Peri-prosthetic bone remodeling of hydroxyapatite-coated compaction short stem was not affected by stem alignment

Abstract

Background

To improve implant survival through accelerated early bone remodeling during total hip arthroplasty (THA), hydroxyapatite (HA) is widely used as a bioactive coating, which is believed to enhance initial fixation by osseointegration. We aimed to investigate the relationship between stem insertion alignment and postoperative bone mineral density (BMD) changes in patients with full hydroxyapatite-coated (HA) compaction short stem and short tapered-wedge stem.

Methods

This retrospective cohort study enrolled 115 consecutive patients (115 joints) undergoing THA using the full HA compaction short (nā€‰=ā€‰59) and short tapered-wedge (nā€‰=ā€‰56) stems. Stem alignment, including anteversion, valgus, and anterior tilt were measured by a three-dimensional template using computed tomography data. Post-operative peri-prosthetic BMD was measured by dual-energy X-ray absorptiometry. The relationship between stem alignment and BMD changes in the stems was analyzed.

Results

Patterns of peri-prosthetic BMD changes were similar in both groups. Stem insertion alignments of anteversion, valgus, and anterior tilt were different between the two stem types. Stem alignment of valgus and anterior tilt did not affect peri-prosthetic BMD in either of the stem type. An absolute anteversion difference between stem anteversion and original canal anteversion caused significant peri-prosthetic BMD loss in Gruen zones one and seven in the tapered-wedge stem. However, stem alignment of absolute anteversion difference did not affect BMD changes in the HA compaction stem.

Conclusions

Peri-prosthetic bone remodeling remained unaffected by stem alignment after THA with the new short full HA compaction stem.

Background

Hydroxyapatite (HA) is widely used in total hip arthroplasty (THA) to improve implant survival through accelerated early bone remodeling [1, 2]. This bioactive coating is believed to enhance initial fixation by osseointegration [1, 2]. The ACTIS stem (DePuy Synthes, Warsaw, IN, USA) is a medial-collared, triple-tapered short stem with full HA-coating, and the cancellous bone was compacted without the implant in contact with the femoral cortex. The Tri-Lock bone preservation stem (BPS) (DePuy Synthes) is a short tapered-wedge stem, which potentially preserves more bone stock, improving proximal load transfer and demonstrating lower stress shielding than long stems [3, 4]. The full HA-coated compaction short stem and short tapered-wedge stem were developed for minimally invasive THA and made from the same titanium alloy, but the geometry, type of surface coating, and concept of both stems differ substantially.

Recently, we reported that peri-prosthetic BMD changes was similar in full HA-coated compaction short stems and short tapered-wedge stems [5]. We further demonstrated that age, body mass index (BMI), and daily activity did not affect proximal femoral BMD changes in both stem types, and femoral bone shape affected the BMD changes in the tapered-wedge stem but not in the full HA compaction stem [5]. The contact force is dependent on subject-specific geometry and that specificity influences stress distribution in the peri-prosthetic bone after THA [6, 7]. A finite element method study reported that stress distribution in periprosthetic bone was dependent on stem alignment, especially anteriorā€“posterior and varusā€“valgus alignments [8]. A radiographic study reported that varus alignment caused early stem loosening because of poor seating of the femoral prosthesis with decreased bone ingrowth [9]. The relation between varusā€“valgus alignment and stress distribution in the periprosthetic bone has been established well in biomechanical studies [9,10,11,12,13]. We have reported that excessive anteversion mismatch between anatomical canal and stem anteversions caused postoperative peri-prosthetic proximal BMD loss in a short tapered-wedge stem [14]. Thus, stress distribution in the peri-prosthetic bone was influenced not only by subject-specific geometry but also by stem alignment.

We hypothesized that stem alignment may affect different patterns of BMD change after THA between HA-coated compaction short stems and short tapered-wedge stems. Therefore, in this study we evaluated the relationship between stem insertion alignment and postoperative BMD change in the full HA-coated compaction short stem and short tapered-wedge stem.

Methods

Patientsā€™ characteristics

A total of 159 consecutive patients (159 joints) at our institution underwent THA using the short tapered-wedge stem (Tri-Lock BPS; 95 joints) from January 2016 to April 2017 or the full HA compaction short stem from April 2017 to November 2018. To analyze the three-dimensional stem alignment, postoperative computed tomography (CT) was performed from June 2016 to July 2018. Two methods of THA were used at our institution; from June 2016 to April 2017, THA was conducted using the short tapered-wedge stem (Tri-Lock BPS; 95 joints). Thereafter, the full HA compaction short stem was introduced in Japan and, from then onward, was used at our institution. Hence, between June 2016 and April 2017, patients enrolled in the present study underwent THA using the short tapered-wedge stem (Tri-Lock BPS; DePuy Synthes; 56 joints) (Fig.Ā 1a); then, from April 2017 to July 2018, patients underwent the full HA compaction short stem (ACTIS; DePuy Synthes, Warsaw, IN; 59 joints) (Fig.Ā 1b). Therefore, this retrospective cohort study enrolled 115 consecutive patients (115 joints) from June 2016 to July 2018.

Fig. 1
figure 1

Images of a the short tapered-wedge stem (Tri-Lock BPS) and b the full HA compaction short stem (ACTIS)

The preoperative diagnoses were osteoarthritis (grade 4, according to the Tƶnnis classification) (89 joints), avascular necrosis of the femoral head (22 joints), and rheumatoid arthritis (4 joints). Patients with a distorted anatomy of the proximal femur, osteoporosis (lumbar spine BMDā€‰<ā€‰0.8), metabolic bone disease, and those who underwent bilateral THA were excluded.

All procedures were performed via the mini-anterolateral supine approach by a single senior surgeon. Full weight-bearing was allowed for all the patients, a day after the operation. At the time of surgery, BMI was assessed. The postoperative follow-up included dual-energy X-ray absorptiometry (DEXA) scanning and evaluation of clinical factors, including hip function, which was evaluated using two grading methods: (a) the Japanese Orthopedic Association (JOA) score, which allocates 40 points for pain, 20 points for range of motion, 20 points for walking ability, and 20 points for activities of daily living, with a maximum total score of 100 points [15] and (b) the University of California Los Angeles (UCLA) activity score, which describes subjectsā€™ level of activity from 1 (ā€œno physical activity, dependent on othersā€) to 10 (ā€œregular participation in impact sportsā€). The JOA and UCLA scores and radiographic findings were evaluated two years postoperatively.

Measurement of stem alignment

Pre-operative and postoperative CT scans from the pelvis to the knee joint were performed and transferred to a three-dimensional template software (Zed Hip; Lexi, Tokyo, Japan). Computer-aided design models of the implants were manually adjusted for postoperative multiplanar reconstruction in the CT images (Fig.Ā 2). Stem anteversion and anatomical canal anteversion angles were measured with respect to the femurā€™s posterior condylar line axis [16]. We compared the anatomical canal and postoperative stem anteversions, and the anteversion error was defined as the difference between the stem and anatomical canal anteversions. Patients were divided into outlier and non-outlier of stem alignment groups. The outliers of stem anteversion were defined as having an absolute difference of 5Ā° or more between anatomical canal and postoperative stem anteversions, and the outliers of varusā€“valgus and anteriorā€“posterior tilt angle were defined as having an absolute difference of 3Ā° or more, according to previous studies [9, 14].

Fig. 2
figure 2

Stem alignments are measured by superimposing the templates of the stem data on postoperative images of the femoral component using Zed Hip software

DEXA measurements

A DPX-L scanner (GE Lunar Corporation, Madison, WI, USA) was used for DEXA scanning. Software version 1.35 was used for DPX-L total body scans, which set the appropriate transverse speed at 16Ā cm/s, 8Ā cm/s, or 4Ā cm/s, depending on the participantā€™s height. The patients were positioned supine with their legs in the neutral position with a knee and foot support to facilitate the scanning of the anteriorā€“posterior projection of the proximal femur, including the area distal to the prosthesis, using an edge-detection technique.

Peri-prosthetic BMD was determined postoperatively in seven regions of interest (ROIs) based on the Gruen zones [17]. The seven Gruen zones were positioned based on the distal tip and shoulder of the prosthesis. The two most proximal Gruen zones (1 and 7) were also combined to create a proximal femur ROI, which represents the region where the porous coating of the uncemented stems is normally situated. The values were expressed as area BMD in grams per square centimeter. The BMD around the stem was assessed within 1Ā month (baseline BMD) and 6, 12, 18, and 24Ā months postoperatively by a technician in the radiology department who was blinded to the stem used. The BMD ratios were calculated by dividing each BMD value at 6, 12, 18, and 24Ā months postoperatively by the baseline BMD.

Statistical analysis

All data are expressed as meanā€‰Ā±ā€‰standard deviation (SD) unless otherwise indicated. The differences in patientsā€™ backgrounds and stem alignment between the full HA compaction and tapered-wedge groups were analyzed using the Mannā€“Whitney U test (Table 1). Sequential changes in the BMD within each group were analyzed by one-way analysis of variance using the Tukeyā€™s post-hoc test. Additionally, the BMD at each timepoint was compared between the groups using the Mannā€“Whitney U test (Fig.Ā 2). The differences in peri-prosthetic BMD changes for stem valgus, anterior tilt, and absolute anteversion error alignment between outliers and non-outliers were analyzed using the Mannā€“Whitney U test (Tables 2, 3, 4). The difference in stem alignment and whether stem collar is contacting the femoral calcar or not were analyzed using the Mannā€“Whitney U test (Table 5). The differences in stem alignments among Dorr types were analyzed using the one-way analysis of variance and the Tukeyā€™s post-hoc test (Table 6). Statistical significance was set at pā€‰<ā€‰0.05.

Table 1 Stem alignment
Table 2 Relation between stem valgus alignment and BMD change at 24Ā months post-operatively in the Gruen zones
Table 3 Relation between stem anterior tilt alignment and BMD change at 24Ā months post-operatively in the Gruen zones
Table 4 Relation between stem absolute anteversion error and BMD change at 24Ā months post-operatively in the Gruen zones
Table 5 Comparison of stem alignment whether stem collar contacting to femoral calcar or not
Table 6 Comparison of stem alignment among Door types of femoral bone shape

Post-hoc power analysis was performed using G*Power [18]. For comparison between the groups, for a sample size of 59 versus 56 elements in the two groups, and a type-I error (Ī±) of 0.05, the study is expected to provide a power (1-Ī²) of 0.83 for detecting an effect size of 0.5. Comparing the outliers and non-outliers, we calculated the effect size by means and SDs based on the Cohenā€™s d for each parameter and 95% confidence interval for effect sizes (Tables 2, 3, 4) [19].

Results

Patientsā€™ characteristics

Patientsā€™ characteristics between the full HA compaction short stem and tapered-wedge stem groups was compared. The mean patient age at operation was 66.8ā€‰Ā±ā€‰11.1Ā years in the full HA compaction group, and 67.5ā€‰Ā±ā€‰9.1Ā years in the tapered-wedge group (pā€‰=ā€‰0.800). During clinical evaluation at 24Ā months postoperatively, the mean BMI was 22.8ā€‰Ā±ā€‰2.9Ā kg/m2 in the full HA compaction group and 23.7ā€‰Ā±ā€‰3.7Ā kg/m2 in the tapered-wedge group (pā€‰=ā€‰0.843). The JOA score was 96.1ā€‰Ā±ā€‰4.5 in the full HA compaction group and 95.9ā€‰Ā±ā€‰6.2 in the tapered-wedge group (pā€‰=ā€‰0.825). The UCLA score was 6.4ā€‰Ā±ā€‰1.6 in the full HA compaction group and 6.1ā€‰Ā±ā€‰1.5 in the tapered-wedge group (pā€‰=ā€‰0.268). The mean values for age, BMI, JOA score, and UCLA activity score were not significantly different between the groups.

Stem insertion alignment was affected by stem design

Table 1 demonstrates that no significant differences in stem and canal anteversions were found between the HA compaction and tapered-wedge groups. The mean absolute value of surgical error (postoperative stem anteversionā€“canal anteversion) was 5.2Ā°ā€‰Ā±ā€‰4.8Ā° (HA compaction) and 8.8Ā°ā€‰Ā±ā€‰9.1Ā° (tapered-wedge), and the mean absolute value of anteversion error was significantly higher in the tapered-wedge group than in the HA compaction group (pā€‰=ā€‰0.039) (Table 1). The mean absolute values of valgus error were 3.1Ā°ā€‰Ā±ā€‰2.2Ā° (HA compaction) and 1.8Ā°ā€‰Ā±ā€‰1.8Ā° (tapered-wedge), and the error was significantly higher in the HA compaction group than in the tapered-wedge group (pā€‰<ā€‰0.001) (Table 1). The mean absolute values of anterior tilt error were 6.0Ā°ā€‰Ā±ā€‰2.2Ā° (HA compaction) and 3.1Ā°ā€‰Ā±ā€‰2.1Ā° (tapered-wedge), and the error was significantly higher in the HA compaction group than in the tapered-wedge group (pā€‰<ā€‰0.001) (Table 1).

Peri-prosthetic BMD changes were similar between the groups

Minimal BMD changes were found in the distal femur (Gruen zones 3, 4, and 5) and zone 6 in both stem types (Fig.Ā 3). However, a significant BMD loss was observed in zones 1 and 7 at each timepoint in both groups compared with the BMD value obtained within 1Ā month postoperatively (full HA compaction: zone 1, 24Ā months, pā€‰<ā€‰0.001; zone 7, 24Ā months, pā€‰<ā€‰0.001) (tapered-wedge: zone 1, 24Ā months, pā€‰<ā€‰0.001; zone 7, 24Ā months, pā€‰<ā€‰0.001). On comparing the BMD changes between the groups, we found that the change in BMD of the tapered-wedge stem group was significantly higher in zone 2 at each timepoint. The change in BMD of the full HA compaction group was significantly higher in zone 4 (Fig.Ā 3).

Fig. 3
figure 3

Bone mineral density (BMD) changes at 6, 12, 18, and 24Ā months postoperatively in the Gruen zone 7. Columns represent mean values of BMD changes. The dotted line indicates the value of HA compaction short stem. The solid line indicates the value of the short tapered-wedge stem. ā€» Indicates pā€‰<ā€‰0.05 compared between two groups

Stem alignment affected peri-prosthetic BMD loss in tapered-wedge stem but not in HA compaction stem

Table 2 demonstrates that larger valgus alignment caused significant differences in peri-prosthetic BMD loss in Gruen zone 2 in the tapered-wedge group, but alignment did not affect BMD changes in the HA compaction group. Tables 2 and 3 demonstrate that peri-prosthetic BMD did not change between the outlier and non-outlier groups of anterior tilt in both HA compaction and tapered-wedge stems. Table 4 demonstrates that larger absolute anteversion errors caused significant differences in peri-prosthetic BMD loss in Gruen zones 1 and 7 in the tapered-wedge group. However, stem alignment of absolute anteversion error did not affect BMD changes in the HA compaction group (Table 4).

HA compaction stem and tapered-wedge stem represent different patterns of contact to bone surface

FigureĀ 4a demonstrates the subject-specific differences of representative cases in HA compaction and tapered-wedge stems. In the HA compaction stem, the stem was in contact with the anterior wall, posterior edge of the proximal femur, and distal anterior wall in the sagittal view and widely occupied the antero-posterior area of the proximal femur in the axial view. Meanwhile, in the tapered-wedge stem, the stem contacts only the posterior edge of the proximal femur in the sagittal view and less occupied antero-posterior area of the proximal femur in the axial view (Fig.Ā 4b).

Fig. 4
figure 4

Three-dimensional reconstruction images of a, c, e the HA compaction short stem (ACTIS) and b, d, f the short tapered-wedge stem (Tri-Lock BPS). a, b Coronal view, c, d sagittal view. The yellow triangle indicates contact area of bone and stem e, f axial view

Relationship between stem collar contact and stem alignment

Table 5 demonstrates comparison of stem alignment, whether stem collar contacts the femoral calcar or not in the HA compaction stem. The mean value of stem valgus was significantly higher in the stem collar-calcar contact group than in the non-contact group (Table 5). The mean values of anterior tilt error and absolute value of anteversion error were not significantly changed between the stem collar-calcar contact and non-contact groups (Table 5).

Relationship between femoral bone shape and stem alignment

Table 6 demonstrates comparison of stem alignments, according to the Dorr classification for femoral bone shape. The significant change in mean value of stem valgus error was noted among Dorr types (Table 6). A post-hoc analysis demonstrated that a significant mean value of valgus changes was found between Dorr types B and C (pā€‰=ā€‰0.003). The mean values of anterior tilt error and absolute value of anteversion error were not significantly changed among the Dorr types (Table 6).

Discussion

In this study, we observed similar patterns of peri-prosthetic BMD changes in the full HA compaction short stem and tapered-wedge stem, although stem alignments of anteversion, valgus, and anterior tilt were different between the two groups. Furthermore, the stem alignment of the tapered-wedge stem affects peri-prosthetic BMD loss after THA, but that of the HA compaction stem did not.

Taniguchi et al. compared the stem insertion alignment of metaphyseal filling and short tapered-wedge stems and found that tapered-wedge stems had greater variation in the increase in anteversion than metaphyseal filling stems [20]. Our previous report demonstrated that the insertion alignment of short tapered-wedge stems was similar to that of straight stems during mini-invasive THA [21]. The current study demonstrated that the alignments of anteversion, valgus, and anterior tilt were different between HA compaction and tapered-wedge stems. This difference can be explained by stem geometry. The current study demonstrated that the HA compaction stem was in contact with the anterior wall, posterior edge of the proximal femur, and distal anterior wall in the sagittal view and widely occupied the antero-posterior area of the proximal femur in the axial view. The reason for contact pattern of HA compaction stem was explained by the stem profile that anteriorā€“posterior width of HA-compaction stem was thicker than tapered-wedge stems, and the thicker anteriorā€“posterior width of the HA-compaction stem may be fixed at larger anterior tilt and smaller anteversion errors along the femoral canal geometry. Another feature of the HA compaction short stem is the medial collar. In several cases, the medial collar may soon attach to the calcar region, and the stem alignment changes to valgus during bone compaction. We demonstrated that the mean value of stem valgus was significantly higher in the stem collar-calcar contact group than in the non-contact group, and the result can explain the reason for stem alignment changes to valgus during bone compaction in the stem collar-calcar contact group. However, we previously reported a relationship between the stem collar and periprosthetic BMD changes, and no differences was found in any of the Gruen zones between the stem collar-calcar contact and non-contact groups [5]. These findings supported our current result that stem alignment does not affect peri-prosthetic bone remodeling in the HA compaction stem.

Previous studies have investigated the influence of the proximal femoral canal shape on post-operative BMD changes in the femur [5, 22]. The pre-operative femoral canalā€™s morphology did not affect the BMD at two years post-operatively around the ZweymĆ¼ller-type stem [22]. We previously reported that the proximal femoral canalā€™s shape affected post-operative BMD changes in the tapered-wedge stem, but did not in the full HA compaction stem group. The current study demonstrated that stem valgus error of HA compaction stem was higher in Dorr type C than in Dorr type B. These findings with our current result showed that stem alignment does not affect peri-prosthetic bone remodeling in the HA compaction stem.

Our current study showed findings similar to those of a previous report; excessive mismatch of stem and anatomical canal anteversions caused peri-prosthetic proximal BMD loss in short tapered-wedge stems [14]. However, stem insertion alignment did not affect the BMD changes in the HA compaction stem group. Kim et al. reported that the metaphyseal fitting stem, which was particularly characterized by metaphyseal fixation, predominantly indicated proximal load transfer and excellent peri-prosthetic BMD preservation in the proximal region of the stem [23]. The proximal profile of the HA compaction short stem represents a thicker antero-posterior width than the short tapered-wedge stem, and can be expected to be fixed in the proximal femoral cavity and provide rotational stability in addition to the effect of HA-coating for early bone remodeling.

The limitations of this study were as follows: First, this was not a randomized study but a retrospective cohort study. To evaluate the clinical and radiographic outcomes, an analysis of a randomized selection of patients is preferable. The backgrounds of the patients were similar in our study. Secondly, the sample size was small with respect to the evaluation of outcomes. In particular, the sample size for the comparison between outliers and non-outliers was underpowered for analysis. Finally, the preoperative peri-prosthetic BMD was not measured. We did not compare BMD values pre- and postoperatively.

Conclusions

In our study, similar patterns of peri-prosthetic BMD changes were observed between the full HA compaction short stem and tapered-wedge stem. Stem insertion alignments of anteversion, valgus, and anterior tilt were different between the two types of stem groups because of the different stem geometries. Moreover, stem alignment of the tapered-wedge stem affects peri-prosthetic BMD loss after THA, but HA compaction stem did not due to wide occupation of the proximal femoral cavity.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

BMD:

Bone mineral density

BMI:

Body mass index

BPS:

Bone preservation

CT:

Computed tomography

DEXA:

Dual-energy X-ray absorptiometry

HA:

Hydroxyapatite-coated

JOA:

Japanese Orthopedic Association

ROI:

Regions of interest

SD:

Standard deviation

THA:

Total hip arthroplasty

UCLA:

University of California Los Angeles

References

  1. Chen YL, Lin T, Liu A, Shi MM, Hu B, Shi ZL, et al. Does hydroxyapatite coating have no advantage over porous coating in primary total hip arthroplasty? A meta-analysis. J Orthop Surg Res. 2015;10:21. https://doi.org/10.1186/s13018-015-0161-4.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  2. Vidalain JP. Twenty-year results of the cementless Corail stem. Int Orthop. 2011;35:189ā€“94. https://doi.org/10.1007/s00264-010-1117-2.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  3. Molli RG, Lombardi AV Jr, Berend KR, Adams JB, Sneller MA. A short tapered stem reduces intraoperative complications in primary total hip arthroplasty. Clin Orthop Relat Res. 2012;470:450ā€“61. https://doi.org/10.1007/s11999-011-2068-7.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  4. Patel RM, Smith MC, Woodward CC, Stulberg SD. Stable fixation of short-stem femoral implants in patients 70 years and older. Clin Orthop Relat Res. 2012;470:442ā€“9. https://doi.org/10.1007/s11999-011-2063-z.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  5. Hayashi S, Hashimoto S, Kuroda Y, Nakano N, Matsumoto T, Kamenaga T, et al. Hydroxyapatite-coated compaction short stem represents a characteristic pattern of peri-prosthetic bone remodelling after total hip arthroplasty. Arch Orthop Trauma Surg. 2021. https://doi.org/10.1007/s00402-021-04140-3.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  6. Jonkers I, Sauwen N, Lenaerts G, Mulier M, Van der Perre G, Jaecques S. Relation between subject-specific hip joint loading, stress distribution in the proximal femur and bone mineral density changes after total hip replacement. J Biomech. 2008;41:3405ā€“13. https://doi.org/10.1016/j.jbiomech.2008.09.011.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  7. Lenaerts G, De Groote F, Demeulenaere B, Mulier M, Van der Perre G, Spaepen A, et al. Subject-specific hip geometry affects predicted hip joint contact forces during gait. J Biomech. 2008;41:1243ā€“52. https://doi.org/10.1016/j.jbiomech.2008.01.014.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  8. Bah MT, Nair PB, Taylor M, Browne M. Efficient computational method for assessing the effects of implant positioning in cementless total hip replacements. J Biomech. 2011;44:1417ā€“22. https://doi.org/10.1016/j.jbiomech.2010.12.027.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  9. Vresilovic EJ, Hozack WJ, Rothman RH. Radiographic assessment of cementless femoral components. Correlation with intraoperative mechanical stability. J Arthroplasty. 1994;9:137ā€“41. https://doi.org/10.1016/0883-5403(94)90062-0.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  10. Anglin C, Masri BA, Tonetti J, Hodgson AJ, Greidanus NV. Hip resurfacing femoral neck fracture influenced by valgus placement. Clin Orthop Relat Res. 2007;465:71ā€“9. https://doi.org/10.1097/BLO.0b013e318137a13f.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  11. Davis ET, Olsen M, Zdero R, Waddell JP, Schemitsch EH. Femoral neck fracture following hip resurfacing: the effect of alignment of the femoral component. J Bone Joint Surg Br. 2008;90:1522ā€“7. https://doi.org/10.1302/0301-620X.90B11.20068.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  12. Nabavi A, Yeoh KM, Shidiac L, Appleyard R, Gillies RM, Turnbull A. Effects of positioning and notching of resurfaced femurs on femoral neck strength: a biomechanical test. J Orthop Surg (Hong Kong). 2009;17:47ā€“50. https://doi.org/10.1177/230949900901700111.

    ArticleĀ  CASĀ  Google ScholarĀ 

  13. Radcliffe IA, Taylor M. Investigation into the effect of varus-valgus orientation on load transfer in the resurfaced femoral head: a multi-femur finite element analysis. Clin Biomech (Bristol Avon). 2007;22:780ā€“6. https://doi.org/10.1016/j.clinbiomech.2007.03.011.

    ArticleĀ  CASĀ  Google ScholarĀ 

  14. Hayashi S, Hashimoto S, Matsumoto T, Takayama K, Nishida K, Ishida K, et al. Stem anteversion mismatch to the anatomical anteversion causes loss of periprosthetic bone density after THA. J Orthop Surg (Hong Kong). 2017;25:2309499017739478.

    ArticleĀ  Google ScholarĀ 

  15. Hasegawa Y, Iwata H, Mizuno M, Genda E, Sato S, Miura T. The natural course of osteoarthritis of the hip due to subluxation or acetabular dysplasia. Arch Orthop Trauma Surg. 1992;111:187ā€“91. https://doi.org/10.1007/BF00571474.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  16. Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P. A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am. 2009;91:2598ā€“604. https://doi.org/10.2106/JBJS.H.01225.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  17. Gruen TA, McNeice GM, Amstutz HC. ā€˜Modes of failureā€™ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17ā€“27. https://doi.org/10.1097/00003086-197906000-00002.

    ArticleĀ  Google ScholarĀ 

  18. Faul F, Erdfelder E, Buchner A, Lang AG. Statistical power analyses using G*Power 3.1: tests for correlation and regression analyses. Behav Res Methods. 2009;41:1149ā€“60. https://doi.org/10.3758/BRM.41.4.1149.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  19. Durlak JA. How to select, calculate, and interpret effect sizes. J Pediatr Psychol. 2009;34:917ā€“28. https://doi.org/10.1093/jpepsy/jsp004.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  20. Taniguchi N, Jinno T, Koga D, Hagino T, Okawa A, Haro H. Cementless hip stem anteversion in the dysplastic hip: a comparison of tapered wedge vs metaphyseal filling. J Arthroplasty. 2017;32:1547ā€“52. https://doi.org/10.1016/j.arth.2016.12.020.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  21. Hayashi S, Fujishiro T, Hashimoto S, Kanzaki N, Kuroda R, Kurosaka M. The contributing factors of tapered wedge stem alignment during mini-invasive total hip arthroplasty. J Orthop Surg Res. 2015;10:52. https://doi.org/10.1186/s13018-015-0192-x.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  22. Nakamura S, Minoda Y, Ohta Y, Sugama R, Yamamura K, Ueyama H, et al. Preoperative morphology of the proximal femoral canal did not affect the postoperative bone mineral density change around the Zweymuller-type stem. Orthopedics. 2019;42:e449ā€“53. https://doi.org/10.3928/01477447-20190604-04.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  23. Kim YH, Choi Y, Kim JS. Comparison of bone mineral density changes around short, metaphyseal-fitting, and conventional cementless anatomical femoral components. J Arthroplasty. 2011;26:931-940.e1. https://doi.org/10.1016/j.arth.2010.10.001.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

Download references

Acknowledgements

Not applicable

Funding

No funding was received for this study.

Author information

Authors and Affiliations

Authors

Contributions

SH participated in the study design, drafting of the manuscript, and data collection. YK carried out data collection and drafting of the manuscript. NN participated in the data collection and drafting of the manuscript. TM participated in the study design and helped to revise the manuscript. TK participated in the data collection and drafting of the manuscript. TM carried out data collection and drafting of the manuscript. TN participated in the data collection and drafting of the manuscript. RK participated in the study design and helped to revise the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Shinya Hayashi.

Ethics declarations

Ethics approval and consent to participate

The study protocol was approved by the Institutional Review Board on September 8, 2011 (No. 1220), and patients provided informed consent for participation.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hayashi, S., Kuroda, Y., Nakano, N. et al. Peri-prosthetic bone remodeling of hydroxyapatite-coated compaction short stem was not affected by stem alignment. J Orthop Surg Res 17, 131 (2022). https://doi.org/10.1186/s13018-022-03022-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-022-03022-7

Keywords