Comparison of the fixation ability between lag screw and bone plate for oblique metacarpal shaft fracture

Abstract

Background

For oblique metacarpal shaft fracture, if anatomical reduction is achieved through conservative cast immobilization rather than stable fixation, bone malrotation can easily occur, resulting in severe loss in hand prehensile function. However, whether bone plate fixation or only lag screw fixation is more preferable remains unclear. Few studies have evaluated whether screw fixation can provide biomechanical fixation strength similar to bone plate fixation.

Objective

We assessed the difference in fixation strength between fixtation with two lag screws and bone plate for oblique metacarpal shaft fractures.

Materials and methods

We created oblique metacarpal shaft fractures on 21 artificial bones and fixated them using (1) double lag screw (2LS group), (2) regular plate (RP group), or (3) locked plate (LP group). To obtain the force–displacement data, a cantilever bending test was conducted for each specimen through a material testing machine. One-way analysis of variance and a Tukey test were conducted to compare the maximum fracture force and stiffness of the three fixation methods.

Results

The maximum fracture force of the 2LS group (mean + SD: 153.6 ± 26.5 N) was significantly lower than that of the RP (211.6 ± 18.5 N) and LP (227.5 ± 10.0 N) groups (p < 0.001). However, no significant differences were discovered between the RP and LP groups. The coefficient of variation for the maximum fracture force of the 2LS group (17.3%) was more than twice as high as that of the RP (8.7%) and LP (4.4%) groups. In addition, the stiffness of the three fixation methods was similar.

Conclusion

Compared with bone plate fixation, double lag screw fixation yielded slightly lower maximum bearable fracture force but similar stiffness. Therefore, this technique could be used for treating oblique metacarpal shaft fractures. However, using double lag screw fixation alone is technically demanding and requires considerable surgical experiences to produce consistent results.

Introduction

Metacarpal fractures are common, accounting for approximately 36%–42% of hand trauma cases [1]. Metacarpal neck fracture is the most common type of metacarpal fracture, which is twice as common as metacarpal shaft fracture [2]. However, unlike metacarpal neck, which is mostly composed of cancellous bone, the metacarpal shaft mainly comprises cortical bone and thus is more difficult to heal and requires surgical treatment for bone union [3]. According to different injury mechanisms, metacarpal shaft fractures result in different forms: axial loading and direct blow tend to cause transverse or comminuted fractures, whereas torsion leads to oblique or spiral fractures, among which torsion accounts for 75% of the causes of metacarpal shaft fractures [4]. Stable, nondisplaced metacarpal shaft fractures can be treated conservatively through cast immobilization. However, fractures with larger displacement and greater degrees of angulation are unstable because of the continual traction force generated by the interosseous muscle and the limited bony contact area at the fracture site. If conservative treatment is adopted in such cases, satisfactory anatomic reduction is difficult to achieve, the probability of nonunion becomes high, and bone malrotation can easily occur. This leads to scissoring deformity of the fingers during hand grasping and severe loss of hand prehensile function [5]. Therefore, oblique and spiral fractures typically require surgical fixation.

Oblique metacarpal shaft fractures are commonly fixed using bone plate or lag screws [6]. However, which of them is superior remains unclear. In general, lag screw fixation is minimally invasive, cost effective and time saving [7, 8]. Nevertheless, surgeons have concerns over it because of its mechanical stability and is a relatively challenging technique. Surgeons need to overcome the steep learning curve of lag screw fixation technique to reduce postoperative complications. By contrast, bone plate fixation provides superior fixation strength and is easier to operate. It enables earlier range of motion and earlier start for the subsequent rehabilitation program, resulting in a more satisfactory treatment result. However, it is also likely to cause complications, such as metacarpophalangeal joint stiffness, extensor tendon adhesion, and iatrogenic injury of the dorsal cutaneous branch of the ulnar or radial nerve, whether a regular or a locked plate is applied. In addition, the cost of such surgery is high.

Studies have discussed the fixation of transverse metacarpal shaft fractures [9,10,11], but few have explored that of oblique fractures [6]. Therefore, in this mechanicstic study, we compared the stability of double lag screw fixation and bone plate fixation to determine the optimal treatment for oblique metacarpal shaft fractures.

Materials and methods

Specimen preparation

We selected 14 artificial third metacarpal bones (Sawbones, Vashon, WA, USA). A 0.8-mm chainsaw was used to create oblique metacarpal shaft fractures on the artificial metacarpal bones (Fig. 1). Epoxy embedding and fixation were conducted at the proximal end of the artificial metacarpal bones.

Fig. 1
figure 1

Artificial metacarpal bones (a) intact and with oblique metacarpal shaft fracture in b lateral view and c cross-sectional view

Fixation approaches

We created oblique fractures on 21 artificial metacarpal bones in total and evenly divided them into three groups: (1) double lag screw fixation (the 2LS group), (2) regular plate fixation (the RP group), and (3) locked plate fixation (the LP group). The fracture fixation was performed by Yung-Cheng Chiu, a senior hand surgeon.

Double lag screw fixation: After anatomical reduction, seven obliquely fractured artificial bones were fixed by placing two 2.3-mm parallel cortical screws (Stryker, Germany) into the bones from the dorsal cortex and perpendicularly to the fracture line until they penetrated the far cortex to stabilize the fracture (Fig. 2).

Fig. 2
figure 2

Radiographs of the three fixation approaches: regular plate fixation in a anterior–posterior view and b lateral view; locked plate fixation in c anterior–posterior view and d lateral view; and double screw fixation (e) anterior–posterior view and f lateral view

Regular plate fixation: After anatomical reduction, seven obliquely fractured artificial bones were fixed by placing a regular 5-hole plate (Stryker, Germany) on the dorsum of the metacarpal shaft and locking a cortical screw each at the proximal point, in the middle, and at the distal point of the fracture site. The screws were inserted vertically to the anatomical axis of the metacarpal bone until they penetrated the near and far cortex (Fig. 2).

Locked plate fixation: After anatomical reduction, seven obliquely fractured bones were fixed by placing a locked 5-hole plate (Stryker, Germany) on the dorsum of the metacarpal shaft and locking a locked screw each at the proximal point, in the middle, and at the distal point of the fracture site. The screws were inserted vertically to the anatomical axis of the metacarpal bone until they penetrated the near and far cortex.

Biomechanical test

The in in vitro test was conducted as described in our previous studies [7, 8, 12,13,14]. The in vitro mechanical test was conducted using cantilever bending tests. The material testing system employed was JSV-H1000 (Japan Instrumentation System, Nara, Japan) (Fig. 3). A force at a decreasing rate of 10 mm/min was applied to the distal region on the dorsal side of the artificial metacarpal bones until they fractured. The force–displacement curve generated during the force application was recorded, and the maximum fracture force and stiffness were obtained accordingly.

Fig. 3
figure 3

The in vitro biomechanical tests for the three fixation approaches: a double lag screw fixation; b regular plate fixation; and c locked plate fixation

Statistical analysis

The maximum fracture force and stiffness of the three fixation methods for oblique metacarpal shaft fractures are expressed as mean ± standard deviation. One-way analysis of variance with a post hoc Tukey test were used to compare the fixation ability of the three methods. SPSS 19.0 (IBM, Armonk, NY, USA) was used for statistical analysis, and a p value less than 0.05 was set as statistically significant.

Results

The maximum fracture force and stiffness of the three fixation methods are presented in Table 1. The maximum fracture force of the 2LS group (153.6 ± 26.5 N) was significantly lower than that of the RP (211.6 ± 18.5 N; difference: 37.7%) and LP groups (227.5 ± 10.0 N; difference: 48.1%) ( p< 0.05). However, no significant differences were observed between the RP and LP groups. In addition, for the maximum fracture force, the coefficient of variation of the 2LS group (17.3%) was more than twice as high as that of the RP (8.7%) and LP (4.4%) groups. Although the mean stiffness of the 2LS group (57.0 ± 8.2 N/mm) was lower than that of the RP (64.7 ± 10.2 N/mm) and LP (65.4 ± 8.2 N/mm) groups, no significant differences were observed among the three groups.

Table 1 Maximum fracture force and stiffness of the three fixation types for oblique metacarpal shaft fracture

Discussion

Clinically, surgical methods have been proposed for fixation of oblique metacarpal shaft fractures [6]. Bone plate fixation provides superior fixation strength, is easier to perform, and facilitates an earlier range of motion and start for rehabilitation programs, thereby resulting in a more satisfactory treatment outcome. However, postoperative complications, such as metacarpophalangeal joint stiffness, extensor tendon adhesion, and iatrogenic injury of the dorsal cutaneous branch of the ulnar or radial nerve, are likely to occur, and the cost of the surgery is relatively high. By contrast, lag screw fixation reduces these complications and is minimally invasive, cost effective, and time saving [8, 13]. Nevertheless, according to literature and clinical experience, lag screw fixation may not provide adequate biomechanical fixation strength [15,16,17,18]. In the present study, we explored the fixation strength using lag screw and bone plate fixation for oblique metacarpal shaft fractures and observed that the maximum bearable fracture force of double lag screw fixation was slightly lower than that of bone plate fixation, but the stiffness of the two was comparable. Hence, double lag screw fixation is feasible for treating oblique metacarpal shaft fractures. However, use of only the double lag screw fixation technique requires more experience to achieve consistent results.

Oblique and spiral fractures account for a high proportion of metacarpal shaft fractures. Because of the traction force generated from the interosseous muscle, the anatomical parts of oblique and spiral fractures can cause greater displacement at the fractured bone end, angulation deformity, shortening of metacarpal lengths because of the overlapping at the fractured bone end, and malrotation. In addition, such fractures are relatively unstable due to the limited bony contact area at the fracture site. Therefore, if conservative cast immobilization treatment is applied, a longer immobilization time is required. Patients may have to begin their rehabilitation program 6–8 weeks after immobilization, which may result in finger joint stiffness [11, 19, 20], Mal-union, nonunion, pressure skin necrosis, muscle ischemia from tightness [20]. The probability of nonunion, bone malrotation, and scissoring deformity of the fingers also increase. Every additional 5° of metacarpal shaft rotation leads to a 1.5 cm overlap at the fingertips [21]. One degree of metacarpal fracture rotation has been shown to produce 5° of fingertip rotation [1, 19] and severe loss of hand prehensile function, thus negatively affecting the patient’s quality of life. Accordingly, determining the optimal technique for fixation of oblique metacarpal shaft fractures is vital.

Previous studies have adopted in vitro experiments for metacarpal fracture fixation, human cadaveric bones [22], animal bones [7], or artificial bones [8, 12,13,14]. We employed artificial metacarpal bone because fresh human metacarpal bones were difficult to obtain and because some literature and the American Society for Testing and Materials had used artificial bones for experiments. Furthermore, we used a cantilever method to estimate the strength of the three fixation methods explored in the present study and adopted maximum fracture force and stiffness as the indices of fixation strength [8, 12,13,14].

Improvement in materials and techniques have provided orthopedic surgeons with many methods to fix oblique and unstable metacarpal shaft fractures, such as (1) K wire, (2) bone plate, (3) lag screw, and (4) intramedullary headless screw fixation [6]. Preceding biomechanical studies on metacarpal fractures have focused on transverse fractures [9, 10, 15]. However, the results of biomechanical experiments have revealed different findings in terms of fixation for oblique fracture. For example, the stiffness of crossed K-wire fixation was revealed to be equivalent to the combination of intraosseous wiring and oblique K-wiring; the stiffness of dorsal plating, with or without lag screws, was revealed to be higher than that of both aforementioned methods [15]. Accordingly, we compared fixation methods between lag screw fixation and bone plate fixation for oblique metacarpal shaft fractures.

In preceding in vitro experiments for oblique metacarpal shaft fractures, [23] applied cadaver metacarpals and 1.5-mm and 2.0-mm screws to fix oblique metacarpal fractures through bicortical interfragmentary fixation and the lag screw technique, respectively, and observed that the 1.5-mm bicortical interfragmentary screws provided sufficient fixation strength. The load to failure of the experiment was 72.31 N, which was much lower than that of double lag screw fixation in the present study (153.6 N). [22] compared the fixation strength of dorsal plating, lag screw fixation, and headless compression screw fixation and revealed that headless compression screw fixation offered the most satisfactory fixation. The maximum fracture force and stiffness of the double lag screw fixation were 234.1 N and 172.18 N/mm, respectively, which were much higher than our results (153.6 N and 57.0 N/mm, respectively). These inconsistencies were primarily because of the differences in the bone specimens used. The previous two studies employed cadaver metacarpals, which are more realistic, whereas we used artificial bones; each artificial bone specimen was identical, which explains the lower coefficient of variation in the present study. Furthermore, we conducted a cantilever bending test for a mechanical test rather than a three-point bending test employed in preceding studies.

In our mechanical experiment, the maximum fracture force of the 2LS group was nonsignificantly lower than that of the RP and LP groups. This implies that as long as patients avoid strong physical impact after the surgery, the fixation strength provided with double lag screw fixation can be adequate for metacarpal shaft fractures. Literature has also indicated that stiffness should be a more critical index for fixation strength assessment and a more reliable indication of construct stability than the maximum fracture force [14, 24], because during fracture healing, refracture of the metacarpal bone as a result of an extremely strong active or passive force does not and should never occur. In the present study, the RP and LP groups exhibited no significant difference in any aspect. In addition, the coefficient of variation of the 2LS group was more than twice as high as that of the RP and LP groups. Basically, in the same group, the larger the coefficient of variation, the worse the consistency of the experimental results, and because the artificial bone specimens used in this study were the same quality, the difference resulted from the material properties of the artificial bone itself might be excluded, so the large differences in experimental results are mainly due to inconsistency in the skill of fixation approach. Therefore, this suggested that double lag screw fixation was less fault tolerant. Even if the surgery is performed successfully, the stability would not be as high as that of the bone plate fixation. Therefore, more experienced surgeons are required to obtain optimal results when using double lag screw fixation.

Although bone plate fixation provides more stability, double lag screw fixation has other benefits: It is cost effective, is minimally invasive, and can reduce extensor tendon and sensory nerve adhesion. Moreover, our results indicated that it provided comparable stiffness to bone plate fixation. The fixation power it provides can sustain bony union. However, compared with patients treated with bone plate fixation, those who receive only screw fixation require more protection. The surgery may fail if patients receive external force damage that exceeds the maximum fracture force. Surgeons’ experience in performing lag screw fixation is critical, given that the coefficient of variation was significantly high even in the in vitro mechanical experiment where the variables were controlled, fracture pattern was consistent, and surgery was conducted by the same surgeon. Hence, caution must be applied when employing this surgical approach. Each screw should be inserted as perpendicularly to the fracture line as possible.

This study has some limitations. As encountered by many previous researchers [7, 14, 24, 25], fresh human metacarpal bones are difficult to obtain. Therefore, we used artificial bones for the experiment. However, the material properties of artificial bones are different from those of real human bones. Moreover, we used a cantilever bending test to assess fixation strength, but such a load pattern may not comprehensively simulate the actual strained condition in phalanx motions. Future studies should design more comprehensive experiments to obtain more detailed findings.

Conclusion

This study employed artificial metacarpal specimens to explore the fixation strength of different surgical methods for oblique metacarpal shaft fractures. Although the maximum bearable fracture force of double lag screw fixation was slightly lower than that of bone plate fixation, the stiffness of the two methods was comparable. However, double lag screw fixation requires considerable surgical experience for consistent results.

Availability of data and materials

The data sets used and analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

2 LS group:

Double lag screw

RP group:

Regular plate

LP group:

Locked plate

ANOVA:

Analysis of variance

References

  1. Dreyfuss D, Allon R, Izacson N, Hutt D. A comparison of locking plates and intramedullary pinning for fixation of metacarpal shaft fractures. Hand. 2019;14(1):27–33.

    Article  Google Scholar 

  2. Aykut S, Öztürk K, Özcan Ç, Demiroğlu M, Gürün AU, Özden E. Results of surgical treatment in metacarpal shaft fractures using low profile mini plates. Turk J Trauma Emerg Surg. 2015;21(4):279–84.

    Google Scholar 

  3. Curtis BD, Fajolu O, Ruff ME, Litsky AS. Fixation of metacarpal shaft fractures: biomechanical comparison of intramedullary nail crossed K-wires and plate-screw constructs. Orthop Surg. 2015;7(3):256–60.

    Article  Google Scholar 

  4. Low C, Wong H, Low Y, Wong H. A cadaver study of the effects of dorsal angulation and shortening of the metacarpal shaft on the extension and flexion force ratios of the index and little fingers. J Hand Surg Br Eur. 1995;20(5):609–13.

    Article  CAS  Google Scholar 

  5. Heckman JD, McKee M, McQueen MM, Ricci W, Tornetta P III. Rockwood and Green’s fractures in adults. Philadelphia: Lippincott Williams & Wilkins; 2014.

    Google Scholar 

  6. Kawamura K, Chung KC. Fixation choices for closed simple unstable oblique phalangeal and metacarpal fractures. Hand Clin. 2006;22(3):287–95.

    Article  Google Scholar 

  7. Chiu Y-C, Ho T-Y, Ting Y-N, Tsai M-T, Huang H-L, Hsu C-E, Hsu J-T. Effect of oblique headless compression screw fixation for metacarpal shaft fracture: a biomechanical in vitro study. BMC Musculoskelet Disord. 2021;22(1):1–8.

    Article  Google Scholar 

  8. Chiu Y-C, Hsu C-E, Ho T-Y, Ting Y-N, Tsai M-T, Hsu J-T. Bone plate fixation ability on the dorsal and lateral sides of a metacarpal shaft transverse fracture. J Orthop Surg Res. 2021;16(1):1–10.

    Article  Google Scholar 

  9. Massengill JB, Alexander H, Langrana N, Mylod A. A phalangeal fracture model—quantitative analysis of rigidity and failure. J Hand Surg. 1982;7(3):264–70.

    Article  CAS  Google Scholar 

  10. Vanik RK, Weber RC, Matloub HS, Sanger JR, Gingrass RP. The comparative strengths of internal fixation techniques. J Hand Surg. 1984;9(2):216–21.

    Article  CAS  Google Scholar 

  11. Black D, Mann R, Constine R, Daniels A. Comparison of internal fixation techniques in metacarpal fractures. J Hand Surg. 1985;10(4):466–72.

    Article  CAS  Google Scholar 

  12. Chiu Y-C, Hsu C-E, Ho T-Y, Ting Y-N, Tsai M-T, Hsu J-T. Effect of a figure-of-eight cerclage wire with two Kirschner wires on fixation strength for transverse metacarpal shaft fractures: an in vitro study with artificial bone. BMC Musculoskelet Disord. 2021;22(1):1–9.

    Article  Google Scholar 

  13. Chiu YC, Hsu CE, Ho TY, Ting YN, Wei BH, Tsai MT, Hsu JT. Comparison of the fixation ability of headless compression screws and locking plate for metacarpal shaft transverse fracture. Medicine 2021;100:39(e27375).

  14. Chiu Y-C, Tsai M-T, Hsu C-E, Hsu H-C, Huang H-L, Hsu J-T. New fixation approach for transverse metacarpal neck fracture: a biomechanical study. J Orthop Surg Res. 2018;13(1):1–7.

    Article  CAS  Google Scholar 

  15. Kilbourne BC, Paul EG. The use of small bone screws in the treatment of metacarpal, metatarsal, and phalangeal fractures. JBJS. 1958;40(2):375–83.

    Article  Google Scholar 

  16. Rüedi TP, Burri C, Pfeiffer KM. Stable internal fixation of fractures of the hand. J Trauma Acute Care Surg. 1971;11(5):381–9.

    Article  Google Scholar 

  17. Crawford GP. Screw fixation for certain fractures of the phalanges and metacarpals. JBJS. 1976;58(4):487–92.

    Article  CAS  Google Scholar 

  18. Trevisan C, Morganti A, Casiraghi A, Marinoni EC. Low-severity metacarpal and phalangeal fractures treated with miniature plates and screws. Arch Orthop Trauma Surg. 2004;124(10):675–80.

    Article  Google Scholar 

  19. Royle S. Rotational deformity following metacarpal fracture. J Hand Surg Br Eur. 1990;15(1):124–5.

    Article  CAS  Google Scholar 

  20. Taghinia AH, Talbot SG. Phalangeal and metacarpal fractures. Clin Plast Surg. 2019;46(3):415–23.

    Article  Google Scholar 

  21. Freeland AE, Jabaley ME, Hughes JL. Oblique and spiral metacarpal shaft fractures. In: Stable fixation of the hand and wrist. edn.. Springer; 1986: 55–57.

  22. Dyrna FG, Avery DM, Yoshida R, Lam D, Oeckenpöhler S, Cote MP, Obopilwe E, Rodner CM, Mazzocca AD. Metacarpal shaft fixation: a biomechanical comparison of dorsal plating, lag screws, and headless compression screws. BMC Musculoskelet Disord. 2021;22(1):1–8.

    Article  Google Scholar 

  23. Liporace FA, Kinchelow T, Gupta S, Kubiak EN, McDonnell M. Minifragment screw fixation of oblique metacarpal fractures: a biomechanical analysis of screw types and techniques. Hand. 2008;3(4):311–5.

    Article  Google Scholar 

  24. Barr C, Behn AW, Yao J. Plating of metacarpal fractures with locked or nonlocked screws, a biomechanical study: how many cortices are really necessary? Hand. 2013;8(4):454–9.

    Article  Google Scholar 

  25. Elfar J, Stanbury S, Menorca RMG, Reed JD. Composite bone models in orthopaedic surgery research and education. J Am Acad Orthop Surg. 2014;22(2):111.

    PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This research was partially supported by China Medical University, Taiwan (Grant No. CMU110-S-05).

Author information

Authors and Affiliations

Authors

Contributions

YC, TH and JH participated in the design of the study. YC, CH, TH, YT, MT, and JH performed the measurements. YC, TH and JH conducted the statistical analysis. YC, TH and JH conceived of the study, participated in its design and coordination, and drafted the manuscript. All authors have read and approved the final manuscript.

Corresponding author

Correspondence to Jui-Ting Hsu.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors have no financial or personal relationships with people or organizations that could be considered inappropriate for conducting this study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chiu, YC., Ho, TY., Hsu, CE. et al. Comparison of the fixation ability between lag screw and bone plate for oblique metacarpal shaft fracture. J Orthop Surg Res 17, 72 (2022). https://doi.org/10.1186/s13018-022-02963-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-022-02963-3

Keywords