The optimal angle of screw for using cement-screw technique to repair tibial defect in total knee arthroplasty: a finite element analysis

Abstract

Background

The cement-screw technique is a convenient method to repair tibial plateau defects in primary and revision total knee arthroplasty (TKA). However, the optimal angle of screw insertions is unknown. This study aimed to perform a finite element analysis (FEA) to determine the optimal screw angle for the repair of tibial plateau defects in TKA.

Methods

Seven FEA models were set and two common different defects (defect 1: area < 12%, depth < 12 mm; defect 2: area > 12%, depth > 12 mm) were simulated. One screw was used in defect 1, and one or two screws were used in defect 2. Screws were parallel to the proximal cortical bone (oblique screw) or perpendicular to the upper surface (vertical screw) of the tibia. Contact stresses on cancellous bone in different areas were determined. Maximum principal stress on the cancellous bone around each screw was also compared.

Results

The FEA models showed that stresses on the surface of cancellous bone in tibial defect (0.13–0.39 MPa) and stress focus spot (0.45 MPa) around the screw were lower when one vertical screw was used in defect 1. The stresses on the surface of cancellous bone in tibial defect (0.09–0.44 MPa), stresses in the medial tibial plateau (0.14–0.21 MPa), and stress focus spot around the screws were lowest (0.42 MPa and 1.37 MPa) when two vertical screws were used in defect 2, followed by of one vertical and one oblique (0.16–0.48 MPa; 0.15–0.21 MPa; 1.63 MPa and 1.11 MPa). No other statistically significant differences were found.

Conclusions

Either for one or two screws, those perpendicular to the upper surface achieve better stability than those parallel to the proximal cortical bone of the tibia. If two vertical screws cannot be performed, one vertical and one oblique is also acceptable.

Background

The cement-screw technique has been used to repair tibial bone defects in total knee arthroplasty (TKA) for over 40 years [1]. The cancellous screw is the metal reinforcement used on a defect filled with bone cement. Compared to other techniques, the cement-screw technique is less time-consuming, easier to perform, and less expensive [2, 3]. Although successful short-term and long-term follow-up is reported for the cement-screw technique, it may not be that simple [4,5,6]. There remains no consensus about the optimal angle of the screws, which is currently based on personal experience [7, 8]. Some have suggested that the screw should be parallel to the proximal cortical bone, while others believe that it should be perpendicular to the upper surface of the tibia plateau. In a finite element analysis (FEA), Zheng et al. [9] assessed the differences between vertical and oblique screws and demonstrated that vertical screws achieve better stability than oblique screws. However, the aim of the study was to determine whether differences exist between vertical and oblique screws and, therefore, only compared the difference for bones with a 12% defect area and a 12-mm depth. Defects with a larger area and depth needing two or more screws, which occurs in clinical practice, were not compared. The authors also did not indicate whether the oblique screw was parallel to the proximal cortical bone of the tibia, similar to previous reports. This study aimed to perform an FEA to determine the optimal screw angle either when one or two screws are used to provide a more basic reference for clinical practice.

Methods

Solid model

Using computed tomography (CT) images of a healthy volunteer (female, 20 years old, 165 cm in height, and 50 kg in weight), a three-dimensional (3D) model of the tibia was constructed using Mimics 21.0 (Materialise Company, Leuven, Belgium). The model was imported into Geomagic Studio 12.0 (3D Systems, Inc., North Carolina, USA) for optimization and then into SolidWorks 2016 (Dassault Systems SolidWorks Corp., Waltham, Massachusetts, USA) for creating the bone defect and implanting the prosthesis. Based on previous studies, two types of bone defects (defect 1, 9% (< 12%) of a total plateau in area, 8 mm (< 12 mm) in depth, and defect 2, 18% (> 12%) of a total plateau in area, 15 mm (> 12 mm) in depth) were made after performing a horizontal resection 9 mm above the tibial plateau [9, 10].

Tibial component

The tibial component consisted of a baseplate and a polyethylene insert. The tibial baseplate was made of cobalt-chrome alloy (#2 size, U2TM Knee, United Orthopedic, Taiwan), and its geometry was obtained by direct measurement. The posterior-stabilized fixed-bearing ultra-high molecular weight polyethylene (UHMWPE) insert was 9 mm in thickness with a rounded-top with flat-inferior surfaces. The tibial component was positioned at the plateau perpendicular to the mechanical axis of tibia.

Screw and bone cement models

In defect 1, a 4 mm × 14 mm cancellous screw (Smith and Nephew, Memphis, TN, USA) was inserted. In defect 2, one or two screws were implanted. A 1.5 mm thick cement layer was used to fix the tibial tray to the upper surface of the tibia after resection [12]. Bone separated from the defect area was defined as the cement model filling it. Finally, all parts were assembled in SolidWorks software (Figs. 1 and 2).

Fig. 1
figure 1

The anteroposterior and lateral view of whole model, consisted of ultrahigh molecular weight polyethylene (UHMWPE) insert, tibial prothesis, cement, screw and tibia

Fig. 2
figure 2

Different tibial bone defects and one or two screws inserted with different angles

Material properties and boundary and loading conditions

For proper meshing and simulation, the models were imported into ANSYS Workbench 17 (Swanson Analysis Systems, Inc., Houston, Pennsylvania, USA) (Fig. 3). The mechanical properties of the component materials are shown in Table 1 [9, 13]. In an uncemented implant, friction is the only mechanism for the transfer of shear stresses at an interface until bone ingrowth occurs; however, friction is expected to play an important role in a cemented implant only after the initiation of interface debonding. Hence, the contact behavior between the screws and bone was defined as the frictional surface-to-surface with a coefficient of 0.3; the others were defined as fully bonded [14, 15]. All implant components were modeled as linear elastic isotropic materials [16]. A total load of 1100N (2.2 times body weight) was used and applied in a 1:1 ratio between the medial and lateral plateaus [17]. The inferior surface of the distal tibia was fixed in all directions (Fig. 4). A conservatively low value for damage stress (2.8 MPa) of the cancellous bone was adopted, and a resorption threshold of 0.1 MPa was used to assess possible bone resorption [18].

Fig. 3
figure 3

The anteroposterior and lateral view of finite element models of the whole model

Table 1 The mechanical properties of component materials in each model
Fig. 4
figure 4

Load areas on the tibial plateau and fixed areas on the distal tibia

Seven FEA models were set up to determine the differences between screw angles with one or two screws (Table 2). Screws perpendicular to the upper surface of the tibial plateau were defined as vertical; screws parallel to the proximal cortical bone were defined as oblique. The different stresses were compared by measuring 6 points on the midcourt line of the medial and lateral plateaus, 16 points on the surface of cancellous bone in the defect, and 16 points on the surface of cancellous bone in the medullary cavity (Fig. 5). The maximum principal stresses of cancellous bone around each screw were also measured. All data were analyzed using SPSS version 22 (SPSS Inc., Chicago, IL, USA). A Paired Student’s t-test was used to compare the differences in stresses between models, and statistical significance was set at p < 0.05.

Table 2 The composition of each model
Fig. 5
figure 5

Stresses measured at different points on the surface of defect, one the midcourt line of the medial and lateral plateau, and on the surface of cancellous bone in the medullary cavity

Results

Bone defect

As shown in Fig. 6, defects were divided into 4 sections (anterior, posterior, medial, and lateral), and stresses were measured at 4 points on the surfaces of each. The results showed that stresses were significantly lower in model 2 (0.13–0.39 MPa) than in model 1 (0.17–0.46 MPa) (p < 0.05). When compared to model 3 (0.18–-0.55 MPa), model 4 (0.16–0.53 MPa), or model 5 (0.16–0.58 MPa), stresses were significantly lower in model 6 (0.16–0.48 MPa, p < 0.05) and model 7 (0.090.44 MPa, p < 0.05). Although stress shielding was possible, stresses measured on the surface of the defect in model 7 were lowest compared to other models.

Fig. 6
figure 6

Stresses (MPa) at 16 points on the surface of defects. A: anterior; P: posterior; M: medial; L: lateral

Tibial surface

Stresses at the 6 points on the midcourt line of the medial and lateral plateaus are shown in Fig. 7. All stresses measured were within the normal range (0.1–2.8 MPa) [18]. The stresses of the medial plateau in model 7 (0.14–0.21 MPa) were significantly lower than those in model 3 (0.17–0.26 MPa, p = 0.038), model 4 (0.17–0.26 MPa, p = 0.04) or model 5 (0.16–0.26 MPa, p = 0.046). However, no other statistically significant differences were found (p > 0.05).

Fig. 7
figure 7

Stresses at 6 points on the midcourt line of the medial and lateral plateau (MPa). P: point

Tibial tray

Stresses at 16 points (anterior, posterior, medial, and lateral) on the surface of cancellous bone in the medullary cervicitis of all models are shown in Fig. 8. A stress point less than 0.1 MPa was found on the anterior surface of cancellous bone in all models, where local stress shielding may occur. The stresses on the surface of cancellous bone in the medullary cervicitis in model 6 (0.08–0.73 MPa) were significantly lower than those in model 5 (0.08–0.83 MPa, p = 0.037). All the other stresses measured were within the normal range (0.1–2.8 MPa), and no statistically significant differences were found (p > 0.05).

Fig. 8
figure 8

Stresses (MPa) at 16 points on the surface of cancellous bone in the medullary cavity. A: anterior; P: posterior; M: medial; L: lateral

Stress focus spots around the screws

Stresses at the focus points that exist on the cancellous bone around the screws are shown in Table 3. All stresses were within the normal range (0.1–2.8 MPa) except in model 4 (2.91 MPa). In models with a 9% defect area and an 8 mm depth, the use of one vertical screw resulted in significantly lower focused stress (0.45 MPa) than one oblique screw (1.72 MPa). In models with an 18% defect area and a 15 mm depth, higher stresses were found with the use of one screw (2.42 MPa and 2.91 MPa, respectively) than that two (2.22 MPa, 1.63 MPa, and 1.37 MPa, respectively).

Table 3 Stress focus spots around the screws (MPa)

Discussion

Based on the FEA results, we found that (1) in defect 1, stresses on the surface of the cancellous bone were significantly lower when the screw was inserted vertically; (2) in defect 2, compared to one screw or two oblique screws, stresses on the surface of cancellous bone in defect and medial tibia surface were significantly lower when at least one vertical screw was inserted. We also compared the stress focus spots around the screws and found that (1) stress is significantly lower when one vertical screw was used to fill in small defects (defect area < 12%, depth < 10 mm); (2) higher stress occurs when only one screw or two oblique screws are used to fill a large defect (defect area > 12%, depth > 10 mm), which may destroy the cancellous bone around the screws; (3) either two vertical or two screws where one is vertical, and one oblique significantly reduces the stresses around them.

There are many types of basic reconstruction methods to treat defects of the tibial plateau in TKA [19]. For defects with a depth less than 10 mm, bone defects can be completely removed by adding bone resection and downsizing and lateralizing the tibial component without requiring further procedures [20]. However, excessive osteotomy may damage ligamentous structures, increase the stress on the proximal tibia, and require a thicker tibial insert [21]. Traditionally, the cement-only technique was used to fill defects less than 5 mm in depth after proximal tibial resection [19]. In all defects more than 5 mm, the cement-screw technique was suggested [22]. Compared to the cement-only technique, added screws greatly enhance the strength of the cement and reduce the possibility of the prothesis loosening [6]. For defects more than 5 mm, metal augmentation is another choice, which can have immediate stability and effective support for stable biomechanics [11]. However, metal augmentation is greatly limited due to the high price, limited shape, and thickness. In contrast, the cement-screw technique simplifies the operation, shortens operating time, and saves costs [7]. These advantages further promote the development of the cement-screw technique. The cement-screw technique is used to repair uncontained bone defects of the tibia in our clinical practice because of its many advantages.

Since being introduced into the clinic, many studies have demonstrated the efficacy of the cement-screw technique. In 1986, Ritter et al. [1] published a report on 57 cases utilizing screws and cement to fill tibial defects during TKA and obtained satisfactory short-term results. With an additional follow-up period of up to 13 years, they found that there were no signs of failure [6]. Berend et al. [4] performed 14,686 primary TKAs between December 1988 and February 2010; 256 patients received screws and cement for tibial defects. Compared to knees without screws, knees receiving tibial screws had significantly worse tibial bone quality preoperatively but equivalent survival probabilities at the last follow-up. From March 2018 to March 2019, Ozcan et al. [10] performed the cement-screw technique on 37 knees of 28 patients with high body mass index (BMI) (> 30 kg/m2). They also obtained satisfactory mid-term clinical outcomes and showed the efficiency of the cement-screw technique, even in high BMI patients. Despite its great success, the cement-screw technique may not be simple as it involves a lot of debate, such as the optimal number and angle of screws. Previous studies suggested that the number of screws depended upon the magnitude of the defect and its depth [1, 4]. However, for the angle, they did not describe the details. Berend et al. [4] suggested that screws should be inserted parallel to the proximal tibial cortex. In contrast, Liu et al. [8] inserted the screws perpendicular to the upper surface of the tibial plateau. Recently, Zheng et al. [9] performed an FEA to assess the differences between screw number and angle in TKA and showed that the number should be moderated for different defects. For the angle of the screws, they thought that vertical ones could achieve better stability than oblique ones. However, whether the screw was parallel to the proximal tibial cortex or not has not been described. Further, only one oblique screw was used in one model, and the analysis with two was absent.

The current incidence of periprosthetic fracture worldwide following primary TKA is believed to be in the range of 0.6–3.0% [23]. Because of alterations in the material properties of the bone as a result of aging, and thinning of the cortical bone structure, patients with osteoporosis may be particularly at risk of periprosthetic fracture following TKA [24]. Although stress fracture of the medial tibial plateau has been reported more after unicompartmental knee arthroplasty, it may occur after TKA as well [25, 26]. It's worth noting that most periprosthetic fracture was initiated at a cutting block pinhole site with varus collapse of the tibial component [26, 27]. Therefore, both reduced stresses and the angle of screws were very important to escape the stress fracture. The screw is needed not only to distribute the stresses, but also to transmit them effectively to the distal tibia.

There are some limitations associated with our study. First, this study is an FEA with models, and we did not consider screw interference with soft tissue that may occur in vivo. Second, we only compared two common angles reported in previous studies and did not consider the other angles, but it can be a good reference for clinical practice. Third, the cancellous screws used were 4 mm in diameter and 14 mm in length that is used in our clinical practice, and different diameters, materials, or lengths may have different effects on the stresses. Fourth, the models used in this study were based on a healthy patient with no deformity and good bone quality, which does not represent real occurrences in patients needing TKA. Finally, in clinical practice, the interactions of bone and screw include press-fit lock and friction, we only considered the frictional interaction and set a frictional coefficient of 0.3 between bone and screw based on previous studies. The knee prothesis used was cemented and fixed-bearing and we set fully bonded for the other contact pairs. Although the finite element model cannot close to real bone-screw interaction completely, it can provide some basic reference for future clinical practices. We believe this study may provide surgical guidance for performing TKA on patients with tibial bone defects.

Conclusions

Screws perpendicular to the upper surface of the tibial plateau can achieve better stability than screws parallel to the proximal cortical bone of the tibia, either for one or two screws. If two screws cannot be inserted vertically, one vertical and one oblique are acceptable.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

TKA:

Total knee arthroplasty

FEA:

Finite element analysis

CT:

Computed tomography

3D:

Three dimensional

BMI:

Body Mass Index

References

  1. Ritter MA. Screw and cement fixation of large defects in total knee arthroplasty. J Arthroplasty. 1986;1(2):125–9.

    Article  CAS  Google Scholar 

  2. Juvonen T, Nuutinen JP, Koistinen AP, Kroger H, Lappalainen R. Biomechanical evaluation of bone screw fixation with a novel bone cement. Biomed Eng Online. 2015;14:74.

    Article  Google Scholar 

  3. Lombardi AV, Berend KR, Adams JB. Management of bone loss in revision TKA: it’s a changing world. Orthopedics. 2010;33(9):662.

    PubMed  Google Scholar 

  4. Berend ME, Ritter MA, Keating EM, Jackson MD, Davis KE. Use of screws and cement in primary TKA with up to 20 years follow-up. J Arthroplasty. 2014;29(6):1207–10.

    Article  Google Scholar 

  5. Ritter MA, Harty LD. Medial screws and cement: a possible mechanical augmentation in total knee arthroplasty. J Arthroplasty. 2004;19(5):587–9.

    Article  Google Scholar 

  6. Ritter MA, Keating EM, Faris PM. Screw and cement fixation of large defects in total knee arthroplasty. A sequel J Arthroplasty. 1993;8(1):63–5.

    Article  CAS  Google Scholar 

  7. Berend ME, Ritter MA, Keating EM, Jackson MD, Davis KE, Malinzak RA. Use of screws and cement in revision TKA with primary or revision specific prosthesis with up to 17 years follow up. J Arthroplasty. 2015;30(1):86–9.

    Article  Google Scholar 

  8. Liu S, Wang S, Wu T, Fan Z, Zhang Y, Rao Y, et al. Short-term effectiveness of bone cement combined with screws for repairing tibial plateau defect in total knee arthroplasty. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2017;31(9):1055–8.

    PubMed  Google Scholar 

  9. Zheng C, Ma HY, Du YQ, Sun JY, Luo JW, Qu DB, et al. Finite element assessment of the screw and cement technique in total knee arthroplasty. Biomed Res Int. 2020:3718705.

  10. Ozcan O, Yesil M, Yuzuguldu U, Kaya F. Bone cement with screw augmentation technique for the management of moderate tibial bone defects in primary knee arthroplasty patients with high body mass index. Jt Dis Relat Surg. 2021;32(1):28–34.

    PubMed  PubMed Central  Google Scholar 

  11. Lee JK, Choi CH. Management of tibial bone defects with metal augmentation in primary total knee replacement: a minimum five-year review. J Bone Joint Surg Br. 2011;93(11):1493–6.

    Article  CAS  Google Scholar 

  12. Frehill B, Crocombe AD. Finite element assessment of metaphyseal sleeves in total knee arthroplasty. J Orthop. 2020;19:1–8.

    Article  CAS  Google Scholar 

  13. Brihault J, Navacchia A, Pianigiani S, Labey L, De Corte R, Pascale V, et al. All-polyethylene tibial components generate higher stress and micromotions than metal-backed tibial components in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2016;24(8):2550–9.

    Article  Google Scholar 

  14. Shirazi-Adl A, Dammak M, Paiement G. Experimental determination of friction characteristics at the trabecular bone/porous-coated metal interface in cementless implants. J Biomed Mater Res. 1993;27(2):167–75.

    Article  CAS  Google Scholar 

  15. Rancourt D, Shirazi-Adl A, Drouin G, Paiement G. Friction properties of the interface between porous-surfaced metals and tibial cancellous bone. J Biomed Mater Res. 1990;24(11):1503–19.

    Article  CAS  Google Scholar 

  16. Pegg EC, Walter J, Mellon SJ, Pandit HG, Murray DW, D’Lima DD, et al. Evaluation of factors affecting tibial bone strain after unicompartmental knee replacement. J Orthop Res. 2013;31(5):821–8.

    Article  Google Scholar 

  17. Zhao D, Banks SA, D’Lima DD, Colwell CW Jr, Fregly BJ. In vivo medial and lateral tibial loads during dynamic and high flexion activities. J Orthop Res. 2007;25(5):593–602.

    Article  Google Scholar 

  18. Huiskes R, Weinans H, Dalstra M. Adaptive bone remodeling and biomechanical design considerations for noncemented total hip arthroplasty. Orthopedics. 1989;12(9):1255–67.

    Article  CAS  Google Scholar 

  19. Cuckler JM. Bone loss in total knee arthroplasty: graft augment and options. J Arthroplasty. 2004;19(4 Suppl 1):56–8.

    Article  Google Scholar 

  20. Rand JA. Bone deficiency in total knee arthroplasty. Use of metal wedge augmentation. Clin Orthop Relat Res. 1991;271:63–71.

    Article  Google Scholar 

  21. Laskin RS. Total knee arthroplasty in the presence of large bony defects of the tibia and marked knee instability. Clin Orthop Relat Res. 1989;248:66–70.

    Article  Google Scholar 

  22. Daines BK, Dennis DA. Management of bone defects in revision total knee arthroplasty. J Bone Joint Surg Am. 2012;94(12):1131–9.

    Article  Google Scholar 

  23. Singh JA, Jensen M, Lewallen D. Predictors of periprosthetic fracture after total knee replacement: an analysis of 21,723 cases. Acta Orthop. 2013;84(2):170–7.

    Article  Google Scholar 

  24. Russo CR, Lauretani F, Seeman E, Bartali B, Bandinelli S, Di Iorio A, Guralnik J, Ferrucci L. Structural adaptations to bone loss in aging men and women. Bone. 2006;38(1):112–8.

    Article  Google Scholar 

  25. Yokoyama M, Nakamura Y, Egusa M, Doi H, Onishi T, Hirano K, Doi M. Factors related to stress fracture after unicompartmental knee arthroplasty. Asia Pac J Sports Med Arthrosc Rehabil Technol. 2018;15:1–5.

    PubMed  PubMed Central  Google Scholar 

  26. Sartawi MM, Rahman H, Kohlmann JM. Medial tibial plateau stress fracture following navigated total knee arthroplasty: two case reports. Am J Case Rep. 2021;22:e933005.

    Article  Google Scholar 

  27. Kim K, Kim YH, Park WM, Rhyu KH. Stress concentration near pin holes associated with fracture risk after computer navigated total knee arthroplasty. Comput Aided Surg. 2010;15:98–103.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The authors thank AiMi Academic Services (www.aimieditor.com) for the English language editing and review services.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

JM and SY contributed to the conceptualization and methodology. GZ contributed to the validation and data curation. JW contributed to the modeling and visualization, and was a major contributor in writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Jianpeng Wang.

Ethics declarations

Ethics approval and consent to participate

The computed tomography images, according to the ethical guidelines of the Helsinki Declaration, were approved by the Human Ethics Committee of Honghui hospital. Written informed consent was obtained from individual or guardian participants.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhao, G., Yao, S., Ma, J. et al. The optimal angle of screw for using cement-screw technique to repair tibial defect in total knee arthroplasty: a finite element analysis. J Orthop Surg Res 17, 363 (2022). https://doi.org/10.1186/s13018-022-03251-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-022-03251-w

Keywords