Biomechanical comparison of different fixation methods in tibiotalocalcaneal arthrodesis: a cadaver study

Abstract

Background

Various fixation methods are available for tibiotalocalcaneal arthrodesis: nail, plate, or screws. An intramedullary bone stabilization system within a balloon catheter has not previously been used in tibiotalocalcaneal arthrodesis. The aim of this study was to compare the stability of these techniques.

Methods

Twenty-four lower legs from fresh-frozen human cadavers were used. Tibiotalocalcaneal arthrodesis was performed with a retrograde nail, a lateral locking plate, three cancellous screws, or an intramedullary bone stabilization system. The ankles were loaded cyclically in plantarflexion and dorsiflexion.

Results

For cyclic loading at 125 N, the mean range of motion was 1.7 mm for nail, 2.2 mm for plate, 6.0 mm for screws, and 9.0 mm for the bone stabilization system (P < .01). For cyclic loading at 250 N, the mean range of motion was 4.4 mm for nail, 7.5 mm for plate, 12.1 mm for screws, and 14.6 mm for the bone stabilization system (P < .01). The mean cycle of failure was 4191 for nail, 3553 for plate, 3725 for screws, and 2132 for the bone stabilization system (P = .10).

Conclusions

The stability of the tibiotalocalcaneal arthrodesis differs depending on the fixation method, with nail or plate showing the greatest stability and the bone stabilization system the least. When three screws are used for tibiotalocalcaneal arthrodesis, the stability is intermediate. As the biomechanical stability of the bone stabilization system is low, it cannot be recommended for tibiotalocalcaneal arthrodesis.

Background

Tibiotalocalcaneal arthrodesis is used for operative treatment of severe degenerative and traumatic pathologies of the ankle and hindfoot. Implants commonly used for tibiotalocalcaneal arthrodesis include nail, plate, and screws [1,2,3,4].

An intramedullary bone stabilization system based on a light-curable monomer inside a balloon catheter was developed specifically for osteoporotic bone. Initial studies investigating the intramedullary bone stabilization system are available, but it has not yet been used for tibiotalocalcaneal arthrodesis [5,6,7,8].

Previous biomechanical studies have compared a retrograde nail with different plates (lateral blade plate, lateral locking plate, lateral locking plate with an augmentation screw, posterolateral locking plate) for tibiotalocalcaneal arthrodesis in fresh-frozen cadavers [9,10,11,12,13]. Some of these studies reported greater stability for a plate, while other studies did not observe any difference in stability between nail and plate for tibiotalocalcaneal arthrodesis [9,10,11,12,13]. A few published biomechanical studies have compared a retrograde nail with different screw constructs for tibiotalocalcaneal arthrodesis [14,15,16]. In fresh-frozen cadavers, Berend et al. observed significantly greater stiffness in the tibiotalocalcaneal arthrodesis with a retrograde nail in comparison with two cancellous screws [14]. In Sawbone studies, similar stability in the tibiotalocalcaneal arthrodesis has been reported with an uncompressed nail and a three-screw construct [15, 16].

The aim of the present biomechanical study was to determine whether there are any differences in stability between a retrograde nail, a lateral locking plate, three cancellous screws, or an intramedullary bone stabilization system for tibiotalocalcaneal arthrodesis. It was hypothesized that there would be no differences in stability between these techniques. The primary aim was to compare the range of motion during cyclic loading. A secondary aim was to compare the cycles of failure.

Mehods

The study was conducted with the approval of the local ethics committee (ref. no. 15–364). Fourteen matched pairs of fresh-frozen human cadaver lower legs from consenting informed donors were used for the study (Table 1). The bone mineral density in each specimen was measured in the posterior third of the calcaneus using dual X-ray absorptiometry (DEXA; Lunar iDXA, GE Healthcare, Chicago, USA) [17, 18]. Radiographs of the specimens in anteroposterior and lateral projections were taken in order to exclude any osseous pathology. Four of the 28 specimens had to be excluded owing to osseous pathologies (Table 1).

Table 1 Study population

The tibia and fibula were shortened to a length of 30 cm measured from the ankle joint [19, 20]. Disarticulation of the Chopart joint was performed [19, 20]. Soft tissues were removed except for the membrana interossea, the syndesmosis, and the ankle ligaments [19, 20]. The joint surfaces were left intact [19, 20]. The specimens were wrapped in saline-soaked compresses and stored at − 20 °C between experiments. For the experiments, the specimens were thawed at room temperature for 12 h.

The 24 specimens were assigned equally to one of four groups (n = 6 for each group): Nail, plate, screws, or bone stabilization system (Table 1). Tibiotalocalcaneal arthrodesis was carried out by the second author. In the nail group, the medullary cavity of the tibia was reamed up to a diameter of 11 mm via the calcaneus and talus. An intramedullary nail (Trigen Hindfoot Fusion Nail, 10 × 200 mm, Smith and Nephew, Memphis, Tennessee, USA) was inserted retrogradely and locked statically via the drill guide using four 5.0-mm locking screws. In the plate group, the distal 10 cm of the fibula was resected. A locking plate (Peri-Loc Ankle Fusion Large Fragment System, length 120 mm, Smith & Nephew, Memphis, Tennessee, USA) was positioned laterally and fixed with one 4.5-mm cortical screw and eight 4.5-mm locking screws in neutral mode. In the screws group, three 6.5-mm partially threaded cancellous screws (DePuy Synthes, Zuchwil, Switzerland; steel) were used for tibiotalocalcaneal arthrodesis. In accordance with the method described by Mückley et al. in Sawbones, two screws were inserted in parallel from the ventral tibia to the dorsal calcaneus, and the third screw in between from the ventral calcaneus to the dorsal tibia [16]. In the bone stabilization system group, the medullary cavity of the tibia was reamed up to a diameter of 11 mm via the calcaneus and talus, as in the nail group. Using the drill guide from the nail group and a 4.0-mm drill bit, four holes were drilled at the same positions as in the nail group. After cleaning of the medullary cavity with a brush, the bone stabilization system (IlluminOss Medical, East Providence, Rhode Island, USA; size 13 × 180 mm; www.illuminoss.com) was positioned in the medullary cavity. The photodynamic liquid monomer was infused into the balloon and cured using the light system for 600 s. After drilling of the previously created four holes through the cured monomer, the bone stabilization system was locked with four 5.0-mm locking screws, as also used in the nail group. In all of the groups, radiographs were taken in anteroposterior and lateral projections to verify correct implant positioning (Fig. 1).

Fig. 1
figure 1

Radiographs in anteroposterior and lateral projections after tibiotalocalcaneal arthrodesis with nail, plate, screws, or the bone stabilization system

The experimental set-up for mechanical testing is based on the method described by Richter et al. and Evers et al. [19, 20]. The base of the calcaneus and the proximal 15 cm of the remaining tibia and fibula were embedded in cold curing resin (Technovit® 4004, Heraeus Kulzer GmbH, Wehrheim, Germany) [19,20,21]. Each specimen was tested in plantarflexion and dorsiflexion, with load application via a lever of 8 cm in length with a materials testing machine (Z010, Zwick Roell, Ulm, Germany) (Fig. 2) [19, 21, 22]. The preload was 5 N. The specimens were loaded at a displacement rate of 10 mm/s for 1500 cycles from − 125 N to 125 N and subsequently for 3500 cycles from − 250 N to 250 N [19, 20]. Every 250 cycles, radiographs were taken in anteroposterior and lateral projections to verify correct implant positioning [19, 20]. Failure was defined as displacement (plantarflexion or dorsiflexion) of 10 mm [19]. The mode of failure was noted.

Fig. 2
figure 2

The experimental set-up for mechanical testing

Force displacement curves were analyzed in Microsoft Excel 2007 (Microsoft Corporation, Redmond, Washington, USA). Statistical analysis was carried out using IBM SPSS Statistics for Windows, version 22.0 (IBM Corporation, Armonk, New York, USA). The Shapiro–Wilk test was used to check the normal distribution of data. The results for the four groups were compared using analysis of variance (ANOVA) for normally distributed data and the Kruskal–Wallis test for nonnormally distributed data. The correlation between bone mineral density and cycle of failure was evaluated using Pearson correlation. Statistical significance was set at the 5% level (P ≤ 0.05).

Results

Table 2 shows the results relative to bone mineral density and mechanical testing. The mean bone mineral density did not differ significantly between the four groups, at 0.64 g/cm2 (range 0.49–0.77 g/cm2) in the nail group, 0.56 g/cm2 (range 0.38–0.78 g/cm2) in the plate group, 0.58 g/cm2 (range 0.41–0.78 g/cm2) in the screws group, and 0.64 g/cm2 (range 0.49–0.83 g/cm2) in the bone stabilization system group (P = 0.61).

Table 2 Results (mean and range) relative to bone mineral density and mechanical testing

The mean range of motion during 125 N loading differed significantly between the four groups, at 1.7 mm (range 1.1–2.4 mm) for nail, 2.2 mm (range 1.2–3.3 mm) for plate, 6.0 mm (range 3.4–14.1 mm) for screws, and 9.0 mm (range 5.2–11.5 mm) for the bone stabilization system (P < 0.01). There was a significantly higher range of motion during 125 N loading in the bone stabilization system group in comparison with the nail group and the plate group (P < 0.01). Figure 3 shows the medians and quartiles for the range of motion during 125 N loading in the four groups.

Fig. 3
figure 3

Box plot for the range of motion during 125 N loading, showing medians and quartiles (nail group n = 5; other groups n = 6 per group)

The mean range of motion during 250 N loading differed significantly between three of the groups, at 4.4 mm (range 2.0–6.1 mm) for nail, 7.5 mm (range 2.8–12.0 mm) for plate, and 12.1 mm (range 8.3–15.8 mm) for screws (P < 0.01). There was a significantly higher range of motion during 250 N loading in the screws group in comparison with the nail group (P < 0.01). In the bone stabilization system group, the mean range of motion during 250 N loading was 14.6 mm (range 13.1–16.0 mm). Due to prior failure, it could only be calculated for two specimens in the bone stabilization system group. The bone stabilization system group was therefore not included in the statistical analysis of 250 N loading. Figure 4 shows the medians and quartiles for the range of motion during 250 N loading in the four groups.

Fig. 4
figure 4

Box plot for the range of motion during 250 N loading, showing medians and quartiles (bone stabilization system group n = 2; other groups n = 5 per group)

The mean cycle of failure did not differ significantly between the four groups, at 4191 (range 146–5000) for nail, 3553 (range 1674–5000) for plate, 3725 (range 501–5000) for screws, and 2132 (range 520–4501) for the bone stabilization system (P = 0.10). Specimens failed as a result of implant loosening. In the bone stabilization system group, the screws broke out of the cured monomer. Table 3 shows the correlation between bone mineral density and cycle of failure. The Pearson correlation coefficient was 0.55 for all groups (P < 0.01) and 0.85 in the screws group (P = 0.03). The other groups did not show any significant correlation between bone mineral density and cycle of failure.

Table 3 Correlation between bone mineral density and cycle of failure

Discussion

This cadaver study compared the biomechanical properties after tibiotalocalcaneal arthrodesis with a retrograde nail, a lateral locking plate, three cancellous screws, and an intramedullary bone stabilization system based on a light-curable monomer inside a balloon catheter.

The present results show that a high degree of stiffness in the tibiotalocalcaneal arthrodesis can be achieved with a retrograde nail or a lateral locking plate. Ohlson et al. did not observe any differences in the rigidity of the tibiotalocalcaneal arthrodesis in fresh-frozen cadavers in comparison with a locking plate with a nail [12]. The locking plate investigated by Ohlson et al. used more screws for fixation in the talus and the calcaneus compared with the locking plate investigated in the present study [12]. In fresh-frozen cadavers, O’Neill et al. noted greater rigidity in the tibiotalocalcaneal arthrodesis when using a locking plate with an additional augmentation screw in comparison with a nail [11]. This result can be explained by the additional augmentation screw from the talus to the tibia used by O’Neill et al. [11]. In fresh-frozen cadavers, Gutteck et al. studied a posterolateral plate for tibiotalocalcaneal arthrodesis and detected greater stiffness in comparison with a nail [13]. The reason for this result might be the posterolateral position of the plate investigated by Gutteck et al. [13]. In the present study, a low level of stiffness in the tibiotalocalcaneal arthrodesis was observed with the bone stabilization system. This was the first use of the bone stabilization system for tibiotalocalcaneal arthrodesis. Biomechanical data for comparison are therefore not available. Nevertheless, the characteristics of the intramedullary bone stabilization system, which include infusion of a photodynamic liquid monomer into a balloon and curing by light, probably did not achieve the stability needed for tibiotalocalcaneal arthrodesis. The stiffness of a tibiotalocalcaneal arthrodesis with three screws was intermediate. For 250 N loading, it was significantly lower in comparison with the nail. Similarly, Berend et al. reported significantly less stiffness for two cancellous screws in comparison with a nail for tibiotalocalcaneal arthrodesis in fresh-frozen cadavers [14]. Based on the present results and the study by Berend et al., it can be concluded that cancellous screws provide lower stability in tibiotalocalcaneal arthrodesis compared with a nail [14].

The mean cycle of failure for a tibiotalocalcaneal arthrodesis with the bone stabilization system was 2132. This result was lower in comparison with a nail, with a mean cycle of failure of 4191, but the difference was not significant. After tibiotalocalcaneal arthrodesis with a nail, one of six specimens failed during cyclic loading. In contrast, all six specimens failed after tibiotalocalcaneal arthrodesis with the bone stabilization system. However, it should be borne in mind here that the specimen after tibiotalocalcaneal arthrodesis with a nail failed at cycle no. 146 out of a total of 5000 possible cycles. This was the first specimen in all four of the groups that failed during cyclic loading. One possible explanation for this might be the bone mineral density of 0.49 g/cm2, the lowest of all specimens in the nail group.

Biomechanical properties are influenced by bone mineral density, which was therefore measured in every specimen used in the present study. The aim with specimen distribution was to achieve a similar mean bone mineral density in the groups, and this was accomplished in this study. There was a significant correlation between bone mineral density and cycle of failure in all groups and in the screws group. A similar result was reported by Klos et al. for tibiotalocalcaneal arthrodesis with a nail in fresh-frozen cadavers [21]. Gutteck et al. observed a linear relationship between stiffness and bone mineral density for tibiotalocalcaneal arthrodesis with a posterolateral plate in fresh-frozen cadavers [13].

An important question is what level of stability is needed in tibiotalocalcaneal arthrodesis to achieve bone healing. There is no exact answer for this question in the literature. Nevertheless, it is assumed that the bone stabilization system cannot achieve bone healing in tibiotalocalcaneal arthrodesis because of its low stability.

The present study has several limitations. An important one is the relatively small number of specimens investigated. This is a problem with most biomechanical studies, as there are only limited numbers of fresh-frozen cadaver specimens available. This biomechanical study investigated the primary stiffness after tibiotalocalcaneal arthrodesis. The effects of bone healing cannot be determined in a biomechanical cadaver model. The results of the study are therefore only applicable to the day of operation.

Conclusions

In conclusion, the stability of the tibiotalocalcaneal arthrodesis differs depending on the fixation method, with nail or plate showing the greatest stability and the bone stabilization system the least. When three screws are used for tibiotalocalcaneal arthrodesis, the stability is intermediate. As the biomechanical stability of the bone stabilization system is low, it cannot be recommended for tibiotalocalcaneal arthrodesis. There was a significant correlation between bone mineral density and cycle of failure in all groups and in the screws group.

Availability of data and materials

The datasets used and analysed during the current study are available from the corresponding author on reasonable request.

References

  1. Rammelt S, Pyrc J, Agren PH, Hartsock LA, Cronier P, Friscia DA, et al. Tibiotalocalcaneal fusion using the hindfoot arthrodesis nail: a multicenter study. Foot Ankle Int. 2013;34:1245–55.

    Article  PubMed  Google Scholar 

  2. Franceschi F, Franceschetti E, Torre G, Papalia R, Samuelsson K, Karlsson J, et al. Tibiotalocalcaneal arthrodesis using an intramedullary nail: a systematic review. Knee Surg Sports Traumatol Arthrosc. 2016;24:1316–25.

    Article  PubMed  Google Scholar 

  3. Mulligan RP, Adams SB Jr, Easley ME, DeOrio JK, Nunley JA 2nd. Comparison of posterior approach with intramedullary nailing versus lateral transfibular approach with fixed-angle plating for tibiotalocalcaneal arthrodesis. Foot Ankle Int. 2017;38:1343–51.

    Article  PubMed  Google Scholar 

  4. Gong JC, Zhou BH, Tao X, Yuan CS, Tang KL. Tibiotalocalcaneal arthrodesis with headless compression screws. J Orthop Surg Res. 2016;11:91.

    Article  PubMed  PubMed Central  Google Scholar 

  5. McSweeney AL, Zani BG, Baird R, Stanley JRL, Hayward A, Markham PM, et al. Biocompatibility, bone healing, and safety evaluation in rabbits with an IlluminOss bone stabilization system. J Orthop Res. 2017;35:2181–90.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  6. Vegt P, Muir JM, Block JE. The photodynamic bone stabilization system: a minimally invasive, percutaneous intramedullary polymeric osteosynthesis for simple and complex long bone fractures. Med Devices. 2014;7:453–61.

    Google Scholar 

  7. Zani BG, Baird R, Stanley JR, Markham PM, Wilke M, Zeiter S, et al. Evaluation of an intramedullary bone stabilization system using a light-curable monomer in sheep. J Biomed Mater Res B Appl Biomater. 2016;104:291–9.

    Article  PubMed  CAS  Google Scholar 

  8. Surke C, Meier R, Haug L, Vogelin E. Osteosynthesis of fifth metacarpal neck fractures with a photodynamic polymer bone stabilization system. J Hand Surg Eur. 2020;45:119–25.

    Article  Google Scholar 

  9. Chiodo CP, Acevedo JI, Sammarco VJ, Parks BG, Boucher HR, Myerson MS, et al. Intramedullary rod fixation compared with blade-plate-and-screw fixation for tibiotalocalcaneal arthrodesis: a biomechanical investigation. J Bone Joint Surg Am. 2003;85:2425–8.

    Article  PubMed  Google Scholar 

  10. Alfahd U, Roth SE, Stephen D, Whyne CM. Biomechanical comparison of intramedullary nail and blade plate fixation for tibiotalocalcaneal arthrodesis. J Orthop Trauma. 2005;19:703–8.

    Article  PubMed  Google Scholar 

  11. O’Neill PJ, Logel KJ, Parks BG, Schon LC. Rigidity comparison of locking plate and intramedullary fixation for tibiotalocalcaneal arthrodesis. Foot Ankle Int. 2008;29:581–6.

    Article  PubMed  Google Scholar 

  12. Ohlson BL, Shatby MW, Parks BG, White KL, Schon LC. Periarticular locking plate vs intramedullary nail for tibiotalocalcaneal arthrodesis: a biomechanical investigation. Am J Orthop. 2011;40:78–83.

    PubMed  Google Scholar 

  13. Gutteck N, Schilde S, Reichel M, Holtz A, Sahmel O, Hinz P, et al. Posterolateral plate fixation with Pantalarlock is more stable than nail fixation in tibiotalocalcaneal arthrodesis in a biomechanical cadaver study. Foot Ankle Surg. 2020;26:328–33.

    Article  PubMed  CAS  Google Scholar 

  14. Berend ME, Glisson RR, Nunley JA. A biomechanical comparison of intramedullary nail and crossed lag screw fixation for tibiotalocalcaneal arthrodesis. Foot Ankle Int. 1997;18:639–43.

    Article  PubMed  CAS  Google Scholar 

  15. Bennett GL, Cameron B, Njus G, Saunders M, Kay DB. Tibiotalocalcaneal arthrodesis: a biomechanical assessment of stability. Foot Ankle Int. 2005;26:530–6.

    Article  PubMed  Google Scholar 

  16. Muckley T, Eichorn S, Hoffmeier K, von Oldenburg G, Speitling A, Hoffmann GO, et al. Biomechanical evaluation of primary stiffness of tibiotalocalcaneal fusion with intramedullary nails. Foot Ankle Int. 2007;28:224–31.

    Article  PubMed  Google Scholar 

  17. Trost M, Bredow J, Boese CK, Loweg L, Schulte TL, Scaal M, et al. Biomechanical comparison of fixation with a single screw versus two kirschner wires in distal chevron osteotomies of the first metatarsal: a cadaver study. J Foot Ankle Surg. 2018;57:95–9.

    Article  PubMed  Google Scholar 

  18. Trost M, Knieps M, Loweg L, Neiss WF, Eysel P, Frey S, et al. Biomechanical comparison of conventional versus modified technique in distal chevron osteotomies of the first metatarsal: a cadaver study. Foot Ankle Surg. 2019;25:665–9.

    Article  PubMed  Google Scholar 

  19. Richter M, Evers J, Waehnert D, Deorio JK, Pinzur M, Schulze M, et al. Biomechanical comparison of stability of tibiotalocalcaneal arthrodesis with two different intramedullary retrograde nails. Foot Ankle Surg. 2014;20:14–9.

    Article  PubMed  Google Scholar 

  20. Evers J, Lakemeier M, Wahnert D, Schulze M, Richter M, Raschke MJ, et al. 3D optical investigation of 2 nail systems used in tibiotalocalcaneal arthrodesis: a biomechanical study. Foot Ankle Int. 2017;38:571–9.

    Article  PubMed  Google Scholar 

  21. Klos K, Wahnert D, Gueorguiev B, Schwieger K, Hofmann GO, Windolf M, et al. Development of a technique for cement augmentation of nailed tibiotalocalcaneal arthrodesis constructs. Clin Biomech (Bristol, Avon). 2010;25:576–81.

    Article  PubMed  Google Scholar 

  22. Muckley T, Hoffmeier K, Klos K, Petrovitch A, von Oldenburg G, Hofmann GO. Angle-stable and compressed angle-stable locking for tibiotalocalcaneal arthrodesis with retrograde intramedullary nails - Biomechanical evaluation. J Bone Joint Surg Am. 2008;90:620–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Open Access funding enabled and organized by Projekt DEAL. No funding was received for conducting this study.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by Matthias Trost, Suzan Yarkin and Matthias Knieps. The first draft of the manuscript was written by Matthias Trost and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Matthias Trost.

Ethics declarations

Ethics approval and consent to participate

The study was conducted with the approval of the local ethics committee (ref. no. 15-364). Informed consent to participate was obtained before death from all donors of the human cadavers included in the study.

Consent for publication

Informed consent for publication was obtained before death from all donors of the human cadavers included in the study.

Competing interests

The authors did not receive any financial support for research on, authorship of, or publication of this article. S. Gick received consultancy payments and lecture fees from IlluminOss Medical during the years 2014 to 2017.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Trost, M., Yarkin, S., Knieps, M. et al. Biomechanical comparison of different fixation methods in tibiotalocalcaneal arthrodesis: a cadaver study. J Orthop Surg Res 18, 971 (2023). https://doi.org/10.1186/s13018-023-04444-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-023-04444-7

Keyword