Injectable bone cement cannulated pedicle screw for lumbar degenerative disease in osteoporosis - clinical follow-up of over 5 years

Abstract

Objective

The aim of this study is to evaluate the clinical efficacy of injectable cemented hollow pedicle screw (CICPS) in the treatment of osteoporotic lumbar degenerative diseases through a large sample long-term follow-up study. Additionally, we aim to explore the risk factors affecting interbody fusion.

Methods

A total of 98 patients who underwent CICPS for transforaminal lumbar interbody fusion (TLIF) for osteoporotic lumbar degenerative disease from March 2011 to September 2017 were analyzed. X-ray and electronic computed tomography (CT) imaging data were collected during preoperative, postoperative, and follow-up periods. The data included changes in intervertebral space height (ΔH), screw failure, cement leakage (CL), and intervertebral fusion. The patients were divided into two groups based on their fusion status one year after surgery: satisfied group A and dissatisfied group B. Surgical data such as operation time, intraoperative bleeding volume and surgical complications were recorded, and visual analog scale (VAS) and Oswestry disability index (ODI) were used to evaluate the improvement of lumbar and leg pain.

Results

The mean follow-up time was 101.29 months (ranging from 70 to 128 months). A total of 320 CICPS were used, with 26 screws (8.13%) leaking, 3 screws (0.94%) experiencing cement augmentation failure, and 1 screw (0.31%) becoming loose and breaking. The remaining screws were not loose or pulled out. Female gender, decreased bone density, and CL were identified as risk factors affecting interbody fusion (P < 0.05). Early realization of interbody fusion can effectively prevent the loss of intervertebral space height (P < 0.05) and maintain the surgical treatment effect. Both VAS and ODI scores showed significant improvement during the follow-up period (P < 0.05). Binary logistic regression analysis revealed that decreased bone density and cement leakage were risk factors for prolonged interbody fusion.

Conclusions

The results of long-term follow-up indicate that PMMA enhanced CICPS has unique advantages in achieving good clinical efficacy in the treatment of osteoporosis lumbar degenerative diseases. Attention should be paid to identify female gender, severe osteoporosis, and CL as risk factors affecting interbody fusion.

Introduction

With the advancement of science and technology and the enhancement of living standards, people’s life expectancy is extended. The aging population has inevitably become a focal point of the world. The incidence of lumbar degenerative diseases in the elderly, including lumbar slip, lumbar disc herniation, and lumbar spinal stenosis, has increased year by year. The segment instability and corresponding nerve root compression of the back can lead to lower limb pain, affecting patients’ health and daily life. If conservative treatment is ineffective, surgery is often necessary to relieve pressure and alleviate symptoms.

Although the research of some scholars has provided a valuable tool for the treatment of emerging non-fusion techniques in recent years [1], the posterior internal fixed fusion is still the gold standard of treatment. Decompressive fusion surgery combined with pedicle screw fixation is a viable solution for spinal problems and can enhance stability and fusion rates [2]. However, osteoporosis (OP) can impact the trabecular bone in the vertebral body, leading to a decrease in bone density and thinning of the bone cortex. This can result in reduced screw holding force and fixation failure [3, 4]. To improve the strength of screws in osteoporotic vertebrae, many new methods have been reported, including extending the length and increasing the diameter of the screw, modifying the trajectory of the screw, using expandable pedicle screw and using cement to enhance the pedicle screw [5,6,7,8,9]. Biomechanical experiments show that the cement-bone interface is formed, which improves the holding force and stable line of the screw.

Polymethylmethacrylate (PMMA) possesses high stability, high curing strength, and rapid curing [10, 11], making it the most effective choice for enhancing pedicle screws to achieve stable fixation [12,13,14,15,16]. However, the disadvantages of high heat polymerization temperature, monomer toxicity, and lack of bone conductivity cannot be ignored during use [17]. In order to utilize this technique, we developed and designed CICPS, outlined the screw design characteristics in the literature [18], and implemented clinical treatment for osteoporotic degenerative spinal diseases in 2011.

The purpose of this study is to evaluate the clinical efficacy of CICPS and explore the risk factors affecting interbody fusion in a long-term clinical follow-up study of patients with osteoporotic lumbar degenerative disease treated with transforaminal lumbar interbody fusion (TLIF).

Materials and methods

Research design

Our study retrospectively analyzed 98 patients with osteoporotic spinal degeneration who received CICPS from March 2011 to September 2017. The inclusion criteria for the study were: 1. Preoperative diagnosis of osteoporosis (two-photon bone mineral density test: T-2.5 SD) is confirmed. The patient also presents with a clear TLIF surgical indication after conservative treatment.<3 surgical segments; 4. Follow-up over 5 years. Exclusion criteria: 1. preoperative coagulation abnormalities or severe cardiopulmonary disease; 2.preoperative local or systemic infection; 3. allergic to any endoplant; 4. Incomplete data during follow-up was a limitation of this study. This study was approved by the Ethics Committee of Southwest Hospital (KY2024037) and conducted in accordance with the ethical guidelines of the Army Military Medical University. All the patients participating in this study signed the informed consent form.

Surgical methods

The surgical area was exposed along the posterior median incision in the prone position of the patient under general anesthesia. CICPS with appropriate specifications - diameter 6.5–7.0 mm, length 40–45 mm, slightly larger than the traditional pedicle screw - was selected according to the preoperative plan.1.5 mL of cement was injected with each screw in the C-arm. If bone cement leakage was found during the injection, the injection was stopped immediately. After the cement injection, TLIF was performed using lever locking screws. According to our previous treatment strategy [19], unilateral cement augmentation is recommended for lumbar spondylolisthesis below II degrees, lumbar disc herniation, and lumbar spinal stenosis, while bilateral cement augmentation is recommended for lumbar spondylolisthesis above III degrees. Due to the need for opening and pressurizing the decompression side, it is preferred to strengthen the decompression side of bone cement.

Post-operative management

Postoperative drainage was placed and the drainage was monitored to be less than 50 ml per 24 h; intravenous antibiotics were administered to prevent infection within 3 days. It was recommended to start walking on the ground on the second day and continue until 3 months after surgery. Routine anti-osteoporosis therapy, including calcium, vitamin D, and phosphate supplementation, was also prescribed.

Observational parameters

Record the operation time, intraoperative bleeding volume, surgical complications, and other surgical data. The vertebral space height (H) was recorded by Choi JY [20] through the standing neutral lateral X-ray (Fig. 1a). The grade of graft fusion was determined by CT scan using the Christoph J. Siepe [21] evaluation system. Only grades I and II levels were defined as satisfactory fusion (Fig. 1b). A translucent shadow around the screw indicates that the screw is loose. Bone cement leakage (CL) is divided into 3 types according to the standards of Yeom [22], including type S (segmental vein leakage), type B (vertebral vein leakage), and type C (cortical defect leakage) (Fig. 2). Symptoms improvement was assessed using the Visual.

Fig. 1
figure 1

The figure shows the vertebral space height H= (H1 + H2) / 2 (A) and the satisfactory fusion achieved in the vertebral space (B)

Fig. 2
figure 2

Type S: segment vein leakage (A); Type C: Cortical defect leakage (B); Type B: vertebral vein leakage (C)

Statistical analysis

Data analysis was conducted using SPSS 25.0 software (IBM, USA). Continuous values are presented as mean ± standard deviation. The T-test is used when the data conforms to a normal distribution, while the rank sum test is used if the data do not meet normality. Pearson’s χ2 test, continuity correction, or Fisher’s exact test is used for counting data. Binary logistic regression was used to analyze the risk factors affecting interbody fusion. A p-value of < 0.05 was considered to indicate statistically significant differences.

Results

General results

This study included a total of 98 patients, with a mean age of 61.29 ± 7.30, consisting of 14 males and 84 females. The patients had a bone mineral density of -3.21 ± 0.61. Diagnoses among the patients included lumbar spondylolisthesis (64), lumbar disc herniation (26), and lumbar spinal stenosis (8) (Table 1).

Table 1 Basic patient information and surgery

According to the study protocol, 78 people were in Group A (satisfied) and 20 people were in Group B (dissatisfied). There were significant differences in gender and bone density between the two groups (P < 0.05), and no significant difference in age, operation time, and blood loss (P > 0.05) (Table 2).

Table 2 Basic information of the two groups

Radiographic evaluation

There was no significant difference in vertebral space height between the two groups during the study (P > 0.05, Table 3). The increase in vertebral space height until the last follow-up in group A (3.41 ± 2.11 mm) was more significant than the increase in group B (2.21 ± 1.19 mm) (P < 0.05, Table 4).

Table 3 Vertebral space height (mm)
Table 4 Δ H change in two groups (mm)

Clinical efficacy

Preoperative VAS and ODI scores in group A of 4.33 ± 1.35 and 36.61 ± 2.17 were higher than the last follow-up score. In group B, preoperative VAS and ODI scores of 3.6 ± 1.05 and 37.14 ± 2.17 were also higher than the last follow-up score, indicating a statistically significant improvement in pain symptoms (P < 0.05, Table 5).

Table 5 VAS, ODI (%) scores of two groups

Correlation study

There were no significant differences in decompression methods, surgical segment, or disease type (P > 0.05). However, there was a significant effect of cement leakage on fusion (P < 0.05) as shown in Table 6.

Table 6 χ2 test for surgical factors

Further analysis of bone mineral density and cement leakage revealed a significant difference in their impact on interbody fusion (P < 0.05) (Table 7).

Table 7 Presents the results of the logistic regression analysis on the fusion factors influencing the outcome

Complication

A total of 150 CISPS experienced CL in 98 patients: 22 patients (22.45%), 26 screws (8.13%), including 10 S (3.13%), 10 type C (3.13%), and 6 type B (1.87%). Three screws (0.94%) had bone cement reinforcement failure, and 1 screw (0.31%) was loose and fractured at 2 years of follow-up (Fig. 3). The remaining screws were not loose, pulled out, or broken. Four patients developed osteoporotic compression fractures about 5 years after surgery, and one patient gradually developed scoliosis deformity. Orthopcorrection was performed 10 years after surgery (Fig. 4). Two patients underwent debridement and suturing for superficial infections, while another patient was instructed to consciously take screws and perform internal fixation and removal (Fig. 5).

Fig. 3
figure 3

A-B: A 57-year-old female with L4/5-L5/S1 postoperative X-ray showed failure of left S1 screw cement augmentation and a 2-year screw fracture (red arrow). During the follow-up period, a 54-year-old woman with L5 spondylolisthesis postoperative underwent a lateral X-ray which showed failure of left S1 screw cement augmentation, with no screw loosening, extraction, or fracture (red arrow) present

Fig. 4
figure 4

Shows a 68-year-old woman with L4 anterior spondylolisthesis, Isthmus crack type, who underwent PVP for a T12 vertebral fracture. The postoperative X-ray showed T12 vertebral cement diffusion with no obvious leakage, complete fusion of the L4/5 vertebral space, and a highly satisfactory vertebral space. C-D: After 10 years, there was no loosening of the intraoperative L4 and L5 screws. The postoperative X-ray showed restored spine physiological curvature, no obvious cement leakage, and no significant changes in L4/5 vertebral space height

Fig. 5
figure 5

A-B: A 55-year-old male with L4/5-L5/S1 level intervertebral disc herniation. The posterior anterior-lateral X-ray revealed that the bone cement augmentation with the left S1 screw had failed. There were no signs of loosening, pulling out, or breakage of the screws during the follow-up (red arrows). C-D: Sinus tract present for 10 years, but the MRI examination showed no obvious signs of infection in the surgical area. There was no significant degeneration of the superior disc, only a local superficial soft tissue infection (red arrow). E-F: Debridement and suture + removal of internal fixation. No screw fracture occurred during the surgery. The lateral X-ray indicated intact removal of bilateral screws with no damage to the bone cement and good vertebral integrity (red arrow)

Discussion

The treatment of lumbar degenerative diseases aims to achieve stable fixation of the spinal column by providing adequate decompression to relieve symptoms of nerve root compression, promoting fusion with interbody bone grafting, and implanting pedicle screws [23]. In our study, all patients underwent TLIF surgery for spinal decompression, used a Polyether ether ester (PEEK) cage for graft fusion, and received PMMA-enhanced CICPS fixation. There were no significant differences in operation time, intraoperative bleeding, decompression method, surgical segment, or disease type among the patients (refer to Tables 2 and 4).

Realizing that good intervertebral fusion is based on the stability of the surgical segment, PMMA increased CICPS provides strong stability for three vertebrae in osteoporosis. However, the cytotoxicity of bone cement itself [24] has a certain impact on the surrounding local metabolism, and there is a risk of embolizing vertebral vessels, thus reducing the endplate blood supply and affecting intervertebral fusion [25]. In our study, we examined the amount of cement used and whether leakage occurred. The results showed no statistical difference in the amount of cement used on interbody fusion, but there was a significant negative correlation between CL and promoting intervertebral fusion (Table 6). The binary Logistic regression analysis also showed that CL is an independent risk factor affecting intervertebral fusion (Table 7). Although CL will not affect the final fusion in our study, as all patients achieved full level fusion in the surgical segment within 3 years postoperatively, it did increase the time required for satisfactory fusion in the surgical segment. A longer fusion time indicates less stability in the surgical section later on, which can have a negative impact on the treatment outcome. Therefore, we recommend avoiding the occurrence of CL as much as possible. This is not only to prevent potential local or systemic effects of leakage but also as a crucial step in improving the rate of interbody fusion and ensuring clinical efficacy.

Cage plays an important role in improving the intervertebral space height, promoting intervertebral body fusion, and restoring the stability of unstable segments [26]. However, the subsidence of cage is the most common postoperative complication, and its most direct effect is the decrease of vertebral space height, the weakening of the surgical effect, and the generation of new compression symptoms [27]. Choi et al. [20] conducted a study on the postoperative subsidence stage, evaluating the relationship between interbody fusion, symptom recurrence, and subsidence development through imaging data. They found no direct relationship between subsidence and symptom recurrence or imaging fusion. In contrast, Marchi et al. [27] demonstrated that cage subsidence could lead to early pain symptoms, which gradually improved with the stabilization of interbody fusion. This is similar to our study. While the height of the two groups was not different during follow-up (Table 3), the changes between the last follow-up and the postoperative vertebral space height were significantly different (Table 4). Additionally, the pain symptoms in Group A with less vertebral height loss showed more relief, and the decrease in VAS and ODI scores was also statistically significant (Table 5). Therefore, we believe that effectively preventing the subsidence of Cage should be the focus of attention. In future research development, the direction may be to find a rate that can effectively improve interbody fusion and avoid stress occlusion, similar to PEEK Cage.

Many studies have shown that decreased bone mass or osteoporosis, measured by DEXA or CT methods, may increase the risk of Cage subsidence [28,29,30,31,32,33]. The study by Chen et al. [34] noted that women and body mass index (BMI) are also risk factors for Cage subsidence. We also analyzed age and showed no significant difference in the age composition between the two groups (Table 2). Therefore, great attention should be paid to the treatment of women and patients with severe osteoporosis. Patients should be informed of the possible risks to prevent the occurrence of postoperative adverse events related to surgery.

A follow-up study of 79 patients who underwent spinal fusion surgery showed that settlement was no longer observed when intervertebral fusion occurred and new bone was able to withstand the load at the Cage-endplate interface [35]. In our study, the loss of vertebral space height at 1 year after surgery in group A was slightly higher than in group B, and the loss of vertebral space height at 3 years after surgery in group B was much higher than in group A (Table 4), with both comparisons being statistically significant. The reason for the analysis may be that good fusion closely connects the Cage surface to the endplate. The faster the early osteogenesis rate, the faster the decline of the intervertebral space. Prolonged vertebral fusion can lead to dissolution and absorption of the implanted bone. After osteolysis absorption, the stress between Cage and end plate increases. When the stress is too large, Cage subsidence occurs, causing excessive loss of intervertebral space height. The instability of the intervertebral space will further affect the time of intervertebral fusion. Therefore, it is important to improve the early fusion of the intervertebral space, prevent the loss of intervertebral space height, and maintain the clinical effect of surgery.

0.6–11% of the fusion failures are due to loosening [36,37,38,39] of pedicle screws. The study by Frankel BM et al. [40] showed a pedicle screw loosening rate of 4.1–12.9% in osteoporotic spinal degenerative disease. In our study, three S1 screws failed to provide reinforcement in 320 PMMA-enhanced CICPS. This failure may have been caused by the improper insertion of the screws into the S1 vertebral cortical bone during the operation and the blockage of cement from exiting the side hole. One screw loosened during the follow-up period, with a loosening rate of 0.31%, and ultimately broke in the second year of follow-up. This breakage may be attributed to the significant shear force generated by the anatomical structure of the S1 vertebral body (Fig. 3).

It has been reported in the literature that reinforced pedicle screws can improve the stability of internal fixation and achieve satisfactory fusion in spinal bones with poor bone mass. Fusion rates have been reported to range from 92.50 to 100% [36, 41,42,43]. In our study, the vertebral fusion rate at 1 year after surgery was 79.59%, mainly occurring in the Cage, indicating the importance of stability at the local fusion interface. By 3 years after surgery, the fusion rate reached 100%, with most of the bone graft around the Cage being dissolved and absorbed. This resulted in true full-level fusion over the three years, with no significant changes in the height of the intervertebral space once full-level fusion was achieved.

The results of Martin-Fernandez M et al. [44] showed that most postoperative infections were found in revision surgery or diabetes patients. Additionally, most infected patients could achieve satisfactory clinical results after conservative treatment. Early surgical intervention should be performed in patients with deep tissue infection. The results also showed that there was no significant correlation between the use of PMMA and the postoperative infection, and it did not need to be cleared during the operation. In our study, two patients developed superficial lumbar soft tissue infection and sinus tract formation during the seventh and tenth years of follow-up. There were no instances of screw area or deep soft tissue infection, and satisfactory results were achieved after debridement (Fig. 5).

Many studies have shown that pedicle screws augmented with PMMA can be safely and completely removed during surgery, but more torque is required to unscrew the screw. Throughout the removal procedure, the vertebral cement will remain intact, and the vertebral body itself will not sustain any damage [15,16,17,18,19,20,21,22,23,24,25,26,27,28,29,30,31,32,33,34,35,36,37,38,39,40,41,42,43,44,45,46,47]. In our study, one patient wanted to obtain internal fixation for 10 years after surgery, during hospital treatment. No screw loosening was seen during the removal of the screw along the long axis of the screw, and the cement and vertebral body were not damaged. Despite fewer revision procedures, it is reasonable to believe that PMMA-enhanced CICPS are safe and feasible in revision surgery (Fig. 5).

Many researchers have suggested that the use of PMMA may contribute to the occurrence of new vertebral fractures in adjacent segments [48,49,50]. However, a study by Aquarius R et al. [51] showed that PMMA itself does not lead to new fractures in adjacent vertebrae. In our study, we observed four patients who experienced vertebral fractures 4–8 years postoperatively. The new vertebral fractures were mainly concentrated in the thoracic and lumbar areas, while our surgical focus was primarily on the lumbosacral area. Combined with our 3–5 years of anti-osteoporosis treatment, we believe that this phenomenon is caused by the worsening of osteoporosis. The thoracic and lumbar areas are known stress concentration areas, leading to the majority of new fractures occurring in this region. Therefore, we recommend the continuous use of anti-osteoporosis drugs throughout the treatment process.

This study also has certain limitations. Firstly, it is a retrospective study with a long-term follow-up, spanning a wide time period, involving a large number of patients, and requiring difficult return visits. These factors may impact the accuracy of the study results. Secondly, it is a single-center study that lacks the credibility of further verification of the results of the control group. Thirdly, a prospective study is necessary to further confirm the observed differences.

Conclusions

Promoting vertebral fusion and preventing the loss of vertebral space height is a diverse and complex process. In women, severe osteoporosis is a risk factor that can affect vertebral fusion. It is important to identify this risk factor and try to avoid the occurrence of compression fractures to minimize the impact on vertebral fusion. Long-term follow-up results have shown that using PMMA enhanced CICPC in the treatment of osteoporosis-related lumbar degenerative disease can achieve good clinical outcomes and has unique advantages.

Data availability

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CICPS bone cement:

injectable cannulated pedicle screws

CL:

cement leakage

CT:

computed tomography

PMMA:

polymethylmethacrylate

TLIF:

transforaminal approach to lumbar interbody fusion

References

  1. Khalaf K, Nikkhoo M. Comparative biomechanical analysis of rigid vs. flexible fixation devices for the lumbar spine: a geometrically patient-specific poroelastic finite element study. Comput Methods Programs Biomed. 2021;212:106481–92.

    Article  PubMed  Google Scholar 

  2. Lonstein JE, Denis F, Perra JH, Pinto MR, Smith MD, Winter RB. Complications associated with pedicle screws. J Bone Joint Surg Am. 1999;81:1519–28.

    Article  PubMed  Google Scholar 

  3. Di Silvestre M, Parisini P, Lolli F, Bakaloudis G. Complications of thoracic pedicle screws in scoliosis treatment. Spine (Phila Pa 1976). 2007;32:1655–61.

    Article  PubMed  Google Scholar 

  4. Reitman CA, Nguyen L, Fogel GR. Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine. J Spinal Disord Tech. 2004;17:306–11.

    Article  PubMed  Google Scholar 

  5. Polly DW Jr, Orchowski JR, Ellenbogen RG. Revision pedicle screws. Bigger, longer shims–what is best. Spine (Phila Pa 1976). 1998;23:1374–9.

    Article  PubMed  Google Scholar 

  6. Kuhns CA, Reiter M, Pfeiffer F, Choma TJ. Surgical strategies to improve fixation in the osteoporotic spine: the effects of tapping, cement augmentation, and screw trajectory. Global Spine J. 2014;4:47–54.

    Article  PubMed  Google Scholar 

  7. Pfeiffer FM, Abernathie DL, Smith DE. A comparison of pullout strength for pedicle screws of different designs: a study using tapped and untapped pilot holes. Spine (Phila Pa 1976). 2006;31:E867–70.

    Article  PubMed  Google Scholar 

  8. Cook SD, Salkeld SL, Stanley T, Faciane A, Miller SD. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4:402–8.

    Article  PubMed  Google Scholar 

  9. Cameron HU, Jacob R, Macnab I, Pilliar RM. Use of polymethylmethacrylate to enhance screw fixation in bone. J Bone Joint Surg Am. 1975;57:655–6.

    Article  PubMed  Google Scholar 

  10. Yi S, Rim DC, Park SW, Murovic JA, Lim J, Park J. Biomechanical comparisons of pull out strengths after Pedicle Screw Augmentation with Hydroxyapatite, Calcium phosphate, or polymethylmethacrylate in the cadaveric spine. World Neurosurg. 2015;83:976–81.

    Article  PubMed  Google Scholar 

  11. Kiyak G, Balikci T, Heydar AM, Bezer M. Comparison of the pullout strength of different pedicle screw designs and augmentation techniques in an osteoporotic bone model. Asian Spine J. 2018;12:3–11.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Elder BD, Lo SF, Holmes C, Goodwin CR, Kosztowski TA, Lina IA, Locke JE, Witham TF. The biomechanics of pedicle screw augmentation with cement. Spine J. 2015;15:1432–45.

    Article  PubMed  Google Scholar 

  13. Bostelmann R, Keiler A, Steiger HJ, Scholz A, Cornelius JF, Schmoelz W. Effect of augmentation techniques on the failure of pedicle screws under cranio-caudal cyclic loading. Eur Spine J. 2017;26:181–8.

    Article  PubMed  Google Scholar 

  14. Hamasaki T, Tanaka N, Kim J, Okada M, Ochi M, Hutton WC. Pedicle screw augmentation with polyethylene tape: a biomechanical study in the osteoporotic thoracolumbar spine. J Spinal Disord Tech. 2010;23:127–32.

    Article  PubMed  Google Scholar 

  15. Chao KH, Lai YS, Chen WC, Chang CM, McClean CJ, Fan CY, Chang CH, Lin LC, Cheng CK. Biomechanical analysis of different types of pedicle screw augmentation: a cadaveric and synthetic bone sample study of instrumented vertebral specimens. Med Eng Phys. 2013;35:1506–12.

    Article  PubMed  Google Scholar 

  16. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, Redl H. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17:1462–9.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Lieberman IH, Togawa D, Kayanja MM. Vertebroplasty and kyphoplasty: filler materials. Spine J. 2005;5:S305–16.

    Article  Google Scholar 

  18. Liu Y, Xu J, Sun D, Luo F, Zhang Z, Dai F. Biomechanical and finite element analyses of bone cement-injectable cannulated pedicle screw fixation in osteoporotic bone. J Biomed Mater Res B Appl Biomater. 2016;104:960–7.

    Article  PubMed  Google Scholar 

  19. Dai F, Liu Y, Zhang F, Sun D, Luo F, Zhang Z, Xu J. Surgical treatment of the osteoporotic spine with bone cement-injectable cannulated pedicle screw fixation: technical description and preliminary application in 43 patients. Clin (Sao Paulo). 2015;70:114–9.

    Article  Google Scholar 

  20. Choi JY, Sung KH. Subsidence after anterior lumbar interbody fusion using paired stand-alone rectangular cages. Eur Spine J. 2006;15:16–22.

    Article  PubMed  Google Scholar 

  21. Siepe CJ, Stosch-Wiechert K, Heider F, Amnajtrakul P, Krenauer A, Hitzl W, Szeimies U, Stäbler A, Mayer HM. Anterior stand-alone fusion revisited: a prospective clinical, X-ray and CT investigation. Eur Spine J. 2015;24:838–51.

    Article  PubMed  Google Scholar 

  22. Yeom JS, Kim WJ, Choy WS, Lee CK, Chang BS, Kang JW. Leakage of cement in percutaneous transpedicular vertebroplasty for painful osteoporotic compression fractures. J Bone Joint Surg Br. 2003;85:83–9.

    Article  PubMed  Google Scholar 

  23. Park SJ, Lee CS, Chung SS, Kang SS, Park HJ, Kim SH. The Ideal cage position for achieving both indirect neural decompression and Segmental Angle Restoration in lateral lumbar Interbody Fusion (LLIF). Clin Spine Surg. 2017;30:E784–784790.

    Article  PubMed  Google Scholar 

  24. Stürup J, Nimb L, Kramhøft M, Jensen JS. Effects of polymerization heat and monomers from acrylic cement on canine bone. Acta Orthop Scand. 1994;65:20–3.

    Article  PubMed  Google Scholar 

  25. van der Werf M, Lezuo P, Maissen O, van Donkelaar CC, Ito K. Inhibition of vertebral endplate perfusion results in decreased intervertebral disc intranuclear diffusive transport. J Anat. 2007;211:769–74.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Chen L, Yang H, Tang T. Cage migration in spondylolisthesis treated with posterior lumbar interbody fusion using BAK cages. Spine (Phila Pa 1976). 2005;30:2171–5.

    Article  PubMed  Google Scholar 

  27. Marchi L, Abdala N, Oliveira L, Amaral R, Coutinho E, Pimenta L. Radiographic and clinical evaluation of cage subsidence after stand-alone lateral interbody fusion. J Neurosurg Spine. 2013;19:110–8.

    Article  PubMed  Google Scholar 

  28. Agarwal N, Faramand A, Alan N, Tempel ZJ, Hamilton DK, Okonkwo DO, Kanter AS. Lateral lumbar interbody fusion in the elderly: a 10-year experience. J Neurosurg Spine. 2018;29:525–9.

    Article  PubMed  Google Scholar 

  29. Jung JM, Chung CK, Kim CH, Yang SH. Clinical and radiologic outcomes of single-level direct lateral lumbar interbody fusion in patients with osteopenia. J Clin Neurosci. 2019;64:180–6.

    Article  PubMed  Google Scholar 

  30. Okano I, Jones C, Salzmann SN, Reisener MJ, Sax OC, Rentenberger C, Shue J, Carrino JA, Sama AA, Cammisa FP, Girardi FP, Hughes AP. Endplate volumetric bone mineral density measured by quantitative computed tomography as a novel predictive measure of severe cage subsidence after standalone lateral lumbar fusion. Eur Spine J. 2020;29:1131–40.

    Article  PubMed  Google Scholar 

  31. Samtani RG, Bernatz JT, Harrison R, Roy S, Gupta S, O’Brien JR. The Effect of Alendronate on Subsidence after lateral transpsoas Interbody Fusion: a preliminary Report. Int J Spine Surg. 2019;13:289–95.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Satake K, Kanemura T, Yamaguchi H, Segi N, Ouchida J. Predisposing factors for Intraoperative Endplate Injury of Extreme lateral Interbody Fusion. Asian Spine J. 2016;10:907–14.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Tempel ZJ, Gandhoke GS, Okonkwo DO, Kanter AS. Impaired bone mineral density as a predictor of graft subsidence following minimally invasive transpsoas lateral lumbar interbody fusion. Eur Spine J. 2015;24(Suppl 3):414–9.

    Article  PubMed  Google Scholar 

  34. Chen E, Xu J, Yang S, Zhang Q, Yi H, Liang D, Lan S, Duan M, Wu Z. Cage Subsidence and Fusion Rate in Extreme lateral Interbody Fusion with and without fixation. World Neurosurg. 2019;122:e969–969977.

    Article  PubMed  Google Scholar 

  35. Lee YS, Kim YB, Park SW. Risk factors for postoperative subsidence of single-level anterior cervical discectomy and fusion: the significance of the preoperative cervical alignment. Spine (Phila Pa 1976). 2014;39:1280–7.

    Article  PubMed  Google Scholar 

  36. Krag MH, Johnson C. Repair of failed transpedicle screw fixation. A biomechanPfeifer BAical study comparing polymethylmethacrylate, milled bone, and matchstick bone reconstruction. Spine (Phila Pa 1976). 1994;19:350–3.

    Article  PubMed  Google Scholar 

  37. Esses SI, Sachs BL, Dreyzin V. Complications associated with the technique of pedicle screw fixation. A selected survey of ABS members. Spine (Phila Pa 1976). 1993;18:2231–8. discussion 2238-9.

    Article  PubMed  Google Scholar 

  38. Okuyama K, Abe E, Suzuki T, Tamura Y, Chiba M, Sato K. Can insertional torque predict screw loosening and related failures? An in vivo study of pedicle screw fixation augmenting posterior lumbar interbody fusion. Spine (Phila Pa 1976). 2000;25:858–64.

    Article  PubMed  Google Scholar 

  39. Wittenberg RH, Shea M, Swartz DE, Lee KS, White AA 3rd, Hayes WC. Importance of bone mineral density in instrumented spine fusions. Spine (Phila Pa 1976). 1991;16:647–52.

    Article  PubMed  Google Scholar 

  40. Frankel BM, Jones T, Wang C. Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery. 2007;61:531–7. discussion 537-8.

    Article  PubMed  Google Scholar 

  41. Amendola L, Gasbarrini A, Fosco M, Simoes CE, Terzi S, De Iure F, Boriani S. Fenestrated pedicle screws for cement-augmented purchase in patients with bone softening: a review of 21 cases. J Orthop Traumatol. 2011;12:193–9.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Wittenberg RH, Lee KS, Shea M, White AA 3rd, Hayes WC. Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res. 1993:278–87.

  43. Paré PE, Chappuis JL, Rampersaud R, Agarwala AO, Perra JH, Erkan S, Wu C. Biomechanical evaluation of a novel fenestrated pedicle screw augmented with bone cement in osteoporotic spines. Spine (Phila Pa 1976). 2011;36:E1210–4.

    Article  PubMed  Google Scholar 

  44. Martín-Fernández M, López-Herradón A, Piñera AR, Tomé-Bermejo F, Duart JM, Vlad MD, Rodríguez-Arguisjuela MG, Alvarez-Galovich L. Potential risks of using cement-augmented screws for spinal fusion in patients with low bone quality. Spine J. 2017;17:1192–9.

    Article  PubMed  Google Scholar 

  45. Cho W, Wu C, Zheng X, Erkan S, Suratwala SJ, Mehbod AA, Transfeldt EE. Is it safe to back out pedicle screws after augmentation with polymethyl methacrylate or calcium phosphate cement? A biomechanical study. J Spinal Disord Tech. 2011;24:276–9.

    Article  PubMed  Google Scholar 

  46. Bullmann V, Schmoelz W, Richter M, Grathwohl C, Schulte TL. Revision of cannulated and perforated cement-augmented pedicle screws: a biomechanical study in human cadavers. Spine (Phila Pa 1976). 2010;35:E932–9.

    Article  PubMed  Google Scholar 

  47. Blattert TR, Glasmacher S, Riesner HJ, Josten C. Revision characteristics of cement-augmented, cannulated-fenestrated pedicle screws in the osteoporotic vertebral body: a biomechanical in vitro investigation. Technical note. J Neurosurg Spine. 2009;11:23–7.

    Article  PubMed  Google Scholar 

  48. Boger A, Heini P, Windolf M, Schneider E. Adjacent vertebral failure after vertebroplasty: a biomechanical study of low-modulus PMMA cement. Eur Spine J. 2007;16:2118–25.

    Article  PubMed  PubMed Central  Google Scholar 

  49. Belkoff SM, Mathis JM, Erbe EM, Fenton DC. Biomechanical evaluation of a new bone cement for use in vertebroplasty. Spine (Phila Pa 1976). 2000;25:1061–4.

    Article  PubMed  Google Scholar 

  50. Kinzl M, Benneker LM, Boger A, Zysset PK, Pahr DH. The effect of standard and low-modulus cement augmentation on the stiffness, strength, and endplate pressure distribution in vertebroplasty. Eur Spine J. 2012;21:920–9.

    Article  PubMed  Google Scholar 

  51. Aquarius R, van der Zijden AM, Homminga J, Verdonschot N, Tanck E. Does bone cement in percutaneous vertebroplasty act as a stress riser. Spine (Phila Pa 1976). 2013;38:2092–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This study was supported by the New Clinical Technology of the Army Medical University (No. 2015–2 A-1-5 and 2019-3-1-34).

Author information

Authors and Affiliations

Authors

Contributions

Conception and Design: F.D. Acquisition of Data: C.L., R.Z., X.Y. Analysis and Interpretation of Data: C.L., Y.J., J.X. Drafting the Article: C.L. Critically Revising the Article: L.S. Reviewed submitted version of manuscript: L.S. Approved the final version of the manuscript on behalf of all authors: F.D. Statistical analysis: W.W., Y.Z.

Corresponding authors

Correspondence to Lei Song or Fei Dai.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Ethics Committee of Southwest Hospital (KY2024037) and conducted in accordance with the ethical guidelines of the Army Military Medical University. All patients participating in this study signed the informed consent form.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it.The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Cc., Yang, Xl., Wu, Ww. et al. Injectable bone cement cannulated pedicle screw for lumbar degenerative disease in osteoporosis - clinical follow-up of over 5 years. J Orthop Surg Res 19, 440 (2024). https://doi.org/10.1186/s13018-024-04934-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-024-04934-2

Keywords