Abstract

Background

Nerve compression symptoms and spinal instability, resulting from spinal metastases, significantly impact the quality of life for patients. A 3D-printed vertebral body is considered an effective approach to reconstruct bone defects following en bloc resection of spinal tumors. The advantage of this method lies in its customized shape and innermost porous structure, which promotes bone ingrowth and leads to reduced postoperative complications.

Objective

The purpose of this study is to assess the effectiveness of 3D-printed auto-stable artificial vertebrae in the en bloc resection and reconstruction of thoracolumbar metastases.

Methods

This study included patients who underwent en bloc resection of thoracolumbar metastases based on the Weinstein-Boriani-Biagini surgical staging system, between January 2019 and April 2021. The patients were divided into two groups: the observation group, which was reconstructed using 3D-printed auto-stable vertebral bodies, and the control group, treated with titanium cages and allograft bone. Evaluation criteria for the patients included assessment of implant subsidence, instrumentation-related complications, VAS score, and Frankel grading of spinal cord injury.

Results

The median follow-up period was 21.8 months (range 12–38 months). Among the patients, 10 received a customized 3D-printed artificial vertebral body, while the remaining 10 received a titanium cage. The observation group showed significantly lower operation time, intraoperative blood loss, and postoperative drainage compared to the control group (P < 0.05). At the final follow-up, the average implant subsidence was 1.8 ± 2.1 mm for the observation group and 5.2 ± 5.1 mm for the control group (P < 0.05). The visual analog scale (VAS) scores were not statistically different between the two groups at preoperative, 24 h, 3 months, and 1 year after the operation (P < 0.05). There were no statistically significant differences in the improvements of spinal cord functions between the two groups.

Conclusion

The utilization of a 3D-printed auto-stable artificial vertebra for reconstruction following en bloc resection of thoracolumbar metastases appears to be a viable and dependable choice. The low occurrence of prosthesis subsidence with 3D-printed prostheses can offer immediate and robust stability.

Introduction

Spinal tumors encompass primary tumors originating in the spine and metastatic tumors arising from other sites [1]. The spine is the most common site of bone metastasis for malignant tumors, with significantly higher incidence compared to primary tumors. Spinal metastases occur in about 30–70% of patients with malignant tumors [2, 3]. These metastases can lead to severe back pain, limb numbness, weakness, and/or paralysis. Spinal cord compression affects 5–10% of patients with spinal metastases [4, 5], greatly impacting patients' quality of life, survival, and placing a substantial burden on society.

Treatment options for spinal metastases include radiotherapy, chemotherapy, surgery, and biological therapy. Surgery is a crucial approach for spinal tumors [6]. En bloc resection is a widely used surgical method, aiming to completely remove tumors with sufficient margins. When the stability of the spine is disrupted and the spinal cord nerves are compressed by tumors, en bloc resection is feasible. The choice of intraoperative implants plays a pivotal role in the surgery. Conventional implants for reconstruction include titanium cages, artificial vertebrae, autologous bone, and bone cement combined with an anterior nail plate or posterior pedicle screw system. However, these implants have drawbacks, such as lack of individualization, long operation time, and increased bleeding. Additionally, issues like internal fixation loosening and breakage are not uncommon, leading to postoperative spinal instability. 3D printing technology, a product of advancements in computer technology and image digitization, offers unique advantages in orthopedic implant design and manufacturing. In spinal surgery, the 3D-printed auto-stable vertebra closely matches the upper and lower vertebral bodies, addressing anatomical and physiological demands while providing immediate postoperative spinal stability with fewer instrumentation-related complications.

Therefore, this study retrospectively analyzes clinical data from patients with thoracolumbar metastases treated using either 3D-printed auto-stable artificial vertebrae or titanium cages. The aim is to evaluate the efficacy of 3D printing technology in treating spinal metastases.

Materials and methods

Study design and population

This study involved a retrospective analysis conducted at a single center. It received approval from the institutional review board of the ethics committee, and all patients provided written informed consent. The study included a total of 20 patients with thoracolumbar metastases who underwent en bloc resection between January 2019 and April 2021. These patients were divided into two groups: the observation group (using 3D-printed auto-stable artificial vertebral bodies) and the control group (not using 3D-printed implants). Each group comprised 10 patients. The inclusion criteria were as follows: (1) Confirmed solitary thoracolumbar metastases through histological diagnosis; (2) Postoperative follow-up exceeding 12 months with available imaging data; (3) Tomita score less than 4 and Tokuhashi score more than 8; (4) Presence of nerve compression symptoms or aggressive bone damage. Patients with extensive visceral metastases or those unable to tolerate general anesthesia and surgery were excluded from the study. The patient characteristics and treatment-related data were gathered from electronic medical records. Imaging data, including radiographs, CT scans, MRI images, and emission computed tomography, were obtained from the picture archiving and communication system. The patients who underwent surgery were regularly followed up.

3D-printed auto-stable artificial vertebrae

Similar to the approach used in our previous study [7], the observation group utilized Mimics 17.0 software to process the preoperative imaging examination's DICOM format files for prosthesis design. A 3D model was established through computer calculations, and 3D reconstruction images were obtained via image fusion technology. The target areas encompassed tumors, bones, blood vessels, and nerves. The models were created using multi-segmentation masks of different colors, restoring bones with corresponding blood vessels and nerves to register the 3D model. A 1:1 multisegment solid spinal tumor model was 3D-printed using polylactic acid. Additionally, the 3D-printed auto-stable artificial vertebrae featured a porous surface and innermost scaffold structure, mimicking cancellous bone, which allowed for bony ingrowth and regulation of elastic modulus. The prosthesis had pore diameters of 700 ± 80 µm and wire diameters of 300 ± 100 µm, with an average porosity of 73%.

To determine the best insertion point and corresponding needle catheter, the spinal vertebral model and corresponding nail channel data were imported into 3-matic 9.0 reverse engineering software. A reverse template was constructed in conjunction with the injection catheter to form a personalized needle guide plate. Subsequently, the pin guide plate underwent in vitro verification and was handled through low-temperature plasma disinfection. Due to the varying organs located in front of different spinal segments, personalized perioperative treatment measures and surgical plans were developed. The 3D printing model accurately depicted the surgical area's anatomical structure, allowing for precise determination of the lesion's location and clarification of its morphology and resection scope. Moreover, the 3D model of the guide plate facilitated accurate screw insertion, effectively simulating the angle and depth of the nail inlet, providing guidance for surgical screw selection (Fig. 1).

Fig. 1
figure 1

3D-printed auto-stable artificial vertebrae

Surgical procedure

For preoperative evaluation, X-ray, CT, MRI, and positron emission CT (PET-CT) scans were routinely conducted. Additionally, when tumors closely involved major vascular structures, CT angiography was performed. In both groups, all patients received intravenous general anesthesia with endotracheal intubation, followed by tumor resection via a posterior approach and internal fixation. During the procedure, the surgeon implanted the 3D-printed prosthesis in the observation group and performed posterior reconstruction using pedicle screws and transverse connectors. The posterior instrumentation was slightly adjusted to compress the inserted prosthesis. On the other hand, in the control group, the surgeon performed the procedure using a titanium cage with autologous iliac bone implantation.

Observed outcomes

Implant subsidence was assessed on midsagittal reconstructed CT images and quantified as the loss of segment height. This measurement involved determining the distance between the midpoint of the upper endplate of the vertebral body above the resection site and the midpoint of the lower endplate of the vertebral body below the resection site. Prosthesis subsidence was defined as a reduction in segment height from the immediate postoperative measurement to the last follow-up measurement. Fusion time was determined using CT imaging. To provide evidence of osseointegration at the bone-metal contact surfaces, we analyzed the density change inside the prosthesis by measuring CT HU. For CT HU measurement, four 2 × 2 mm areas were selected on the upper and lower ends of the prosthesis and both sides of the prosthesis. The average value of these measurements was taken as the patient's postoperative CT HU value.

We evaluated the VAS score and Frankel grading at preoperative, 24 h, 3 months, and 1 year after the operation. Additionally, we reviewed all postoperative imaging to identify any instrumentation-related complications, such as implant loosening, breakage, prosthesis migration, or other noticeable complications detectable on imaging during the follow-up period.

Statistical analysis

Statistical calculations were conducted using SPSS version 28.0. Measurements that followed a normal distribution were presented as mean ± standard deviation, and we used the independent sample t test for these data. For count data, we employed the chi-square test. A significance level of P < 0.05 was considered statistically significant.

Results

Demographic and clinical characteristics

Twenty patients underwent en bloc resection of thoracolumbar metastasis. Table 1 provides details of their baseline demographic and clinical characteristics, such as comorbidities, symptoms, duration of symptoms, radiological assessment, and treatments. There were no significant differences in general data between the two groups, including age, height, weight, and histological grade (P > 0.05).

Table 1 Comparison of general data between the two groups

The primary site of cancer in the study participants was as follows: breast (7 patients), lung (4 cases), prostate (4 cases), renal (1 case), stomach (3 cases), and colorectum (1 case). Among them, 13 cases had lesions in the thoracic spine, and the remaining 7 cases had lesions in the lumbar spine. Ten patients received 3D-printed artificial vertebral bodies, while the other 10 received titanium cages. The median follow-up for all patients was 21.8 months (range 12–38 months). There was no significant difference in survival between the two groups (P > 0.05).

During the perioperative period, the observation group showed significantly lower operation time and intraoperative blood loss compared to the control group (P < 0.05). After surgery, the observation group had significantly lower postoperative drainage rates and shorter extubation times than the control group (P < 0.05). Refer to Table 2 for detailed data.

Table 2 Comparison of surgery in the two groups

Comparison of the correlation scores

After surgery, both groups of patients experienced relief from their preoperative symptoms. VAS scores were assessed at preoperative, 24 h, 3 months, and 1 year after the operation. Before surgery, the VAS score for the observation group was 6.3 ± 1.5, and 24 h after surgery, it was 4.9 ± 1.4. There was no statistically significant difference compared to the control group (t = – 0.447, P = 0.660; t = – 0.848, P = 0.408). However, at 3 months after surgery, the VAS score for the observation group was 3.3 ± 0.7, and at 1 year after surgery, it was 2.1 ± 1.0 points. The difference was statistically significant compared to the control group (t = 0.632, P = 0.535; t = 0.511, P = 0.616) (Table 3).

Table 3 Comparison of the correlation scores

At the last follow-up, 9 patients in the observation group and 7 patients in the control group regained their spinal function according to the Frankel scale. However, there was no statistical difference in the improvement of spinal cord function between the two groups (P = 0.582).

Comparison of adverse reactions

Both groups were followed after treatment, and the complications were assessed. In the observation group, the occurrences of nerve paralysis, lower limb weakness, hypoesthesia, CSF leak, and infection were all lower compared to the control group (Table 4).

Table 4 Comparison of complications after surgery

During the postoperative follow-up, the fusion time and implant subsidence in the observation group were lower than those in the control group, and the difference was statistically significant (P < 0.05) (Table 5).

Table 5 Comparison of the postoperative biofusion conditions

Typical cases

Case 1: A 36-year-old male was admitted to the hospital with a history of nasopharyngeal cancer, thoracic 10 metastasis, and previous radiotherapy and chemotherapy. He presented with back pain for 3 months, as well as numbness and weakness in both lower extremities for 2 weeks. Examination revealed nerve compression, and the patient's paralysis was progressively worsening. Based on CT, MRI, and other auxiliary examinations, the diagnosis was confirmed as nasopharyngeal cancer spinal metastasis. The patient underwent en bloc resection of the tumor vertebral body with neuroelectrophysiological detection, followed by reconstruction using pedicle screw internal fixation and 3D-printed prosthesis implantation. After surgery, the patient showed a good recovery (Fig. 2).

Fig. 2
figure 2

Case 1

Case 2: A 52-year-old female with a history of right breast cancer presented with 2 weeks of low back pain and numbness in the right lower extremity. Examination revealed nerve compression, and further diagnostic imaging, including CT, MRI, and PET-CT, confirmed spinal metastasis of breast cancer. The patient underwent en bloc resection of the tumor vertebral body with neuroelectrophysiological detection, followed by reconstruction using pedicle screw internal fixation and 3D-printed prosthesis implantation. The patient showed a favorable recovery after the surgery (Fig. 3).

Fig. 3
figure 3

Case 2

Discussion

The spine has now become the third most common site of tumor metastasis after the liver and lungs. Surgical resection remains the primary treatment for spinal tumors, involving tumor removal, spinal cord decompression, and stability reconstruction which are fundamental aspects of whole spine resection. Since the entire vertebral segment, along with ligament and muscle tissues, is removed during the procedure, the anterior, middle, and posterior pillars are disrupted, necessitating stability reconstruction [8]. The objective of reconstruction is to achieve long-term biological fusion. While titanium cages have provided effective support, concerns have arisen about their long-term reliability. Chen [9] followed up with 300 cases of cervical titanium cage reconstruction and observed that titanium cage subsidence was a common phenomenon associated with postoperative complications. Yoshioka [10] also found that internal fixation failure after total spinal resection of the thoraco-lumbar spine was a common occurrence. Due to the subsidence of the titanium cage and the collapse of the intervertebral space, maintaining the segment angle becomes challenging, leading to loss of biological stability, concentration of rod stress, and an increased risk of fracture. Park [11] identified perioperative radiotherapy as an independent risk factor for postoperative rod fracture, as radiotherapy affects bone mass and bone fusion, having a negative impact on spine stability. Moreover, the small contact surface of the titanium cage may cause the spinal force line to deviate from the bone contact surface, resulting in an unfavorable angle for bone fusion and internal fixation failure. However, artificial vertebral bone offers a larger contact surface, conducting the force line through the vertebral body via facial contact, which provides better stability [12].

Biological fusion is essential for maintaining long-term stability in spine reconstruction. The porous structure inside the 3D-printed artificial vertebra facilitates tissue fluid flow, promotes bone cell migration and proliferation, and a porosity of 70–80% is considered ideal for bone penetration. The complex microscopic rough structure on the surface encourages the recruitment of anti-inflammatory factors and osteoblast differentiation, creating a special cellular environment for bone formation [13, 14]. McGilvray [15] conducted in vitro fusion experiments in sheep lumbar spines and found that porous titanium alloy vertebrae exhibited superior bone penetration compared to PEEK material. Similarly, the use of a titanium alloy cage in anterior cervical surgery has demonstrated better osseointegration rates [16]. CT scans during patient follow-up in the observation group showed bone ingrowth in the bone-prosthesis interface and excellent osseointegration ability.

The length of the 3D-printed vertebral body is determined based on preoperative CT data, ensuring an accurate match with bone defects and allowing for the setup of autofixation or lateral fixation devices to connect adjacent vertebrae, resulting in better stability. In comparison to modular artificial vertebrae with autostabilization devices, the 3D-printed artificial vertebral autostabilization device is more personalized, facilitating intraoperative procedures for different patients [17]. Its bone contact surface has a rough structure imitating trabecular pores, rather than a spike-like device that can embed in the bone surface and reduce subsidence of the artificial vertebral body. Mobbs [18] reconstructed C2 with individualized artificial vertebrae, and at the 9-month postoperative follow-up, there was no loosening or displacement of the prosthesis. Doyoung [19] achieved accurate hemisacral reconstruction using 3D-printed prosthesis, and one year after the operation, the prosthesis remained in good position, with osseointegration occurring with the surrounding bone.

This study demonstrates that en bloc resection and reconstruction with 3D-printed artificial vertebrae offer advantages such as shorter operation time, reduced bleeding, and faster patient recovery. Compared to traditional titanium cages, 3D-printed artificial vertebrae maintain segment height more effectively and have a lower settlement rate. Their porous structure and rough bone contact surface are more conducive to osteogenesis. However, this study is limited by the number of cases and lacks long-term follow-up. Future research is needed to improve the therapeutic efficacy of spinal metastases.

References

  1. Bone Oncology Group, Bone Science Branch of Chinese Medical Association. Guidelines for the surgical treatment of spinal metastases. Chin J Orthop. 2019;39(12):717–26.

    Google Scholar 

  2. Barzilai O, et al. Integrating evidence-based medicine for treatment of spinal metastases into a decision framework: neurologic, oncologic, mechanicals stability, and systemic disease. J Clin Oncol. 2017;35(21):2419–27.

    Article  CAS  PubMed  Google Scholar 

  3. Barzilai O, et al. Predictors of quality of life improvement after surgery for metastatic tumors of the spine: prospective cohort study. Spine J. 2018;18(7):1109–15.

    Article  PubMed  Google Scholar 

  4. Moussazadeh N, Laufer I, Yamada Y, Bilsky MH. Separation surgery for spinal metastases: effect of spinal radiosurgery on surgical treatment goals. Cancer Control. 2014;21(2):168–74.

    Article  PubMed  Google Scholar 

  5. Witham TF, Khavkin YA, Gallia GL, Wolinsky JP, Gokaslan ZL. Surgery insight: current management of epidural spinal cord compression from metastatic spine disease. Nat Clin Pract Neurol. 2006;2(2):87–94.

    Article  PubMed  Google Scholar 

  6. Laufer I, et al. The noms framework: approach to the treatment of spinal metastatic tumors. Oncologist. 2013;18(6):744–51.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Cao Y, Qing P, Deng H, Xu J. 3D printing in multilevel spinal tumor surgery. Chin J Bone Jt Inj. 2020;35(11):1161–3.

    Google Scholar 

  8. Kawahara N, Tomita K, Murakami H, Demura S. Total en bloc spondylectomy for spinal tumors: surgical techniques and related basic background. Orthop Clin N Am. 2009;40(1):47–63.

    Article  Google Scholar 

  9. Chen Y, et al. Subsidence of titanium mesh cage: a study based on 300 cases. J Spinal Disord Tech. 2008;21(7):489–92.

    Article  PubMed  Google Scholar 

  10. Yoshioka K, et al. Risk factors of instrumentation failure after multilevel total en bloc spondylectomy. Spine Surg Relat Res. 2017;1(1):31–9.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Park SJ, et al. Rod fracture and related factors after total en bloc spondylectomy. Spine J. 2019;19(10):1613–9.

    Article  PubMed  Google Scholar 

  12. Alsaadi G, Quirynen M, Komárek A, van Steenberghe D. Impact of local and systemic factors on the incidence of late oral implant loss. Clin Oral Implants Res. 2008;19(7):670–6.

    Article  PubMed  Google Scholar 

  13. Olivares-Navarrete R, et al. Implant materials generate different peri-implant inflammatory factors: poly-ether-ether-ketone promotes fibrosis and microtextured titanium promotes osteogenic factors. Spine (Phila Pa 1976). 2015;40(6):399–404.

    Article  PubMed  Google Scholar 

  14. Olivares-Navarrete R, et al. Rough titanium alloys regulate osteoblast production of angiogenic factors. Spine J. 2013;13(11):1563–70.

    Article  PubMed  Google Scholar 

  15. McGilvray KC, et al. Bony ingrowth potential of 3d-printed porous titanium alloy: a direct comparison of interbody cage materials in an in vivo ovine lumbar fusion model. Spine J. 2018;18(7):1250–60.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Cabraja M, Oezdemir S, Koeppen D, Kroppenstedt S. Anterior cervical discectomy and fusion: comparison of titanium and polyetheretherketone cages. BMC Musculoskelet Disord. 2012;13:172.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Lin Y, Fang G, Wu J, et al. Application of 3D-printed metal artificial vertebrae in giant cell tumor of the spinal bone: 1 case. Bioorthop Mater Clin Res. 2020;17(1):31–2.

    Google Scholar 

  18. Mobbs RJ, Coughlan M, Thompson R, Sutterlin CE, Phan K. The utility of 3d printing for surgical planning and patient-specific implant design for complex spinal pathologies: case report. J Neurosurg Spine. 2017;26(4):513–8.

    Article  Google Scholar 

  19. Kim D, et al. Sacral reconstruction with a 3d-printed implant after hemisacrectomy in a patient with sacral osteosarcoma: 1-year follow-up result. Yonsei Med J. 2017;58(2):453–7.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgments

The authors would like to thank the Health Commission of Sichuan Province, for supporting our project.

Funding

This study is funded by the Health Commission of Sichuan Province, the 3D printing technology combined with simple posterior total spinal resection for spinal tumors (17PJ139).

Author information

Authors and Affiliations

Authors

Contributions

YC provided conception and design of study, YC and NY drafted the manuscript, SBW and CW provided writing and illustration support, QH, QFW and YYZ provided writing assistance and revised and proof read the article.

Corresponding author

Correspondence to Yun Cao.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the institutional review board of ethics committee of Sichuan Science City Hospital.

Consent for publication

Informed consent was obtained from all individual participants included in the study.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cao, Y., Yang, N., Wang, S. et al. The application of 3D-printed auto-stable artificial vertebral body in en bloc resection and reconstruction of thoracolumbar metastases. J Orthop Surg Res 18, 638 (2023). https://doi.org/10.1186/s13018-023-04135-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-023-04135-3

Keywords