Effects of contact stress on patellarfemoral joint and quadriceps force in fixed and mobile-bearing medial unicompartmental knee arthroplasty

Abstract

Background

Unicompartmental knee arthroplasty (UKA) is an effective treatment for end-stage, symptomatic unicompartmental osteoarthritis of the knee joint. However, patellofemoral joint degeneration is a contraindication to medial UKA. Therefore, the objective of this study was to evaluate the biomechanical effect of medial UKA using fixed-bearing (FB) and mobile-bearing (MB) design prostheses on the patellofemoral joint.

Methods

A three-dimensional finite-element model of a normal knee joint was developed using medical image data. We performed statistical analysis for each model. The differences in contact stress on the patellofemoral joint and the quadriceps force between the FB and MB designs were evaluated under a deep-knee-bend condition.

Results

At an early flexion angle, the results of contact stress showed no significant difference between the FB and MB medial UKA models compared with the intact model. However, at high flexion angles, we observed a significant increase in contact stress with the FB models compared with the intact model. On the contrary, in the case of the MB models, we found no statistically significant increment compared with the intact model. A larger quadriceps force was needed to produce an identical flexion angle for both the FB and MB UKA designs than for the intact model. At high flexion angles, a significant increase quadriceps force whit the FB model compared with the intact model.

Conclusions

Our results indicate that with medial UKA, the contact stress increased and greater quadriceps force was applied to the patellofemoral joint. However, performing UKA on a patellofemoral joint with osteoarthritis should not be difficult, unless anterior knee pain is present, because the increase in contact stress is negligible.

Background

Unicompartmental knee arthroplasty (UKA) is a surgical treatment alternative to total knee arthroplasty for isolated medial compartmental arthritis of the knee joint. The benefits of UKA include fewer complications, faster recovery, improved functional outcomes, and cost-effectiveness [1,2,3,4]. Therefore, medial UKA had been increasingly used for the treatment of medial compartmental osteoarthritis (OA) over the past two decades [5]. Historically, patellofemoral (PF) joint degeneration, and more specifically, advanced lateral PF joint facet degeneration, along with anterior knee pain, has been considered as an exclusion criterion for medial UKA [6, 7]. However, PF joint degeneration has been reported recently to have no influence on the clinical outcomes after UKA [8, 9]. In addition, whether preexisting PF joint degeneration is a contraindication to UKA is controversial. Thein et al. recently performed a study to determine the effect of medial fixed-bearing (FB) UKA on postoperative PF joint congruence and analyzed the effect of preoperative PF joint degeneration on the clinical outcome [10]. No correlation was observed between preoperative PF joint congruence or degeneration severity and the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores at 2-year follow-up [10]. Preoperative PF joint congruence and degenerative changes do not affect the clinical outcomes after UKA [10]. However, multiple studies that used the Oxford knee system indicated that neither preoperative anterior knee pain nor moderate radiological PF osteoarthritic changes affected the long-term clinical outcomes and survivorship of patients after mobile-bearing (MB) UKA [9, 11, 12]. One study suggested that MB UKA provides better restoration of normal knee kinematics, which theoretically translates to better patellar tracking and long-term outcomes [13]. Although several studies have revealed no significant differences in clinical outcomes and complication rates between the FB and MB UKA designs, the mode of failure often differs [14]. In addition, research on the biomechanical effect of medial UKA on the PF joint is lacking. The biomechanical effect on the PF joint can be investigated by performing a finite-element (FE) analysis to evaluate the contact stress and quadriceps force after medial UKA [15]. Accurate in silico evaluations of knee joint replacements are useful for clinical assessment [15].

Therefore, the objective of this study was to evaluate the biomechanical effects of medial UKA using FB and MB design prostheses on the PF joint. The differences in contact stress on the PF joint and quadriceps force between the FB and MB designs were evaluated under a deep-knee-bend condition. We hypothesized that applying medial UKA would not be difficult even with OA of the PF joint (unless accompanied by anterior knee pain) because the differences in biomechanical effect on the PF joint are negligible between the UKA and normal knee joint models.

Methods

Normal knee joint model

In this study, an existing three-dimensional non-linear FE model of the knee joint based on data from four male subjects (subject 1: age, 36 years; height, 178 cm; mass, 75 kg; subject 2: age, 34 years; height, 173 cm; mass, 83 kg; subject 3: age, 32 years; height, 182 cm; mass, 79 kg; subject 4: age, 34 years; height, 173 cm; mass, 71 kg) and one female subject (subject 5: age, 26 years; height, 163 cm; mass, 65 kg) was used. The FE model was developed using computed tomography and magnetic resonance imaging data with a slice thikness of 0.1 mm and 0.4 mm slice, respectively [16, 17] and included the bony structures of the knee joint and the soft tissues of the PF and tibiofemoral (TF) joint anatomies. The articular cartilage and menisci were defined as isotropic linearly elastic materials and transversely isotropic and linearly elastic materials, respectively [18]. The material properties of the articular cartilage and menisci are presented in Table 1.

Table 1 Material properties of the articular cartilage and menisci

All the ligaments were modeled with nonlinear and tension-only spring elements [19, 20]. Mesh convergence tests were performed to complete the simulation. Convergence was obtained if the relative change between two adjacent meshes was < 5%. The average element sizes were 0.8 mm for the cartilage and menisci, respectively. The details of the meshes used in the FE model were described in Table 2. The interfaces between the cartilage and bones were modeled to be fully bonded. Contact was applied between the femoral cartilage and meniscus, meniscus and tibial cartilage, and femoral cartilage and tibial cartilage for both the medial and lateral sides [16].

Table 2 Details of meshes used in the FE model

Medial UKA model

A FB UKA prosthesis (Zimmer, Inc., Warsaw, IN, USA) and a MB UKA prosthesis of the Oxford knee system (Biomet, Warsaw, IN) were virtually implanted in the medial compartment of the normal knee joint model. (Fig. 1) The bone models were imported and appropriately positioned, trimmed, and meshed with rigid elements according to the surgical techniques [16]. The tibial component was defined as a square (0°) inclination in the coronal plane, with a 5° posterior slope. The rotating axis was defined as a line parallel to the lateral edge of the tibial component passing through the center of the femoral component peg. A femoral component distal cut perpendicular to the mechanical axis of the femur and parallel to the tibial cut was reproduced. The height of the PE insert was identical to the anatomy in a sagittal plane aligned with the mechanical axis of the tibia and positioned at the medial edge of the tibia. The materials used for the femoral component, tibial insert, tibial baseplate, and bone cement were cobalt chromium molybdenum alloy, ultrahigh-molecular-weight polyethylene, titanium alloy, and polymethyl methacrylate, respectively (Table 3) [17, 21, 22].

Fig. 1
figure 1

FE model used in this study for a UKA model, b FB UKA componenet design, and c MB UKA componenet design

Table 3 Material properties of implant

The femoral component and tibial insert were in contact with a coefficient of friction of 0.04 [21]. The FE simulation involved two types of loading conditions corresponding to the loads used in the model validation experiment and to predictions of loading scenarios in daily activities. An axial loading of 1150 N was applied to the model to evaluate the contact stresses and compare them with those reported in previous studies [23] (Fig. 2). The second loading condition corresponded to a deep knee bend, and squat loading was applied to evaluate the knee joint mechanics. A computational analysis was performed using an anteroposterior force applied to the femur that was based on the compressive load applied to the hip with constrained femoral internal-external (IE) rotation, free medial-lateral translation, and knee flexion, for a combination of vertical hip and quadriceps loads. Therefore, a 6-degrees-of-freedom (DOF) TF joint was developed [24, 25]. A proportional-integral-derivative controller was incorporated into the computational model to control the quadriceps in a manner similar to that in previous experiments [26]. A control system was used to calculate the instantaneous displacement of the quadricep muscles to match the target flexion profile used in the experiment. Furthermore, IE and varus-valgus torques were applied to the tibia, while the remaining tibial DOF were constrained [24, 25].

Fig. 2
figure 2

Loading condition of the UKA FE model

The FE models were analyzed using Abaqus version 6.11 software (Simulia, Providence, RI, USA). The contact stress and quadriceps force on the PF joint were evaluated for the FB and MB medial UKA designs.

Statistical analyses

We performed the test at 11 time points (0.0 to 1.0 phases) for single cycles of deep-knee-bend loading conditions. To assess the two models, FB and MB, the condition of each model was compared with that of the normal knee in a pairwise manner by using non-parametric repeated-measures Friedman tests at each phase of the cycle. In this study, we used Wilcoxon rank test with Holm correction for post hoc comparisons to control the familywise error rate for the tests conducted within each phase of the cycle. Statistical analyses were performed using SPSS version 20.0.0 for Windows (SPSS Inc., Chicago, IL, USA). Statistical significance was set at P < .05.

Results

The results of the five subject-specific FE models were compared with previous results of the same model for model validation [23]. The mean contact stresses on the medial and lateral menisci in the present and previous studies are presented in Table 4.

Table 4 Comparison of the average contact stresses on the menisci for the validation of the model under an axial loading condition

The minor differences may be due to the variations in geometry, such as the thicknesses of the cartilage and meniscus, between the studies. However, the consistency between the results confirms the ability of the FE model to produce reasonable results [23]. Figure 3 shows the contact stresses on the PF joint with the FB and MB medial UKA designs under the deep-knee-bend condition. No significant difference in contact stress on the PF joint was observed between the FB and MB medial UKA models and the intact model at an early flexion angle. At a larger flexion angle, the contact stress showed a significant increase of 7% (on average), which is a small but significant increase, for the FB model compared with the intact model. For the MB models, we observed an increase of 4% (on average) in contact stress. However, we found no statistically significant increment.

Fig. 3
figure 3

Differences in contact stress on the PF joint with intact, FB, and MB design medial UKA under the deep-knee-bend condition (*P<.05)

The quadriceps forces exerted on the PF joint with the FB and MB medial UKA designs under the deep-knee-bend condition are shown in Fig. 4. A larger quadriceps force was needed to produce an identical flexion angle for both the FB and MB UKA designs than for the intact model. The quadriceps force rapidly increased the flexion of the knee joint in all the models. On average, the maximum quadriceps force significantly ranged from 2710 N for the MB UKA design to 2830 N for the FB UKA design. At a mid-flexion angle, the quadriceps forces were smaller for the FB and MB UKA designs than for the intact model. In addition, a lower quadriceps force was needed to produce identical flexion angles with the MB UKA design than with the FB UKA design. And at high flexion angles, a significant increase quadriceps force whit the FB model compared with the intact model. The FB and MB UKA designs required 12% and 8% (on average) more quadriceps force, respectively, than the intact model.

Fig. 4
figure 4

Differences in the quadriceps muscle force for intact, FB, and MB tibial insert materials under the deep-knee-bend condition (*P < .05)

Discussion

The most important finding of this study was that the contact stress on the PF joint increased less with the MB UKA design than with the FB UKA design; however, no significant difference in contact stress on the PF joint was found between the medial UKA and intact knee joints. The quadriceps force needed to produce the same flexion angle with the MB UKA design was lower than that with the FB UKA design. UKA can be performed with either an FB or MB design. In a prospective study that involved 48 patients, who were randomly assigned to either FB or MB UKA prostheses, Li et al. observed better knee kinematics and a lower incidence of radiolucency with the MB design, but the Knee Society, WOMAC, and SF-36 scores were equivalent between the two designs [27]. In another study, the range of motion, limb alignment, patient-reported outcomes, incidence of aseptic loosening, and reoperation rate were identical between the FB and MB UKA designs [28]. However, the time to reoperation and failure mode differed. Early failure due to bearing dislocation occurred with the MB design, whereas late failure due to polyethylene wear occurred with the FB design. A previous study indicated that during a ≥ 15-year follow-up period, some type of revision arthroplasty was required for 12 (15%) of 77 knees in the case of FB UKA (Miller-Galante; Zimmer) and for 10 (12%) of 79 knees in the case of MB UKA (Oxford; Biomet) [29]. No significant differences were observed in the number of knees with progressive lateral OA that required revision arthroplasty between the FB and MB UKA designs [29]. Thus, many arguments have emerged regarding the biomechanical issues of the FB and MB UKA designs. In previous studies on the progression of OA after UKA, the radiological assessment was neither blinded nor randomized [27].

The advantage of FE analysis is that the impact of the UKA design can be determined without external variables [30]. Most in vitro biomechanical studies have involved evaluations using aged cadaveric subjects with loosening between the specimen and the device, and tissue attenuation, which can occur owing to the successive loading in mechanical testing [26]. An intact joint model was the foundation of this study and involved FEM validation steps. The results exhibited good agreement with those of previous computational studies [23, 31]. Therefore, the UKA models used in the present study and related analyses are considered reliable.

Kozinn and Scott proposed that UKA should not be offered to patients with PF joint arthritis for optimal results [32]. This sparked a contentious debate on PF joint disease because other authors demonstrated only a weak correlation between PF degenerative changes and anterior knee pain [11, 33].

In addition, owing to the differences in the design and biomechanics of the FB UKA model, damage to the PF joint has traditionally been a contraindication. Lim et al. recently showed that the presence of significant preoperative radiological PF disease does not affect long-term implant success, and patients had excellent postoperative functional outcomes for 10 years [34]. In the present study, the MB UKA design produced a smaller increase in contact stress on the PF joint than the intact joint and FB UKA design models. The results of previous studies on MB UKA indicated that the presence of PF degeneration does not compromise clinical outcomes because the implant is believed to be more patella friendly owing to better kinematics, which supports our results [34, 35]. Moreover, although the contact stress on the PF joint increased with both the FB and MB UKA designs, the increase was not statistically significant. Biomechanical studies have indicated that the progression of arthritis of the PF joint typically does not necessitate revision.

The quadriceps force needed to produce a squatting motion was greater for the FB design than for the MB design (by as much as 120 N for knee flexion angles > 100°). Thus, increased quadriceps strength leads to improved functional performance [36]. As patients who have undergone OA and knee arthroplasty experience significant quadriceps weakness, the FB UKA design, which increases the required quadriceps force, can result in more difficulty for patients to walk, kneel, or perform a deep knee bend [37]. This agrees with the results of a previous in vitro study in which a UKA model required less quadriceps force at a mid-flexion angle than an intact model [27].

From a biomechanical viewpoint, our results indicate that the risk of progressive OA of the PF joint can be reduced with the MB UKA design because it preserves the normal biomechanical effect, in contrast to the FB UKA design. In addition, the MB UKA design requires a lower quadriceps force and makes it easier for recipients to kneel, squat, or rise from a chair.

The three strengths of our study should be highlighted. First, a well-validated setup that accounted for numerous previous results was used. Second, in contrast to previous UKA studies, the present study included the tibia, femur, and related soft tissues in the FE model. Third, in contrast to the current biomechanical UKA models, the model used in this study included the deep-knee-bend and squat loading conditions, rather than the simple vertical static loading condition.

Despite these strengths, this study had certain limitations. First, the results did not predict clinical results or patient satisfaction. Second, the computational model was developed using data from four male subjects and one female subject. Using data from subjects of various ages would improve the validity of the results, as it would increase the diversity of the knee joint geometry. However, in this study, our objective was to evaluate the biomechanical effect of UKA in young individuals. Third, the bony structures were assumed to be rigid. In reality, the bone is composed of cortical and cancellous tissues. However, the main purpose of the study was not to evaluate the effects of different prostheses on bone. In addition, this assumption had a minimal influence on the results of the study because the bone stiffness exceeds that of the relevant soft tissues [23]. Finally, the simulation only involved a deep knee bend; thus, simulations involving rising from or sitting on chairs, climbing/descending stairs, and squatting should be performed in future investigations.

In conclusion, this study provides biomechanical evidence that degenerative changes in the PF joint should not be considered an absolute contraindication to treatment with medial UKA. In addition, UKA is not problematic even in PF joints with OA, unless accompanied by anterior knee pain, because no significant difference in contact stress was observed.

Availability of data and materials

Not applicable.

Abbreviations

UKA:

Unicompartmental knee arthroplasty

OA:

Osteoarthritis

PF:

Patellofemoral

FB:

Fixed-bearing

MB:

Mobile-bearing

TF:

Tibiofemoral

IE:

Internal-external

DOF:

Degree of freedom

References

  1. Siman H, Kamath AF, Carrillo N, Harmsen WS, Pagnano MW, Sierra RJ. Unicompartmental Knee Arthroplasty vs Total Knee Arthroplasty for Medial Compartment Arthritis in Patients Older Than 75 Years: Comparable Reoperation, Revision, and Complication Rates. J Arthroplasty. 2017;32(6):1792–7.

    Article  Google Scholar 

  2. Liddle AD, Judge A, Pandit H, Murray DW. Adverse outcomes after total and unicompartmental knee replacement in 101,330 matched patients: a study of data from the National Joint Registry for England and Wales. Lancet (London, England). 2014;384(9952):1437–45.

    Article  Google Scholar 

  3. Kim MS, Koh IJ, Choi YJ, Lee JY, In Y. Differences in Patient-Reported Outcomes Between Unicompartmental and Total Knee Arthroplasties: A Propensity Score-Matched Analysis. J Arthroplasty. 2017;32(5):1453–9.

    Article  Google Scholar 

  4. Shankar S, Tetreault MW, Jegier BJ, Andersson GB, Della Valle CJ. A cost comparison of unicompartmental and total knee arthroplasty. The Knee. 2016;23(6):1016–9.

    Article  Google Scholar 

  5. Borus T, Thornhill T. Unicompartmental knee arthroplasty. J Am Acad Orthop Surg. 2008;16(1):9–18.

    Article  Google Scholar 

  6. Foran JR, Brown NM, Della Valle CJ, Berger RA, Galante JO. Long-term survivorship and failure modes of unicompartmental knee arthroplasty. Clin Orthop Relat Res. 2013;471(1):102–8.

    Article  Google Scholar 

  7. Kozinn SC, Scott RD. Surgical treatment of unicompartmental degenerative arthritis of the knee. Rheum Dis Clin North Am. 1988;14(3):545–64.

    CAS  PubMed  Google Scholar 

  8. Berend KR, Lombardi AV Jr. Liberal indications for minimally invasive oxford unicondylar arthroplasty provide rapid functional recovery and pain relief. Surg Technol Int. 2007;16:193–7.

    PubMed  Google Scholar 

  9. Beard DJ, Pandit H, Ostlere S, Jenkins C, Dodd CA, Murray DW. Pre-operative clinical and radiological assessment of the patellofemoral joint in unicompartmental knee replacement and its influence on outcome. J Bone Joint Surg Br. 2007;89(12):1602–7.

    Article  CAS  Google Scholar 

  10. Thein R, Zuiderbaan HA, Khamaisy S, Nawabi DH, Poultsides LA, Pearle AD. Medial unicondylar knee arthroplasty improves patellofemoral congruence: a possible mechanistic explanation for poor association between patellofemoral degeneration and clinical outcome. J Arthroplasty. 2015;30(11):1917–22.

    Article  Google Scholar 

  11. Hamilton TW, Pandit HG, Maurer DG, Ostlere SJ, Jenkins C, Mellon SJ, et al. Anterior knee pain and evidence of osteoarthritis of the patellofemoral joint should not be considered contraindications to mobile-bearing unicompartmental knee arthroplasty: a 15-year follow-up. Bone Joint J. 2017;99-b(5):632–9.

    Article  CAS  Google Scholar 

  12. Kim KT, Lee S, Lee JS, Kang MS, Koo KH. Long-term clinical results of unicompartmental knee arthroplasty in patients younger than 60 years of age: minimum 10-year follow-up. Knee Surg Relat Res. 2018;30(1):28–33.

    Article  Google Scholar 

  13. Argenson JN, Parratte S. The unicompartmental knee: design and technical considerations in minimizing wear. Clin Orthop Relat Res. 2006;452:137–42.

    Article  Google Scholar 

  14. Gleeson RE, Evans R, Ackroyd CE, Webb J, Newman JH. Fixed or mobile bearing unicompartmental knee replacement? A comparative cohort study. Knee. 2004;11(5):379–84.

    Article  CAS  Google Scholar 

  15. Kwon OR, Kang KT, Son J, Suh DS, Baek C, Koh YG. Importance of joint line preservation in unicompartmental knee arthroplasty: Finite element analysis. J Orthop Res. 2017;35(2):347–52.

    Article  CAS  Google Scholar 

  16. Park KK, Koh YG, Park KM, Park JH, Kang KT. Biomechanical effect with respect to the sagittal positioning of the femoral component in unicompartmental knee arthroplasty. Bio-med Mater Eng. 2019;30(2):171–82.

    Article  Google Scholar 

  17. Koh YG, Lee JA, Lee HY, Chun HJ, Kim HJ, Kang KT. Anatomy-mimetic design preserves natural kinematics of knee joint in patient-specific mobile-bearing unicompartmental knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2019;28(5):1465–72.

  18. Haut Donahue TL, Hull M, Rashid MM, Jacobs CR. How the stiffness of meniscal attachments and meniscal material properties affect tibio-femoral contact pressure computed using a validated finite element model of the human knee joint. J Biomech. 2003;36(1):19–34.

    Article  Google Scholar 

  19. Mesfar W, Shirazi-Adl A. Biomechanics of the knee joint in flexion under various quadriceps forces. The Knee. 2005;12(6):424–34.

    Article  CAS  Google Scholar 

  20. Takeda Y, Xerogeanes JW, Livesay GA, Fu FH, Woo SL. Biomechanical function of the human anterior cruciate ligament. Arthroscopy. 1994;10(2):140–7.

    Article  CAS  Google Scholar 

  21. Godest AC, Beaugonin M, Haug E, Taylor M, Gregson PJ. Simulation of a knee joint replacement during a gait cycle using explicit finite element analysis. J Biomech. 2002;35(2):267–75.

    Article  CAS  Google Scholar 

  22. Inoue S, Akagi M, Asada S, Mori S, Zaima H, Hashida M. The Valgus Inclination of the Tibial Component Increases the Risk of Medial Tibial Condylar Fractures in Unicompartmental Knee Arthroplasty. J Arthroplasty. 2016;31(9):2025–30.

    Article  Google Scholar 

  23. Pena E, Calvo B, Martinez MA, Palanca D, Doblare M. Why lateral meniscectomy is more dangerous than medial meniscectomy. A finite element study. J Orthop Res. 2006;24(5):1001–10.

    Article  Google Scholar 

  24. Kutzner I, Heinlein B, Graichen F, Bender A, Rohlmann A, Halder A, et al. Loading of the knee joint during activities of daily living measured in vivo in five subjects. J Biomech. 2010;43(11):2164–73.

    Article  CAS  Google Scholar 

  25. Halloran JP, Clary CW, Maletsky LP, Taylor M, Petrella AJ, Rullkoetter PJ. Verification of predicted knee replacement kinematics during simulated gait in the Kansas knee simulator. J Biomech Eng. 2010;132(8):081010.

    Article  Google Scholar 

  26. Kang KT, Koh YG, Son J, Kim SJ, Choi S, Jung M, et al. Finite element analysis of the biomechanical effects of 3 posterolateral corner reconstruction techniques for the knee joint. J Arthroscopy. 2017;33(8):1537–50.

    Article  Google Scholar 

  27. Heyse TJ, El-Zayat BF, De Corte R, Scheys L, Chevalier Y, Fuchs-Winkelmann S, et al. Biomechanics of medial unicondylar in combination with patellofemoral knee arthroplasty. Knee. 2014;21(Suppl 1):S3–9.

    Article  Google Scholar 

  28. Cheng T, Chen D, Zhu C, Pan X, Mao X, Guo Y, et al. Fixed- versus mobile-bearing unicondylar knee arthroplasty: are failure modes different? Knee Surg Sports Traumatol Arthrosc. 2013;21(11):2433–41.

    Article  Google Scholar 

  29. Parratte S, Pauly V, Aubaniac JM, Argenson JN. No long-term difference between fixed and mobile medial unicompartmental arthroplasty. Clin Orthopaed Relat Res. 2012;470(1):61–8.

    Article  Google Scholar 

  30. Kang KT, Koh YG, Son JH, Kwon OR, Lee JS, Kwon SK. Influence of increased posterior tibial slope in total knee arthroplasty on knee joint biomechanics: a computational simulation study. J Arthroplasty. 2018;33(2):572–9.

  31. Kang KT, Kim SH, Son J, Lee YH, Chun HJ. Computational model-based probabilistic analysis of in vivo material properties for ligament stiffness using the laxity test and computed tomography. J Mater Sci Mater Med. 2016;27(12):183.

    Article  Google Scholar 

  32. Kozinn SC, Scott R. Unicondylar knee arthroplasty. J Bone Joint Surg Am. 1989;71(1):145–50.

    Article  CAS  Google Scholar 

  33. Song EK, Park JK, Park CH, Kim MC, Agrawal PR, Seon JK. No difference in anterior knee pain after medial unicompartmental knee arthroplasty in patients with or without patellofemoral osteoarthritis. Knee Surg, Sports Traumatol Arthrosc. 2016;24(1):208–13.

    Article  Google Scholar 

  34. Lim JW, Chen JY, Chong HC, Pang HN, Tay DKJ, Chia SL, et al. Pre-existing patellofemoral disease does not affect 10-year survivorship in fixed bearing unicompartmental knee arthroplasty. Knee Surg Sports Traumatol arthrosc. 2019;27(6):2030–6.

    Article  Google Scholar 

  35. Argenson JN, O'Connor JJ. Polyethylene wear in meniscal knee replacement. A one to nine-year retrieval analysis of the Oxford knee. J Bone Joint Surg Br. 1992;74(2):228–32.

    Article  CAS  Google Scholar 

  36. Hurley MV, Scott DL. Improvements in quadriceps sensorimotor function and disability of patients with knee osteoarthritis following a clinically practicable exercise regime. Br J Rheumatol. 1998;37(11):1181–7.

    Article  CAS  Google Scholar 

  37. Mizner RL, Petterson SC, Stevens JE, Vandenborne K, Snyder-Mackler L. Early quadriceps strength loss after total knee arthroplasty. The contributions of muscle atrophy and failure of voluntary muscle activation. J Bone Joint Surg Am. 2005;87(5):1047–53.

    Article  Google Scholar 

Download references

Funding

There is no funding source.

Author information

Authors and Affiliations

Authors

Contributions

HMK designed the study and drafted the paper. JAL developed the three-dimensional model. YGK evaluated the FEA results. KKP validated the data. KTK supervised the study and analyzed the data. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Kyoung-Tak Kang.

Ethics declarations

Ethics approval and consent to participate

Approval was not required, as neither human participants nor animals were involved in this study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kwon, H.M., Lee, JA., Koh, YG. et al. Effects of contact stress on patellarfemoral joint and quadriceps force in fixed and mobile-bearing medial unicompartmental knee arthroplasty. J Orthop Surg Res 15, 517 (2020). https://doi.org/10.1186/s13018-020-02047-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-020-02047-0

Keywords