A comparison of EFECE systems with tension band wiring for patella fracture fixation in cadavers

Abstract

Background

EFECE systems are newly defined internal fixation systems, which are suitable for patella fracture fixation. The aim of this study was to compare the fixation strength of EFECE Systems with tension band wiring for transverse patellar fracture simulation on fresh frozen cadaver models.

Methods

Quadriceps tendon-patella-patellar tendon (QT-P-PT) complex was prepared from human cadavers. After simulation of a transverse patella fracture, in group 1, 5 patella were fixed with a pair of 1.2 mm EFECE wires and 4 EFECE devices. In group 2, 5 patella were fixed with a pair of 1.2 mm Kirschner wires (K-wire) and a cerclage wire according to the tension band technique.

Using a testing device with custom-made jaws, increasing distraction force was applied to these QT-P-PT complexes. Extension of these complexes with the distraction forces was observed. The maximum distraction force and the elongation during maximum force were evaluated.

Results

After 5 experiments with the EFECE systems, there was no EFECE wire breakage or EFECE wire-EFECE device catching failure. The median maximum force was 740 N (720-810 N). During maximum distraction force the median extension was 2.5 mm (1.6-2.5 mm). After 5 experiments with the tension band technique, there was no K-wire breakage. The median maximum force was 330 N (240-510 N). During this maximum distraction force the median extension was 3.4 mm (2.2-3.8 mm).

Conclusions

Based on the biomechanical advantages, patella fracture treatment with EFECE systems may constitute a reasonable alternative in the treatment of patella fractures.

Introduction

The patella is the largest sesamoid bone in the human body and fractures of this bone account for 1% of all skeletal system fractures [1]. Conservative treatment is indicated to fractures with minimal displacement and those with no articular surface incongruency. If fragment displacement is more than 2 mm or there is articular incongruency, surgical treatment is indicated [2]. For the surgical treatment of patella, several techniques and implant materials have been described [3].

The tension band technique, defined by Müller et al. in 1979, became the standard technique for patella fracture fixation [4]. In this open fixation technique, 2 parallel Kirschner wires (K-wire) and a cerclage wire compress the fracture line and cause compression of the articular surface with distraction forces.

EFECE systems are newly defined and patented, internal fixation implants, which also have the indication scale for patella fractures. EFECE systems comprise an EFECE device, EFECE wire and surgical tools. The EFECE device functions as a screw head on the EFECE wire to compress and fix the fracture line.

The aim of this study was to compare the conventional tension band technique with the new EFECE systems to determine whether the new EFECE systems could resist greater distraction force and cause less displacement during maximum force.

Material and method

EFECE systems comprise an EFECE device, EFECE wire and surgical tools. The EFECE device (EFECE, Akım Metal, 2015, Istanbul, Turkey) is cylinder shaped with a 6-mm radius and a 5-mm length that features a hole for the insertion of a 1.2-mm EFECE wire (EFECE, Akım Metal, 2015, Istanbul, Turkey). The EFECE device contains 2 threaded pieces that connect to each other. The top piece functions as a cap, whereas the second piece contains 3 gloves for the insertion of 3 balls. These balls are 1.5 mm in diameter. The locking mechanism is aided by the balls in the cone-shaped gloves. In forward movements, the balls move back to the base of the cone, and during pulling movements, the balls move to the narrow part of the cone and lock the EFECE wire [5] (Fig. 1). The surgical kit for EFECE systems (EFECE, Akım Metal, 2015, Istanbul, Turkey) is for compression of the fracture line, locking and unlocking the EFECE device.

Fig. 1
figure 1

The EFECE implant locking mechanism is applied with the help of balls in the cone-shaped gloves. In forward movement to compress the fractured fragments across the K-wire, the balls move back to the base of the cone and release the K-wire. When pulled back, the balls move to the narrow part of the cone and lock the K-wire

With the approval from our institution’s institutional review board (KUGOKAEK) numbered 2020/89, a total of 11 fresh cadavers (7 female and 4 male; age range, 55-94 years; mean age, 74.5 years) which were acquired from the Acıbadem Mehmet Ali Aydinlar University Anatomy Department were used in this study. None of the knees had undergone surgery or trauma to the extensor mechanism. The dissections were performed in Acıbadem Mehmet Ali Aydinlar University Atasehir, Istanbul, Turkey. The study was conducted according to the guiding principles of the Declaration of Helsinki.

A quadriceps tendon-patella-patellar tendon (QT-P-PT) complex was prepared from each cadaver. Transverse patellae fractures were created in the same localization in all the samples. In group 1, 5 patella were fixed with a pair of 1.2 mm EFECE wires and 4 EFECE devices. The EFECE devices were driven mutually on EFECE wires until contact was made with the bone cortex c. Then, these devices were compressed and tightened using the surgical tools.

In group 2, 5 patella were fixed with a pair of 1.2 mm K-wires and a cerclage wire according to the tension band technique.

Using a servo hydraulic testing machine with custom-made jaws (Instron Inc, Norwood, Massachusetts, USA), increasing distraction force (1 cm/min) was applied to these QT-P-PT complexes (Fig. 2). The maximum loads and the elongations during maximum load were observed and recorded with the Nexygen software (Fig. 3).

Fig. 2
figure 2

QT-P-PT complexes fixed with EFECE systems or tension band wiring were attached to the jaws of the material testing system servo hydraulic testing machine and increasing distraction force was applied to these systems

Fig. 3
figure 3

Evaluation of the maximum distraction force and the total extension during maximum force

All statistical analyses were performed using IBM SPSS for Windows version 20.0 (IBM Corp., Armonk, NY, USA). The Kolmogorov-Smirnov test was applied to test the normality of data distribution. Continuous variables were expressed as median values (25th-75th percentiles). Comparisons of continuous variables between the groups were performed using the Mann-Whitney U test. A two-sided value of p < 0.05 was considered statistically significant.

Results

After 5 experiments with EFECE system fixation, the median maximum load was 740 N (720-810 N). During these maximum loads, the median extension was 2.5 mm (1.6-2.5 mm).

After 5 experiments with tension band wiring, the median maximum load was 330 N (240-510 N). During these maximum loads, the median extension was 3.4 mm (2.2-3.8 mm).

The maximum load level of the EFECE system fixation was determined to be more than 2 times greater than that of tension band wiring (Table 1, P = 0.008). No statistically significant difference was determined between the EFECE system and tension band wiring in respect of extension (Table 1, P = 0.095).

Table 1 Maximum load resistance of the fixation and the extension with that maximum load

Discussion

Selecting the best surgical approach, appropriate method and materials for treating patellar fractures is a tough challenge for orthopedists [6,7,8,9,10,11]. The main goal with surgical treatment of patella fractures is to provide a congruent articular surface and maintain rigid fixation. Open reduction and tension band wiring technique widely accepted by the orthopaedic trauma surgeons because of mostly achieving these goals [12, 13].

Although open reduction and internal fixation techniques are the standard method for patellar fracture surgical treatment, there are many reports describing disadvantages and complications associated with traditional surgical treatments [10, 14]. Technique modifications such as cannulated screws, cannulated screws with tension band wire and plate fixation have been defined to enhance the surgical outcome [15,16,17,18]. And also for biomechanical comparison of these techniques, scoring methodology has been developed to assess the quality of biomechanical studies [19].

LeBrun et al. evaluated functional outcomes after surgically treatment of patella fractures [20]. In this study, with evaluation of validated outcome measures as well as objective physical findings it was concluded that functional deficits and significant symptomatic complaints persist after patella fracture operative treatment. These results have encouraged scientists to define “new and minimally invasive” implant technologies.

With the patented EFECE systems, fractures can be treated with an internal fixation technique. EFECE systems include an EFECE device, EFECE wire and surgical kit. The EFECE device is cylinder shaped with a hole for the insertion of the EFECE wire. The surgical kit for EFECE systems comprises a sleeve, working cannula, screwdrivers, wire tensioner, wire cutter and magnet. The surgical kit is designed for use in the percutaneous surgical technique. After reduction of the fracture and EFECE wire placement as in the conventional technique, using the sleeve, working cannula and screwdrivers, two reciprocal EFECE devices are pushed forward on EFECE wires until contact is made with the bone cortex. At this stage, there is no need to rotate the screwdriver on the EFECE wire and there is no need to measure the diameter of the implants, which are time-consuming procedures for screw insertion. Compression force can then be applied on the fracture fragments with adjustment of the EFECE wire tension using the wire tensioner. Once the implants are locked with the screwdrivers, these devices do not allow forward or backward movement on the EFECE wires. The remaining part of the wire should be cut with the percutaneous wire cutter.

As the balls in the locking mechanism are magnetically active, these implants can be removed with magnets. This implant removal technique is also an easy and new approach in orthopaedic surgery.

There are limitations of the EFECE systems. The EFECE system needs at least two reciprocal EFECE devices for wire fixation. The counter part of the EFECE wire that goes through the fragments and leaves the bone cortex needs to be prepared for EFECE insertion. The fixation strength of the EFECE system is dependent on the mechanical properties of a thin EFECE wire.

The findings of this study suggest that osteosynthesis of patella fractures with EFECE systems provides sufficient mechanical stability to prevent fracture separation with postoperative functional loading. Compared with tension band wiring, the EFECE systems demonstrated greater fixation strength and no difference in fracture gap distraction. The tension band wiring failed at significantly lower loads than the EFECE systems (P = 0.008).

In this study, the failure mechanism with EFECE systems was seen to be due to failure of the bone with sinking of the EFECE device into the patella. There were no failures in the EFECE device–EFECE wire interspace. The failure mechanism with tension band wiring was due to slippage of the cerclage wire and then sinking of the K-wire into the patella.

Conclusion

Patella fracture fixation with EFECE systems provides a biomechanical stable fixation. Compared with tension band wiring, EFECE systems fixation resulted in higher mechanical strength and reduced fragment dislocation under loading. In comminuted patella fractures, more than 2 EFECE systems in different alignments can be used to compress and fix the fractured fragments with the percutaneous technique. Based on these biomechanical findings, EFECE systems may constitute a reasonable alternative to tension band wiring in the treatment of patella fractures.

Availability of data and materials

All data generated or analysed during this study are included in this published article.

Abbreviations

QT-P-PT:

Quadriceps tendon-patella-patellar tendon

K-Wire:

Kirschner wire

N:

Newton

References

  1. Lotke PA, Ecker ML. Transverse fractures of the patella. Clin Orthop Relat Res. 1981;158:180–4.

    Google Scholar 

  2. Gosal HS, Singh P, Field RE. Clinical experience of patellar fracture fixation using metal wire or non-absorbable polyester: a study of 37 cases. Injury. 2001;32:129–35.

    Article  CAS  Google Scholar 

  3. Heusinkveld MHG, den Hamer A, Traa WA, Oomen PJA, Maffulli N. Treatment of transverse patellar fractures: a comparison between metallic and non-metallic implants. Br Med Bull. 2013;107:69–85.

    Article  Google Scholar 

  4. Muller ME, Allgöwer M, Schneider R, Willenegger H., (1979) Manual of internal fixation: techniques recommended by the AO Group. 2nd ed. Berlin, etc: Springer-Verlag,.

  5. Emre Karadeniz (2011, Oct 17). efece video 1, video, https://m.youtube.com/watch?v = AloRcFcae6g.

  6. Benjamin J, Bried J, Dohm M, McMurtry M. Biomechanical evaluation of various forms of fixation of transverse patellar fractures. J Orthop Trauma. 1987;1:219–22.

    Article  CAS  Google Scholar 

  7. Fortis AP, Milis Z, Kostopoulos V, Tsantzalis S, Kormas P, Tzinieris N, Boudouris T. Experimental investigation of the tension band in fractures of the patella. Injury. 2002;33:489–93.

    Article  CAS  Google Scholar 

  8. John J, Wagner WW, Kuiper JH. Tension-band wiring of transverse fractures of patella: the effect of site of wire twists and orientation of stainless steel wire loop: a biomechanical investigation. Int Orthop. 2007;31:703–7.

    Article  CAS  Google Scholar 

  9. Schnabel B, Scharf M, Schwieger K, Windolf M, Pol B, Braunstein V, Appelt A. Biomechanical comparison of a new staple technique with tension band wiring for transverse patella fractures. Clin Biomech (Bristol, Avon). 2009;24:855–9.

    Article  Google Scholar 

  10. Smith ST, Cramer KE, Karges DE, Watson JT, Moed BR. Early complications in the operative treatment of patella fractures. J Orthop Trauma. 1997;11:183–7.

    Article  CAS  Google Scholar 

  11. Wu CC, Tai CL, Chen WJ. Patellar tension band wiring: a revised technique. Arch Orthop Trauma Surg. 2001;121:12–6.

    Article  CAS  Google Scholar 

  12. Carpenter JE, Kasman R, Matthews LS. Fractures of the patella. Instr Course Lect. 1994;43:97–108.

    CAS  PubMed  Google Scholar 

  13. Kaylor KL. Injuries to the patella and extensor mechanism. In: Kasser JR, editor. Orthopaedic knowledge update 5, vol. 5. Rosemont, IL: AAOS; 1996. p. 153–8.

    Google Scholar 

  14. Carpenter JE, Kasman RA, Patel N, Lee ML, Goldstein SA. Biomechanical evaluation of current patella fracture fixation techniques. J Orthop Trauma. 1997;11:351–6.

    Article  CAS  Google Scholar 

  15. Tandogan RN, Demirors H, Tuncay CI, Cesur N, Hersekli M. Arthroscopic-assisted percutaneous screw fixation of select patellar fractures. Arthroscopy. 2002;18:156–62.

    Article  Google Scholar 

  16. Tian Y1, Zhou F, Ji H, Zhang Z, Guo Y. Cannulated screw and cable are superior to modified tension band in the treatment of transverse patella fractures. Clin Orthop Relat Res. 2011;469(12):3429–35.

    Article  Google Scholar 

  17. Cho JH. Percutaneous cannulated screws with tension band wiring technique in patella fractures. Knee Surg Relat Res. 2013;25(4):215–9.

    Article  Google Scholar 

  18. Wurm S, Bühren V, Augat P. Treating patella fractures with a locking patella plate - first clinical results. Injury. 2018;49(Suppl 1):51–5.

    Article  Google Scholar 

  19. Traa WA, Oomen PJA, den Hamer A, Heusinkveld MHG, Maffulli N. Biomechanical studies on transverse olecranon and patellar fractures: a systematic review with the development of a new scoring method. Br Med Bull. 2013;108:131–57.

    Article  Google Scholar 

  20. LeBrun CT, Langford JR, Sagi HC. Functional outcomes after operatively treated patella fractures. J Orthop Trauma. 2012;26(7):422–6.

    Article  Google Scholar 

Download references

Acknowledgements

We thank Canan Baydemir for assistance in the statistical analysis of our data and preparation of table for this article.

Funding

Emre Karadeniz received funding from Akım Metal for this research. Elif Nedret Keskinoz has no funding source.

Author information

Authors and Affiliations

Authors

Contributions

The author Emre Karadeniz is the patent owner, study designer and writer of this manuscript. The author Elif Nedret Keskinoz is the study designer, writer of this manuscript and also responsible for cadaver preparation. The authors read and approved the final manuscript.

Corresponding author

Correspondence to Emre Karadeniz.

Ethics declarations

Ethics approval and consent to participate

This article does not contain any studies with human participants or animals performed by any of the authors.

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Karadeniz, E., Keskinoz, E.N. A comparison of EFECE systems with tension band wiring for patella fracture fixation in cadavers. J Orthop Surg Res 15, 256 (2020). https://doi.org/10.1186/s13018-020-01781-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-020-01781-9

Keywords