Fig. 1
figure 1

Flow chart of the patients included in the study. TKR: Total knee arthroplasty; PS: posterostabilized; CCK: constrained condylar knee; RHK: rotating hinge knee

Demographic data were collected for each patient, including age at surgery, sex, reason for revision, prior implant, number of comorbidities expressed as Charlson index, and degree of bone defect according to the Anderson Orthopaedic Research Institute (AORI) Classification[17]. Standard follow-up of each patient includes clinical and radiographic reviews at 3 months, 6 months, one year, and every 2 years thereafter. Any complication or reoperation was recorded at any time during the follow-up. Clinical evaluation was performed using the Knee Society Clinical Score (KSCS), the Knee Society Functional Score (KSFC), and range of motion (ROM). Radiographic evaluation was performed by an orthopaedic surgeon unrelated to the surgery, including under load anteroposterior views, lateral views, and Merchant skyline views in all patients. The presence or location of radiolucence lines was determined according to the modification of the Knee Society TKA radiographic evaluation system for long-stemmed revision prostheses [18]. All data were retrospectively collected from our database.

All procedures were carried out following standard revision TKA procedures. In both cases, conventional stem lengths were used (120 mm in the Endomodel group and 145 mm in the LCCK group). In both groups, fixation of both components was cemented, differentiating pandiaphyseal cementation in patients with rotary hinges, compared to those with metaphyseal cementation using press-fit stems in the group of patients with CCK implants. In AORI III defect trabecular metal metaphyseal cones were used when primary stability was not obtained.

Statistical analysis

Statistical analysis was performed using SPSS version 24.0 for MacOs (IBM Corp., Armonk, New York, USA). All data were checked for normality using the Shapiro–Wilk W test. Patient demographics were described using medians and standard deviation. The t test was used to compare scores for the normal distribution between the preoperative and postoperative data. Chi-square test and Fisher exact test were used to compare qualitative variables. Statistical significance was set at p < 0.05. Prosthesis survival was analysed using the Kaplan–Meier method. Survival analysis was performed using revision due to any reason, aseptic loosening, or septic revision as endpoints, and 95% confidence intervals were calculated.

Results

The mean follow-up duration was 13 ± 1.475 years. No significant differences were seen in age (p = 0.6), sex (p = 0.82), body mass index (p = 0.54), preoperative comorbidities measured with the Charlson index (p = 0.75), and ASA (p = 0.81) between both groups. However, patients in the hinged group had poorer preoperative range of motion (p = 0.001), greater preoperative varus deformity (p = 0.001), lower preoperative KSS (p = 0.035), and lower preoperative KOOS (p = 0.031). The time between the primary implant and revision was 64.8 ± 38.5 months, and this was lower among patients in the hinged group (61.3 ± 36.9 months vs 67.1 ± 43.6 months, p = 0.021), though in this group the percentage of patients with previous revision implants was significantly higher (25% vs 8.1%, p = 0.037). The main cause of revision was aseptic loosening (53%), followed by periprosthetic infection (30.7%), with a greater number of septic revisions in the hinged group (39.3%) than in the CCK implant group (22.9%), though this difference was not significant (p = 0.11). The demographic characteristics of the patients are summarized in Table 1.

Table 1 Patient’s demographic

The mean surgery time was 106.4 ± 27.5 min, and this was significantly lower in patients implanted with a rotating hinge (96 min vs 115 min, p = 0.02) (Table 2). No difference was observed in the number of cones used in both group (8 patients; 7 tibial and 6 femoral in CCK group, vs 6 patients; 6 tibial and 3 femoral in RHK group. p = 0.24). At the end of the follow-up period, 37.6% of patients experienced some type of complication. These are summarized in Table 3. Aseptic loosening (12.8%), periprosthetic infection (5.1%) and periprosthetic fractures (5.1%) were the most common complications—the number of complications was significantly higher in the CCK implant group (p = 0.001) as shown in Table 3.

Table 2 Comparison of the results of both groups at final follow-up
Table 3 Comparison of complication between patients with constrained condylar knee prostheses and those with rotating hinge knee prostheses

A significant difference was seen between preoperative and postoperative values in the range of motion, KSS, KSFS, and KOOS. The mean range of motion gain was 9.2° (p = 0.001). In patients in whom a hinged implant was used, the joint range of motion achieved (99.1 ± 25.9° vs 93.5° ± 15, p = 0.023) as well as the gain in range of motion (14.9° vs 6.8°, p = 0.001) was significantly higher than in patients with constrained condylar implants (Table 2). The mean gain in the different clinical scores was 42 (KSS), 25.6 (KSFS), and 27 points (KOOS) (p = 0.001). The postoperative KSS results (83.7 ± 9.8 vs 78.7 ± 12.1; p = 0.015), KSFS (58.7 ± 9 vs 56 ± 8.9; p = 0.043) and KOOS (68.5 ± 13.2 vs 64.1 ± 16; p = 0.031) were significantly higher in patients with hinged implants (Table 2).

According to the Knee Society criteria, radiolucencies were observed in 29.9% of patients. Out of these, 23.9% occurred at the tibial level and 17.9% at the femoral level. At the end of follow-up, 15 patients (12.8%) had to undergo repeat surgery for implant loosening. The group of patients with CCK implant had significantly more radiolucencies (45.9% vs 12.5%, p < 0.001). (Table 4). At the end of follow-up, a greater number of implants had to be revised in this group for aseptic loosening (p = 0.021). Table 4.

Table 4 Radiolucency around the implants at final follow-up

The mean implant survival in both groups at the end of follow-up was 75.2%. No significant differences were seen in survival in both groups when this was defined as the performance of any type of intervention. However, implant survival, defined as the need for repeat surgery to replace the implant, was significantly greater in patients with hinged implants (91.1%) as compared to patients with CCK implants (67.2%) (p = 0.003) (Fig. 2A, B). When aseptic loosening was considered as the cause of implant failure, survival was significantly higher in the group of patients with hinged implants (94.6%) than in the group of patients treated with CCK implants (80,3%) (p = 0.02) (Fig. 2C). No differences were seen in survival due to sepsis between both groups (96.4% RHK vs 95.1% CCK, p = 0.73) (Fig. 2D).

Fig. 2
figure 2

A Long-term Kaplan–Meier survival curves comparing CCK implant group and the RHK implant group using any new surgery as the endpoint. B Long-term Kaplan–Meier survival curves using implant revision as endpoint C Long-term Kaplan–Meier survival curves using aseptic loosening surgery as the endpoint. D Long-term Kaplan–Meier survival curves using septic revision surgery as the endpoint

Discussion

Despite continuous improvements in implants and surgical techniques, the number and complexity of knee revisions continues to increase. The constraint/loosening ratio has relegated the use of theoretically more constrained implants (rotating hinges) in elderly patients with low functional demands, or in situations where another type of implant cannot be used [1, 5, 19]. However, in recent years, based on the good results obtained with specific models of rotating hinges in primary and revision surgery [20,21,22,23,24], some authors have questioned this marginal use of rotating hinges [25, 26]. This allows physiological rotation with flexion/extension, restricted by most CCK implants. Unfortunately, all attempts to compare these two types of implants were unable to show differences in either of these characteristics, probably because of the high number of different implants available (particularly in the hinge group), the short follow-up period, and the existence of a high heterogeneity in the different groups. To the best of our knowledge, this is the longest follow-up series ever reported that compared a single constrained condylar implant and a single hinged implant.

In this series, we observed a poorer preoperative score on the different scales, as well as a lower range of motion in the group of patients that underwent surgery with hinged implants, despite the lack of differences in comorbidities before surgery. The higher percentage of cases of septic origin and severe instability may also explain the presence of these poorer preoperative outcomes [27]. Hossain et al. [28] compared the mid-term results of 343 knee revisions performed with 3 different types of constraint (PS, CC and RH). In their population, patients undergoing surgery with RH implants had a lower range of motion (1°) and KSCS score (3 points); their findings are similar to those seen in this series (2.6° of ROM, 3 points on KSCS and 2 points on KSFS).

One of the main limitations to the widespread use of hinged implants is the theoretical increased risk of complications observed with their use. Pour et al. [5] reported medical complications rates of 18% and surgical complications rate of 20.9% after prosthetic revision in 43 knees with a third-generation condylar hinged implant, recommending its use only in selected cases. However, the indications selected were 23 massive bone defects, 10 instabilities, 4 periprosthetic fractures, and three comminuted supracondylar fractures, these cases had high complexity which could involve bias. To the best of our knowledge, there is no study that directly compares the complication rate between hinged and semi-constrained implants in a homogeneous group of patients. Shen et al. [1] found no increase in the rate of infections or aseptic loosening with the use of hinges in AORI II and III defects in septic or aseptic revisions. Similar data were recently been published by Malcon et al. and Yoon et al. [27, 29]. Our results differ from those of these authors, who found a significantly higher complication rate with the use of semi-constrained implants (CCK) (32/61) as compared to rotary hinged implants (12/56). After excluding minor complications (haematoma or partial lesions of the extensor apparatus), the complication rate remained higher in patients in whom semi-constrained implants were used as a result of a greater number of aseptic loosening (19.6% vs 5.3%), ligament instability (4.9% vs 0%), and periprosthetic fractures (8.2% vs 1.8%) (Table 3).

There is considerable controversy regarding the difference in clinical outcomes between hinge implants and semi-constrained CCK implants. Walker et al. [30] in one of the first articles published on this subject, questioned the theoretical clinical superiority of CCK implants over RHK implants. The clinical results of 56 RHK and 33 CCK implants were subsequently compared, and lower residual laxity was seen with RHK implants, and this led to better results in the KSCS, but had no impact on the range of motion or KSFS result. Hossain et al. [28] in 2010 found a greater range of motion with the use of hinged implants (111.7° vs 106°) but Dwivedi et al. [31] did not observed any difference in ROM, OKS, and KSS. However, their mean follow-up duration was less than 5 years. Malcon et al. [27] in the first published meta-analysis analysing 544 CCK vs 254 RHK implants, and recently Yoon Jung-Ro et al. [29] with 775 CCK vs 402 RHK implants, reported that no differences were found in the postoperative range of motion between both types of implants; however, both authors observed a minimal higher postoperative clinical score in the CCK group, but in this group, the preoperative clinical score was also higher. In our series, a significantly greater range of motion and better clinical results (KSCS, KSFS, and KOOS) were seen with the use of RHK implants at the end of follow-up. This observed difference in the range of motion can be explained by the intrinsic characteristics of the RHK implant chosen (intracondylar hinge, in which collateral ligaments are sectioned for the correct biomechanics of the design, etc.) which makes it easier to achieve a high degree of mobility [20, 26]. The clinical differences observed between this series and previous studies may be explained by the high homogeneity of both groups, using only one implant per group versus the high heterogeneity in other series that mixed different implants with completely different biomechanical designs.

Along with the theoretical increased risk of complications, the fear of decreased implant survival due to its theoretical greater constraint is the second limiting factor for expanded use of rotating hinges. Gehrke et al. [20] and Samguietei et al. [26] showed a 90% and 93.3% survival in complex primary and revision surgery, respectively, with the same RHK implant after 13.5 years of follow-up. To the best of our knowledge, all studies published to date have been unable to find survival differences between semi-constrained and constrained implants [1, 14, 22, 27,28,29, 31,32,33,34,35]. In our series, we observed a lower survival of the semi-constrained implant versus the constrained implant (67.2% vs 91.1%), mainly due to the greater number of revisions for loosening and instability. The high level of radiolucencies at the end of follow-up in CCK implants (45.9%) is particularly striking compared to those in patients with hinged implants (12.5%), which in our opinion could reflect the deleterious effect on implant fixation of the high degree of rotational constraint of certain CCK implants (Table 4).

This study has some limitations. First, patients were operated on by different surgeons. However, they were all experienced surgeons, with the same philosophy as regards selection of the degree of constraint. Second, the low number of patients in each group may have influenced the lack of statistical significance in some of the results observed. However, the high homogeneity of both groups, as only a single prosthetic implant was used, allowed us to observe statistically significant differences despite this low number of patients. Third, the results observed in this study can only be extrapolated to the included implants, due to their special mechanical characteristics (high rotational constraint in LCCK and intracondylar rotating hinge design in Endomodel). However, the choice of these implants was based on the existence of good results with each of them [20, 36] and their wide use in different countries, being the most commonly used implants in their group according to different registries [37, 38]. Fourth, we used two different fixation technique, full cemented stem in RHK group and hybrid cemented (metaphyseal cemented + press-fit diaphyseal stem), and this may explain the differences observed in aseptic loosening rate. However, there are a lot of reports [39,40,41] that show similar results using both techniques with the same bone defect and similar implant. In our series, no difference was observed in the number of cones used in both groups to treat severe bone defect. Fifth, the patients included in the group of constrained implants had poorer clinical outcomes and a greater number of surgeries and infections, which could affect the results obtained. However, despite this “limitation” or possible bias, patients with hinged implants had better results and survival at the end of follow-up.

Conclusion

Our results suggest that the use of one specific, more constrained implant, such as an intracondylar rotating hinge, does not worsen clinical outcomes, but has better results and longer long-term survival. We hypothesize that this difference could be the result of the lower rotational constraint of these implants; however, more experimental and clinical test are necessary to support this affirmation.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

RHK:

Rotation hinge knee

CCK:

Condylar constrained knee

KSCS:

Knee Society Clinical Score

KSFS:

Knee Society Functional Score

KOOS:

Knee Injury and Osteoarthritis Outcome Score

TKA:

Total knee arthroplasty

ROM:

Range of motion

AORI:

Anderson Orthopedic Research Institute

KSS:

Knee Society Score

PS:

Postero stabilized

CC:

Condilar constrained

RH:

Rotating hinge

ASA:

American Society of Anesthesiologists Score

HKA:

Hip-knee angle

References

  1. Shen C, Lichstein PM, Austin MS, Sharkey PF, Parvizi J. Revision knee arthroplasty for bone loss: choosing the right degree of constraint. J Arthroplasty. 2014;29:127–31. https://doi.org/10.1016/j.arth.2013.04.042.

    Article  PubMed  Google Scholar 

  2. Rossi SMP, Ivone A, Ghiara M, Jannelli E, Sangaletti R, Perticarini L, et al. A ligament tensor-guided extramedullary alignment technique for distal femoral cut in total knee replacement: results at a minimum 3 years follow-up. Arch Orthop Trauma Surg. 2021;141:2295–302. https://doi.org/10.1007/s00402-021-04115-4.

    Article  PubMed  Google Scholar 

  3. Aletto C, Zara A, Notarfrancesco D, Maffulli N. Computer assisted total knee arthroplasty: 2.5 years follow-up of 200 cases. Surg J R Coll Surg Edinb Irel. 2021;19:e394-401. https://doi.org/10.1016/j.surge.2020.12.003.

    Article  Google Scholar 

  4. Tornese D, Robustelli A, Ricci G, Rancoita PMV, Maffulli N, Peretti GM. Predictors of postoperative hospital length of stay after total knee arthroplasty. Singapore Med J. 2021. https://doi.org/10.11622/smedj.2021142.

    Article  PubMed  Google Scholar 

  5. Pour AE, Parvizi J, Slenker N, Purtill JJ, Sharkey PF. Rotating hinged total knee replacement: use with caution. J Bone Joint Surg Am. 2007;89:1735–41. https://doi.org/10.2106/JBJS.F.00893.

    Article  PubMed  Google Scholar 

  6. McAuley JP, Engh GA. Constraint in total knee arthroplasty: when and what? J Arthroplasty. 2003;18:51–4. https://doi.org/10.1054/arth.2003.50103.

    Article  PubMed  Google Scholar 

  7. Naudie DD, Rorabeck CH. Managing instability in total knee arthroplasty with constrained and linked implants. Instr Course Lect. 2004;53:207–15.

    PubMed  Google Scholar 

  8. Al-Jabri T, Brivio A, Maffulli N, Barrett D. Management of instability after primary total knee arthroplasty: an evidence-based review. J Orthop Surg. 2021;16:729. https://doi.org/10.1186/s13018-021-02878-5.

    Article  Google Scholar 

  9. Sharkey PF, Lichstein PM, Shen C, Tokarski AT, Parvizi J. Why are total knee arthroplasties failing today—has anything changed after 10 years? J Arthroplasty. 2014;29:1774–8. https://doi.org/10.1016/j.arth.2013.07.024.

    Article  PubMed  Google Scholar 

  10. Touzopoulos P, Drosos GI, Ververidis A, Kazakos K. Constrained implants in total knee replacement. Surg Technol Int. 2015;26:307–16.

    PubMed  Google Scholar 

  11. Akbari Shandiz M, Boulos P, Saevarsson SK, Yoo S, Miller S, Anglin C. Changes in knee kinematics following total knee arthroplasty. Proc Inst Mech Eng. 2016;230:265–78. https://doi.org/10.1177/0954411916632491.

    Article  Google Scholar 

  12. Boguszewski DV, Joshi NB, Yang PR, Markolf KL, Petrigliano FA, McAllister DR. Location of the natural knee axis for internal-external tibial rotation. Knee. 2016;23:1083–8. https://doi.org/10.1016/j.knee.2015.11.003.

    Article  PubMed  Google Scholar 

  13. Freeman MAR, Pinskerova V. The movement of the normal tibio-femoral joint. J Biomech. 2005;38:197–208. https://doi.org/10.1016/j.jbiomech.2004.02.006.

    Article  CAS  PubMed  Google Scholar 

  14. Vasso M, Beaufils P, Schiavone PA. Constraint choice in revision knee arthroplasty. Int Orthop. 2013;37:1279–84. https://doi.org/10.1007/s00264-013-1929-y.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Samiezadeh S, Bougherara H, Abolghasemian M, D’Lima D, Backstein D. Rotating hinge knee causes lower bone-implant interface stress compared to constrained condylar knee replacement. Knee Surg Sports Traumatol Arthrosc Off J ESSKA. 2019;27:1224–31. https://doi.org/10.1007/s00167-018-5054-8.

    Article  Google Scholar 

  16. Sabatini L, Risitano S, Rissolio L, Bonani A, Atzori F, Massè A. Condylar constrained system in primary total knee replacement: our experience and literature review. Ann Transl Med. 2017;5:135. https://doi.org/10.21037/atm.2017.03.29.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Engh GA, Ammeen DJ. Bone loss with revision total knee arthroplasty: defect classification and alternatives for reconstruction. Instr Course Lect. 1999;48:167–75.

    CAS  PubMed  Google Scholar 

  18. Murray PB, Rand JA, Hanssen AD. Cemented long-stem revision total knee arthroplasty. Clin Orthop. 1994;309:116–23.

    Google Scholar 

  19. Kouk S, Rathod PA, Maheshwari AV, Deshmukh AJ. Rotating hinge prosthesis for complex revision total knee arthroplasty: a review of the literature. J Clin Orthop Trauma. 2018;9:29–33. https://doi.org/10.1016/j.jcot.2017.11.020.

    Article  PubMed  Google Scholar 

  20. Gehrke T, Kendoff D, Haasper C. The role of hinges in primary total knee replacement. Bone Jt J. 2014;96B:93–5. https://doi.org/10.1302/0301-620X.96B11.34143.

    Article  Google Scholar 

  21. Cottino U, Abdel MP, Perry KI, Mara KC, Lewallen DG, Hanssen AD. Long-term results after total knee arthroplasty with contemporary rotating-hinge prostheses. J Bone Joint Surg Am. 2017;99:324–30. https://doi.org/10.2106/JBJS.16.00307.

    Article  PubMed  Google Scholar 

  22. Felli L, Coviello M, Alessio-Mazzola M, Cutolo M. The Endo-Model(®) rotating hinge for rheumatoid knees: functional results in primary and revision surgery. Orthopade. 2016;45:446–51. https://doi.org/10.1007/s00132-015-3193-x.

    Article  CAS  PubMed  Google Scholar 

  23. Gudnason A, Milbrink J, Hailer NP. Implant survival and outcome after rotating-hinge total knee revision arthroplasty: a minimum 6-year follow-up. Arch Orthop Trauma Surg. 2011;131:1601–7. https://doi.org/10.1007/s00402-011-1330-5.

    Article  PubMed  Google Scholar 

  24. Joshi N, Navarro-Quilis A. Is there a place for rotating-hinge arthroplasty in knee revision surgery for aseptic loosening? J Arthroplasty. 2008;23:1204–11. https://doi.org/10.1016/j.arth.2007.10.016.

    Article  PubMed  Google Scholar 

  25. Petrou G, Petrou H, Tilkeridis C, Stavrakis T, Kapetsis T, Kremmidas N, et al. Medium-term results with a primary cemented rotating-hinge total knee replacement. A 7- to 15-year follow-up. J Bone Joint Surg Br. 2004;86:813–7. https://doi.org/10.1302/0301-620x.86b6.14708.

    Article  CAS  PubMed  Google Scholar 

  26. Sanguineti F, Mangano T, Formica M, Franchin F. Total knee arthroplasty with rotating-hinge Endo-Model prosthesis: clinical results in complex primary and revision surgery. Arch Orthop Trauma Surg. 2014;134:1601–7. https://doi.org/10.1007/s00402-014-2061-1.

    Article  CAS  PubMed  Google Scholar 

  27. Malcolm TL, Bederman SS, Schwarzkopf R. Outcomes of varus Valgus constrained versus rotating-hinge implants in total knee arthroplasty. Orthopedics. 2016;39:e140-148. https://doi.org/10.3928/01477447-20151228-07.

    Article  PubMed  Google Scholar 

  28. Hossain F, Patel S, Haddad FS. Midterm assessment of causes and results of revision total knee arthroplasty. Clin Orthop. 2010;468:1221–8. https://doi.org/10.1007/s11999-009-1204-0.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Yoon J-R, Cheong J-Y, Im J-T, Park P-S, Park J-O, Shin Y-S. Rotating hinge knee versus constrained condylar knee in revision total knee arthroplasty: a meta-analysis. PLoS ONE. 2019;14: e0214279. https://doi.org/10.1371/journal.pone.0214279.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  30. Walker PS, Manktelow AR. Comparison between a constrained condylar and a rotating hinge in revision knee surgery. Knee. 2001;8:269–79. https://doi.org/10.1016/s0968-0160(01)00121-1.

    Article  CAS  PubMed  Google Scholar 

  31. Dwivedi DD, Thomas S, Verma N, Kapil N, Singh KK, et al. Similar outcomes with usage of varus valgus constrained and a rotating hinge knee: a review of 108 cases. J Arthrosc Jt Surg. 2020;7:59–63. https://doi.org/10.1016/j.jajs.2020.06.005.

    Article  Google Scholar 

  32. Bali K, Naudie DD, Howard JL, McCalden RW, MacDonald SJ, Teeter MG. Comparison of Tibial insert polyethylene damage in rotating hinge and highly constrained total knee arthroplasty: a retrieval analysis. J Arthroplasty. 2016;31:290–4. https://doi.org/10.1016/j.arth.2015.07.007.

    Article  PubMed  Google Scholar 

  33. Fuchs S, Sandmann C, Gerdemann G, Skwara A, Tibesku CO, Bottner F. Quality of life and clinical outcome in salvage revision total knee replacement: hinged vs total condylar design. Knee Surg Sports Traumatol Arthrosc Off J ESSKA. 2004;12:140–3. https://doi.org/10.1007/s00167-003-0401-8.

    Article  Google Scholar 

  34. Hommel H, Wilke K, Kunze D, Hommel P, Fennema P. Constraint choice in revision knee arthroplasty: study protocol of a randomised controlled trial assessing the effect of level of constraint on postoperative outcome. BMJ Open. 2017;7: e012964. https://doi.org/10.1136/bmjopen-2016-012964.

    Article  PubMed  PubMed Central  Google Scholar 

  35. Luttjeboer JS, Bénard MR, Defoort KC, van Hellemondt GG, Wymenga AB. Revision total knee arthroplasty for instability-outcome for different types of instability and implants. J Arthroplasty. 2016;31:2672–6. https://doi.org/10.1016/j.arth.2016.06.062.

    Article  PubMed  Google Scholar 

  36. Cholewinski P, Putman S, Vasseur L, Migaud H, Duhamel A, Behal H, et al. Long-term outcomes of primary constrained condylar knee arthroplasty. Orthop Traumatol Surg Res OTSR. 2015;101:449–54. https://doi.org/10.1016/j.otsr.2015.01.020.

    Article  CAS  PubMed  Google Scholar 

  37. National Joint Registry for England Wales and Northern Ireland Prostheses used in hip, knee, ankle, elbow and shoulder replacement procedures 2017; 2018.

  38. The Swedish Knee Arthroplasty Register: Annual Report 2017; 2018.

  39. Alipit V, Kirk A, Scholl D, Schmidig G, Springer BD, Lee G-C. Micromotion analysis of various tibial constructs in moderate tibial defects in revision total knee arthroplasty. J Arthroplasty. 2021;36:362-367.e1. https://doi.org/10.1016/j.arth.2020.07.013.

    Article  PubMed  Google Scholar 

  40. Kosse NM, van Hellemondt GG, Wymenga AB, Heesterbeek PJC. Comparable stability of cemented vs press-fit placed stems in revision total knee arthroplasty with mild to moderate bone loss: 6.5-year results from a randomized controlled trial with radiostereometric analysis. J Arthroplasty. 2017;32:197–201. https://doi.org/10.1016/j.arth.2016.06.003.

    Article  PubMed  Google Scholar 

  41. Greene JW, Reynolds SM, Stimac JD, Malkani AL, Massini MA. Midterm results of hybrid cement technique in revision total knee arthroplasty. J Arthroplasty. 2013;28:570–4. https://doi.org/10.1016/j.arth.2012.08.010.

    Article  PubMed  Google Scholar 

Download references

Funding

No funding was received to assist with the preparation of this manuscript.

Author information

Authors and Affiliations

Authors

Contributions

PSR planned the study, collected the data, examined the patients, evaluated the radiology findings, performed the statistical analysis, and wrote the manuscript, VELR collected the data, examined the patients and evaluated the radiology findings. MVM help in the planned study. JAMD, MVM and JV examined on the patients and reviewed the manuscript. All the authors read and approved the final version of the manuscript.

Corresponding author

Correspondence to Pablo Sanz-Ruiz.

Ethics declarations

Ethics approval and consent to participate

This was retroactive study that reviewed a prospectively collected database and was performed in accordance with the principles of the 1964 Declaration of Helsinki as revised in 2013 and was approved by the research ethics committee of our centre (cod: VERLR-COT-2017). All participants gave their informed consent to participate in the study and for their clinical and radiological data to be reproduced.

Consent for publication

All participants gave their write consent to participate in the study and for any possible publication.

Competing interests

The authors declare that they have no competing interests, and no external funding was received.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sanz-Ruiz, P., León-Román, V.E., Matas-Diez, J.A. et al. Long-term outcomes of one single-design varus valgus constrained versus one single-design rotating hinge in revision knee arthroplasty after over 10-year follow-up. J Orthop Surg Res 17, 135 (2022). https://doi.org/10.1186/s13018-022-03026-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-022-03026-3

Keywords