Cemented or uncemented fixation: Which allows a more acceptable prosthetic femoral version in total hip arthroplasty?

Abstract

Background

Three-dimensional computed-tomography (3D-CT) planning for primary Total Hip Arthroplasty (THA) typically uses the external femoral surface; as a result, it is difficult to predict the prosthetic femoral version (PFV) for uncemented femoral stems that press-fit to the internal surface of the bone. Cemented fixation allows the surgeon to adjust the version independent of the internal femoral anatomy. We aimed to better understand the effect of the fixation type on PFV.

Methods

This was a case series study including a total of 95 consecutive patients (106 hips), who underwent uncemented (n = 81 hips) and cemented (n = 25 hips) primary THA using the posterior approach. The surgeon aimed for a PFV of 20°. Our primary objective was to compare PFV in both groups; our secondary objective was to evaluate the clinical outcomes.

Results

The mean (± SD) PFV was 13° (± 9°) and 23° (± 8°) for the uncemented and cemented THA groups (P < 0.001), respectively. In the uncemented THA group, 36% of the patients had a PFV of < 10°. In the cemented THA group, this clinically important threshold dropped to 8%. Similarly, the Bland–Altman (BA) plots showed wider 95% limits of agreement for the uncemented group. Satisfactory clinical outcomes were recorded.

Conclusion

We found that the PFV was more clinically acceptable, for the posterior surgical approach, in the cemented group when compared to the uncemented group. Both THA groups reported high variability indicating the need to develop surgical tools to guide the PFV closer to the surgical target.

Background

Previous computed-tomography (CT) studies have reported a high variability of prosthetic femoral version (PFV) in primary total hip arthroplasty (THA), ranging from − 17° to 72° [1,2,3,4,5,6,7,8,9,10,11,12,13]. The accuracy in measuring version angles using Two-Dimensional (2D) cross-sectional CT images is lower when compared to Three-Dimensional (3D)-CT model-based measurements [14,15,16]. Existing literature has highlighted a lower discrepancy between 3D-CT and dry bone measurements than using single 2D cross-sectional scans, concluding that the 3D-CT method is the virtual equivalent of the reference standard (dry bone measurements) [14]. Studies using 3D-CT analysis have highlighted an increased incidence of prosthetic femoral retroversion and a wide range of PFV (− 23° to 43°) [16,17,18,19,20,21,22,23,24,25,26] (Table 1).

Table 1 CT-measured PFV in previous studies

The PFV of an uncemented femoral stem is partly dictated by the stem design and the highly variable internal morphology of the proximal femur [27]. Consequently, the final stem position is a compromise of best-fitting a straight stem down to the proximal femur, leaving the surgeon with minimal control over the PFV [23, 28, 29]. Contrastingly, in cemented femoral stems, the surgeon can intra-operatively adjust the version of the femoral stem to a desired position within the variable thickness of the cement mantle [17, 30, 31].

Suboptimal placement of the femoral stem with regards to its version is associated with the biomechanical instability of the reconstructed hip joint [20, 32, 33]. Impingement has been reported common in uncemented femoral stems with a PFV of < 5° [30], and a low PFV is associated with increased dislocation rate via the posterior approach [9, 27]. Furthermore, a revision rate of 40% has been reported among stems with a PFV of < 10° [20].

Considering the lower revision and dislocation rates that have been reported in the cemented femoral stems when compared to the uncemented femoral stems [34,35,36] and the relationship between the PFV and potential adverse clinical effects, we aimed to better understand the effect of the fixation type on PFV. Our primary objective was to measure the PFV in uncemented and cemented THA. Our secondary objective was to measure clinical outcomes. Our hypothesis was that cemented fixation, using a collarless double-tapered femoral stem, offers greater control of PFV than uncemented straight femoral stems.

Materials and methods

Study design

This was a retrospective study including 95 patients (106 hips) who underwent primary THA between February 2017 and June 2021 due to osteoarthritis. We divided the patients into two groups based on the stem fixation technique adopted, uncemented and cemented (Table 2). The surgery was performed through a posterior approach by a single consultant orthopaedic surgeon who specialises in hip arthroplasty.

Table 2 Study group characteristics

A single hemispheric press-fit Hydroxyapatite (HA) coated cup was used in both groups, two stem designs were used, straight-tapered in the uncemented and collarless double-tapered in the cemented group.

The outcome measures were as follows:

  1. 1.

    PFV angles.

  2. 2.

    Cup version angles

  3. 3.

    Clinical outcomes.

Pre-operative radiology and 3D software planning

All patients underwent pre-operative CT scanning of the hip region and the knee joint according to a standard protocol. A PS femoral neck osteotomy guide was designed, using the pre-operative CT data. In addition, 3D pre-operative planning was performed to establish the optimal acetabular and femoral implant size and position (MyHip, Medacta International SA, Castel San Pietro, Switzerland). During the operation, the surgeon aimed for a PFV of 20°.

3D-CT models of the pre-operative femurs were generated to measure NFV using Simpleware ScanIP software (Version 2021.03; Synopsys, Inc., Mountain View, USA) [16].

Surgical approach

All surgeries were performed through a posterior approach by a single consultant orthopaedic surgeon.

Prosthetic components

In the uncemented THA group, a straight-tapered femoral stem was used (Quadra-H; Medacta International SA, Castel San Pietro, Switzerland); in the cemented THA group, a collarless double-tapered femoral stem (X-Acta system; Medacta International SA, Castel San Pietro, Switzerland) was used. A hemispheric press-fit Hydroxyapatite (HA) coated cup was used for both groups (Mpact system; Medacta International SA, Castel San Pietro, Switzerland).

Patient-specific instrumentation (PSI)

During the surgery, a PS femoral neck osteotomy guide was used to perform the osteotomy. The guide was 3D-printed to fit the contours of the femoral head-neck junction. During the surgery, the sterilised PS cutting jig was positioned on the femoral head-neck junction and two pins secured its position. The surgeon then performed the osteotomy with the oscillating saw blade flush on the surface of the guide. The femoral neck osteotomy plane was defined as a plane inclined by 45° to the long axis of the proximal femur.

Post-operative radiology

All patients underwent post-operative CT scanning of the hip region and the knee joint that was done according to a standard protocol. Post-operative evaluation took place; number of fractures and dislocations was recorded. Oxford hip score (OHS) of cases reporting complications was recorded. 3D models of the post-operative femurs and prosthetic components were generated, using Simpleware ScanIP software (Version 2021.03; Synopsys, Inc., Mountain View, USA). PFV angles were measured, to assess the impact of the fixation method on the PFV.

Prosthetic femoral version (PFV) and acetabular cup version

Post-operative PFV was measured, as the angle between the axis of the neck of the femoral stem and the posterior condylar axis (PCA) projected on a plane perpendicular to the mechanical axis of the reconstructed femur (femoral stem) (Fig. 1) [19]. The stem neck axis was defined as the line connecting the centre of the head with the top mark of the stem [26].

Fig. 1
figure 1

Illustration of the coordinate system used to measure the PFV

Post-operative acetabular cup version was measured in the radiographic definition using the anterior pelvic plane (APP) [37, 38]. The cup plane was computed as the best-fitted plane based on 10 points chosen on the cup rim.

Reproducibility and reliability analysis

We measured the reproducibility of our CT measurement method using intra and interobserver analysis. For the intraobserver analysis, the same user measured PFV twice for 30 randomly selected cases, while for the interobserver analysis, a second user ran the test twice for 20 randomly selected cases.

Measurements of PFV were also obtained using an independent commercially available software (ZedHip, LEXI Co, Ltd, Tokyo, Japan).

Statistical analysis

SPSS software was used to perform the statistical analysis (version 28, SPSS, Chicago, USA). The Shapiro–Wilk test was used to evaluate the normality of the data in both groups and Mann–Whitney U test was implemented to evaluate differences between the two groups with regard to the study group characteristics. The median and interquartile range (IQR) were estimated for NFV angles. The mean, median, IQR and standard deviation (SD) were estimated for PFV angles. We compared the NFV and PFV for each case using a Bland–Altman (BA) plot.

The data describing the PFV angles were of different sample size and variance. Therefore, we performed the Welch’s test to assess if the mean values of PFV in both THA groups were statistically different.

For the reproducibility and reliability analysis, mean and SD of differences between the measurements of the same and different users were reported. Intraclass correlation coefficient (ICC) was obtained for both intra- and interobserver reliability.

Results

Discrepancy between NFV and PFV of individual cases in both THA groups

A BA plot of the discrepancy between NFV and PFV showed a mean discrepancy of − 1° and 10° of version for the uncemented and cemented groups, respectively. Furthermore, the uncemented THA cases had 95% limits of agreement between − 17 and 15°. This was wider than the cemented THA cases, where the 95% limits of agreement were − 2 and 22.5° (Fig. 2).

Fig. 2
figure 2

BA plot of the comparison between NFV and PFV of individual cases in (A) the uncemented THA group and (B) the cemented THA group

PFV in uncemented and cemented THA

The uncemented THA had a mean (± SD) and median (IQR) PFV of 13° (± 9°) and 13° (8–17°), respectively. The cemented THA group had mean (± SD) and median (IQR) PFV of 23° (± 8°) and 24° (18–28°), respectively. We found a statistically significant difference between the mean values of PFV in the uncemented and cemented THA groups (P < 0.001) (Fig. 3).

Fig. 3
figure 3

Box plot illustrating the PFV in the uncemented and cemented THA groups (*P < 0.001)

In the uncemented THA group, PFV measurements ranged from -18° retroversion to 33° anteversion. Five patients in the uncemented THA group had retroverted PFV. Four of these patients had a PFV of − 5° ± 1° and one had a PFV of − 18°. The NFV of these patients was 17°, 1°, 3°, 2° and − 13°, respectively. In addition, the absolute difference between the PFV and NFV was 21°, 6°, 9°, 8° and 5°, respectively.

In the cemented THA group, PFV values were ranged between 5° and 34°. There were no retroverted femoral stems in this group.

In the uncemented THA group, a PFV of between 5° and 10° and between 10° and 15° was reported in 16% and 25% of the femoral stems, respectively. Twenty-one per cent (21%) of the uncemented femoral stems had a PFV of between 15° and 20° and 7% had a PFV of between 20° and 25°. Finally, 11% of the uncemented femoral stems were anteverted of between 25° and 35° and 20% were anteverted of less than 5° (Fig. 4).

Fig. 4
figure 4

Distribution of PFV in uncemented and cemented THA groups

Concerning the distribution of PFV in the cemented THA group, 8% of the femoral stems reported a PFV between 5° and 10° and between 10° and 15°. A PFV of between 15° and 20° and between 20° and 25° was reported in 16% and 20% of the cemented femoral stems, respectively. Forty-eight per cent (48%) of the cemented femoral stems were anteverted more than 25° (Fig. 4).

Acetabular cup version in uncemented and cemented THA

The uncemented THA had a mean (± SD) and median (IQR) acetabular cup version of 23° (± 8°) and 23° (17–28°), respectively. The cemented THA group had mean (± SD) and median (IQR) acetabular cup version of 26° (± 7°) and 25° (20–30°), respectively.

Reproducibility and reliability analysis

We achieved excellent intraobserver repeatability and interobserver reproducibility in PFV. In any case, the ICC was more than 0.99, while the mean (± SD) difference between the same and different raters was 0.01 ± 1° and − 0.4 ± 2°, respectively.

The mean (± SD) difference of PFV measured by the external software (ZedHip, LEXI Co, Ltd, Tokyo, Japan) and our method was − 1 ± 2°.

Clinical outcomes

The median follow-up time was 45 and 21 months for the uncemented and cemented series, respectively. Satisfactory clinical outcome was recorded without any intra-operative fracture and post-operative implant loosening. Two dislocations were reported in the uncemented THA group: one as a result of deep hip flexion and one during yoga (child’s pose), 5 weeks and 12 months after the surgical procedure, respectively. Treatment included one closed reduction procedure with no further surgeries up to 48 months post-operatively. There was no recurrence of the dislocation and the OHS was 48/48 for both cases at 1 year post-operatively. The NFV for the two dislocated cases was 18° and 8°, respectively. Post-surgery, both cases reported a PFV of 9°, while the absolute difference between the PFV and NFV for these cases were 9° and 1°, respectively. In addition, the acetabular cup version was 29° and 8°, while the combined version (CV) was 38° and 17°, respectively.

Discussion

This was the first study to assess the impact of the fixation method on the PFV (the achieved version of the femoral stem) using 3D-CT image analysis. We found that PFV in the cemented THA group was higher when compared to the uncemented THA group (P < 0.001). In the uncemented THA group, 36% of the patients reported a PFV of < 10°, with 5 patients having a retroverted PFV. This percentage dropped to 8% in the cemented THA group. There were not any retroverted femoral stems in the cemented THA group.

Table 1 includes all the studies assessing the PFV using CT scans or 3D-CT analysis. High variability of PFV has been reported, ranging from − 23° to 72°. In this study, the PFV of an uncemented straight-tapered femoral stem ranged between − 18° and 33°.

Previous literature addressing the effect of the fixation method on PFV using Magnetic Resonance Imaging (MRI) scans reported a lower PFV in cemented femoral stems when compared to the findings of the present study (13 ± 8° vs. 23 ± 8° in our study) [39]. They declared 2.3% of retroverted cemented femoral stems and 11.8% of retroverted uncemented femoral stems. In our study, there were no retroverted femoral stems in the cemented THA group, but there were 6% of retroverted uncemented femoral stems. The difference may be explained by the different designs of the femoral stems used.

The findings of this study are consistent with the increased surgeon control of the position of cemented femoral stem designs and highlight that by intra-operatively adjusting the PFV, using the malleable nature of the cement mantle, the surgeons can avoid delivering a retroverted or insufficient PFV. This information is clinically relevant, considering the importance of an adequate PFV to eliminate undesirable events like dislocation or impingement in primary THA [9, 30, 36].

It is true that not only cemented femoral stems can offer intra-operative control of the femoral stem position. Uncemented femoral stems featuring modular necks allow modularity of the femoral stem neck in various configurations of PFV [18]. These components, however, have been linked to a risk of mechanical failure [40]. In contrast, the excellent implant survivorship reported for the cemented femoral stem designs [41] indicates that the cemented fixation is a safe alternative to deliver an adequate PFV in primary THA.

For straight, uncemented femoral stems, the femoral implant is tightly press-fitted into the bone to achieve the so-called best-fit position within the medullary canal of the proximal femur, leaving the surgeon with minimal control over the PFV [29]. For this reason, Dorr et al. [17] have emphasised the importance of delivering a CV (the sum of the acetabular and femoral version angles) between 25° and 50° to avoid dislocation in primary uncemented THA. In this context, when pre-operative planning indicates the risk of a suboptimal PFV due to either an excessive or retroverted patient’s NFV, surgeons can consider adjusting the cup version to compensate for an abnormal PFV, using the approach of the femur first technique [42].

However, despite its importance for avoiding dislocation [5, 17], the concept of a CV within the optimal range does not guarantee an optimal version for the individual prosthetic components (e.g., a case with a PFV of − 10° and a cup version of 40°). Recent studies have highlighted that focusing solely on the cup version to define a universal safe optimum for hip motion is not sufficient [43], and even when the generally accepted optimal range of CV is achieved, dislocations are not infrequent [44].

Current commercial planning solutions cannot predict the PFV of a straight uncemented femoral stem. Despite the high accuracy of 3D pre-operative planning in predicting implant size [45, 46], with the potential to minimise implant inventory [47], the surgical plan does not always deliver the targeted PFV in uncemented THA. PFV demonstrates an increased variability, ranging from − 19° of retroversion to 33° of anteversion [19]. This information implies the need to develop novel designs of intra-operative measuring tools that could potentially measure the version of the femoral broach. Surgeons could then classify the cases where uncemented fixation could not deliver an adequate PFV and choose cemented femoral stem designs instead.

Nevertheless, the greater adjustability that a cemented femoral stem may offer did not seem to lower the SD of the PFV in the cemented THA group. Intra-operative adjustment of PFV using the cement mantle is subjected to the surgeon’s perspective. Intra-operatively guiding the PFV of a cemented femoral stem, using either robotic tools or 3D-printed customised surgical guides, may be considered beneficial.

Particular emphasis should be given to femoral broaching and the implantation of a cemented femoral stem. Differences in the PFV of the femoral broach and of the implanted cemented femoral stem would result in an asymmetrical thickness of the cement column. Existing literature has supported that the thickness of the cement mantle around the femoral stem impacts cement strains [48], stem subsidence [49], micro-movement at the cement–bone interface [50], and the overall long-term radiographic outcome [51]. The presence of defects may affect the fixation interfaces and potentially function as an area of osteolysis [52,53,54]. However, there is still a debate about the optimal cement thickness [55, 56]. Long-term follow-up clinical studies are needed to determine if an asymmetrical cement mantle thickness negatively affects the implant’s survival.

Furthermore, considering that there is a continuing debate around the most appropriate fixation technique in primary THA, the findings of this study suggest that PFV may constitute an additional criterion during the selection process. A limited number of studies so far, have reported higher dislocation, revision and loosening rates in the uncemented primary THA, when compared to the cemented THA [34,35,36]. In this context, it is probable that this increased prevalence of untoward events in uncemented THA may stem from the high variability of PFV [17]. Long-term clinical studies are imperative to identify any significant association between PFV and post-operative clinical outcome in uncemented and cemented THA.

We acknowledge limitations. Firstly, PFV angles were measured based on 3D-CT reconstructed models of the proximal femurs and prosthetic models using a standardised coordinate system; this procedure is considered as a virtual equivalent to the reference standard [16]. The main limitation of this method is the amount of subjectivity induced by landmarks selection. Excellent intra and interobserver analysis proved that the PFV measurements were not significantly influenced by the user input. For the measurement of NFV measurements we used a published method [16]. Our measurements of NFV values are in accordance with those of previous studies [16, 29, 57].

Secondly, the findings may depend on the geometry of the femoral stems used, not fully reflecting the influence of cemented femoral stems of a different geometry on PFV. We compared a fit-fill, uncemented femoral stem of a trapezoidal cross-section with a narrower, highly polished, cemented femoral stem. A potential comparison between uncemented and cemented femoral stems of similar geometry would have resulted in less difference in PFV and the incidence of retroversion.

Lastly, the two THA groups had unequal sample sizes and follow-up time. Considering the fact that dislocation has been reported to occur within the first 12 months of the surgery [58, 59], we assumed that the follow-up time of the cemented THA group (21 months) is considered an adequate follow-up time to detect any clinically adverse effects.

Conclusions

Recent CT studies have reported a high variability of PFV in uncemented THA, suggesting that the internal morphology of the proximal femur may affect the final version of the femoral component. In this study, we found that the use of a cemented fixation technique led to higher PFV, when compared to the uncemented group although both groups reported a similar variability. With cemented fixation, surgeons have greater control of PFV. There is need to develop surgical tools that can intra-operatively measure and/or guide version of the femoral component.

Availability of data and materials

The data is available from the corresponding author on reasonable request.

Abbreviations

APP:

Anterior pelvic plane

BA:

Bland–Altman

CT:

Computed-tomography

CV:

Combined version

HA:

Hydroxyapatite

ICC:

Intraclass correlation coefficient

IQR:

Interquartile range

NFV:

Native femoral version

OHS:

Oxford hip score

PCA:

Posterior condylar axis

PFV:

Prosthetic femoral version

PS:

Patient-specific

PSI:

Patient-specific instrumentation

SD:

Standard deviation

3D-CT:

Three-dimensional computed-tomography

THA:

Total hip arthroplasty

References

  1. Komeno M, Hasegawa M, Sudo A, Uchida A. Computed tomographic evaluation of component position on dislocation after total hip arthroplasty. Orthopaedics. 2006;29:1104.

    Article  Google Scholar 

  2. Suh KT, Kang JH, Roh HL, Moon KP, Kim HJ. True femoral anteversion during primary total hip arthroplasty. J Arthroplast. 2006;21:599–605.

    Article  Google Scholar 

  3. Wines AP, McNicol D. Computed tomography measurement of the accuracy of component version in total hip arthroplasty. J Arthroplast. 2006;21:696–701.

    Article  Google Scholar 

  4. Reikerås O, Gunderson RB. Components anteversion in primary cementless THA using straight stem and hemispherical cup: a prospective study in 91 hips using CT-scan measurements. Orthop Traumatol Surg Res. 2011;97:615–21.

    Article  PubMed  Google Scholar 

  5. Nakashima Y, Hirata M, Akiyama M, Itokawa T, Yamamoto T, Motomura G, et al. Combined anteversion technique reduced the dislocation in cementless total hip arthroplasty. Int Orthop. 2014;38:27–32. https://doi.org/10.1007/s00264-013-2091-2.

    Article  PubMed  Google Scholar 

  6. Hirata M, Nakashima Y, Ohishi M, Hamai S, Hara D, Iwamoto Y. Surgeon error in performing intraoperative estimation of stem anteversion in cementless total hip arthroplasty. J Arthroplast. 2013;28:1648–53.

    Article  Google Scholar 

  7. Fujishiro T, Hayashi S, Kanzaki N, Hashimoto S, Kurosaka M, Kanno T, et al. Computed tomographic measurement of acetabular and femoral component version in total hip arthroplasty. Int Orthop. 2014;38:941–6. https://doi.org/10.1007/s00264-013-2264-z.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Hirata M, Nakashima Y, Itokawa T, Ohishi M, Sato T, Akiyama M, et al. Influencing factors for the increased stem version compared to the native femur in cementless total hip arthroplasty. Int Orthop. 2014;38:1341–6. https://doi.org/10.1007/s00264-014-2289-y.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Fujishiro T, Hiranaka T, Hashimoto S, Hayashi S, Kurosaka M, Kanno T, et al. The effect of acetabular and femoral component version on dislocation in primary total hip arthroplasty. Int Orthop. 2016;40:697–702. https://doi.org/10.1007/s00264-015-2924-2.

    Article  PubMed  Google Scholar 

  10. Okada T, Fukunishi S, Yoshiya S, Tachibana T, Fujihara Y, Masumoto Y, et al. Achievement of optimal implant alignment using taper wedge stems with cup-first THA through the MIS antero-lateral approach. Eur J Orthop Surg Traumatol. 2020;30:1505–14. https://doi.org/10.1007/s00590-020-02696-1.

    Article  PubMed  Google Scholar 

  11. Jackson JB, Martin JR, Christal A, Masonis JL, Springer BD, Mason JB. The direct anterior approach total hip arthroplasty reliably achieves “safe zones” for combined anteversion. Arthroplast Today. 2020;6:651–4.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Hochreiter J, Böhm G, Fierlbeck J, Anderl C, Birke M, Münger P, et al. Femoral antetorsion after calcar-guided short-stem total hip arthroplasty: a cadaver study. J Orthop Res. 2021. https://doi.org/10.1002/jor.25228.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Imai H, Miyawaki J, Kamada T, Takeba J, Mashima N, Miura H. Preoperative planning and postoperative evaluation of total hip arthroplasty that takes combined anteversion. Eur J Orthop Surg Traumatol. 2016;26:493–500. https://doi.org/10.1007/s00590-016-1777-8.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Kim JS, Park TS, Park SB. Measurement of femoral neck anteversion in 3D. Part 1: 3D imaging method. Med Biol Eng Comput. 2000;38:603–9.

    Article  CAS  PubMed  Google Scholar 

  15. Sugano N, Noble PC, Kamaric E. A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr. 1998;22(4):610–4.

    Article  CAS  PubMed  Google Scholar 

  16. Park KK, Tsai T-Y, Dimitriou D, Kwon Y-M. Utility of preoperative femoral neck geometry in predicting femoral stem anteversion. J Arthroplast. 2015;30:1079–84.

    Article  Google Scholar 

  17. Dorr LD, Malik A, Dastane M, Wan Z. Combined anteversion technique for total hip arthroplasty. Clin Orthop. 2009;467:119–27.

    Article  PubMed  Google Scholar 

  18. Sariali E, Mouttet A, Pasquier G, Durante E, Catone Y. Accuracy of reconstruction of the hip using computerised three-dimensional pre-operative planning and a cementless modular neck. J Bone Jt Surg. 2009;91:8.

    Google Scholar 

  19. Sendtner E, Tibor S, Winkler R, Wörner M, Grifka J, Renkawitz T. Stem torsion in total hip replacement: CT measurements in 60 patients. Acta Orthop. 2010;81:579–82. https://doi.org/10.3109/17453674.2010.524596.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Kiernan S, Hermann KL, Wagner P, Ryd L, Flivik G. The importance of adequate stem anteversion for rotational stability in cemented total hip replacement: a radiostereometric study with 10-year follow-up. Bone Jt J. 2013;95-B:23–30. https://doi.org/10.1302/0301-620X.95B1.30055.

    Article  CAS  Google Scholar 

  21. Inoue D, Kabata T, Maeda T, Kajino Y, Fujita K, Hasegawa K, et al. Value of computed tomography-based three-dimensional surgical preoperative planning software in total hip arthroplasty with developmental dysplasia of the hip. J Orthop Sci. 2015;20:340–6.

    Article  PubMed  Google Scholar 

  22. Dimitriou D, Tsai T-Y, Kwon Y-M. The effect of femoral neck osteotomy on femoral component position of a primary cementless total hip arthroplasty. Int Orthop. 2015;39:2315–21. https://doi.org/10.1007/s00264-015-2739-1.

    Article  PubMed  Google Scholar 

  23. Weber M, Messmer B, Woerner M, Grifka J, Renkawitz T. Novel measurement method on plain radiographs to predict postoperative stem anteversion in cementless THA: stem version is predictable in cementless THA. J Orthop Res. 2016;34:2025–30. https://doi.org/10.1002/jor.23202.

    Article  CAS  PubMed  Google Scholar 

  24. Hayashi S, Hashimoto S, Matsumoto T, Takayama K, Nishida K, Ishida K, et al. Stem anteversion mismatch to the anatomical anteversion causes loss of periprosthetic bone density after THA. J Orthop Surg. 2017;25:230949901773947. https://doi.org/10.1177/2309499017739478.

    Article  Google Scholar 

  25. Nodzo SR, Chang C-C, Carroll KM, Barlow BT, Banks SA, Padgett DE, et al. Intraoperative placement of total hip arthroplasty components with robotic-arm assisted technology correlates with postoperative implant position: a CT-based study. Bone Jt J. 2018;100-B:1303–9. https://doi.org/10.1302/0301-620X.100B10-BJJ-2018-0201.R1.

    Article  CAS  Google Scholar 

  26. Belzunce MA, Henckel J, Di Laura A, Hart A. Uncemented femoral stem orientation and position in total hip arthroplasty: a CT study. J Orthop Res. 2020;38:1486–96. https://doi.org/10.1002/jor.24627.

    Article  CAS  PubMed  Google Scholar 

  27. van Erp JHJ, Snijders TE, Weinans H, Castelein RM, Schlösser TPC, de Gast A. The role of the femoral component orientation on dislocations in THA: a systematic review. Arch Orthop Trauma Surg. 2022;142:1253–64. https://doi.org/10.1007/s00402-021-03982-1.

    Article  PubMed  Google Scholar 

  28. Marcovigi A, Ciampalini L, Perazzini P, Caldora P, Grandi G, Catani F. Evaluation of native femoral neck version and final stem version variability in patients with osteoarthritis undergoing robotically implanted total hip arthroplasty. J Arthroplast. 2019;34:108–15.

    Article  Google Scholar 

  29. Worlicek M, Weber M, Craiovan B, Wörner M, Völlner F, Springorum HR, et al. Native femoral anteversion should not be used as reference in cementless total hip arthroplasty with a straight, tapered stem: a retrospective clinical study. BMC Musculoskelet Disord. 2016;17:399. https://doi.org/10.1186/s12891-016-1255-9.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Malik A, Maheshwari A, Dorr LD. Impingement with Total hip replacement. J Bone Jt Surg. 2007;89:1832–42.

    Google Scholar 

  31. Emerson RH. Increased anteversion of press-fit femoral stems compared with anatomic femur. Clin Orthop. 2012;470:477–81.

    Article  PubMed  Google Scholar 

  32. Bergmann G, Graichen F, Rohlmann A. Hip joint loading during walking and running, measured in two patients. J Biomech. 1993;26:969–90.

    Article  CAS  PubMed  Google Scholar 

  33. Gill HS, Alfaro-Adrián J, Alfaro-Adrián C, McLardy-Smith P, Murray DW. The effect of anteversion on femoral component stability assessed by radiostereometric analysis. J Arthroplast. 2002;17:997–1005.

    Article  CAS  Google Scholar 

  34. Dobzyniak M, Fehring TK, Odum S. Early failure in total hip arthroplasty. Clin Orthop. 2006;447:76–8.

    Article  PubMed  Google Scholar 

  35. Zijlstra WP, De Hartog B, Van Steenbergen LN, Scheurs BW, Nelissen RGHH. Effect of femoral head size and surgical approach on risk of revision for dislocation after total hip arthroplasty: an analysis of 166,231 procedures in the Dutch arthroplasty register (LROI). Acta Orthop. 2017;88:395–401. https://doi.org/10.1080/17453674.2017.1317515.

    Article  PubMed  PubMed Central  Google Scholar 

  36. van Stralen GMJ, Struben PJ, van Loon CJM. The incidence of dislocation after primary total hip arthroplasty using posterior approach with posterior soft-tissue repair. Arch Orthop Trauma Surg. 2003;123:219–22. https://doi.org/10.1007/s00402-003-0482-3.

    Article  PubMed  Google Scholar 

  37. Sugano N. Computer-assisted orthopaedic surgery and robotic surgery in total hip arthroplasty. Clin Orthop Surg. 2013;5:1.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Wan Z, Malik A, Jaramaz B, Chao L, Dorr LD. Imaging and navigation measurement of acetabular component position in THA. Clin Orthop. 2009;467:32–42.

    Article  PubMed  Google Scholar 

  39. Fischer T, Stern C, Fritz B, Zingg PO, Pfirrmann CWA, Sutter R. Impact of stem design and cementation on postoperative femoral antetorsion in 227 patients with total hip arthroplasty (THA). Skeletal Radiol. 2020;49:2001–9. https://doi.org/10.1007/s00256-020-03483-z.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Krishnan H, Krishnan SP, Blunn G, Skinner JA, Hart AJ. Modular neck femoral stems. Bone Jt J. 2013;95-B:1011–21. https://doi.org/10.1302/0301-620X.95B8.31525.

    Article  CAS  Google Scholar 

  41. Carrington NC, Sierra RJ, Gie GA, Hubble MJW, Timperley AJ, Howell JR. The Exeter Universal cemented femoral component at 15–17 years. J Bone Jt Surg. 2009;91:8.

    Google Scholar 

  42. Loppini M, Longo UG, Caldarella E, Rocca AD, Denaro V, Grappiolo G. Femur first surgical technique: a smart non-computer-based procedure to achieve the combined anteversion in primary total hip arthroplasty. BMC Musculoskelet Disord. 2017;18:331. https://doi.org/10.1186/s12891-017-1688-9.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Pour AE, Schwarzkopf R, Patel KP, Anjaria M, Lazennec JY, Dorr LD. Is combined anteversion equally affected by acetabular cup and femoral stem anteversion? J Arthroplast. 2021;36:2393–401.

    Article  Google Scholar 

  44. Hernández A, Lakhani K, Núñez JH, Mimendia I, Pons A, Barro V. Can we trust combined anteversion and Lewinnek safe zone to avoid hip prosthesis dislocation? J Clin Orthop Trauma. 2021;21:101562.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Viceconti M, Lattanzi R, Antonietti B, Paderni S, Olmi R, Sudanese A, et al. CT-based surgical planning software improves the accuracy of total hip replacement preoperative planning. Med Eng Phys. 2003;25:371–7.

    Article  CAS  PubMed  Google Scholar 

  46. Sariali E, Mauprivez R, Khiami F, Pascal-Mousselard H, Catonné Y. Accuracy of the preoperative planning for cementless total hip arthroplasty. A randomised comparison between three-dimensional computerised planning and conventional templating. Orthop Traumatol Surg Res. 2012;98:151–8.

    Article  CAS  PubMed  Google Scholar 

  47. Di Laura A, Henckel J, Hothi H, Hart A. Can 3D surgical planning and patient specific instrumentation reduce hip implant inventory? A prospective study. 3D Print Med. 2020;6:25. https://doi.org/10.1186/s41205-020-00077-2.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Fisher DA, Tsang AC, Paydar N, Milionis S, Turner CH. Cement-mantle thickness affects cement strains in total HIP replacement. J Biomech. 1997;30:1173–7.

    Article  CAS  PubMed  Google Scholar 

  49. Takahashi E, Kaneuji A, Tsuda R, Numata Y, Ichiseki T, Fukui K, et al. The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem: when are thick cement mantles detrimental? Bone Jt Res. 2017;6:351–7. https://doi.org/10.1302/2046-3758.65.BJR-2017-0028.R1.

    Article  CAS  Google Scholar 

  50. Ramaniraka N, Rakotomanana L, Leyvraz P. The fixation of the cemented femoral component: effects of stem stiffness, cement thickness and roughness of the cement-bone surface. J Bone Joint Surg. 2000;82-B:297–303.

    Article  Google Scholar 

  51. Ebramzadeh E, Sarmiento A, McKellop H, Llinas A, Gogan W. The cement mantle in total hip arthroplasty. J Bone Jt Surg. 1994;76-A:77–87.

    Article  Google Scholar 

  52. Scheerlinck T, De Mey J, Deklerck R, Noble PC. CT analysis of defects of the cement mantle and alignment of the stem. In vitro comparison of Charnley–Kerboul femoral hip implants inserted line-to-line and undersized in paired femora. J Bone Jt Surg Br. 2006;88-B:19–25. https://doi.org/10.1302/0301-620X.88B1.16715.

    Article  Google Scholar 

  53. Breusch SJ, Lukoschek M, Kreutzer J, Brocai D, Gruen TA. Dependency of cement mantle thickness on femoral stem design and centralizer. J Arthroplast. 2001;16:648–57.

    Article  CAS  Google Scholar 

  54. Anthony P, Gie GA, Howie CR. Localised endosteal bony lysis in relation to femoral components of cemented total hip arthroplasty. Bone Jt Surg. 1990;72-B:971–9.

    Article  Google Scholar 

  55. Cassar-Gheiti AJ, McColgan R, Kelly M, Cassar-Gheiti TM, Kenny P, Murphy CG. Current concepts and outcomes in cemented femoral stem design and cementation techniques: the argument for a new classification system. EFORT Open Rev. 2020;5:241–52.

    Article  PubMed  PubMed Central  Google Scholar 

  56. Skinner JA, Todo S, Taylor M, Wang JS, Pinskerova V, Scott G. Should the cement mantle around the femoral component be thick or thin? J Bone Jt Surg Br. 2003;85-B:45–51. https://doi.org/10.1302/0301-620X.85B1.13055.

    Article  Google Scholar 

  57. Folinais D, Thelen P, Delin C, Radier C, Catonne Y, Lazennec JY. Measuring femoral and rotational alignment: EOS system versus computed tomography. Orthop Traumatol Surg Res. 2013;99:509–16.

    Article  CAS  PubMed  Google Scholar 

  58. Dargel J, Oppermann J, Brüggemann GP, Eysel P. Dislocation following total hip replacement. Deutsches Ärzteblatt Int. 2014;111(51–52):884.

    Google Scholar 

  59. Brooks PJ. Dislocation following total hip replacement. Bone Jt J. 2013;95:3.

    Google Scholar 

Download references

Acknowledgements

This research study was supported by The Arthroplasty for Arthritis Charity, The Maurice Hatter Foundation, the RNOH Charity, the Rosetrees Trust and the Stoneygate Trust, by researchers at the National Institute for Health Research University College London Hospitals Biomedical Research Centre. The authors are grateful to the engineers at Medacta International for their technical support.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

MM, ADL, JH, and AJH contributed to study design. MM, ADL, HH, JH, and AJH contributed to data collection, analysis, and interpretation. MM, ADL, HH, JH, and AJH involved in writing and revision of the manuscript. All co-authors have read the manuscript and agreed with the contents of it.

Corresponding author

Correspondence to Alister J. Hart.

Ethics declarations

Ethics approval and consent to participate

This project was evaluated in accordance with National Research Ethics Service “NRES—defining research” protocol. It was categorised by the NHS RNOH R&D department as service evaluation and as such does not require approval from Research Ethics Committee (REC) or R&D Office. Institutional Review Board Statement: Institutional review board approval NHS RNOH R&D Service Evaluation (SE16.020-11/08/2016).

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article