- Research
- Open access
- Published:
A clinical and biomechanical comparison of INFIX plus single versus double sacroiliac screw fixation for unstable pelvic ring injury
Journal of Orthopaedic Surgery and Research volume 17, Article number: 285 (2022)
Abstract
Background
The aim of this study is to compare the clinical and biomechanical outcome of INFIX plus single with two sacroiliac screw fixation for unstable pelvic fractures of Type C.
Methods
Sixteen cadavers were randomly subjected to INFIX plus single or double sacroiliac screw fixations and then mounted onto the ElectroForce loading machine under different vertical loads. To investigate the clinical outcomes of the two techniques, nineteen patients were retrospectively analyzed. The main outcome measures were postoperative radiographic reduction grading (using the Tornetta and Matta grading system), functional outcome (using the Majeed scoring system), and incidence of complications.
Results
In the biomechanical study, INFIX plus double sacroiliac screw fixation showed better biomechanical stability than fixation with a single sacroiliac screw (p < 0.05). In our clinical case series, all 19 patients had bony union 6 months after the operation. INFIX plus double sacroiliac screw fixation also demonstrated a better functional outcome and a higher radiographic satisfactory rate than INFIX plus single sacroiliac screw fixation (79.25 ± 5.47; 91.33 ± 4.97; p < 0.05), (77.78% vs. 60%; p = 0.05). One patient in INFIX plus single-screw fixation group had screw loosening at 6-month follow-up postoperatively. One case in each group suffered heterotopic ossification and the lateral femoral cutaneous nerve paralysis, and one patient suffered from infection.
Conclusion
INFIX plus double sacroiliac screw fixation demonstrated more stability in cadaveric biomechanical analysis and better clinical outcomes than INFIX plus single sacroiliac screw fixation.
Introduction
Pelvic fractures are common and associated with significant morbidity and mortality. Pelvic ring injury fracture (AO/OTA type C) is severe unstable fracture that is often accompanied by rupture of the posterior ligaments. Researchers have shown that simultaneous fixation of the anterior and posterior pelvic rings can offer adequate biomechanical stability, satisfactory healing, and functional recovery [1, 2]. By using simultaneous fixation, the complications related to single anterior or posterior pelvic ring fixation, such as pain, chronic pelvic instability, and disability, can potentially be avoided.
Traditional methods of anterior pelvic ring fixation include open reduction and internal fixation (ORIF) or external fixation. The risks of ORIF for anterior pelvic ring fractures include vascular nerve injury and inguinal hernia [3]. External fixation is mainly used in the emergency department to achieve hemodynamic stability, with potential complications including poor patient tolerance, postoperative infection, aseptic loosening, and joint dyskinesia [4,5,6]. Anterior subcutaneous internal fixation (INFIX) has recently been proposed by several scholars as a minimally invasive technique to treat anterior pelvic ring injury, with proposed benefits including shorter operative times, smaller wound size, less periosteal stripping, and adequate functional outcomes compared to traditional techniques [7,8,9,10,11].
For posterior pelvic ring injury, percutaneous sacroiliac screw fixation has emerged as a preferred technique [12]. Proposed benefits of this technique include reduced blood loss, shorter operative times, reduced tissue trauma, and fewer complications when compared to internal fixation [13, 14]. A double-screw fixation technique has been shown to provide optimal biomechanical stability compared to a single-screw fixation technique in cadaver model studies [15,16,17], but these findings have not been replicated in clinical studies [18]. Studies that have investigated the outcomes of combining INFIX and sacroiliac screw fixation to treat anterior and posterior pelvic ring fractures (AO/OTA type C) have shown varied and inconclusive outcomes.
The aim of our study was to compare the biomechanical stability and clinical outcomes of combining INFIX with single versus double sacroiliac screw fixation for Type C unstable pelvic fractures to determine whether a single-screw technique provides comparable outcomes to those of a double-screw technique.
Materials and methods
Cadaveric study
For the cadaveric research, 16 fresh human adult cadaver pelvises (8 males and 8 females; mean age ± SD: 55 ± 12.8 years) were selected for biomechanical testing at Southern Medical University and get donor families’ consent. We obtained approval from our southern medical university’s ethics committee (Number: XHEC-D-2015-112). Soft tissue was removed from each pelvis, retaining the following ligaments: anterior and posterior sacroiliac, sacrospinous, and sacrotuberous, and bones: L4, L5, sacrum, and 15 cm of each proximal femur. In addition, the specimens were scanned by X-ray to exclude pathologies such as tumors, tuberculosis, and osteoporosis. The specimens were stored in a Freezer at − 30 °C.
Fracture model creation
Models of AO/OTA type C1.3 unstable pelvic fractures were created by sawing each right upper and lower pubic ramus vertically with an electric pendulum saw and then sawing each ipsilateral sacrum vertically through the Denis I region (Fig. 1). The specimens were randomly divided into two treatment groups (groups A and B) of eight specimens each, with group A receiving INFIX plus single sacroiliac screw fixation at S1 and group B receiving INFIX plus double sacroiliac screw fixation at S1 and S2. All implants were placed by the same operator, and all implants were provided by Synthes. Cannulated screws were used to fix S1 (screw dimension: 7.3 mm × 105 mm) and S2 (screw dimensions: 7.3 mm × 65 mm). Two multiaxial pedicle screws (6.5 mm × 70 mm) were inserted through the point of both anterior inferior iliac spine and fixed with a titanium rod (6.0 mm × 400 mm). Then, two Kirschner wires (1.0 mm) were inserted into the vertical horizontal line at each end of the pubic ramus fracture to measure the displacement between the two wires during pelvic compression.
Biomechanical testing
All biomechanical experiments were conducted in the Department of Anatomy in Southern Medical University’s Engineering Institute. The L5 vertebra and the distal part of each femur were embedded and immobilized within a self-congealing resin denture powder and installed in the biomechanical testing machine, ElectroForce® 3510 (Bach. Co., USA) (Fig. 1). Axial compression was applied to the base of the upper sacrum at a loading rate of 20 N/S and maintained for 30 s when the load reached 200, 400, 600, and 800 N. The distance between the two Kirschner wires on each specimen was recorded, and each loading cycle was repeated at least three times.
Clinical study
We conducted a retrospective consecutive case series of patients treated in the Department of Orthopedics at the Fifth Affiliated Hospital, Southern Medical University, from February 2018 to January 2020. Nineteen patients fulfilled inclusion criteria (a minimum follow-up period of 12 months) with OTA/AO type C pelvic ring fractures and received INFIX plus sacroiliac screw fixation during the period in review. Of these patients, group C (4 females and 6 males) received INFIX plus single sacroiliac screw fixation and group D (3 females and 6 males) received INFIX plus double sacroiliac screw fixation. Patients’ demographic data, injury severity score (ISS), injury mechanism, time to surgery, procedure time, and estimated blood loss were recorded. At each 12–24-month follow-up, the following outcome measures were recorded: postoperative radiographic reduction grading, functional outcome using the Majeed scoring system [8], and incidence of complications. Reduction quality and implant position were assessed on radiographs (anteroposterior, inlet, and outlet views of the pelvis) (Fig. 2d–f) using the Tornetta and Matta grading system, at monthly intervals for the first 6 months.
Statistical analysis
All statistical analyses were performed using SPSS software, version 23.0 (IBM SPSS Statistics for Windows, Armonk, NY, USA). The measurement data were presented as mean ± SD and analyzed using a paired t test. The Chi-square test was used for enumeration type data, such as the incidence of complications. Statistical significance was set at p < 0.05.
Results
Cadaveric study
There was no significant difference in specimen height or weight between the two groups (p > 0.05). As shown in Table 1, no significant difference was found in anterior pelvic ring stability between the single- and double-screw fixation techniques when axial loading was less than 200 N. In both groups, anterior pelvis displacement increased proportionally with increased axial loading. Group B showed significantly greater stability under axial loading with 400 N (mean displacement differences: group A = 1.07 ± 0.39 mm; group B = 0.21 ± 0.09 mm; p < 0.001). In addition, group B showed significantly better anterior pelvis stability under axial loading with 800 N (mean displacement difference: group A = 2.37 ± 0.12 mm; group B = 1.29 ± 0.14 mm; p = 0.012). These findings suggest that fixation outcomes are superior when using a double-screw technique.
Clinical study
Baseline characteristics of patients
Nineteen patients met the inclusion criteria. Group C included 10 cases: 6 males and 4 females, mean age = 40.88 ± 11.43 years (range 28–56 years), and mean follow-up time = 20.5 months (range 12–24 months). Group D included 9 cases: 6 males and 3 females, mean age = 37.50 ± 12.36 years (range 26–50 years), and mean follow-up time = 23.2 months (range 12–27 months). The baseline characteristics of the patients are shown in Table 3. No significant inter-group differences were observed in terms of age, sex, ISS scores, fracture type, and injury mechanism (Table 2). All surgeries were performed by one group of surgeons, including a senior orthopedic pelvic trauma surgeon (Zhong Hua) and two general trauma surgeons (Chen Hongfen and Ding Chao), and a radiologist.
Postoperative radiographic and functional outcomes
The follow-up rate was 100%, with a mean follow-up time of 24.4 months (range 12–27 months). Surgical parameters between the groups were similar in terms of time to surgery and estimated blood loss (p > 0.05) (Table 3). Group C was superior to group D in terms of procedure time (p < 0.05). All patients’ fractures healed at 6 months post-surgery, and the mean time to hardware removal was 20.4 weeks. Postoperative radiographic reduction grading showed that group D had higher “satisfactory” Tornetta and Matta ratings (all ratings of “excellent” and “good,” divided by total number of patients) than group C (77.78% vs. 60%, p = 0.046) [19]. Moreover, group D had no “poor” reduction ratings of the anterior ring (p = 0.019).
Group D received a higher Majeed rating, which was statistically significant at the 6-month follow-up visit (p = 0.029) (Table 3). However, no statistically significant difference was found at the 12-month follow-up visit (Table 3).
Complications
In group C, one patient developed non-union of the pubic ramus fracture, while all other patients achieved bony union at 3 months after surgery. Three patients (30%) in group C had postural pelvic tilt and sacroiliac joint instability at their 3-month post-surgery follow-up visit (Fig. 2). All three patients achieved bone healing after 3 months of strict bed rest. Two patients in each group experienced lateral femoral cutaneous nerve paralysis with symptom resolution achieved within 1 month after hardware removal. Two patients in each group developed asymptomatic heterotopic ossification. One patient developed a superficial infection due to the poor condition of soft tissue at the surgical site but achieved wound healing after debridement and dressing changes.
Discussion
The aim of surgical treatment of unstable pelvic fractures is to correct the deformity, restore the anatomical structure of the pelvic ring, and promote early functional exercise. Previous studies have shown that the anterior and posterior rings account for 40% and 60% of the stability of the pelvis, respectively [20]. Therefore, simultaneous fixation of the anterior and posterior pelvic rings is often necessary to treat unstable pelvic fractures.
Treatment of posterior pelvic ring fractures
Surgical fixation options for managing traumatic disruptions to the posterior pelvic ring included percutaneous sacroiliac screw fixation, plate fixation, and lumbosacral iliac screws. However, the most popular operative treatment option at present is percutaneous sacroiliac screw fixation, as it is associated with less trauma, less blood loss, and earlier mobilization than other treatments [21]. The fixation of double screws into the sacrum is associated with an increased risk of nerve injury due to significant variation in sacral anatomy [22].
Sacroiliac joint screw placement in the S1 vertebral body is the standard technique for posterior ring fixation [19, 23, 24]. Mears et al. showed that a single sacroiliac screw fixation can restore biomechanical stability similar to that of the complete pelvis in cadaveric models, under 10–350 N of vertical compressive load [25]. However, Yinger suggested that in unstable pelvic fractures, the placement of two sacroiliac joint screws increases stability against rotation and vertical displacement when the pelvic ring is loaded to 1000 N [26]. It is likely that the study by Mears et al. did not detect a difference between the techniques due to the use of inadequate vertical compressive load. In our study, we found no difference in anterior and posterior pelvic ring stability between the single- and double-screw fixation techniques when axial loading was less than 200 N; there are no differences between the single- and double-screw fixation in posterior stability under loads of 400 N and 800 N. However, for anterior pelvic ring stability, with loads of 400 N and 800 N, significant differences emerged between the two techniques, suggesting the superiority of the double-screw fixation technique.
Treatment of anterior pelvic ring fractures
Traditionally, open reduction and plate fixation is the optimal treatment for anterior pelvic ring fractures, providing excellent stability and early mobilization. However, disadvantages include long operation times, large wound sizes and periosteal stripping areas, and increased bleeding. In addition, the risk of infection and re-operation exists, especially in obese patients or patients with a history of abdominal surgery. Furthermore, another technique employed in the treatment of these fractures is channel screw fixation, but it requires extensive training and a high level of surgical skill.
To address these treatment challenges, in 2009, Kuttert et al. performed the first anterior ring fixation for unstable pelvic fractures using INFIX [27]. INFIX has since become a popular treatment technique for unstable pelvic fractures, with reported benefits including reduced soft tissue injury, less blood loss, and low incidence of intraoperative iatrogenic nerve injury [7,8,9]. Studies investigating the biomechanics, anatomy, and clinical outcomes of INFIX have found that INFIX may provide adequate pelvic stability and achieve good clinical outcomes, despite its associated complications [27,28,29,30].
There are no data on the functional outcome of INFIX plus single versus INFIX plus double sacroiliac screw fixation technique for unstable pelvic ring injury (AO/OTA type C). We found that INFIX plus double sacroiliac screw fixation offered significantly better stability in the anterior pelvis than INFIX plus single sacroiliac screw fixation, under 400–800 N of axial loading. This result further supports the conclusion that the addition of a second screw improved the stability of the pelvis, which is consistent with the results of previous studies [18, 26, 31].
In addition, to investigate whether the INFIX plus double sacroiliac screw fixation technique improved functional outcomes clinically, 19 patients with unstable pelvic ring injury (AO/OTA type C) were retrospectively analyzed and followed up. We found that all fractures healed by the 6-month follow-up visit, with one patient who received INFIX plus single sacroiliac screw fixation experiencing non-union of the pubic ramus fracture.
In terms of joint stability, we did not find greater joint instability with single-screw fixation in our biomechanics study, but we did observe greater joint instability in the retrospective clinical analysis. In terms of nerve injury, two patients developed anterolateral numbness in the affected thigh, which is resolved by 3 months post-implant removal. Fang et al. reported an LFCN paralysis rate of 48.3% [32], which was much higher than the rate in our study. The reason for our finding may be improved surgical execution of the INFIX technique in the last decade and our use of pedicle screws with a smaller diameter (6.5 mm), compared with those used in Fang et al.’s study (7.3–10 mm). The limitations of this study are a retrospective study with a small sample size. A multicenter prospective studies with large sample size should be conducted in future study.
Conclusion
INFIX plus double sacroiliac screw fixation showed greater stability in biomechanical analysis and better functional clinical outcomes in the treatment of unstable pelvic ring injury (AO/OTA type C) than INFIX plus single sacroiliac screw fixation.
Availability of data and materials
All data generated or analyzed during this study are included in this published article.
References
Riemer BL, Butterfield SL, Diamond DL, Young JC, Raves JJ, Cottington E, Kislan K. Acute mortality associated with injuries to the pelvic ring: the role of early patient mobilization and external fixation. J Trauma. 1993;35:671–5. https://doi.org/10.1097/00005373-199311000-00003.
Gansslen A, Pohlemann T, Paul C, Lobenhoffer P, Tscherne H. Epidemiology of pelvic ring injuries. Injury. 1996;27(Suppl 1):S-A13-S-A20. https://doi.org/10.1016/S0020-1383(96)90106-0.
Sagi HC, Papp S. Comparative radiographic and clinical outcome of two-hole and multi-hole symphyseal plating. J Orthop Trauma. 2008;22:373–8. https://doi.org/10.1097/BOT.0b013e31817e49ee.
Tucker MC, Nork SE, Simonian PT, Routt ML Jr. Simple anterior pelvic external fixation. J Trauma. 2000;49:989–94. https://doi.org/10.1097/00005373-200012000-00002.
Mason WT, Khan SN, James CL, Chesser TJ, Ward AJ. Complications of temporary and definitive external fixation of pelvic ring injuries. Injury. 2005;36:599–604. https://doi.org/10.1016/j.injury.2004.11.016.
Vaidya R, Colen R, Vigdorchik J, Tonnos F, Sethi A. Treatment of unstable pelvic ring injuries with an internal anterior fixator and posterior fixation: initial clinical series. J Orthop Trauma. 2012;26:1–8. https://doi.org/10.1097/BOT.0b013e318233b8a7.
Kumbhare C, Meena S, Kamboj K, Trikha V. Use of INFIX for managing unstable anterior pelvic ring injuries: a systematic review. J Clin Orthop Trauma. 2020;11:970–5. https://doi.org/10.1016/j.jcot.2020.06.039.
Yin Y, Luo J, Zhang R, Li S, Jiao Z, Zhang Y, Hou Z. Anterior subcutaneous internal fixator (INFIX) versus plate fixation for pelvic anterior ring fracture. Sci Rep. 2019;9:2578. https://doi.org/10.1038/s41598-019-39068-7.
Patel S, Aggarwal S, Jindal K, Kumar V, Sharma S. Outcomes and complications of the INFIX technique for unstable pelvic ring injuries with high-velocity trauma: a systematic review and meta-analysis. Arch Orthop Trauma Surg. 2021. https://doi.org/10.1007/s00402-020-03742-7.
Vaidya R, Tonnos F, Nasr K, Kanneganti P, Curtis G. The anterior subcutaneous pelvic fixator (INFIX) in an anterior posterior compression type 3 pelvic fracture. J Orthop Trauma. 2016;30(Suppl 2):S21-22. https://doi.org/10.1097/BOT.0000000000000583.
Muller FJ, Stosiek W, Zellner M, Neugebauer R, Fuchtmeier B. The anterior subcutaneous internal fixator (ASIF) for unstable pelvic ring fractures: clinical and radiological mid-term results. Int Orthop. 2013;37:2239–45. https://doi.org/10.1007/s00264-013-2032-0.
Moed BR, Geer BL. S2 iliosacral screw fixation for disruptions of the posterior pelvic ring: a report of 49 cases. J Orthop Trauma. 2006;20:378–83. https://doi.org/10.1097/00005131-200607000-00002.
Pieske O, Landersdorfer C, Trumm C, Greiner A, Wallmichrath J, Gottschalk O, Rubenbauer B. CT-guided sacroiliac percutaneous screw placement in unstable posterior pelvic ring injuries: accuracy of screw position, injury reduction and complications in 71 patients with 136 screws. Injury. 2015;46:333–9. https://doi.org/10.1016/j.injury.2014.11.009.
Suzuki T, Hak DJ, Ziran BH, Adams SA, Stahel PF, Morgan SJ, Smith WR. Outcome and complications of posterior transiliac plating for vertically unstable sacral fractures. Injury. 2009;40:405–9. https://doi.org/10.1016/j.injury.2008.06.039.
Nork SE, Jones CB, Harding SP, Mirza SK, Routt ML Jr. Percutaneous stabilization of U-shaped sacral fractures using iliosacral screws: technique and early results. J Orthop Trauma. 2001;15:238–46. https://doi.org/10.1097/00005131-200105000-00002.
Osterhoff G, Dodd AE, Unno F, Wong A, Amiri S, Lefaivre KA, Guy P. Cement augmentation in sacroiliac screw fixation offers modest biomechanical advantages in a cadaver model. Clin Orthop Relat Res. 2016;474:2522–30. https://doi.org/10.1007/s11999-016-4934-9.
Suero EM, Greiner A, Becker CA, Cavalcanti Kussmaul A, Weidert S, Pfeufer D, Woiczinski M, Braun C, Flatz W, Bocker W, Kammerlander C. Biomechanical stability of sacroiliac screw osteosynthesis with and without cement augmentation. Injury. 2020. https://doi.org/10.1016/j.injury.2020.01.043.
Khaled SA, Soliman O, Wahed MA. Functional outcome of unstable pelvic ring injuries after iliosacral screw fixation: single versus two screw fixation. Eur J Trauma Emerg Surg. 2015;41:387–92. https://doi.org/10.1007/s00068-014-0456-x.
Matta JM, Tornetta P 3rd. Internal fixation of unstable pelvic ring injuries. Clin Orthop Relat Res. 1996;329:129–40. https://doi.org/10.1097/00003086-199608000-00016.
Bi C, Wang Q, Nagelli C, Wu J, Wang J. Treatment of unstable posterior pelvic ring fracture with pedicle screw-rod fixator versus locking compression plate: a comparative study. Med Sci Monit. 2016;22:3764–70. https://doi.org/10.12659/msm.900673.
Shetty AP, Bosco A, Perumal R, Dheenadhayalan J, Rajasekaran S. Midterm radiologic and functional outcomes of minimally-invasive fixation of unstable pelvic fractures using anterior internal fixator(INFIX) and percutaneous iliosacral screws. J Clin Orthop Trauma. 2017;8:241–8. https://doi.org/10.1016/j.jcot.2017.05.009.
Sagi HC, Ordway NR, DiPasquale T. Biomechanical analysis of fixation for vertically unstable sacroiliac dislocations with iliosacral screws and symphyseal plating. J Orthop Trauma. 2004;18:138–43. https://doi.org/10.1097/00005131-200403000-00002.
Cole JD, Blum DA, Ansel LJ. Outcome after fixation of unstable posterior pelvic ring injuries. Clin Orthop Relat Res. 1996;329:160–79. https://doi.org/10.1097/00003086-199608000-00020.
Moed BR, Whiting DR. Locked transsacral screw fixation of bilateral injuries of the posterior pelvic ring: initial clinical series. J Orthop Trauma. 2010;24:616–21. https://doi.org/10.1097/BOT.0b013e3181df97eb.
Mears SC, Sutter EG, Wall SJ, Rose DM, Belkoff SM. Biomechanical comparison of three methods of sacral fracture fixation in osteoporotic bone. Spine (Phila Pa 1976). 2010;35:E392-395. https://doi.org/10.1097/BRS.0b013e3181cb4fcd.
Yinger K, Scalise J, Olson SA, Bay BK, Finkemeier CG. Biomechanical comparison of posterior pelvic ring fixation. J Orthop Trauma. 2003;17:481–7. https://doi.org/10.1097/00005131-200308000-00002.
Kuttner M, Klaiber A, Lorenz T, Fuchtmeier B, Neugebauer R. The pelvic subcutaneous cross-over internal fixator. Unfallchirurg. 2009;112:661–9. https://doi.org/10.1007/s00113-009-1623-0.
Vigdorchik JM, Esquivel AO, Jin X, Yang KH, Onwudiwe NA, Vaidya R. Biomechanical stability of a supra-acetabular pedicle screw internal fixation device (INFIX) vs external fixation and plates for vertically unstable pelvic fractures. J Orthop Surg Res. 2012;7:31. https://doi.org/10.1186/1749-799X-7-31.
Apivatthakakul T, Rujiwattanapong N. “Anterior subcutaneous pelvic internal fixator (INFIX), Is it safe?” A cadaveric study. Injury. 2016;47:2077–80. https://doi.org/10.1016/j.injury.2016.08.006.
Vaidya R, Martin AJ, Roth M, Tonnos F, Oliphant B, Carlson J. Midterm radiographic and functional outcomes of the anterior subcutaneous internal pelvic fixator (INFIX) for pelvic ring injuries. J Orthop Trauma. 2017;31:252–9. https://doi.org/10.1097/BOT.0000000000000781.
McDonald E, Theologis AA, Horst P, Kandemir U, Pekmezci M. When do anterior external or internal fixators provide additional stability in an unstable (Tile C) pelvic fracture? A biomechanical study. Eur J Trauma Emerg Surg. 2015;41:665–71. https://doi.org/10.1007/s00068-014-0482-8.
Fang C, Alabdulrahman H, Pape HC. Complications after percutaneous internal fixator for anterior pelvic ring injuries. Int Orthop. 2017;41:1785–90. https://doi.org/10.1007/s00264-017-3415-4.
Acknowledgements
We thank Department of Anatomy in Southern Medical University’s Engineering Institute for technical assistance.
Funding
This study was supported by Medical Science and Technology Research Fund of Guangdong Province (NO: A2018178 and B2021252), the National Natural Science Foundation of China (No. 81501196), and Dean fund of the Fifth Affiliated Hospital, Southern Medical University (NO:YZ7202ZX02). The fund launch for the National Natural Science Foundation of China (No.YZ2020ZX021).
Author information
Authors and Affiliations
Author notes
Hongfen Chen, Chao Ding, Yongqiang Liu, the first three author contribute equally to this paper
Contributions
HZ and YY designed the research; HC and CD analyzed the data; YL, ZK, and FH performed research; YY wrote the paper; and QG developed software necessary to perform and record experiments. All authors read and approved the final manuscript.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
The present study was approved by The Fifth Affiliated Hospital, Southern Medical University.
Consent for publication
Not applicable.
Informed consent
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Chen, H., Ding, C., Liu, Y. et al. A clinical and biomechanical comparison of INFIX plus single versus double sacroiliac screw fixation for unstable pelvic ring injury. J Orthop Surg Res 17, 285 (2022). https://doi.org/10.1186/s13018-022-03133-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13018-022-03133-1