The optimal screw-hole positions of the eccentric revision cup based on a morphological study

Abstract

Background

Bridging bone defects in revision total hip arthroplasty is a challenge to orthopedic surgeons. The eccentric revision cup is a progression of jumbo cup. Our aim is to confirm the optimal screw-hole positions of the eccentric revision cup by morphological measurements of three-dimensional pelvic reconstruction.

Methods

Eighty CT images were converted to virtual three-dimensional bones. After simulating the surgery procedure, all available screw holes were inserted with the screws in virtual. By measuring the length of the screw in the pelvic bone, we determined the rich bone stock area. Then the screw holes were designed according to the characteristics of bone stock distribution. The peripheral screw-hole cluster and inner screw-hole cluster were studied respectively.

Results

For peripheral screw-hole cluster, five screw holes were evenly distributed between point A and point B in the thicker rim. For inner screw-hole cluster, screw hole 1 and screw hole 2 are the recommended inner screw holes.

Conclusion

The eccentric revision cup has inherited the strengths of jumbo cup besides several unique advantages, including using the peripheral screws enhancing primary stability; decreasing the shift of hip rotation center and restoring biomechanical function; reducing the risk of dislocation because of the smaller head-cup differences; increasing the contact area between the outer cup and the host bone while maintaining a normal inclination of the inner cup. In this study, we confirmed the optimal screw-hole positions of the eccentric revision cup by surgical simulation and morphological measurement. However, biomechanical tests are still being further explored.

Introduction

Bridging bone defects in revision total hip arthroplasty (rTHA) is a major challenge to orthopedic surgeons [1]. Among the techniques reconstructing extensive acetabular defects, many studies have proven the validity of Jumbo cups [2,3,4,5,6]. The advantages include providing a large absolute contact area with the host bone which allows for long-term implant stability; reducing the need for structural allografts or augments because of the larger cup size; and making the surgery a relatively easier procedure with a similar surgical technique of primary total hip arthroplasty. However, hip center elevation [7, 8], limited screw fixation options [9], and large head-cup differences [10, 11] may limit the use of jumbo cups.

The eccentric revision cup -built on the strengths of jumbo cup and improved the problems mentioned above- is a progression of jumbo cup (Fig. 1). Each eccentric cup has an outer and an inner hemisphere (outer cup and inner cup) with different diameters. These two hemispheres are snapped together and close to each other on one side, forming an eccentric structure. A 20-degree angle is between the planes of outer cup and inner cup. And two types of screw-hole clusters exist in the eccentric revision cup: peripheral screw-hole cluster located in the thicker rim of the cup and inner screw-hole cluster in the inner cup.

Fig. 1
figure 1

The Eccentric Revision Cup. a: top view of the virtual eccentric revision cup; b: side view of the virtual eccentric revision cup; c: top view of the eccentric revision cup; d: side view of the eccentric revision cup; Red arrow: outer cup; Blue arrow: inner cup

In revision THA, screws are an essential supplement for acetabular shell fixation. Together with the press-fit technique, they increase the primary stability of the cup, which is basic for long-term osseointegration [12,13,14]. The optimal screw-hole positions could guide screws into the deepest bone stock avoiding injuring the neurovascular structures [15, 16]. The purpose of this article is to confirm the optimal screw-hole positions of the eccentric revision cup by morphological measurements of three-dimensional pelvic reconstruction.

Materials and methods

Eighty (40 males and 40 females) CT images (Brilliance iCT; Philips Healthcare, Cleveland, OH, USA) were obtained from a database composed of normal skeletal adults who underwent pelvic CT scans for non-hip diseases. The basic information was shown in Table 1. To conduct computer simulation, all CT scans were transferred to the Mimics medical imaging program (Materialise, Leuven, Belgium). Then the CT data were converted to virtual three-dimensional bones. As the methodology of designing the left and right components are the same, only the right acetabulum of each patient was selected in this study.

Table 1 Demographic characteristics

The average native acetabular diameter was 54.24 ± 3.98 mm (ranged from 45 to 62 mm). We categorized them into 4 groups by the native acetabular size. Group 1, with diameter from 45 to 51 mm (22 cases); Group 2, with diameter from 52 to 54 mm (19 cases). Group 3, with diameter from 55 to 56 mm (19 cases); Group 4, with diameter from 57 to 62 mm (20 cases). To simulate revision THAs using extra-large acetabular components, for each group, the biggest native acetabular diameter added 12 mm (in primary THA, the mean difference between the implanted cup size and the native acetabular size is 2 mm[17]; and the jumbo cup is a revision cup that is 10 mm greater than the cup size used for primary THA[18]) was the representative eccentric revision cup size and then implanted into the virtual three-dimensional bone. The diameters of the representative eccentric revision cup of the four groups were 64 mm, 66 mm, 68 mm, and 74 mm respectively.

Simulating surgical procedure, the thinner cup rim was aligned to the inferior acetabular rim with the inner cup placed at 40° inclination and 20° anteversion. The predesigned peripheral screw holes in the thicker rim of the cup were a string of 6 mm locking-screw holes right next to each other. And the positions of inner screw holes are similar to a conventional multi-hole revision cup. There are eight inner screw holes in two latitudes. The connecting line between the inner screw hole and the center of the inner cup forms an angle of 60 degrees and 25 degrees respectively with the plane of the inner cup. Then screws were virtually implanted into the cup.

Peripheral screw-hole cluster

Based on the assembly of the virtual eccentric revision cup and the virtual bone, all available peripheral screw holes were inserted with the locking screws. The length of the screw outside the shell and in the pelvic bone was measured respectively. If the length was bigger than 15 mm, the screw hole was considered valid and marked. Among these marked screw holes, the two most marginal ones were marked A and B (Fig. 2). The center of the outer cup opening plane was labeled O, and the symmetrical axis of the eccentric cup through O intersected the superior edge of the outer cup at the point C. Then the angles of AOB, AOC, BOC were measured.

Fig. 2
figure 2

Measurement of Peripheral Screw-hole Positions. O: the center of the outer cup opening plane; A and B: the outermost available screw holes; C: the point that the symmetrical axis of the eccentric cup intersected the superior edge of the outer cup

Inner screw-hole cluster

The study of optimal inner screw-hole cluster also started from the assembly of the virtual eccentric revision cup and the virtual bone. To avoid injuring the neurovascular structures during screw placement, the inner screw holes out of “safe zone” were excluded[19]. As shown in Fig. 3, there are four inner screw holes marked 1, 2, 3 and 4 within the “safe zone”. Then cancellous screws were implanted into these inner screw holes. Same to the method of peripheral screw, the length of the screw in the pelvic bone was measured respectively.

Fig. 3
figure 3

Measurement of Inner Screw-hole Positions. O’: the center of the inner cup opening plane; Red arrow: pointing to the anterior superior iliac spine; 1, 2, 3 and 4: the inner screw holes within the “safe zone”

Statistical analysis

The mean values and ranges were calculated for demographic data and presented using mean ± standard deviation with ranges. Variance analysis was used to determine the differences of the angles among different groups. Statistical significance was defined as p < 0.05. All statistical analyses were conducted with SPSS version 26.0 (IBM Inc., Armonk, New York).

Results

Our aim is to validate the optimal screw-hole positions of the eccentric revision cup through simulating surgical procedure and morphological measurement. The result of optimal peripheral screw-hole position is shown in Table 2. No difference was found among the four groups. It indicates that the areas (in pelvic of different sizes) where the peripheral screws purchase into the deepest bone stock have a consistent distribution. Within this area, five screw holes were evenly designed in the thicker rim of the eccentric revision cup. Seeing the mean value of AOB was 83.39 ± 2.12°, the angle between every two adjacent screw holes is 20 degrees. All peripheral screw holes were in the “safe zone”. The mean value of AOC and BOC was 26.99 ± 3.44° and 56.40 ± 3.29°, respectively. It shows that the relatively large amount of bone stock is in the posterosuperior part. According to our design, three peripheral screw holes were located posterior to the symmetrical axis and two were located anterior.

Table 2 The result of optimal peripheral screw-hole position

The lengths of inner screws are shown in Table 3. The optimal inner screw holes were located in the place where the inner screws could gain the deepest bone stock. According to the results, screw hole 1 and screw hole 2 are recommended for operation. The connecting line between the screw hole 1 and the center of the inner cup is 60 degrees and 17 degrees (posterior to the symmetry plane) respectively with the plane of the inner cup and the symmetry plane of the inner cup. And the connecting line between the screw hole 2 and the center of the inner cup is 25 degrees and 13 degrees (anterior to the symmetry plane) respectively with the plane of the inner cup and the symmetry plane of the inner cup.

Table 3 The mean length of inner screw

Discussion

In this article, we located the rich bone stock area to study the optimal positions of screw holes. For peripheral screw-hole cluster, five screw holes were evenly distributed between point A and point B in the thicker rim. Three holes were located posterior to the symmetrical axis and two located anterior. For inner screw-hole cluster, screw hole 1 and screw hole 2 are the optimal inner screw holes.

Bone deficiency is a challenge in rTHA. To solve this problem, several reconstruction strategies are performed. In most patients, stability can be achieved using an uncemented normal-sized hemispheric acetabular shell or a Jumbo cup[20,21,22]. Other strategies include structural allografts, augments, cages and reinforcement rings, oblong acetabular components and custom triflange components[23]. However, graft resorption and nonunion[24, 25], independent preparation for augments[26], breakage or loosening of cages and rings[27], absence of biologic fixation[28], and wide exposure[29] may limit the use of these methods.

Comparing with other methods, jumbo cup has become a preferable way with its unique advantages. First, it has a larger absolute contact area with the host bone, which is basic for long-term biological fixation. Second, the technique of jumbo cup is a relatively easy way (similar to primary THA). Third, jumbo cup provides an alternative to placing a cup into the superior defect or using augments in some certain cases. However, the elevation of hip rotation center has aroused wide concerns using jumbo cups[7, 8]. A vertical hip center shift alters hip biomechanics and potentially causes insufficiency of the abductor muscles, abnormal gait, and increased risk of dislocation from impingement[30,31,32]. In a computer simulating study of Nwankwo et al.[7], they found that the hip center shifted 0.27 mm superiorly and 0.02 mm anteriorly for every 1 mm increase in reamer diameter using a jumbo cup. Facing with this problem, Ries et al.[9] invented an offset COR acetabular shell designed to maintain the center of rotation closer to its anatomic position. Through radiographic evaluation, they reported that the mean vertical COR displacement of the test group was reduced by 3.5 mm[33]. Despite lack of follow-up results, it provides an effective way to address the elevation of hip rotation center. For our eccentric revision cup, in theory, the hip center would reconstruct closer to anatomic COR than using the offset COR cup because of the presence of a 20-degree angle between the planes of inner cup and outer cup. But future studies are needed to confirm this theory.

Limited screw fixation option is another vital problem using jumbo cups. Besides conventional dome screw fixation, the offset COR cup allows peripheral screws to be fixed into the posterior column of the pelvis[9]. However, the distribution of rich bone stock area in revision THA is different from that in primary THA, and the rich bone stock area is asymmetrical along the axis of the cup. The screw-hole design of the offset COR cup failed to reflect the characteristics of bone stock distribution. In our study, the peripheral screw-hole design of eccentric revision cup was based on morphological measurements. We found that most of the bone stock is located between 26.99 degrees anterior to the symmetrical axis and 56.40 degrees posterior to it. After that, we tested the relative location between the screw holes and the “safe zone”. The posterosuperior quadrant and the posteroinferior quadrant of acetabulum are safe for placing long screws[19, 34,35,36], avoiding injuring vital vessels nor nerves by screw trajectories. All peripheral screw holes were in the “safe zone”. In addition, we selected locking screws as peripheral screws for greater stability. Various mechanical tests have proved that locking screws have significantly greater stiffness and yield strength than non-locking screws (13).

In summary, the eccentric revision cup has inherited the strengths of jumbo cup besides several unique advantages as follows. First, this design decreases the shift of hip rotation center restoring biomechanical function. Second, the peripheral screw holes are designed in the thicker rim of the eccentric revision cup enhancing primary stability of the cup by locking screws[37]. Third, the smaller head-cup differences reduce the risk of dislocation[10, 11]. Fourth, it increases the contact area between the outer cup and the host bone while maintaining a normal inclination of the inner cup. The appropriate inclination angles may reduce the stress on the bearing surface and benefit for long-term results[38, 39].

Limitation

There are several limitations of this study. First, our study was based on normal pelvic and acetabular anatomy. However, it may not represent the variation in individual anatomy encountered in different revision THA settings. Second, this study was a 3D CT reconstruction morphological study and biomechanical tests are still being further explored. Third, because of individual variation, the peripheral screw holes do not always perfectly match the rich bone stock area in every patient. However, initial stability was achieved with 2 or 3 screw fixations in most jumbo cups. In our design, besides inner screws, one or two peripheral screws may enough for most patients. In the eccentric revision cup, there are 5 screw holes in the thick rim of the cup, which are enough to place screws pursuing the initial stability. Fourth, our study focused on screw-hole design on acetabular cups. However, improved design will not substitute for good surgical techniques. The proper surgical approach, adequate exposure, preparation of the acetabulum, and correct position of the acetabular cup are equally important.

Conclusion

Based on surgical simulation and morphological measurement, five peripheral screw holes evenly distribute between point A and point B in the thicker rim; screw hole 1 and screw hole 2 are the recommended inner screw holes. Although this novel acetabular design has many theoretical advantages, we recommend more studies be performed to determine that it achieves durable fixation and improved clinical outcomes before its widespread adoption. The costs and risks associated with new implant designs must still be justified by studies that evaluate implant durability and patient-reported outcome scores.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

THA:

Total hip arthroplasty

COR:

Center of rotation

References

  1. Volpin A, Konan S, Biz C, Tansey RJ, Haddad FS. Reconstruction of failed acetabular component in the presence of severe acetabular bone loss: a systematic review. Musculoskelet Surg. 2019;103(1):1.

    Article  CAS  Google Scholar 

  2. Brooks PJ. The jumbo cup: the 95% solution. Orthopedics. 2008;31(9):913.

    PubMed  Google Scholar 

  3. Lachiewicz PF, Soileau ES. Fixation, survival, and dislocation of jumbo acetabular components in revision hip arthroplasty. J Bone Joint Surg Am. 2013;95(6):543.

    Article  Google Scholar 

  4. Moon JK, Ryu J, Kim Y, Yang JH, Hwang KT, Kim YH. Acetabular revision arthroplasty using press-fitted jumbo cups: an average 10-year follow-up study. Arch Orthop Trauma Surg. 2019;139(8):1149.

    Article  Google Scholar 

  5. von Roth P, Abdel MP, Harmsen WS, Berry DJ. Uncemented jumbo cups for revision total hip arthroplasty: a concise follow-up, at a mean of twenty years, of a previous report. J Bone Joint Surg Am. 2015;97(4):284.

    Article  Google Scholar 

  6. Zhang J, Huang Y, Zhou B, Zhou Y. Mid-term follow-up of acetabular revision arthroplasty using jumbo cups. Orthop Surg. 2019;11(5):811.

    Article  Google Scholar 

  7. Nwankwo C, Dong NN, Heffernan CD, Ries MD. Do jumbo cups cause hip center elevation in revision THA? A computer simulation. Clin Orthop Relat Res. 2014;472(2):572.

    Article  Google Scholar 

  8. Nwankwo CD, Ries MD. Do jumbo cups cause hip center elevation in revision THA? A radiographic evaluation. Clin Orthop Relat Res. 2014;472(9):2793.

    Article  Google Scholar 

  9. Faizan A, Black BJ, Fay BD, Heffernan CD, Ries MD. Comparison of head center position and screw fixation options between a jumbo cup and an offset center of rotation cup in revision total hip arthroplasty: a computer simulation study. J Arthroplasty. 2016;31(1):307.

    Article  Google Scholar 

  10. Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res. 1998;355:163.

    Article  Google Scholar 

  11. Peter R, Lübbeke A, Stern R, Hoffmeyer P. Cup size and risk of dislocation after primary total hip arthroplasty. J Arthroplasty. 2011;26(8):1305.

    Article  Google Scholar 

  12. Hadjari MH, Hollis JM, Hofmann OE, Flahiff CM, Nelson CL. Initial stability of porous coated acetabular implants: the effect of screw placement, screw tightness, defect type, and oversize implants. Clin Orthop Relat Res. 1994;307:117.

    Google Scholar 

  13. Curtis MJ, Jinnah RH, Wilson VD, Hungerford DS. The initial stability of uncemented acetabular components. J Bone Joint Surg Br. 1992;74(3):372.

    Article  CAS  Google Scholar 

  14. Lachiewicz PF, Suh PB, Gilbert JA. In vitro initial fixation of porous-coated acetabular total hip components: a biomechanical comparative study. J Arthroplasty. 1989;4(3):201.

    Article  CAS  Google Scholar 

  15. Darmanis S, Pavlakis D, Papanikolaou A, Apergis E. Neurovascular injury during primary total hip arthroplasty caused by a threaded acetabulum cup. J Arthroplasty. 2004;19(4):520.

    Article  Google Scholar 

  16. Nachbur B, Meyer RP, Verkkala K, Zürcher R. The mechanisms of severe arterial injury in surgery of the hip joint. Clin Orthop Relat Res. 1979;141:122.

    Google Scholar 

  17. Odri GA, Padiolleau GB, Gouin FT. Oversized cups as a major risk factor of postoperative pain after total hip arthroplasty. J Arthroplasty. 2014;29(4):753.

    Article  Google Scholar 

  18. Whaley AL, Berry DJ, Harmsen WS. Extra-large uncemented hemispherical acetabular components for revision total hip arthroplasty. J Bone Joint Surg Am. 2001;83(9):1352.

    Article  CAS  Google Scholar 

  19. Meldrum R, Johansen RL. Safe screw placement in acetabular revision surgery. J Arthroplasty. 2001;16(8):953.

    Article  CAS  Google Scholar 

  20. Templeton JE, Callaghan JJ, Goetz DD, Sullivan PM, Johnston RC. Revision of a cemented acetabular component to a cementless acetabular component: a ten to fourteen-year follow-up study. J Bone Joint Surg Am. 2001;83(11):1706.

    Article  CAS  Google Scholar 

  21. Konan S, Duncan CP, Masri BA, Garbuz DS. Porous tantalum uncemented acetabular components in revision total hip arthroplasty: a minimum ten-year clinical, radiological and quality of life outcome study. Bone Joint J. 2016;98(6):767.

    Article  Google Scholar 

  22. Gustke KA, Levering MF, Miranda MA. Use of jumbo cups for revision of acetabulae with large bony defects. J Arthroplasty. 2014;29(1):199.

    Article  Google Scholar 

  23. García-Cimbrelo E, García-Rey E. Bone defect determines acetabular revision surgery. Hip Int. 2014;24(10Suppl):S33.

    Article  Google Scholar 

  24. Issack PS, Nousiainen M, Beksac B, Helfet DL, Sculco TP, Buly RL. Acetabular component revision in total hip arthroplasty. Part II: management of major bone loss and pelvic discontinuity. Am J Orthop Belle Mead NJ. 2009;38(11):550.

    PubMed  Google Scholar 

  25. Peng KT, Hsu WH, Shih HN, Chen CC, Yeh JH. Revision total hip arthroplasty for large medial defects with witch’s hat-shaped structural allografts–minimum 10-year follow-up. J Arthroplasty. 2014;29(2):428.

    Article  Google Scholar 

  26. Jenkins DR, Odland AN, Sierra RJ, Hanssen AD, Lewallen DG. Minimum five-year outcomes with porous tantalum acetabular cup and augment construct in complex revision total hip arthroplasty. J Bone Joint Surg Am. 2017;99(10): e49.

    Article  Google Scholar 

  27. Malahias MA, Ma QL, Gu A, Ward SE, Alexiades MM, Sculco PK. Outcomes of acetabular reconstructions for the management of chronic pelvic discontinuity: a systematic review. J Arthroplasty. 2020;35(4):1145.

    Article  Google Scholar 

  28. Chen WM, Engh CA Jr, Hopper RH Jr, McAuley JP, Engh CA. Acetabular revision with use of a bilobed component inserted without cement in patients who have acetabular bone-stock deficiency. J Bone Joint Surg Am. 2000;82(2):197.

    Article  CAS  Google Scholar 

  29. Holt GE, Dennis DA. Use of custom triflanged acetabular components in revision total hip arthroplasty. Clin Orthop Relat Res. 2004;429:209.

    Article  Google Scholar 

  30. Dou Y, Zhou Y, Tang Q, Yang D, Liu J. Leg-length discrepancy after revision hip arthroplasty: are modular stems superior? J Arthroplasty. 2013;28(4):676.

    Article  Google Scholar 

  31. Delp SL, Wixson RL, Komattu AV, Kocmond JH. How superior placement of the joint center in hip arthroplasty affects the abductor muscles. Clin Orthop Relat Res (328): 137, 1996

  32. Gustke KA. Jumbo cup or high hip center: is bigger better? J Arthroplasty. 2004;19(4 Suppl 1):120.

    Article  Google Scholar 

  33. Khlopas A, Chughtai M, Elmallah RK, Hip-Flores D, Malkani AL, Harwin SF, Mont MA, Ries MD. Novel acetabular cup for revision tha improves hip center of rotation: a radiographic evaluation. Clin Orthop Relat Res. 2018;476(2):315.

    Article  Google Scholar 

  34. Keating EM, Ritter MA, Faris PM. Structures at risk from medially placed acetabular screws. J Bone Joint Surg Am. 1990;72(4):509.

    Article  CAS  Google Scholar 

  35. Wasielewski RC, Cooperstein LA, Kruger MP, Rubash HE. Acetabular anatomy and the transacetabular fixation of screws in total hip arthroplasty. J Bone Joint Surg Am. 1990;72(4):501.

    Article  CAS  Google Scholar 

  36. Barrack RL. Neurovascular injury: avoiding catastrophe. J Arthroplasty. 2004;19(4 Suppl 1):104.

    Article  Google Scholar 

  37. Milne LP, Kop AM, Kuster MS. Polyaxial locking and compression screws improve construct stiffness of acetabular cup fixation: a biomechanical study. J Arthroplasty. 2014;29(5):1043.

    Article  Google Scholar 

  38. Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13(5):530.

    Article  CAS  Google Scholar 

  39. Del Schutte H, Jr Lipman AJ, Bannar SM, Livermore JT, Ilstrup D, Morrey BF. Effects of acetabular abduction on cup wear rates in total hip arthroplasty. J Arthroplasty. 1998;13(6):621.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

The authors declare that no funds, grants, or other support were received during the preparation of this manuscript.

Author information

Authors and Affiliations

Authors

Contributions

YZ contributed to the study conception and design. Material preparation, data collection performed by HM and YL and YD. JS and BZ performed Analysis. The first draft of the manuscript was written by YZ and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Yonggang Zhou.

Ethics declarations

Ethics approval

This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Ethics Committee of Medical School of Chinese PLA.

Consent to publish

Not applicable.

Competing interests

The authors have no relevant financial or non-financial interests to disclose.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, Y., Ma, H., Liu, Y. et al. The optimal screw-hole positions of the eccentric revision cup based on a morphological study. J Orthop Surg Res 17, 386 (2022). https://doi.org/10.1186/s13018-022-03260-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-022-03260-9

Keywords