Table 1 Maximum load resistance of the fixation and the extension with that maximum load

From: A comparison of EFECE systems with tension band wiring for patella fracture fixation in cadavers

 

EFECE (n = 5)

Tension band wiring (n = 5)

P

Median (25th-75th per.)

Median (25th-75th per.)

Load (N)

740.00 (720.00-790.00)

330.00 (265.00-425.00)

0.008

Extension (mm)

2.50 (2.00-2.50)

3.40 (2.60-3.75)