Biomechanical evaluation of a healed acetabulum with internal fixators: finite element analysis

Abstract

Background

Treatment of complicated acetabular fracture with internal fixation usually has high risk of failure because of unbefitting fixation. However, evaluation of the biomechanical effect of internal fixation under physiological loading for fracture healing is still generally rarely performed. The purpose of this study is to analyze the biomechanical characteristics of a healed acetabulum with designed internal fixators under gait and to explore the biomechanical relationship between the healed bone and the internal fixator.

Methods

A patient-specific finite element model of whole pelvis with designed internal fixators was constructed based on the tomographic digital images, in which the spring element was used to simulate the main ligaments of the pelvis. And the finite element analysis under both the combination loading of different phases and the individual loading of each phase during the gait cycle was carried out. The displacement, von Mises stress, and strain energy of both the healed bone and the fixation were calculated to evaluate the biomechanical characteristics of the healed pelvis.

Results

Under the combination loading of gait, the maximum difference of displacement between the left hip bone with serious injury and the right hip bone with minor injury is 0.122 mm, and the maximum stress of the left and right hemi-pelvis is 115.5 MPa and 124.28 MPa, respectively. Moreover, the differences of average stress between the bone and internal fixators are in the range of 2.3–13.7 MPa. During the eight phases of gait, the stress distribution of the left and right hip bone is similar. Meanwhile, based on the acetabular three-column theory, the strain energy ratio of the central column is relatively large in stance phases, while the anterior column and posterior column of the acetabular three-column increase in swing phases.

Conclusions

The acetabular internal fixators designed by according to the anatomical feature of the acetabulum are integrated into the normal physiological stress conduction of the pelvis. The design and placement of the acetabular internal fixation conforming to the biomechanical characteristics of the bone is beneficial to the anatomical reduction and effective fixation of the fracture, especially for complex acetabular fracture.

Introduction

The acetabulum is one of the most important bone in the human body, which undertakes the complex physiological function of weight-bearing walking. Complicated acetabular fractures caused by high energy and severe violence are often accompanied by high disability and mortality. At present, open reduction and internal fixation are often used to treat fracture in clinic [1, 2], and according to specific fractures, different internal fixation methods, such as screws, plates, and intramedullary nail, are applied. Because the acetabulum is a unique structure with irregular shape and complicated surrounding anatomical relation, and deep fracture often occurs with various injuries, the treatment of complicated acetabular fractures usually has high risk of failure. The loss of reduction and complications caused by improper internal fixation often occur, which seriously affect the life quality of patients [3,4,5]. The main reason for improper fracture fixation is the unfitting structure of internal fixator and its fixation method. In terms of biomechanical view, the design and placement of internal fixator should conform to the physiological force of the bone, so that the broken bone can be anatomical reduction and stable fixation, and restore the stress conduction of the bone. Hence, to improve treatment effect of fracture, it is of great significance to study biomechanical characteristic of fracture acetabulum with internal fixation.

Existing biomechanical researches on fracture internal fixation are usually using either experimental testing or finite element analysis. Biomechanical studies of different fixation methods are important to determine optimal fixation techniques and test novel implants in a preclinical setting [6]. Cai et al. [7], Fan et al. [8] and Marmor et al. [9] adopt cadaveric bone for biomechanical testing, which leads to difficulty in obtaining experimental samples. Krappinger et al. [10, 11], Alfonso et al. [12] and Wu et al. [13] choose synthetic pelvis models, which leads to high cost. Additionally, there is no existing standardized mechanical experimental method [1, 6]. Finite element analysis is a powerful tool for orthopedic biomechanical research [14, 15]. Kocsis [16], Bignardi et al. [17], Hedelin et al. [18], Zhang et al. [19] and Lei et al. [20] construct three-dimensional model of bone and fracture internal fixation based on CT data and numerical simulation is carried out, which is low cost, high repeatability and accurate results, but involves the problem of how to validate the finite element model and the applied loading. The existing biomechanical study of fracture internal fixation can provide some guidance for fracture treatment, but there are also some common problems, which arise from the following aspects.

  1. 1.

    Most of the physiological loads simulated in the current studies are relatively simple, focusing on one-legged or two-legged stance.

  2. 2.

    The evaluation criteria of fixation methods, is usually the maximum fixation stiffness, which is limited.

  3. 3.

    The relationship between the biomechanical characteristics of fracture internal fixation and the effect of bone healing is still not clear.

To evaluate the implantation effect of fixation method and explore the relationship between the biomechanical characteristics of internal fixation and the effect of bone healing, it is of great value to analyze the internal fixation implants and their fixation methods with good clinical results [21]. Such analysis not only helps understand the principle of fixation, but also provide guidance for design of internal fixation implants and precise treatment of fractures. However, due to the complexity of internal fixation of acetabular fracture, there is little biomechanical analysis and further research on the clinical cured cases. Therefore, this is a novel work presenting the in-depth biomechanical analysis of cured complex acetabular fracture and study on clinical internal fixation implant and fixation method.

A pelvis of a cured patient with complex acetabular fracture is selected as the research object in this paper. Three-dimensional model of healed pelvis with internal fixators was established by using three-dimensional reconstruction technique, and the finite element analysis under both the combination loading of different phases and individual loading of each phase during the gait cycle was carried out. The aims of the study are: (1) analysis of the biomechanical characteristics of acetabular internal fixator in the treatment of acetabular fractures under gait cycle loading, obtaining some relevant mechanical indexes such as deformation, stress and strain energy, and (2) exploration of the relationship between the biomechanical characteristics of internal fixation and the healing effect of acetabular fractures, so as to provide guidance to accurate treatment of acetabular fractures.

Materials and methods

This study was approved by the ethics committee of Changhai Hospital of Shanghai, China. Written informed consent was obtained from the patient prior to the study.

Surgical techniques

A 33-year-old female patient (height 162 cm, weight 55 kg) suffered a serious injury in the left acetabulum, multiple fractures and a slight injury in the right acetabulum due to a car accident. According to the fracture condition of the patient, the acetabular internal fixators were placed and no longer removed. The acetabular internal fixators used in this patient are designed by the variations of pelvic anatomy and anatomical feature of the acetabulum [22]. They are classified into three groups: bow-teeth nails for connection of two displaced fragments (Fig. 1a, b), anterior column or wall fixator (Fig. 1c), and posterior column or wall fixator (Fig. 1d). There are eleven internal fixators which is numbered 1–11 clockwise (Fig. 1e), including seven bow-teeth nails, one sacroiliac joint three-corner fixator connected by iliac arch and sacral arch, two anterior column or wall fixators composed of anterior column arch, arm branch and square-area top panel, and one posterior column or wall fixator composed of posterior column arch, fossa net and square-area back panel. In order to provide good anatomical reduction and stability, the acetabular internal fixators are mostly placed in the force line and point of acetabular anterior column by the doctor’s experience [23] (Fig. 1f), for example, fixators 2 and 3 are placed in the anterior of the sacroiliac joint, and the fixators 5 and 7 are placed at the arcuate line (Fig. 1e). The patient was cured after acetabular internal fixation, and the left acetabulum with severe injury achieved integrity reduction.

Fig. 1
figure 1

Acetabular internal fixators and acetabular force line and placement of internal fixators: a bow-teeth nails, b sacroiliac joint three-corner fixator, c anterior column or wall fixator, d posterior column or wall fixator, e placement of acetabular internal fixators, f force line and point of acetabular anterior column [23]

Geometric modeling of pelvis and internal fixators

The healed pelvis and acetabular internal fixators models were constructed using computer tomography (CT) data (with 1 mm slice thickness, including 376 images). The CT scanning data were imported into Mimics 14.0 software (Materialise, Belgium) to reconstruct the surface geometry models of the healed pelvis and internal fixators by region growing and mask editing. Semi-automatic segmentation of the CT data was performed to identify the boundary of sacrum, left and right hip bone by contour interpolation. The surface geometry models were exported as point cloud format files. Then these files were imported into Geomagic Studio 12 (Geomagic, USA) for smoothing and surface construction and modification, and converted into solid models, respectively.

Finite element model

The solid models of healed pelvis and internal fixators were meshed using HyperMesh 14.0 (Altair, USA). Due to the same CT images, the spatial position of the bone and the internal fixators is fused automatically. Additionally, the model was processed by Boolean operation, so the mesh of the contact between bone and internal fixators were set in the form of common nodes. The bone consists of cancellous and cortical parts in the finite element model. Solid elements are used for cancellous bone and internal fixators. Shell elements are clad around the solid elements to represent the cortical bone. The thickness of the cortical bone is assumed to be an average value of 1.5 mm [24]. The finite element model consists of 11,807,170 tetra4 solids elements and 1,374,437 tria3 shell elements (Table 1), for an average element size of 0.5 mm. The model is defined as linear elastic, homogeneous and isotropic [14, 15, 25]. The material properties are listed in Table 1 [24, 26].

Table 1 Material properties of hip bones and fixators [24, 26]

The pelvis includes the left and right hips and sacrum to form a complete pelvic ring. The pelvis is a stable ring structure due to the strong support of the sacroiliac ligament and the firm ligament connection at the pubic symphysis. The main ligaments, including the anterior and posterior sacroiliac, the superior and arcuate pubic, are modeled as linear spring elements with tension-only stiffness in finite element models, as shown in Fig. 2. The rigid elements are applied to ligament attachment area nodes based on the anatomical feature of the pelvis. The stiffness values for ligaments are estimated from existing literature [27, 28] (Table 2). The tied contact pairs at the sacroiliac joint and pubic symphysis are set in finite element model.

Fig. 2
figure 2

Finite element model of healed pelvis of the patient with acetabular internal fixators: a anterior view, b posterior view, c zoom view

Table 2 Material properties of the pelvic ligaments in the model [27, 28]

Loading and boundary conditions

Gait is one of the most common and essential activities in daily human life, periodic and regular. Each gait cycle consists of a sequence of ordered gait phases. Figure 3 shows the typical gait cycle includes eight phases of a single leg, starting from heel strike of the left leg and continuing until that heel strikes again [29]. Eight phases are initial contact (IC), load response (LR), mid-stance (MST), terminal stance (TST), pre-swing (PSW), initial swing (ISW), mid-swing (MSW), and terminal swing (TSW). The description and proportion of each phase of the gait cycle is listed in Table 3. The hip joint forces experienced by the pelvis for eight phases of gait are obtained using an existing healthy model (62 kg, 173 cm) which is freely available from the AnyBody software (AnyBody Technology, Denmark). It should be noted that the existing healthy model in AnyBody is different from the specific-patient in finite element modeling, however, because the focus of our study is on the distribution rules and the relative values of biomechanical response between the concerning parts, rather than the magnitude of them, it is acceptable for using the hip joint forces directly from the analysis results from AnyBody. The magnitude of the hip joint forces for the eight phases of the gait cycle is listed in Table 4. The direction of the hip joint forces is shown in Fig. 4. From the comparison, it is found that the trend of the hip joint force in the eight phases of the gait cycle obtained here is similar to that of literature [29]. Figure 4 shows the force vectors are applied on the node at location consistent with the femoral head. The node is attached to the acetabulum by the rigid element, which means the forces are applied to the relevant areas of the acetabulum. For the boundary condition, the nodes located on the superior surface of the sacrum were constrained in all DOF (Fig. 4).

Fig. 3
figure 3

Typical gait phases of the human (the left side is in red, the right side is in green)

Table 3 The description of each phase of the gait cycle [29]
Table 4 Applied forces of hip joints
Fig. 4
figure 4

Constraints and the hip contact force vector

Mesh sensitivity analysis

A mesh sensitivity analysis was carried out using the model of right acetabulum with different element size. The result of mesh sensitivity analysis is shown in Fig. 5. When the element size decreases from 3 to 1 mm, the value of displacement and stress increases greatly. However, when the element size decreases from 1 to 0.3 mm, the increase of displacement and stress is small. The sensitivity analysis shows that the convergence is achieved. Considering the geometric complexity and computational efficiency of the bone model, the element size 0.5 mm is selected for the finite element models in our study.

Fig. 5
figure 5

Mesh sensitivity analysis: a convergence curve between the displacement and the element size, b convergence curve between the von Mises stress and the element size

Results

To investigate the effect of internal fixators on acetabular fracture reduction and pelvic stability, the deformation, von Mises stress, and strain energy of the whole pelvis and the healed acetabulum with internal fixators are analyzed under eight gait cycle phases or a combination of all phases of the gait cycle. Note that the combination of all phases is composed of eight different phases in proportion (the corresponding weight factor is 0.11, 0.22, 0.13, 0.04, 0.11, 0.22, 0.13, 0.04, respectively) according to the occurrence of each phase in gait cycle (Table 3 lists the percentage [29]). The main purpose of this study is to analyze the biomechanical characteristics of the healed left hip bone after internal fixation; therefore, the right acetabulum with minor injury is compared as a relatively healthy bone approximately.

Displacement analysis

Figure 6 shows the displacement of pelvis with internal fixators under the combination of all phases of the gait cycle. The maximum displacement in the pelvis is 0.558 mm appears at the left hemi-pelvis near the ischial tuberosity (Fig. 6a). The displacement distribution in the left and right hemi-pelvis shows a slight variance. The maximum displacement of the right hemi-pelvis is 0.436 mm on the pubic symphysis (Fig. 6c). In comparison, the left acetabulum moves slightly upward after healing. In some gait cycle phases, the left hip joint forces are greater than that of the right. The left hemi-pelvis produces more movement along pubis from ischial tuberosity.

Fig. 6
figure 6

The displacement of pelvis with internal fixators under the combination of all phases of the gait cycle: a the pelvis with internal fixators, b the left healed hemi-pelvis, c the right healthy hemi-pelvis

Stress analysis

Figure 7 shows the von Mises stress distribution of the pelvis with internal fixators under the combination of all phases during the gait cycle. The high-stress regions are concentrated in the superior of the acetabulum, sacroiliac joint, the iliac arch of the arcuate line, and the great notch of the ischium (Fig. 7). A maximum stress of 174.89 MPa is observed at fixator 3 intersection with the sacral cortex. The secondary peak stress of 148.48 MPa is shown on the fixator 2 intersection with the iliac cortex. The results indicate that the peak stress is less than the yield strength (1050 MPa) and fatigue strength (315 MPa) of titanium alloy [19], and no failure occurs. The pelvic bone shows the maximum stress of 124.28 MPa at the acetabular fossa of the right hemi-pelvis, while the maximum stress of the left hemi-pelvis is 115.55 MPa on the acetabular fossa, which decreases by roughly 7% compared with that of the right hemi-pelvis.

Fig. 7
figure 7

The von Mises stress distribution of pelvis with internal fixators under the combination of all phases of the gait cycle: a anterior view, b posterior view, c lateral view (from right to left), d lateral view (from left to right)

To compare the stress on the bone and that on the fixator, the elements at the connection of internal fixators and bone are selected to calculate the average stress. Figure 8a shows the selected elements in fixator 5, where the yellow brightening elements are the selected elements in fixator 5, and the white brightening elements are the iliac cortical elements in contact with fixator 5. Figure 8b shows comparison of von Mises stresses on the selected elements under the combination of all phases during the gait cycle. It is found that stress on the internal fixators is higher than that on the bone, because the elastic modulus of the fixator is higher than that of bone. However, the stress differences between the bone and the fixators are not too large; for example, the average stress differences between the fixators and the bone are 13.7 MPa, 9.9 MPa, and 8.4 MPa in the fixator 2, 1, and 5, respectively, and the average stress difference between other fixators and bone are roughly 2.3–5.2 MPa. The results indicate that the position of the fixator impacts the stress distribution, but the stress shielding of acetabular internal fixation is minor. The biomechanical compatibility between fixator and bone is good.

Fig. 8
figure 8

Elements selection example and comparison of von Mises stresses at selected elements under the combination of all phases of the gait cycle: a element selection example, b comparison of von Mises stresses at selected elements under the combination of all phases of the gait cycle

Figure 9a–h shows the von Mises stress distribution of the left hemi-pelvis for each phase of the gait cycle, respectively, with the corresponding gait cycle indicated at the bottom center of each figure. During the stance phase, the left leg changes from flexion to extension. When flexion, the stress distribution on the lunate surface, superior acetabular dome, and iliac arch of the arcuate line are high (Fig. 9a, b). When extension, in addition to the previous observation, the stress distribution also displays on the sacroiliac joint and greater sciatic notch (Fig. 9c–e). During the swing phases, the left leg is all flexion, a noticeable stress reduction is observed around the lunate surface, superior acetabular dome, and the sacroiliac joint (Fig. 9f–h). On the other hand, the high-stress regions decrease during the swing phases. In the TST phase, with the left foot single support, the larger the extension angle, the greater is the joint force (the peak force magnitude of approximately 2169N), thus the maximum stress is 110.97 MPa. In the MSW phase, with the left foot in the air (the force magnitude of approximately 54N), the maximum stress gradually decreases to 6.94 MPa. From the medial view of the pelvis, the stress distribution is mainly transmitted from the acetabulum along the arcuate line upward to sacroiliac joint. From the lateral view of the pelvis, the stress distribution is transmitted from the acetabulum upward to anterior inferior iliac spine during flexion, and along the greater sciatic notch to posterior inferior iliac spine during extension. The results indicate that the stress distribution and conduction of the pelvis under gait are related to the biomechanics of the hip joint during walking. The hip joint plays an important role in the weight-bearing, multi-directional movement and balance function of the lower limbs of the human body.

Fig. 9
figure 9

Von Mises stress distribution of the left hemi-pelvis under each phase of gait cycle: a initial contact phase, b loading response phase, c mid-stance phase, d terminal stance phase, e pre-swing phase, f initial swing phase, g mid-swing phase, h terminal swing phase

Human walking is a coordinated movement of alternating left and right legs. As shown in Fig. 3 and Table 3, when the left leg is at IC phase, the right leg is at PSW phase, and similarly for the other phases. In comparison, Fig. 10a–h shows the von Mises stress distribution of the right hemi-pelvis for each phase of the gait cycle, with the corresponding gait cycle indicated at the bottom center of each figure. Under the PSW phase, it is observed that the left hemi-pelvis and right hemi-pelvis produce a greater variation of stress distribution (Figs. 9e, 10e), and the right hemi-pelvis displays noticeably stress reduction. Similar observation can be found in IC phase (Figs. 9a, 10a). For the rest of the gait phases, small differences can be observed between left and right hemi-pelvis. Because the left and right legs are not completely symmetric in simulation, there are significant differences in obtained right hip joint forces under the IC and PSW phases. The results indicate that the stress distribution of the left hip bone returns to normal during gait cycle.

Fig. 10
figure 10

Von Mises stress distribution of the right hemi-pelvis under each phase of gait cycle: a initial contact phase, b loading response phase, c mid-stance phase, d terminal stance phase, e pre-swing phase, f initial swing phase, g mid-swing phase, h terminal swing phase

Strain energy analysis

In order to guide the reduction and fixation of acetabular fractures, Zhang et al. [30] put forward the acetabular three-column theory from the view of acetabular anatomy and the practical treatment of acetabular fractures and combined with a large number of clinical experiences. According to the acetabular three-column theory, the anterior column is formed with the cortical thickening region of the ilium and the pubic bone migration, the central column is formed with the migration of the acetabular roof in loading-bearing direction, and the posterior column is formed with the sciatic migration. The left acetabulum is divided into the above three regions, as shown in Fig. 11a, in which the red region is the anterior column, the green region is the central column, and the blue region is the posterior column. When a structure is deformed under the action of external force, the strain energy is stored inside the structure. With strain energy as the mechanical index, the load-bearing capacity of structure can be quantitatively analyzed from the view of energy flow. Combined with the acetabular three-column theory, in order to explore the load-bearing capacity of the three-column during the gait, the percentages of the strain energy of the three columns in the total strain energy under eight phases of gait cycle are calculated according to Eq. (1),

$$\begin{aligned} E & = E_{{\text{a}}} + E_{{\text{c}}} + E_{{\text{p}}} \\ \mu_{i} & = \frac{{E_{i} }}{E}(i = {\text{a, c, p}}) \\ \end{aligned}$$
(1)

where Ea, Ec, and Ep are the strain energy of the anterior, central, and posterior columns, respectively. E is the total strain energy of the three columns, and μa, μc, and μp are the strain energy ratios of the anterior, central and posterior columns, respectively.

Fig. 11
figure 11

The acetabular three-column and strain energy ratios: a the components of the acetabulum in 3-column classification are marked with different colors, b strain energy ratio of acetabular three columns in the eight phases of the gait cycle

Strain energy ratios of three columns in the left acetabulum during the eight phases of the gait cycle are calculated by Eq. (1) and shown in Fig. 11b. It is found that the strain energy ratio of the central column decreases, and the anterior and posterior column increase gradually from the gait phase of IC to MSW, and the strain energy ratios of the central column are higher than that of the anterior and posterior column. The strain energy ratio of the central column in IC phase is closer to LR phase with the leg all flexion. From MST to PSW phase, the larger is the extension angle, the less is the strain energy ratio of the central column. In the swing phase, where the left leg changes from extension to flexion, the strain energy ratio of the central column decreases at first and then increases. In the MSW phase with the left foot in the air, the strain energy ratio of the central column is the smallest.

Discussion

Open reduction and internal fixation are the most common method for the treatment of acetabular fractures. However, for complex comminuted acetabular fractures, it is difficult to select single reconstruction plates and screws to achieve effective fixation and functional recovery of acetabular fractures. Zhang et al. creatively developed the acetabular internal fixators according to the anatomical characteristics of the acetabulum and studied the application by retrospective analysis and follow-up of clinical results [22]. Retrospective analysis or follow-up method based on medical imaging data combined with scoring system and criteria to evaluate the recovery of patients is often used by clinicians [31,32,33], which is relatively reliable, but the workload is large. It requires a large number of case statistics to guide fracture treatment by using this method, which is poor in timeliness. Based on the finite element modeling and analysis of a patient-specific pelvis with acetabular internal fixators, this study analyzed the biomechanical characteristics of acetabular internal fixators in the treatment of acetabular fractures and explored the biomechanical relationship between the healed bone and the internal fixator.

The healed pelvis model with internal fixators was established by the finite element method, and cortical bone and cancellous were distinguished. Spring elements were used to simulate the main ligaments of the pelvis. Finite element analysis was carried out for the evaluation of biomechanical response of the acetabular internal fixators under loads of the gait cycle. Watson et al. constructed a series of finite element models of the same human pelvis with different boundary conditions and compared the deformation, stress and strain of the hemi-pelvis under the 15% loading regime (15% of the gait cycle) and 48% loading regime (48% of the gait cycle). For the complete pelvic ring model, the region with the greatest deformation under both conditions was the ischial-pubic bone. The high stresses concentrated in the inferior ilium, along the superior regions of the acetabular rim, and around the sacroiliac joint [24]. Figure 9b, d shows the similar results, which indicates the finite element model in this study is relatively reliable.

Based on the displacement analysis of the pelvis with internal fixators under the combination of all phases during the gait cycle, it was found that there was a difference in the displacement distribution between the left healed hip bone and the right healthy hip bone because the left hip bone after the anatomical reduction is not the same as the right hip bone. However, clinical studies have found that the smaller the postoperative displacement of the pelvic ring, the better the stability of the internal fixation. If the displacement of the pelvic ring is less than the reduction of 1 cm, its long-term function will be improved [19]. This study shows that the maximum displacement of the whole pelvic ring is smaller than 1 mm, and the maximum displacement difference between the left and right acetabulum is 0.122 mm, which indicate that the stability of internal fixation is ideal, and the integrity of the acetabulum is restored.

When acetabular fractures are stabilized to achieve integrity reconstruction and anatomical reduction, the closer is the postoperative bone stress to the natural stress of healthy bone, the better is the clinical effect of the internal fixation [19]. In this study, the stress distribution of the left and right hemi-pelvis under the combination of all phases of the gait cycle is very similar, and the maximum stress difference between the left and right hemi-pelvis is only 7%. During different phases of the gait cycle, the stress distribution of the left and right acetabulum is also similar, only a slight difference in the high-stress region under IC and PSW phases. Furthermore, from the comparison of the average stress between each internal fixator and the left hip bone under the combination of all phases of the gait cycle, it is found that most differences are between 2.3–5.2 MPa, which indicates that the acetabular internal fixators provide a good mechanical environment and conforms to the stress conduction of the bone.

Clinicians usually classify and treat acetabular fractures based on the two-column concept proposed by Judet-Letournel [34]. In the two-column concept of acetabulum, the iliac crest, spina iliace, pubic bone and the anterior acetabular rim are defined as the anterior column, while the ischium, ischial spine, the posterior acetabular rim and the hard ischial notch are defined as the posterior column (Fig. 12). However, with the in-depth study of acetabular fractures, there are limitations for complex acetabular fractures in two-column classification. Zhang et al. have put forward the three-column theory based on the physiological development and anatomical characteristics of the acetabulum and clinical experience, dividing the acetabular roof's weight-bearing area and the bone thickening area above it into the central column. The classification of acetabular fractures based on three-column theory more clearly reflects the severity of fractures and helps to provide guidance for anatomical reduction and fixation of acetabular fractures [30]. This study analyzes the load-bearing capacity of the three columns during gait by strain energy. The results indicate that the strain energy ratio of central column is higher than that of the anterior and posterior columns, which provides support for the acetabular three-column theory from the point of view of numerical quantification. Combined with the stress analysis results, the superior region of acetabulum generates higher stress (Fig. 7). Under different gait phases, the stress conduction of the pelvis shows a certain rule: on the medial side, the stress is mainly transferred upward from acetabulum through the arch line to the sacroiliac joint; on the lateral side the stress is mainly transmitted from the acetabulum forward to the anterior inferior iliac spine or backward through the great notch of the ischium to the posterior inferior iliac spine. These results are consistent with the firmness of the anatomical cortex of the acetabulum and the direction of the pattern. Therefore, the acetabular three-column classification is helpful for the reduction and fixation of acetabular fractures.

Fig. 12
figure 12

Diagram of components of the acetabulum on the two-column classification

The quality of internal fixation and fracture healing are affected by postoperative activity. Any fixation methods of acetabular fractures need to be evaluated with the full pelvic ring under walking motions; thus, it is necessary to study the effects of walking with internal fixators attached so that the internal fixator can be better designed for earlier mobility [29]. However, the existing research is mainly focused on the finite element modeling and stress analysis of healthy pelvis under gait, and the evaluation of internal fixation methods rarely involves gait. Through the analysis of acetabular internal fixation during the gait cycle, it is found that the stress of the central and posterior column of the acetabulum is higher when the hip joint flexion, and the stress of the central and anterior column of the acetabulum is higher when the hip joint extension. Whenever flexion or extension, the maximum stress in the pelvis and internal fixators is less than the yield strength of the material. Therefore, it is suggested that extension training can be carried out earlier after using sacroiliac joint three-corner fixator for acetabular anterior column fractures. Flexion training can be carried out earlier after using the posterior column or wall fixator for acetabular posterior column fractures.

Although our results evaluated the biomechanical characteristics of the healed pelvis with internal fixators, simplifications and limitations of this study exist. First, unfortunately the material properties of the cortical and cancellous bone could not be accurately ascertained from the scan data, therefore homogeneous values from previous studies were used [24, 26]. In the existing studies, linear material is common and acceptable [29]. Second, the actual physiological environment of bone was simplified without considering the effect of cartilage in the model. However, as this study aims to evaluate the biomechanical characteristics of the healed pelvis with internal fixators, the cartilage is not the region of interest. Finally, because it is difficult to simulate the patient’s gait, the approximate joint forces of the left and right hip bones are adopted through the simulation of the healthy human in AnyBody. Additionally, the effect of pelvis muscle forces is not taken into consideration. The muscle appears to reduce stress concentrations, and produce a more even stress distribution on the pelvis [35]. The influence of the muscle forces on the pelvic stress conduction is not so great that it was not included in this study.

Conclusions

In conclusion, this study successfully uses finite element analysis to evaluate the biomechanical characteristics of acetabular internal fixation for the treatment of a complex acetabular fracture. The results show the stress distribution of healed bone is different under each phase of gait cycle, the biomechanical evaluation of healed acetabulum with internal fixators under gait cycle is helpful for acetabular fracture fixation and functional training. It is also found that, under both the combination loading of different phases and the individual loading of each phase during the gait cycle, the stress distribution of the left and right hip bone is very similar, and the average stress difference between the bone and internal fixator is within a certain range. The acetabular internal fixation is integrated into the normal physiological stress conduction of the pelvis. The design and placement of the acetabular internal fixation conforming to the biomechanical characteristics of the bone is beneficial to the anatomical reduction and effective fixation of the fracture, especially for complex acetabular fracture.

It should be noted that although the patient-specific CT data were used to create the finite element model in this study, this is a special case. Although some mechanical rules can be revealed and some similar cases should be able to get more general conclusions, the proposed method is feasible and will be further explored in the future.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

References

  1. Ferrante H, et al. Biomechanical analysis of fixation methods for acetabular fractures: a review. Med Eng Phys. 2021;89:51–62.

    Article  PubMed  Google Scholar 

  2. Cimerman M, et al. Fractures of the acetabulum: from yesterday to tomorrow. Int Orthop. 2021;45(4):1057–64.

    Article  PubMed  Google Scholar 

  3. Gueorguiev B, Lenz M. Why and how do locking plates fail? Inj-Int J Care Inj. 2018;49:S56–60.

    Article  Google Scholar 

  4. Gardner MJ, Evans JM, Dunbar RP. Failure of fracture plate fixation. J Am Acad Orthop Surg. 2009;17(10):647–57.

    Article  PubMed  Google Scholar 

  5. Huang JW, Gao XS, Yang YF. Early prediction of implant failures in geriatric intertrochanteric fractures with single-screw cephalomedullary nailing fixation. Inj-Int J Care Inj. 2022;53(2):576–83.

    Article  Google Scholar 

  6. Hinz N, et al. Biomechanical analysis of fixation methods in acetabular fractures: a systematic review of test setups. Eur J Trauma Emerg Surg. 2022;48(5):3541–60.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Wu H, et al. Double column acetabular fractures fixation using a novel dynamic anterior plate-screw system: a biomechanical analysis. Injury. 2021;52(3):407–13.

    Article  PubMed  Google Scholar 

  8. Wen X. Comparative biomechanical testing of customized three-dimensional printing acetabular-wing plates for complex acetabular fractures. Adv Clin Exp Med Off Organ Wroclaw Med Univ. 2020;29(4):459–68.

    Article  Google Scholar 

  9. Marmor M, Knox R, Huang A, Herfat S. Acetabulum cup stability in an early weight-bearing cadaveric model of geriatric posterior wall fractures. J Orthop Trauma. 2020;34(1):55–61.

    Article  PubMed  Google Scholar 

  10. Le Quang H, et al. Single column plate plus other column lag screw fixation vs. both column plate fixation for anterior column with posterior hemitransverse acetabular fractures—a biomechanical analysis using different loading protocols. Injury. 2021;52(4):699–704.

    Article  PubMed  Google Scholar 

  11. Quang HL, et al. Biomechanical comparison of fixation techniques for transverse acetabular fractures—single-leg stance vs. sit-to-stand loading. Inj-Int J Care Inj. 2020;51(10):2158–64.

    Article  Google Scholar 

  12. Alfonso NA, et al. Fixation of transverse acetabular fractures with precontoured plates alone causes fracture malreduction: a biomechanical assessment. J Orthop Trauma. 2020;34(2):89–94.

    Article  PubMed  Google Scholar 

  13. Cao QY, et al. A biomechanical comparison of different fixation techniques for fractures of the acetabular posterior wall (vol 42, pg 673, 2018). Int Orthop. 2019;43(5):1277–1277.

    Article  PubMed  Google Scholar 

  14. Ouldyerou A, et al. Functionally graded ceramics (FGC) dental abutment with implant-supported cantilever crown: finite element analysis. Compos Commun. 2023;38:101514.

    Article  Google Scholar 

  15. Ouldyerou A, et al. Biomechanical performance of Ti-PEEK dental implants in bone: an in-silico analysis. J Mech Behav Biomed Mater. 2022;134:105422.

    Article  CAS  PubMed  Google Scholar 

  16. Kocsis A, et al. Hybrid solution combining osteosynthesis and endoprosthesis for double column acetabular fractures in the elderly provide more stability with finite element model. Eklem Hastalik Ve Cerrahisi = Joint Dis Relat Surg. 2019;30(2):106–11.

    Article  Google Scholar 

  17. Terzini M, Pietro AD, Aprato A, Artiaco S, Massè A, Bignardi C. Are suprapectineal quadrilateral surface buttressing plates performances superior to traditional fixation? A Finite element analysis. Appl Sci-Basel. 2021;11(2):858.

    Article  CAS  Google Scholar 

  18. Hedelin H, et al. Postoperative stability following a triple pelvic osteotomy is affected by implant configuration: a finite element analysis. J Orthop Surg Res. 2022;17(1):275.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Hu P, et al. Biomechanical comparison of three internal fixation techniques for stabilizing posterior pelvic ring disruption: a 3D finite element analysis. Orthop Surg. 2019;11(2):195–203.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Lei JY, et al. Biomechanical analysis of the fixation systems for anterior column and posterior hemi-transverse acetabular fractures. Acta Orthop Traumatol Turc. 2017;51(3):248–53.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Bohme J, et al. Clinical implementation of finite element models in pelvic ring surgery for prediction of implant behavior: a case report. Clin Biomech. 2012;27(9):872–8.

    Article  CAS  Google Scholar 

  22. Liu XW, et al. Application of a shape-memory alloy internal fixator for treatment of acetabular fractures with a follow-up of two to nine years in China. Int Orthop. 2010;34(7):1033–40.

    Article  PubMed  Google Scholar 

  23. Zhang CC, et al. The treatment of acetabular fracture New concepts and new techniques. Shanghai: Shanghai Science and Technology Press; 2015.

    Google Scholar 

  24. Watson PJ, et al. The effect of boundary constraints on finite element modelling of the human pelvis. Med Eng Phys. 2017;43:48–57.

    Article  CAS  PubMed  Google Scholar 

  25. Ouldyerou A, et al. Biomechanical analysis of printable functionally graded material (FGM) dental implants for different bone densities. Comput Biol Med. 2022;150:106111.

    Article  CAS  PubMed  Google Scholar 

  26. Cun YW, et al. Traditional and bionic dynamic hip screw fixation for the treatment of intertrochanteric fracture: a finite element analysis. Int Orthop. 2020;44(3):551–9.

    Article  PubMed  Google Scholar 

  27. Clarke SG, Phillips ATM, Bull AMJ. Evaluating a suitable level of model complexity for finite element analysis of the intact acetabulum. Comput Methods Biomech Biomed Eng. 2013;16(7):717–24.

    Article  CAS  Google Scholar 

  28. Iqbal T, et al. Development of finite element model for customized prostheses design for patient with pelvic bone tumor. Proc Inst Mech Eng Part H-J Eng Med. 2017;231(6):525–33.

    Article  Google Scholar 

  29. Volinski B, Kalra A, Yang K. Evaluation of full pelvic ring stresses using a bilateral static gait-phase finite element modeling method. J Mech Behav Biomed Mater. 2018;78:175–87.

    Article  PubMed  Google Scholar 

  30. Zhang CC, Xu SG, et al. Clinical significance of acetabular three-column conception and floating classification of acetabular fracture. China J Orthop Trauma. 2007;20(7):433–6.

    Google Scholar 

  31. Rommens PM, et al. Open reduction and internal fixation of acetabular fractures in patients of old age. Int Orthop. 2020;44(10):2123–30.

    Article  PubMed  Google Scholar 

  32. Passaplan C, Simonin A, Gautier E. Posterior wall acetabular fracture in a 13-year-old boy treated by open reduction and mini-plate internal fixation: long-term follow-up of 17 years. Arch Orthop Trauma Surg. 2020;140(9):1163–7.

    Article  PubMed  Google Scholar 

  33. Zhang YT, et al. Comparative study of comminuted posterior acetabular wall fracture treated with the Acetabular Tridimensional Memory Fixation System. Inj-Int J Care Inj. 2014;45(4):725–31.

    Article  Google Scholar 

  34. Letournel E, Judet R. Fractures of the acetabulum. 2nd ed. Berlin: Springer; 1993.

    Book  Google Scholar 

  35. Hao ZX, et al. The effect of boundary condition on the biomechanics of a human pelvic joint under an axial compressive load: a three-dimensional finite element model. J Biomech Eng-Trans ASME. 2011;133(10):101006.

    Article  Google Scholar 

Download references

Acknowledgements

The authors are grateful to Prof. Chuncai Zhang, for his advice and guidance to our work.

Funding

This study was supported by the Support Program for the National Natural Science Foundation of China (Grant No. 51975380) and Natural Science Foundation of Shanghai (Grant No. 22ZR1442800). The funding organization played no role in the study design, conduction, data collection, or statistical analysis.

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Xiaohong Ding.

Ethics declarations

Ethics approval and consent to participate

This article does not contain any studies with human participants or animals performed by any of the authors. This study was approved by the ethics committee of Changhai Hospital of Shanghai.

Consent for publication

Informed consent was obtained from the individual participant included in the study. The participant has consented to the submission of the case report to the journal.

Competing interests

The authors have no competing interests to declare that are relevant to the content of this article.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Duan, P., Ding, X., Xiong, M. et al. Biomechanical evaluation of a healed acetabulum with internal fixators: finite element analysis. J Orthop Surg Res 18, 251 (2023). https://doi.org/10.1186/s13018-023-03736-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-023-03736-2

Keywords