Second-generation bone cement-injectable cannulated pedicle screws for osteoporosis: biomechanical and finite element analyses

Abstract

Background

Biomechanical and finite element analyses were performed to investigate the efficacy of second-generation bone cement-injectable cannulated pedicle screws (CICPS) in osteoporosis.

Methods

This study used the biomechanical test module of polyurethane to simulate osteoporotic cancellous bone. Polymethylmethacrylate (PMMA) bone cement was used to anchor the pedicle screws in the module. The specimens were divided into two groups for the mechanical tests: the experimental group (second-generation CICPS) and control group (first-generation CICPS). Safety was evaluated using maximum shear force, static bending, and dynamic bending tests. Biomechanical stability evaluations included the maximum axial pullout force and rotary torque tests. X-ray imaging and computed tomography were used to evaluate the distribution of bone cement 24 h after PMMA injection, and stress distribution at the screw fracture and screw–cement–bone interface was assessed using finite element analysis.

Results

Mechanical testing revealed that the experimental group (349.8 ± 28.6 N) had a higher maximum axial pullout force than the control group (277.3 ± 8.6 N; P < 0.05). The bending moments of the experimental group (128.5 ± 9.08 N) were comparable to those of the control group (113.4 ± 20.9 N; P > 0.05). The screw-in and spin-out torques of the experimental group were higher than those of the control group (spin-in, 0.793 ± 0.015 vs. 0.577 ± 0.062 N, P < 0.01; spin-out, 0.764 ± 0.027 vs. 0.612 ± 0.049 N, P < 0.01). Bone cement was mainly distributed at the front three-fifths of the screw in both groups, but the distribution was more uniform in the experimental group than in the control group. After pullout, the bone cement was closely connected to the screw, without loosening or fragmentation. In the finite element analysis, stress on the second-generation CICPS was concentrated at the proximal screw outlet, whereas stress on the first-generation CICPS was concentrated at the screw neck, and the screw–bone cement–bone interface stress of the experimental group was smaller than that of the control group.

Conclusion

These findings suggest that second-generation CICPS have higher safety and stability than first-generation CICPS and may be a superior choice for the treatment of osteoporosis.

Introduction

An aging population is increasingly becoming a global problem, and spinal degenerative diseases are now frequently diagnosed in older patients. Pain caused by intervertebral disc herniation, spinal stenosis, and spondylolysis impose a heavy burden on the economy, and seriously affect the health and quality of life of patients [1]. Surgery is indicated when conservative treatment fails to achieve spinal recovery.

Pedicle screw fixation has been widely used in the surgical treatment of spinal degenerative diseases [2,3,4,5]; however, patients with osteoporosis provide a greater challenge. Osteoporosis seriously affects bonding at the bone–screw interface and screw stability owing to the destruction of bone tissue microstructure and reduction of bone mass; this results in screw loosening, displacement, or protrusion [6, 7].

Polymethylmethacrylate (PMMA)-reinforced screws have been proven to be the most effective system for strengthening pedicle screws and achieving stable fixation [2, 8,9,10,11]. PMMA bone cement penetrates the bone trabeculae to form a cement–bone interface in addition to the screw–bone interface, improving the stability of pedicle screws in osteoporotic vertebrae [12,13,14].

Since the initial application of first-generation cement-injectable cannulated pedicle screws (CICPS) enhanced with PMMA in clinical practice in 2011, the incidence of screw loosening, bone cement leakage, and even pulmonary embolism has reduced [15,16,17,18].

In 1975, Liu et al. [19] constructed the first three-dimensional finite element model of the vertebral body for the study of direct shear resistance in the lumbar spine, marking the beginning of finite element analysis in spinal biomechanics. This analysis has been increasingly used with the rapid development of computer technology [20, 21] and involves the creation of a detailed and accurate computer-simulated three-dimensional stereo model [22, 23]. Stress analysis can then be performed using this model, which greatly assists in the correct selection of surgical methods, reducing the rates of surgical failure and complications, and optimizing implant design.

To further optimize screw design and reduce complications, we have improved the design of first-generation CICPS to create second-generation CICPS. This study aimed to evaluate screw safety through biomechanical experiments and analyze stress distribution at the screw fracture and screw–cement–bone interface using finite element analysis.

Methods

Screw design

Second-generation CICPS design features a dual thread with an inner diameter of 1.6 mm, a thread lead of 6 mm, and a pitch of 1.5 mm in the proximal two-fifths and 3 mm in the distal three-fifths. The three side holes in the front two-fifths of the screw are separated from each other by 120° and two threads. From the near to far sides, the holes are round (2 × 2 mm), oval (3 × 2 mm), and U-shaped (4 × 2 mm) and can be designed with different specifications. All pedicle screws used in this study were manufactured by Chongqing FWS Medical Device Co., Ltd. (Chongqing, China) and had a diameter of 6.5 mm and a length of 45 mm (Fig. 1).

Fig. 1
figure 1

The proximal, middle, and distal side holes are shown in the left, middle, and right panels, respectively

Specimen preparation

Commercially available polyurethane (PU; catalog no. 1522-507; open cell 7.5 PCF, 18 × 13 × 4 cm; Sawbones, Pacific Research Laboratories, Vashon Island, WA, USA) was used to simulate the mechanical properties of osteoporotic cancellous bone. An open cone was used to drill holes in the side of the specimen, and a nail hole with a depth of 40 mm was prepared perpendicular to the side of the specimen. After installation of the screw placement instrument, the pedicle screw was slowly screwed vertically into the cancellous bone substitute, until the screw head was in contact with the surface of the specimen. During screw placement, the specimen structure was protected from damage by left–right shaking. All specimens were used self-tapping screws to avoid weakening the screw-holding force (Fig. 2A).

Fig. 2
figure 2

Screw placement was done according to the standard clinical method (A) and was followed by injection of 1.5 mL PMMA into each screw at room temperature (B)

Bone cement injection

At room temperature, 20 g PMMA bone cement and 10 mL solvent were mixed in a stainless-steel bowl, with a strict powder to liquid ratio of 2:1. When the bone cement became a paste, 10 mL was placed into a syringe, and the air was removed from the connecting device and screw. When the bone cement had developed the consistency of toothpaste, 1.5 mL was injected into each screw (Fig. 2B). After allowing the bone cement to set for 24 h, X-ray imaging and computed tomography (CT) were performed to observe the distribution of the bone cement in the module.

Mechanical testing

Static bending test

This was performed using an electronic universal testing machine (model no. CMT-5105). The long axis of the adjusting screw was perpendicular to the bending direction, and the load was applied at a constant rate of 25 mm/min perpendicular to the long axis of the pedicle screw. A computer automatically recorded the continuity value and drew the load–displacement curve until the load dropped significantly (Fig. 3A).

Fig. 3
figure 3

Experimental setups for the static bending test (A) and rotary torque test (B)

Dynamic bending test

This was performed using a fatigue tester (model no. Fd3000-P), with the screw installed in the same manner as in the above test, and the maximum loads set at 65%, 75%, and 80% of the yield load. The loading wave was sinusoidal, R-value was 0.1, and frequency was 10 Hz. Dynamic tests were conducted for each screw in turn, and the load–cycle curves were drawn using a computer analysis of the results until the screw broke or the cycle stopped at 2.5 × 106 cycles.

Rotary torque test

This was performed using a surgical implant electromagnetic dynamic mechanics test system (model no. M-100T). The adjusting screw was perpendicular to the specimen surface and a force of 11.17 N was applied vertically at a constant torsional speed of 3 r/min. The computer automatically recorded the continuity value and drew the torque–angular displacement curve; the recording stopped when all threads had entered the module. Unscrewing was performed in the same manner until the threads had completely exited the module (Fig. 3B).

Maximum axial pullout force test

This was performed using an electronic universal testing machine (model no. CMT-5105). The long axis of the adjusting screw was parallel to the stretching direction. The screw was pulled out along the long axis of the pedicle screw at a constant rate of 5 mm/min. The computer automatically recorded the continuity value and drew the load–displacement curve until the load dropped significantly (Fig. 4A).

Fig. 4
figure 4

Experimental setups for the maximum axial pullout force test (A) and maximum shear force test (B)

Maximum shear force test

The test element was a fatigue tester (model no. FD5000-P) with maximum loads of 1000, 1500, and 2000 N. The loading wave was a sine wave with an R-value of 0.1 and a frequency of 5 Hz. Dynamic tests were carried out for each screw in turn, and the load–cycle curves were drawn by computer analysis of the results until the screw broke or the cycles stopped at 2.5 × 106 cycles (Fig. 4B).

Finite element analyses

The CT scan results at 0.625 mm intervals were imported into SolidWorks geometric model processing software version 1.0 (SolidWorks Corp., Waltham, MA, USA), and ANSYS Workbench software version 19.0 (Ansys, Inc., Canonsburg, PA, USA) was used for the finite element calculation. Screw and screw–cement–bone models were established by meshing. A grid convergence analysis was conducted, and the network type was a 10-node tetrahedral element, as shown in Fig. 5A. The results indicated that the maximum stress difference was less than 5% and that the model converged. Considering the calculation accuracy, speed, and stress distribution of the key parts, a 0.04 mm element size was adopted for mesh division in all calculations, and 0.04 mm was also used for all contact surface elements [24, 25].

Fig. 5
figure 5

Convergence analysis results of the finite element mesh (A), screw model (B), and screw–bone cement–bone model (C)

Screw (Fig. 5B) and screw–bone cement–bone models (Fig. 5C) were established according to the elastic modulus and Poisson’s ratio of each material in the model [26, 27]. The screw model is a homogeneous, continuous, isotropic, and linearly elastic material [17, 18]. Bone is also considered to be homogeneous, isotropic, and linearly elastic [28]. In the screw–bone cement–bone model, friction contact exists between the screw, bone cement, and bone block [18, 29]. The maximum shear force test was used to evaluate the stress distribution at the screw fracture, and the maximum axial pullout test was used to evaluate the stress distribution at the screw–cement–bone interface.

Table 1 lists the mechanical properties of the investigated material [17, 29].

Table 1 Properties of materials used in this study

Statistical analysis

All statistical analyses were performed using SPSS software version 21 (IBM Corp., Armonk, NY, USA); a P value < 0.05 was considered significant. Comparisons of all indices between the two groups were performed using a two-tailed Student’s t-test or analysis of variance.

Results

Mechanical testing

Static bending test

As shown in Fig. 6, no significant differences were noted in bending stiffness (128.5 ± 9.08 vs. 113.4 ± 20.9 N), bending structure stiffness (3.13 ± 0.22 vs. 2.76 ± 0.508 N), and ultimate bending moment (268.9 ± 15.1 vs. 289.0 ± 23.0 N) between the two generations of screws, indicating that the two generations of CICPS had similar bending moments (P > 0.05).

Fig. 6
figure 6

Load–displacement curves of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Dynamic bending test

As presented in Fig. 7, the second-generation CICPS passed 75% of the test of its ultimate bending moment, with fracture at the proximal side hole in 85% and 95% of the tests (Fig. 8A). The first-generation CICPS passed 50% of the test of its ultimate bending moment, and the screw neck broke in 60% and 70% of the tests (Fig. 8B). Second-generation CICPS can, therefore, withstand higher loads and more cycles before failure.

Fig. 7
figure 7

Load–cycle curves of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Fig. 8
figure 8

Second-generation screw fracture (A) and first-generation screw fracture (B)

Rotary torque test

As presented in Fig. 9, the torque of the second-generation CICPS (0.793 ± 0.015 N) was higher than that of the first-generation CICPS (0.577 ± 0.062 N; P < 0.01). The spin-out torque of second-generation CICPS (0.764 ± 0.027 N) was also higher than that of first-generation CICPS (0.612 ± 0.049 N; P < 0.01).

Fig. 9
figure 9

Torque–angular displacement curves of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Maximum axial pullout force test

As shown in Fig. 10, the maximum axial pullout force of second-generation CICPS (349.8 ± 28.6 N) was higher than that of first-generation CICPS (277.3 ± 8.6 N; P < 0.05).

Fig. 10
figure 10

Load–displacement curves of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Maximum shear force test

As presented in Fig. 11, the two generations of screws passed the 1000 N load test and broke in the 1500 and 2000 N tests; however, the second-generation CICPS was able to withstand more cycles.

Fig. 11
figure 11

Load–cycle curves of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Finite element analyses

The results of the simulated dynamic bending test (Fig. 12) show that the overall stress distribution of the second-generation CICPS was more uniform, and the stress distribution of the proximal side hole was equal to the maximum stress. The stress distribution of the first-generation CICPS was concentrated in the neck, and the stress on the proximal side hole was significantly lower than the maximum stress.

Fig. 12
figure 12

Stress cloud diagrams of first-generation (A) and second-generation (B) cement-injectable cannulated pedicle screws

The results of the simulated maximum axial pullout force test (Fig. 13) showed that under the same displacement, the screw–cement interface stress of the second-generation CICPS was smaller than that of the first-generation CICPS and less likely to fail.

Fig. 13
figure 13

Stress cloud diagrams of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Discussion

Transpedicular screw instrumentation is regarded as a proven surgical procedure in the treatment of spinal diseases, achieving rigid stability in the anterior, middle, and posterior layers of the spine. However, surgical failure due to loosening, breaking, or pulling of the screw can occur, with screw pullout being one of the most common clinical complications [30, 31].

Many studies have reported methods to improve pedicle screw stability, including changing the shape of the screw, redesigning the screw threads, increasing the screw diameter, and reducing the screw pitch [12, 32,33,34,35,36,37]. Brasiliense et al. [38] used human vertebrae and PU foam blocks to compare screw insertion torque and maximum pulling force between conventional and double-threaded screws. Their results showed that the use of double-threaded screws significantly increased the insertion torque. Ramaswamy et al. [13] used PU at densities of 0.32, 0.24 and 0.16 g/cm3 to simulate normal bone, reduced bone mass, and osteoporotic bone, respectively, to compare screw fixation stability. They concluded that screw fixation strength significantly correlated with bone density and thread design. An osteoporosis-simulating PU bone block was used in our study. The second-generation screw adopts a double-thread design because of its smaller pitch and greater number of threads, thus, resulting in a larger contact area with the bone block, greater influence of friction at the screw–bone interface, and higher screw-in torque than that of the first-generation screw. A higher torque also means a stronger fixation effect, reducing the probability of screw pullout and further improving screw stability.

In the static bending experiment, the limit bending moment of the second-generation screw was lower than that of the first-generation screw, indicating that the first-generation screw should have better ductility. However, in the dynamic bending experiment, the force of the limit bending moment of the first-generation screw was less than that of the second-generation screw; in the maximum shear force experiment, the second-generation screw could withstand more cycles under the same force. Unlike conventional screw breakage that occurs in the neck, breakage of the second-generation screw occurs in the proximal lateral hole of the screw. We have reason to believe that the cause of this phenomenon is that the double-thread design changed the original stress distribution, making the stress distribution of the second-generation screw more uniform. Screw fracture occurred in the proximal side hole because it is closer to the thread change; structural changes are more susceptible to the influence of mechanics and are also more likely to exhibit morphological changes.

Many studies have reported the use of PMMA to improve the stability of osteoporotic spinal pedicle screws [39,40,41,42,43,44,45]. Liu et al. [45] found a significant positive correlation between screw stability and PMMA volume. However, an excessive increase in PMMA volume does not significantly improve screw stability but rather increases the risk of cement leakage. Therefore, 1.5 mL is the optimal volume for PMMA injection to ensure the stability of the screw and safety of the cement. In the first-generation CICPS, the cement was concentrated on the side of the screw, which increased the leakage rate of the cement. Therefore, we improved the design such that the three side holes were evenly distributed on the surface of the screw, aiming to alter the distribution of the cement. In our study, the screws were assessed by X-ray imaging and CT after injection of 1.5 mL PMMA. PMMA showed a spherical distribution in the second-generation CICPS (Fig. 14A) and a conical distribution in the first-generation CICPS (Fig. 14B). In the second-generation CICPS, PMMA was more evenly distributed and broader along the screw axis. In the maximum pullout force experiment, the second-generation screw required a greater pullout force, indicating that under the same volume, the distribution form and scope of the cement can change the maximum pullout force of the screw, enabling the screw to achieve greater stability. After pullout, the cement was tightly connected to the screw, without loosening or cracking (Fig. 15).

Fig. 14
figure 14

CT scans of PMMA distribution around the second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

Fig. 15
figure 15

Removal from the bone block of second-generation (A) and first-generation (B) cement-injectable cannulated pedicle screws

In the finite element analysis, the simulated maximum shear force test showed that the neck maximum stress value of 1029.7 N for the first-generation screw was much higher than the maximum stress value of the proximal side hole at 424, 27 N, whereas the neck maximum stress value of 1046.2 N for the second-generation screw was not significantly different from the maximum stress value of the proximal side hole at 1002.9 N. This result is compatible with the mechanical experiments, which further shows that the double-thread design can change the stress distribution and failure form of the screws. The simulated maximum pullout force test of the screw–cement–bone interface as a whole showed that the force of second-generation CICPS (157.26 N) is less than the force of first-generation CICPS(282.95 N). The results show that more evenly distributed cement allows the screw–cement–bone interface to bear more uniform forces and results in greater stability between the internal structures.

Limitations

First, simulated osteoporotic cancellous bone was selected for use in this study; however, the actual vertebral body is composed of cancellous bone and skin mass, which may result in some deviations from the experimental data. Second, the stress process was the result of a series of complex and comprehensive effects, and mechanical tests were completed under specific conditions and parameter sets; thus, the experimental results are not fully applicable to clinical practice. Finally, it is difficult to accurately simulate the real properties of materials using their geometric properties, which means that further experiments are required.

Future studies should collect more data in animal models of osteoporosis to further verify the safety and stability of second-generation CICPS prior to clinical application.

Conclusions

The double-thread design results in a stronger pullout force, and the evenly arranged side holes make the cement package more uniform, resulting in less screw–cement–bone interface stress. The experimental results suggest that second-generation CICPS are safer and more stable than first-generation CICPS and represent a better choice for the surgical treatment of osteoporosis and spinal degenerative diseases.

Availability of data and materials

The datasets supporting the conclusions of this article are included within the article.

Abbreviations

CICPS:

Cement-injectable cannulated pedicle screws

CT:

Computed tomography

PMMA:

Polymethylmethacrylate

PU:

Polyurethane

References

  1. Koh K, Yamada K, Enomoto T, Kawai A, Hamaoka S, Chiba S, Iseki M. Sex-specific impact of pain severity, insomnia, and psychosocial factors on disability due to spinal degenerative disease. Pain Res Manag. 2020;2020:8496527. https://doi.org/10.1155/2020/8496527.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Elder BD, Lo SF, Holmes C, Goodwin CR, Kosztowski TA, Lina IA, Locke JE, Witham TF. The biomechanics of pedicle screw augmentation with cement. Spine J. 2015;15(6):1432–45. https://doi.org/10.1016/j.spinee.2015.03.016.

    Article  PubMed  Google Scholar 

  3. Gaines RW Jr. The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am. 2000;82(10):1458–76. https://doi.org/10.2106/00004623-200010000-00013.

    Article  PubMed  Google Scholar 

  4. Nakashima H, Yukawa Y, Imagama S, Kanemura T, Kamiya M, Yanase M, Ito K, Machino M, Yoshida G, Ishikawa Y, Matsuyama Y, Ishiguro N, Kato F. Complications of cervical pedicle screw fixation for nontraumatic lesions: a multicenter study of 84 patients. J Neurosurg Spine. 2012;16(3):238–47. https://doi.org/10.3171/2011.11.spine11102.

    Article  PubMed  Google Scholar 

  5. Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ. Pedicle screw loosening: a clinically relevant complication? Eur Spine J. 2015;24(5):1005–16. https://doi.org/10.1007/s00586-015-3768-6.

    Article  PubMed  Google Scholar 

  6. Di Silvestre M, Parisini P, Lolli F, Bakaloudis G. Complications of thoracic pedicle screws in scoliosis treatment. Spine (Phila Pa 1976). 2007;32(15):1655–61. https://doi.org/10.1097/BRS.0b013e318074d604.

    Article  PubMed  Google Scholar 

  7. Reitman CA, Nguyen L, Fogel GR. Biomechanical evaluation of relationship of screw pullout strength, insertional torque, and bone mineral density in the cervical spine. J Spinal Disord Tech. 2004;17(4):306–11. https://doi.org/10.1097/01.bsd.0000090575.08296.9d.

    Article  PubMed  Google Scholar 

  8. Bostelmann R, Keiler A, Steiger HJ, Scholz A, Cornelius JF, Schmoelz W. Effect of augmentation techniques on the failure of pedicle screws under cranio-caudal cyclic loading. Eur Spine J. 2017;26(1):181–8. https://doi.org/10.1007/s00586-015-3904-3.

    Article  PubMed  Google Scholar 

  9. Hamasaki T, Tanaka N, Kim J, Okada M, Ochi M, Hutton WC. Pedicle screw augmentation with polyethylene tape: a biomechanical study in the osteoporotic thoracolumbar spine. J Spinal Disord Tech. 2010;23(2):127–32. https://doi.org/10.1097/BSD.0b013e31819942cd.

    Article  PubMed  Google Scholar 

  10. Chao KH, Lai YS, Chen WC, Chang CM, McClean CJ, Fan CY, Chang CH, Lin LC, Cheng CK. Biomechanical analysis of different types of pedicle screw augmentation: a cadaveric and synthetic bone sample study of instrumented vertebral specimens. Med Eng Phys. 2013;35(10):1506–12. https://doi.org/10.1016/j.medengphy.2013.04.007.

    Article  PubMed  Google Scholar 

  11. Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, Redl H. Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J. 2008;17(11):1462–9. https://doi.org/10.1007/s00586-008-0769-8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, Chen WJ. Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord. 2011;12:33. https://doi.org/10.1186/1471-2474-12-33.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Ramaswamy R, Evans S, Kosashvili Y. Holding power of variable pitch screws in osteoporotic, osteopenic and normal bone: are all screws created equal? Injury. 2010;41(2):179–83. https://doi.org/10.1016/j.injury.2009.08.015.

    Article  PubMed  Google Scholar 

  14. Goost H, Deborre C, Wirtz DC, Burger C, Prescher A, Fölsch C, Pflugmacher R, Kabir K. PMMA-augmentation of incompletely cannulated pedicle screws: a cadaver study to determine the benefits in the osteoporotic spine. Technol Health Care. 2014;22(4):607–15. https://doi.org/10.3233/thc-140815.

    Article  CAS  PubMed  Google Scholar 

  15. Liu YY, Xiao J, Yin X, Liu MY, Zhao JH, Liu P, Dai F. Clinical efficacy of bone cement-injectable cannulated pedicle screw short segment fixation for lumbar spondylolisthesis with osteoporosise. Sci Rep. 2020;10(1):3929. https://doi.org/10.1038/s41598-020-60980-w.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Wang Z, Liu Y, Rong Z, Wang C, Liu X, Zhang F, Zhang Z, Xu J, Dai F. Clinical evaluation of a bone cement-injectable cannulated pedicle screw augmented with polymethylmethacrylate: 128 osteoporotic patients with 42 months of follow-up. Clinics (Sao Paulo). 2019;74:e346. https://doi.org/10.6061/clinics/2019/e346.

    Article  PubMed  Google Scholar 

  17. Amaritsakul Y, Chao CK, Lin J. Biomechanical evaluation of bending strength of spinal pedicle screws, including cylindrical, conical, dual core and double dual core designs using numerical simulations and mechanical tests. Med Eng Phys. 2014;36(9):1218–23. https://doi.org/10.1016/j.medengphy.2014.06.014.

    Article  PubMed  Google Scholar 

  18. Zhou J, Wang Y. Finite element analysis of application and research progress in orthopedics (in Chinese). Pract Hosp Clin J. 2018;15(01):205–8.

    Google Scholar 

  19. Liu YK, Ray G, Hirsch C. The resistance of the lumbar spine to direct shear. Orthop Clin North Am. 1975;6(1):33–49.

    Article  CAS  PubMed  Google Scholar 

  20. Masni A, Tanaka M. Biomechanical investigation on the influence of the regional material degeneration of an intervertebral disc in a lower lumbar spinal unit: a finite element study. Comput Biol Med. 2018;98:26–38.

    Article  Google Scholar 

  21. Salvatore G, Berton A, Giambini H, et al. Biomechanical effects of metastasis in the osteoporotic lumbar spine: a finite element analysis. BMC Musculoskelet Disord. 2018;19(1):38.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Campbell JQ, Petrella AJ. An automated method for landmark identification and finite-element modeling of the lumbar spine. IEEE Trans Biomed Eng. 2015;62(11):2709–16.

    Article  PubMed  Google Scholar 

  23. Zheng J, Yang Y, Lou S, et al. Construction and validation of a three-dimensional finite element model of degenerative scoliosis. J Orthop Surg Res. 2015;10:189.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Grewal AS, Sabbaghian M. Load distribution between threads in threaded connections. J Press Vessel Tech-Trans Asme. 1997;119:91–5.

    Article  Google Scholar 

  25. Macdonald KA, Deans WF. Stress-analysis of drillstring threaded connections using the finite-element method. Eng Fail Anal. 1995;2:1–30.

    Article  CAS  Google Scholar 

  26. Yan YB, Teo EC, Qiu TX, Wu ZX, Qi W, Liu D, Lei W. Finite element study on the amount of injection cement during the pedicle screw augmentation. J Spinal Disord Tech. 2013;26:29–36.

    Article  PubMed  Google Scholar 

  27. Hashemi A, Bednar D, Ziada S. Pullout strength of pedicle screws augmented with particulate calcium phosphate: an experimentalstudy. Spine J. 2009;9:404–10.

    Article  PubMed  Google Scholar 

  28. Ouldyerou A, Mehboob H, Merdji A, Aminallah L, Mehboob AM, Mukdadi O. Biomechanical analysis of printable functionally graded material (FGM) dental implants for different bone densities. Comput Biol Med. 2022;150:106111. https://doi.org/10.1016/j.compbiomed.2022.106111.

    Article  CAS  PubMed  Google Scholar 

  29. Zheng C, Ma HY, Du YQ, Sun JY, Luo JW, Qu DB, Zhou YG. Finite element assessment of the screw and cement technique in total knee arthroplasty. Biomed Res Int. 2020;2020:3718705. https://doi.org/10.1155/2020/3718705.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  30. Renner SM, Lim TH, Kim WJ, Katolik L, An HS, Andersson GB. Augmentation of pedicle screw fixation strength using an injectable calcium phosphate cement as a function of injection timing and method. Spine. 2004;29(11):E212-216. https://doi.org/10.1097/00007632-200406010-00020.

    Article  PubMed  Google Scholar 

  31. Dickman CA, Fessler RG, Macmillan M, Haid RW. Transpedicular screw-rod fixation of the lumbar spine: Operative technique and outcome in 104 cases. J Neurosurg. 1992;77:860–70. https://doi.org/10.3171/jns.1992.77.6.0860.

    Article  CAS  PubMed  Google Scholar 

  32. McLain RF, Fry MF, Moseley TA, Sharkey NA. Lumbar pedicle screw salvage: pullout testing of three different pedicle screw designs. J Spinal Disord. 1995;8(1):62–8.

    CAS  PubMed  Google Scholar 

  33. Mehta H, Santos E, Ledonio C, Sembrano J, Ellingson A, Pare P, et al. Biomechanical analysis of pedicle screw thread differential design in an osteoporotic cadaver model. Clin Biomech. 2012;27(3):234–40.

    Article  CAS  Google Scholar 

  34. Yaman O, Demir T, Arslan AK, Iyidiker MA, Tolunay T, Camuscu N, et al. The comparison of pullout strengths of various pedicle designs on synthetic foams and ovine vertebrae. Turk Neurosurg. 2015;25(4):532–238. https://doi.org/10.5137/1019-5149.JTN.8907-13.1.

    Article  PubMed  Google Scholar 

  35. Bianco RJ, Arnoux PJ, Wagnac E, Mac-Thiong JM, Aubin CE. Minimizing pedicle screw pullout risks: a detailed biomechanical analysis of screw design and placement. Clin Spine Surg. 2017;30(3):E226–32. https://doi.org/10.1097/BSD.0000000000000151.

    Article  PubMed  Google Scholar 

  36. Abshire BB, McLain RF, Valdevit A, Kambic HE. Characteristics of pullout failure in conical and cylindrical pedicle screws after full insertion and back-out. Spine J. 2001;1(6):408–14. https://doi.org/10.1016/s1529-9430(01)00119-x.

    Article  CAS  PubMed  Google Scholar 

  37. Tsai WC, Chen PQ, Lu TW, Wu SS, Shih KS, Lin SC. Comparison and prediction of pullout strength of conical and cylindrical pedicle screws within synthetic bone. BMC Musculoskelet Disord. 2009;10:44. https://doi.org/10.1186/1471-2474-10-44.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Brasiliense LB, Lazaro BC, Reyes PM, Newcomb AG, Turner JL, Crandall DG, et al. Characteristics of immediate and fatigue strength of a dual-threaded pedicle screw in cadaveric spines. Spine J. 2013;13(8):947–56. https://doi.org/10.1016/j.spinee.2013.03.010.

    Article  PubMed  Google Scholar 

  39. Cook SD, Salkeld SL, Stanley T, et al. Biomechanical study of pedicle screw fixation in severely osteoporotic bone. Spine J. 2004;4:402–8. https://doi.org/10.1016/j.spinee.2003.11.010.

    Article  PubMed  Google Scholar 

  40. Frankel BM, D’Agostino S, Wang C. A biomechanical cadaveric analysis of polymethylmethacrylate-augmented pedicle screw fixation. J Neurosurg Spine. 2007;7:47–53. https://doi.org/10.3171/SPI-07/07/047.

    Article  PubMed  Google Scholar 

  41. Chang MC, Liu CL, Chen TH. Polymethylmethacrylate augmentation of pedicle screw for osteoporotic spinal surgery: a novel technique. Spine. 2008;33:E317-24. https://doi.org/10.1097/BRS.0b013e31816f6c73.

    Article  PubMed  Google Scholar 

  42. Chen YL, Chen WC, Chou CW, et al. Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws. Med Eng Phys. 2014;36:1416–20. https://doi.org/10.1016/j.medengphy.2014.05.003.

    Article  PubMed  Google Scholar 

  43. Fölsch C, Goost H, Figiel J, et al. Correlation of pullout strength of cement-augmented pedicle screws with CT-volumetric measurement of cement. Biomed Tech (Berl). 2012;57:473–80. https://doi.org/10.1515/bmt-2012-0012.

    Article  PubMed  Google Scholar 

  44. Liu D, Zhang XJ, Liao DF, et al. Biomechanical comparison of pedicle screw augmented with different volumes of polymethylmethacrylate in osteoporotic and severely osteoporotic synthetic bone blocks in primary implantation: an experimental study. Biomed Res Int. 2016. https://doi.org/10.1155/2016/9627504.articleID9627504.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Liu D, Zhang B, Xie QY, et al. Biomechanical comparison of pedicle screw augmented with different volumes of polymethylmethacrylate in osteoporotic and severely osteoporotic cadaveric lumbar vertebrae: an experimental study. Spine J. 2016;16:1124–32. https://doi.org/10.1016/j.spinee.2016.04.015.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This study was supported by the New Clinical Technology of the Army Medical University (No. 2015-2A-1-5 and 2019-3-1-34).

Author information

Authors and Affiliations

Authors

Contributions

LS, JX, WW, YJ, RZ, and FD contributed to conceptualization. LS performed data curation. FD performed funding acquisition. LS, JX, WW, YJ, RZ, and FD performed investigation. CL and FD provided methodology. CL, RZ, and FD performed validation. CL performed writing-original draft. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Rui Zhou or Fei Dai.

Ethics declarations

Ethics approval and consent to participate

This study did not involve in human samples.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, C., Song, L., Xiao, J. et al. Second-generation bone cement-injectable cannulated pedicle screws for osteoporosis: biomechanical and finite element analyses. J Orthop Surg Res 18, 343 (2023). https://doi.org/10.1186/s13018-023-03752-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-023-03752-2

Keywords