Computer simulation of optimal lipped polyethylene liner orientation against prosthetic impingement

Abstract

Background

Lipped or elevated acetabular liners are to improve posterior stability and are widely used in hip arthroplasty. However, concerns of increasing impingement exist when using such liners and optimal orientation of the elevated rim remains unknown. We aimed to identify the impact of lipped liner on the range of motion (ROM) before impingement and propose its optimal orientation.

Methods

An isochoric three-dimensional model of a general hip-replacement prosthesis was generated, and flex-extension, add-abduction and axial rotation were simulated on a computer. The maximum ROM of the hip was measured before the neck impinged on the liner. Different combinations of acetabular anteversion angles ranging from 5 to 30 degrees, and lipped liner orientations from posterior to anterior were tested.

Results

When acetabular anteversion was 10 or 15 degrees, placing the lip of the liner in the posterosuperior of the acetabulum allowed satisfactory ROM in all directions. When acetabular anteversion was 20 degrees, extension and external rotation were restricted. Adjusting the lip to the superior restored satisfactory ROM. When acetabular anteversion was 25 degrees, only placing the lip into the anterosuperior could increase extension and external rotation to maintain satisfactory ROM.

Conclusions

This study showed that optimal lipped liner orientation should depend on acetabular anteversion. When acetabular anteversion was smaller than 20 degrees, placing lip in the posterior allowed an optimally ROM. When acetabular anteversion was greater than 20 degrees, adjusting lip to the anterior allowed a comprehensive larger ROM to avoid early impingement.

Introduction

Dislocation is the first reason for revision total hip arthroplasty (THA) and recurrent dislocation occurs in 60% of patients after a first dislocation and more than half of them need re-revision [1, 2], which seriously deteriorates joint function and increases social financial burdens [3, 4]. Therefore, surgeons must be aware and deal well with factors which may bring postoperative joint instability, including patient general condition, surgical technique and implant characters, to avoid dislocation as much as possible after primary THA [5].

Impingement is the contact between metal femoral neck and cup liner or bone-to-bone contact and is related to polyethylene wear and subsequent dislocation [6]. Accurate acetabulum position is integral to avoid these complications and is greatly studied [7,8,9]. However, previous studies often ignored polyethylene liner which was also an important part of the whole acetabular system. Lipped liner has an elevated rim of 10–20 degrees, which is designed to be placed in the posterosuperior acetabulum to decrease posterior dislocation rate by restricting femoral head movement through a hemisphere over 180 degrees and is widely used in primary THAs [10, 11]. However, although it is generally accepted that lipped liner should be placed posterosuperiorly, it has not been proven insufficiently [12, 13]. Besides, concerns of component impingement, wear and future loosing exist when using such liners with over 81% of retrieved lipped liners showing impingement signs [10, 14, 15].

As the inability to identify liner radiographically, computer simulation is used as a protocol to trace liner impingement [16]. Since impingement causes joint instability and continuing movement leads to dislocation, range of motion (ROM) before impingement reflects safe motion zone. Although dislocation risk is expected to be reduced with lipped liner increasing femoral head jumping distance, it happens after impingement and acquiring a wider ROM before impingement is important to decrease possible polyethylene wear. Thus, we intended to use computer simulation to investigate the influence of different combinations of acetabulum anteversion degrees and lipped liner orientations on ROM before impingement. We tried to answer: (1) Would the ROM before impingement meet the requirements of daily activities when a lipped liner was used? (2) When acetabulum anteversion was changed, where should lipped liner be placed to obtain a maximum ROM before impingement?

Materials and methods

Our institutional review board approved this study. The isochoric three-dimensional model of a general hip-replacement prosthesis (Stryker Orthopaedics, Kalamazoo, Michigan, USA) was initially generated with the use of Rhino (Rhinoceros 6.0, Robert McNeel & Associates, USA) via sweeping and lofting (Fig. 1). The geometry parameters of the prosthesis used are listed in Table 1. A 10-degree lipped polyethylene liner was used. The model was of the same size and geometry characters with the hip prosthesis system.

Fig. 1
figure 1

a The hip prothesis system to be modeled. The stem was Secur-Fit and the head was Biolox. The cup was Trident PSL and the liner was Trident X3 with a 10 degrees polyethylene elevated rim. b The isochoric three-dimensional model based on the prothesis

Table 1 Geometry characters of the prothesis

The three-dimensional model was imported into Materialise 3-matic Medical 11.0 (Materialise, Leuven, Belgium) to mimic different motions. The coordinate system and the axes around which motion was described were determined according to the recommendations made by the International Society of Biomechanics [17]. The origin of hip axes was located at the center of the head and it was also the center of the acetabulum. With the hip in the neutral position, the x-axis pointed anteriorly and was perpendicular to the vertical axis of the stem. The y-axis pointed superiorly and was parallel to the vertical axis of the stem. The z-axis pointed to the patient’s right side (Fig. 2). Flex-extension of the hip was described around the z-axis, with axial rotation around the y-axis and add-abduction around the x-axis. Acetabular anatomical anteversion and abduction was adopted [18]. The femoral anteversion was measured as the angle between the femoral neck axis and the y-z axial plane. The orientation of the elevated rim was determined by the angle formed by the midline of the elevated rim and the line intersected by the acetabular opening plane and the y–z axial plane (Fig. 2). Negative degrees represented the lipped liner pointing posteriorly, and positive degrees represented the lipped liner pointing anteriorly. We used − 30, 0 and 30 degrees to represent lipped liner pointing to the posterosuperior, superior and anterosuperior, respectively.

Fig. 2
figure 2

a The coordinate system determined by the International Society of Biomechanics. b The view of the acetabular opening plane to show the orientation of the elevated rim. One line connected the center of the acetabular circle and the midpoint of the elevated rim. The other line was intersected by the acetabular opening plane and the y–z axial plane. The angle ω formed by these two lines was the orientation of the elevated rim

The ROM was measured by moving the femur in the desired direction until the neck visually impinged on the liner of the cup (Fig. 3). This method was consisted with the study before and proved to have enough accuracy [19]. The computer software was capable of detecting this degree, and the maximum ROM before impingement was defined as the number of it. The acetabular anteversion angles of 5, 10, 15, 20, 25 and 30 degrees were studied in combination with the lipped liner pointing at − 30, 0, and 30 degrees. Flexion, extension, add-abduction, axial rotation and internal rotation at hip 90 degrees flexion were measured for all above combinations. The cup abduction was fixed to 50 degrees for an approximate 40 degrees total acetabulum abduction when using a 10-degree lipped liner and the stem anteversion was 15 degrees without changing during the whole test.

Fig. 3
figure 3

a Flexion was simulated and was observed from side. b Impingement on the liner of the cup during flexion was detected visually from a horizontal view and the flexion angle was recorded once impingement being visible. The red block showed the impingement area and the red arrow pointed to the contact between the femoral neck and the cup liner

To determine the effect of different combinations of prosthetic positions on ROM, each ROM was classified according to two zones: excellent and poor. An excellent ROM allows greater than 110 degrees of flexion, 30 degrees of extension, 30 degrees of adduction–abduction, 30 degrees of internal-external rotation and 30 degrees of internal rotation at 90 degrees hip flexion. A poor ROM means less than 110 degrees of flexion, 30 degrees of extension, 30 degrees of adduction–abduction, 30 degrees of internal-external rotation and 30 degrees of internal rotation at 90 degrees hip flexion. The excellent zone met the requirements of daily activities [20]. We used Prism (GraphPad Prism 8 for Mac, GraphPad Software, San Diego, CA, USA) to generate graphs.

Results

Overall, increasing acetabular anteversion angles increased hip flexion, internal rotation, internal rotation at hip 90 degrees flexion and decreased extension, external rotation and adduction–abduction. Placing the lip of the liner in the anterosuperior increased extension, external rotation, adduction and decreased hip flexion, internal rotation. Placing the lip anterosuperiorly did not have much influence on abduction (Fig. 4).

Fig. 4
figure 4

Range of motion under different combinations of acetabular anteversion degrees and lipped liner orientations

When the lip of the liner was placed in the posterosuperior quadrant of the acetabular component and the acetabular anteversion was 10 or 15 degrees, it allowed hip flexion over 110 degrees, extension over 30 degrees and internal rotation over 80 degrees at hip 90 degrees flexion. This ROM met the demand of daily activities and the use of a lipped liner did not restrict hip movements.

At 10 or 15 degrees of acetabular anteversion, placing the lip of the liner in the posterosuperior brought an excellent ROM. If the lip was placed in the superior or anterosuperior, although extension and external rotation was increased, hip flexion was decreased to less than 110 degrees, which made the ROM poor.

However, at 20 degrees of acetabular anteversion, if the lip was still placed in the posterosuperior, extension and external rotation would be restricted to less than 30 degrees. If the lip was adjusted to the superior quadrant of the acetabulum in the above situation, although hip flexion was slightly decreased, it still reached over 110 degrees and extension and external rotation was restored to over 30 degrees, which made the ROM excellent again.

When the acetabular anteversion was 25 degrees or even larger, posterosuperiorly placed lip even decreased extension and external rotation to less than 20 degrees. When the lip was placed in the superior, extension and external rotation reached slightly over 30 degrees. Only when the lip was placed in the anterosuperior, extension and external rotation could be restored to over 30 degrees and flexion still reached over 110 degrees. It turned the ROM excellent again.

In above situations, hip adduction–abduction was over 40 degrees and internal rotation was more than 80 degrees. Internal rotation at hip 90 degrees flexion was all over 70 degrees. These three motions met the requirements of daily activities and were not included into analysis.

Discussion

Our study was designed to characterize impingement patterns associated with various components’ locations and orientations. Prosthetic impingement determines the functional end point of the stable ROM after a hip arthroplasty and is associated with wear, loosing and long-term failure [6]. Therefore, optimum positioning of the components is necessary to avoid a decrease in the stable ROM due to prosthetic impingement.

Lipped liner, as an important part of the acetabular system, is widely used to decrease the tendency of dislocation. However, there are concerns of elevated rim polyethylene liner increasing rate of impingement and there are few studies about it [11]. Our study initially put elevated rim polyethylene liner into consideration and found that lipped liner placed in the posterosuperior quadrant of the acetabular component did not decrease the ROM before impingent when the acetabular anteversion was 10 or 15 degrees. When the acetabular anteversion was 15 degrees, posterosuperiorly placed lipped liner allowed over 110 degrees of flexion and over 30 degrees of extension, which met the demand of daily activities [20]. Our result probably explained some clinical outcomes that posterosuperiorly placed lipped liner did not increase the cumulative probability of revision because of loosening compared to standard liner [15, 21].

The elevated rim of the liner should be placed in the posterosuperior conventionally. Lipped liners were designed by Charnley in the early 1970s to decrease the tendency for posterior dislocation of the femoral head [22]. The lipped lip increases stability by increasing the displacement required for the prosthetic head to dislocate [21], and it becomes apparent to place the liner in the posterior quadrant when using a posterolateral approach. Using a 15-degree liner in the posterior quadrant increases the internal rotation range of movement by 8.9 degrees without causing dislocation [23]. Our result suggested that when the acetabular component was in 10 to 15 degrees of anteversion, placing the lipped liner in the posterosuperior achieved enough ROM in all directions without increasing impingement risk compared to lipped liner placed in other directions. A relatively small acetabular anteversion made posterior dislocation easy [24] and our result supported the theory of placing the liner in the posterosuperior to increase stability by achieving the largest ROM before impingement. It proved the feasibility of placing the elevated rim in the posterosuperior quadrant of the acetabulum under common situation.

But our result also suggested that we should not place the lipped liner in the posterosuperior in all situations without changing. The optimum position of the elevated rim was depended on the orientation of the acetabular component. When the acetabular anteversion was more than 20 degrees, the ROM before impingement would be restricted by extension and external rotation less than 30 degrees which might cause anterior dislocation if the lipped liner was still placed in the posterosuperior. When the acetabular anteversion was more than 25 degrees, it became worse by extension and external rotation less than 25 degrees. However, if the lipped liner was transferred to the superior or anterosuperior in above situations, extension and external rotation would be restored to over 30 degrees while it had little effect on ROM on other directions. It suggested that when the acetabulum was placed in larger anteversion than common, an anteriorly or superiorly placed lipped liner might bring lager ROM before impingement. It has been reported that impingement was more frequently associated with an elevated anti-dislocation rim [25] liner and the predictive risk for dislocation was 6.6 times larger compared to standard liners when using an elevated rim liner for primary THA [26]. This might be that in these retrieved cases, the position of the elevated rim was not adjusted with the orientation of the acetabular component and was placed in the posterosuperior routinely. Yamaguchi et al. found an association between excessive cup anteversion combined with posteriorly placed elevated rim and impingement in retrieved liners [27]. Krushell et al. tested prosthetic ROM with acetabular component in different alignments and found elevated rim liners in satisfactorily positioned acetabular components offered no demonstrable benefit and the primary indication for these implants appeared to be in cases of instability due to acetabular malposition [28]. It was consistent with our study that the optimum position of the elevated rim depended on the orientation of the acetabular component.

Component-on-component impingement, followed by levering out of the femoral head, is by far the most common mode of dislocation in total hip arthroplasty [29]. Although ROM before dislocation does not equal to that before impingement [30], impingement should be minimized to reduce possible wear and loosing. Moreover, multiple confounding factors and sources of variability made it difficult to identify specific parameters influencing ROM before impingement. For example, implant position changes such as stem anteversion, neck-shaft angle and prosthetic head size have been proved to have an impact on no-impingement ROM [31] and it becomes more complex when taking in vivo factors into consideration. Our study first controlled other variables and explored the influence of the elevated rim of liner on ROM before impingement. The influence of the elevated rim of liner on dislocation deserves to be further studied.

There are a number of limitations associated with this study. First, this study focused on prosthetic impingement, while bony impingement also matters and there are other factors having an impact on prosthetic impingement including head size and head/neck ratio. However, given that there are so many confounding factors and bony anatomy characters vary, it is too complex to take all variables into consideration. We emphasized on the impact of the lipped liner on ROM by controlling other factors the same, and we offered a new insight into the optimum position of the elevated rim of the liner. Second, although this study was simulated on computer and might be different from in vivo conditions when taking soft tissue into consideration, it gave a general suggestion which might be still useful in vivo. Third, this study used only one kind of prosthesis. However, the prosthesis used in this study is of classical design and was applied widely. We believed that although values might be slightly changed in other kinds of prosthesis, the tendency of movement and the conclusion would be same.

Conclusion

An optimally placed lipped polyethylene liner should depend on acetabular anteversion in total hip arthroplasty. When the acetabular anteversion is relatively small for about less than 20 degrees, placing the liner in the posterior quadrant of the acetabulum is acceptable. When the acetabular anteversion is more than 20 degrees, it is better to place the liner in the superior or anterior to achieve a comprehensive ROM to avoid early impingement.

Availability of data and materials

The datasets used or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ROM:

Range of motion

THA:

Total hip arthroplasty

References

  1. Bozic KJ, Kamath AF, Ong K, Lau E, Kurtz S, Chan V, Vail TP, Rubash H, Berry DJ. Comparative epidemiology of revision arthroplasty: failed THA poses greater clinical and economic burdens than failed TKA. Clin Orthop Relat Res. 2015;473(6):2131–8.

    Article  PubMed  Google Scholar 

  2. Kotwal RS, Ganapathi M, John A, Maheson M, Jones SA. Outcome of treatment for dislocation after primary total hip replacement. J Bone Joint Surg Br. 2009;91(3):321–6.

    Article  CAS  PubMed  Google Scholar 

  3. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital cost of dislocation after primary total hip arthroplasty. J Bone Joint Surg Am. 2006;88(2):290–4.

    Article  PubMed  Google Scholar 

  4. Abdel MP, Cross MB, Yasen AT, Haddad FS. The functional and financial impact of isolated and recurrent dislocation after total hip arthroplasty. Bone Joint J. 2015;97(8):1046–9.

    Article  PubMed  Google Scholar 

  5. Rowan FE, Benjamin B, Pietrak JR, Haddad FS. Prevention of dislocation after total hip arthroplasty. J Arthroplasty. 2018;33(5):1316–24.

    Article  PubMed  Google Scholar 

  6. Malik A, Maheshwari A, Dorr LD. Impingement with total hip replacement. J Bone Joint Surg Am. 2007;89(8):1832–42.

    PubMed  Google Scholar 

  7. McCarthy TF, Alipit V, Nevelos J, Elmallah RK, Mont MA. Acetabular cup anteversion and inclination in hip range of motion to impingement. J Arthroplasty. 2016;31(9 Suppl):264–8.

    Article  PubMed  Google Scholar 

  8. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60(2):217–20.

    Article  CAS  PubMed  Google Scholar 

  9. Wera GD, Ting NT, Moric M, Paprosky WG, Sporer SM, Della Valle CJ. Classification and management of the unstable total hip arthroplasty. J Arthroplasty. 2012;27(5):710–5.

    Article  PubMed  Google Scholar 

  10. Cobb TK, Morrey BF, Ilstrup DM. The elevated-rim acetabular liner in total hip arthroplasty: relationship to postoperative dislocation. J Bone Joint Surg Am. 1996;78(1):80–6.

    Article  CAS  PubMed  Google Scholar 

  11. Behery OA, Long WJ. The prevalence of elevated-rim polyethylene liner use in primary total hip arthroplasty in the New York State metropolitan area. Arthroplast Today. 2019;5(4):486–8.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Sculco PK, Cottino U, Abdel MP, Sierra RJ. Avoiding Hip instability and limb length discrepancy after total hip arthroplasty. Orthop Clin North Am. 2016;47(2):327–34.

    Article  PubMed  Google Scholar 

  13. Rogers M, Blom AW, Barnett A, Karantana A, Bannister GC. Revision for recurrent dislocation of total hip replacement. Hip Int. 2009;19(2):109–13.

    Article  PubMed  Google Scholar 

  14. Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA. Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty. 2005;20(4):427–35.

    Article  PubMed  Google Scholar 

  15. Cobb TK, Morrey BF, Ilstrup DM. Effect of the elevated-rim acetabular liner on loosening after total hip arthroplasty. J Bone Joint Surg Am. 1997;79(9):1361–4.

    Article  CAS  PubMed  Google Scholar 

  16. Vigdorchik JM, Sharma AK, Madurawe CS, Elbuluk AM, Baré JV, Pierrepont JW. Does prosthetic or bony impingement occur more often in total hip arthroplasty: a dynamic preoperative analysis. J Arthroplasty. 2020;35(9):2501–6.

    Article  PubMed  Google Scholar 

  17. Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part I: ankle, hip, and spine. Int Soc Biomech J Biomech. 2002;35(4):543–8.

    Article  Google Scholar 

  18. Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75(2):228–32.

    Article  CAS  PubMed  Google Scholar 

  19. Zhang J, Wei J, Mao Y, Li H, Xie Y, Zhu Z. Range of hip joint motion in developmental dysplasia of the hip patients following total hip arthroplasty with the surgical technique using the concept of combined anteversion: a study of crowe I and II patients. J Arthroplasty. 2015;30(12):2248–55.

    Article  PubMed  Google Scholar 

  20. D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82(3):315–21.

    Article  CAS  PubMed  Google Scholar 

  21. Insull PJ, Cobbett H, Frampton CM, Munro JT. The use of a lipped acetabular liner decreases the rate of revision for instability after total hip replacement: a study using data from the New Zealand Joint Registry. Bone Joint J. 2014;96-b(7):884–8.

    Article  CAS  PubMed  Google Scholar 

  22. Charnley J. Low friction arthroplasty of the hip: theory and practice. New York: Springer; 1979.

    Book  Google Scholar 

  23. Sultan PG, Tan V, Lai M, Garino JP. Independent contribution of elevated-rim acetabular liner and femoral head size to the stability of total hip implants. J Arthroplasty. 2002;17(3):289–92.

    Article  CAS  PubMed  Google Scholar 

  24. Biedermann R, Tonin A, Krismer M, Rachbauer F, Eibl G, Stöckl B. Reducing the risk of dislocation after total hip arthroplasty: the effect of orientation of the acetabular component. J Bone Joint Surg Br. 2005;87(6):762–9.

    Article  CAS  PubMed  Google Scholar 

  25. Marchetti E, Krantz N, Berton C, Bocquet D, Fouilleron N, Migaud H, Girard J. Component impingement in total hip arthroplasty: frequency and risk factors. A continuous retrieval analysis series of 416 cup. Orthop Traumatol Surg Res. 2011;97(2):127–33.

    Article  CAS  PubMed  Google Scholar 

  26. Girard J, Kern G, Migaud H, Delaunay C, Ramdane N, Hamadouche M. Primary total hip arthroplasty revision due to dislocation: prospective French multicenter study. Orthop Traumatol Surg Res. 2013;99(5):549–53.

    Article  CAS  PubMed  Google Scholar 

  27. Yamaguchi M, Akisue T, Bauer TW, Hashimoto Y. The spatial location of impingement in total hip arthroplasty. J Arthroplasty. 2000;15(3):305–13.

    Article  CAS  PubMed  Google Scholar 

  28. Krushell RJ, Burke DW, Harris WH. Elevated-rim acetabular components. Effect on range of motion and stability in total hip arthroplasty. J Arthroplasty. 1991;6:S53–8.

    Article  PubMed  Google Scholar 

  29. Brown TD, Callaghan JJ. Impingement in total hip replacement: mechanisms and consequences. Curr Orthop. 2008;22(6):376–91.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Scifert CF, Brown TD, Pedersen DR, Callaghan JJ. A finite element analysis of factors influencing total hip dislocation. Clin Orthop Relat Res. 1998;355:152–62.

    Article  Google Scholar 

  31. Widmer KH. The impingement-free, prosthesis-specific, and anatomy-adjusted combined target zone for component positioning in THA depends on design and implantation parameters of both components. Clin Orthop Relat Res. 2020;478(8):1904–18.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This study was sponsored by National Nature Science Foundation of China (Grant No. 31900941), the Interdisciplinary Program of Shanghai Jiao Tong University (project No. ZH2018QNA06) and Project of Biobank from Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine (project No. YBKA201911).

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study conception and design. Data extraction and assessment were performed by JWZ, YH, XHZ. Research coordination was done by HQ, ZAZ, HWL. The first draft of the manuscript was written by JWZ, YH. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Huiwu Li or Jingwei Zhang.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the ethics committee of Ninth People's Hospital of Shanghai Jiao Tong University School of Medicine.

Consent for publication

All authors have seen the manuscript and approved it to submit to your journal.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1

. The cross section of the neck was not a standard circle but an ellipse.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hu, Y., Zhou, X., Qiao, H. et al. Computer simulation of optimal lipped polyethylene liner orientation against prosthetic impingement. J Orthop Surg Res 17, 204 (2022). https://doi.org/10.1186/s13018-022-03093-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-022-03093-6

Keywords