Skip to main content
  • Research article
  • Open access
  • Published:

Three-dimensional technology assisted trabecular metal cup and augments positioning in revision total hip arthroplasty with complex acetabular defects

Abstract

Background

Revision total hip arthroplasty (THA) with large acetabular defect remains a challenge. Though trabecular metal (TM) cup and augments have been introduced in defect reconstruction with good result, the accurate positioning of implant is important to avoid complications. Therefore, we aimed to evaluate the usefulness of three-dimensional (3D) simulation and 3D model in assisting implant positioning during complex revision THA.

Methods

Sixteen patients (18 hips) who underwent revision THA with a Paprosky type III acetabular defect were analyzed retrospectively. Placement of acetabular cup and TM augments was simulated with 3D simulation software and 3D model preoperatively. Cup anteversion, abduction angle, and hip center were measured in each case preoperatively and postoperatively. Primary outcome was the percentage of outliers according to Lewinnek safe zone and Harris hip score (HHS). Secondary outcome was the correlation between the 3D planned and the postoperative value.

Results

The percentage of outliers was significantly corrected from 77.78% (14/18) preoperatively to 38.88% (7/18) postoperatively (p = 0.04). There was a significant correlation between mean planned cup anteversion and postoperative value (13.39 vs 11.99, r = 0.894; p < 0.001). There was a significant correlation between mean planned abduction and postoperative value (42.67 vs 44.91, r = 0.921, p < 0.001). The number of planned and used augments was the same in all the cases. In 15 cases (83.33%), the size of planned and used TM augments was the same. The HHS was significantly improved at final follow-up (80.94 vs 27.50, p < 0.001). No cases presented dislocation or radiological signs of loosening.

Conclusion

Preoperative 3D simulation and model were considered the useful method to assist implant positioning in revision THA with complex acetabular defect, with moderate to high accuracy and satisfied clinical outcome.

Background

As the aging population is coming, more and more primary total hip arthroplasty (THA) will be performed annually [1]. As a result, the revision THA is estimated to increase exponentially in the future. Compared with primary THA, the revision THA is more difficult with complications, such as the intraoperative fracture, dislocation risk, and aseptic loosening [2, 3].

In revision THA, acetabular defect is not uncommon and regarded as a challenge. Normally, the type of bone loss and remaining bone quality determine the way of acetabular reconstruction [4]. The trabecular metal cups and augments have been commonly used together for individualized acetabular defect reconstruction with satisfied outcome [5, 6]. Characteristics of TM implant including high porosity and low modulus of elasticity could provide strong primary stability and promote a deep bony in-growth [7].

The outcome of revision surgery strongly depends on the accurate position of implant [8]. Malposition of the acetabular component increases the dislocation risk and other complications [9]. Previous study showed preoperative 3D planning technology improved the accuracy of implant position in primary THA [10]. However, application of 3D simulation and model to assist TM cup and augments positioning in revision THA was rarely reported [11, 12]. We aimed to evaluate the advantages and disadvantages of this technology and to provide suggestions for its clinical application in revision THA with complex acetabular defect.

Patients and methods

Patient population

Ethics approval was obtained from our hospital institutional review board. We performed a retrospective study using our hospital electronical medical records, including clinical notes, imaging, and clinical scores. We identified all patients who underwent revision THA with using the 3D simulation plan and 3D model from May 2013 to July 2017.

The inclusion criteria are as follows: (1) revision THA with acetabular defect, Paprosky type III; (2) using both TM cup and augments in acetabulum reconstruction; (3) using preoperative 3D simulation plan and 3D model based on the preoperative computed tomography (CT) scans.

The exclusion criteria are as follows: (1) primary THA; (2) the reasons of first THA were tumor or infection.

A total of 16 patients (18 hips) undergoing acetabular revision procedures were included in the study.

Preoperative planning protocol

Three-dimensional pelvis model

Before the surgery, patients underwent computer tomography (CT) scan. Scans covered the pelvis and femur with a 1.25-mm slice thickness (General Electric Company, USA). The CT scans were performed at the same medical imaging center with the same parameters.

Then, the CT scan data was imported into radiological post-processing software Materialise Mimics software (version 15.0; Materialize, Leuven, Belgium) to create a simulated model of the pelvis. After reconstruction, the 3D images of pelvis were simultaneously displayed in software and data was transported to the 3D printer (Formlab, America); then, 3D pelvis model was printed using the resin material. Finally, we could get a 3D pelvis model in real size with all defects, which helped to assess the acetabular bone defect. In our study, we classified the acetabular defect according to Paprosky classification [13].

Simulation of cup and TM augments positioning

The acetabular cup and TM augment positioning process were simulated in Materialise Mimics software. The acetabular cup (Zimmer, USA) and TM augment templates (Zimmer, USA) were placed in 3D virtual acetabulum step by step (Fig. 1a, b). The cup orientation was determined by preoperatively set abduction and anteversion angles. The distance from the cup edge to acetabular edge was measured relative to the anterior, posterior walls as well as to the superior edge in software, which was given to surgeons in operation. In addition, the plan of 3D simulation was rechecked by positioning cup and augments in pelvis model again (Fig. 1c). If it was proved to be practicable and suitable, the final views of the cup and trabecular augments within the acetabulum were given to the surgeon at the time of surgery.

Fig. 1
figure 1

a, b Cup and trabecular metal augments positioning with 3D simulation technology. c Cup and trabecular metal augments were positioned in 3D plaster model. d Cup and trabecular augments were placed based on the 3D plan

Surgical protocol

The 3D printed pelvis model was sterilized and prepared for using in surgery. The revision THA was performed by the same medical team using a posterolateral approach. First of all, we removed the previous failed hip implant, debrided the remaining acetabular cavity, and reassessed acetabula bony defect comparing with the 3D pelvis model. The acetabulum was gently reamed until contact was made with the bleeding host bone, then inserted the new cup and TM augments based on the parameters from preoperative 3D planning (Fig. 1d). We tried to make the distance from cup edge to acetabular edge as the same as the values of 3D planning by using sterilized flexible ruler in operation.

Rehabilitation after operation

On the first day after surgery, patients were encouraged to move the knee along with static quadriceps. Partial weight-bearing walking with crutches was requested until 6 weeks after surgery; a full weight-bearing gait was permitted at 6 weeks postoperatively.

Parameters on X-ray

In all patients, preoperative and postoperative angle measurements were performed on standardized radiographs of the Picture Archiving and Communication System (PACS). The radiographs comprised an anteroposterior view of the pelvis centered over the pubic symphysis with the hips at 15° of internal rotation. The instructions of parameter are shown in Fig. 2.

Fig. 2
figure 2

Measurement with anteroposterior view of an X-ray. a measures the angle of anteversion, anteversion angle = arcsin (D1/D2); D1: short red line of the ellipse; D2: long blue line. b measures the angle of abduction. c shows the vertical and horizontal position of COR. A: vertical position of COR; B: horizontal position of COR

Inter- and intra-observer reliability was assessed by intraclass correlation coefficients (ICCs) for measurements of angle parameters. For the interobserver reliability, all the angle parameters were measured by two independent authors (X.P.C; Q.T.L). For the intra-observer reliability, the same radiograph was measured by each observer after 1 month. The ICCs larger than 0.8 were considered to have excellent reliability of measurement. Excellent reliability of measurement was found in our study (ESM-1), so radiological measurements from only one investigator were included in analysis.

Acetabular cup anteversion angle

We measure the acetabular cup anteversion angle on anterior-posterior pelvic plain radiograph. According to Lewinnek, anteversion angle = arcsin (D1/D2) [14]. D1 is the distance of the short axis of the ellipse; D2 is the long axis of the acetabular component, reflecting the maximum diameter of the cup (Fig. 2a).

Acetabular cup abduction angle

The abduction angle was the angle between acetabular axis and trans-ischial in the coronal plane (Fig. 2b).

Position of center of rotation

The center of rotation of hip was confirmed by using a concentric circular region of interest that was digitally drawn to best fit the acetabular cup. Vertical position of the center of rotation (COR) was defined as the distance between the COR and the inter-teardrop line. The horizontal position of COR was defined as the distance from the COR to the floor of the acetabular teardrop (Fig. 2c).

Safe zone

The “safe zone” for acetabular inclination and anteversion in total hip arthroplasty (THA) was firstly defined by Lewinnek et al. in 1978 [14]. They defined the ideal inclination angle of acetabular cup was 30° to 50°, and the optimal anteversion angle was 5° to 25°.

Evaluation criteria

The main evaluation criterion was the percentage of postoperative outliers according to the Lewinnek safe zone. The second criterion was the correlation between the preoperative plan and the postoperative measurement. The third criterion was the difference between the number and size of planned and postoperative used metal augments. Patients were assessed clinically according to the Harris hip score before surgery and at final follow-up.

Statistical analysis

Kolmogorov-Smirnov test was used for testing normality. Data of abduction angle was non-normally distributed; data of anteversion angle, hip center, and HHS was normally distributed. Paired T test was used to assess the difference between preoperative/postoperative; 3D planned/postoperative anteversion angle. Wilcoxon test was used for abduction angle. Chi-square test was used for comparing preoperative/postoperative percentage of outliers. The Pearson coefficient test was used to assess the correlation between planned and postoperative anteversion angle and Spearman test for assessing correlation of abduction angle. Statistical significance was defined as a two-tailed p value < 0.05. The analysis was performed using SPSS 20 (Chicago, IL, USA).

Results

Demographic data

From May 2013 to July 2017, 18 patients (20 hips) underwent acetabular reconstruction using trabecular metal cup and augments with 3D preoperative planning. After reviewing medical records, 2 patients were excluded due to primary THA. Finally, 16 patients (18 hips) were included in the study and the demographic data is demonstrated in Table 1.

Table 1 Patient characteristics

Patients included 11 men and 5 women, with mean age of 58.06 ± 8.29 years at the time of surgery. Paprosky type IIIA defect was found in 13 hips, whereas a Paprosky type IIIB defect was present in 5 hips. The reasons for the primary THA were hip osteoarthritis (n = 3), hip fractures (n = 4), avascular necrosis of femoral head (n = 9), and ankylosing spondylitis (n = 2). The mean time from the primary THA to the revision was 16.78 ± 7.46 years. The etiology requiring revision THA was aseptic loosening in 15 hips and septic loosening in 3 hips.

Intraoperative data

The mean operation time was 254 ± 91 min; blood loss was 891 ± 423 ml. The mean blood transfusion was 860 ± 400 ml.

Radiologic outcome

Based on the Lewinnek safe zone, 11 of 18 (61.1%) patients were positioned within the safe zone (Fig. 3). Of the 7 outliers, 3 were out of the abduction safe zone, 5 were out of the anteversion safe zone, and 1 was out of both safe zones. Percentage of outliers was corrected from 77.78% (14/18) preoperatively to 38.89% (7/18) postoperatively, with statistical significance (p = 0.040).

Fig. 3
figure 3

Scatter plot of each hip’s abduction and anteversion angles within the safe zone of Lewinnek et al.

As shown in Table 2, ratio of vertical position of COR in surgical site/contralateral site was corrected from 1.15 ± 0.19 to 1.09 ± 0.20 postoperatively (p = 0.185). Ratio of horizontal position of COR in surgical site/contralateral site was changed from 0.97 ± 0.21 to 1.00 ± 0.18 postoperatively (p = 0.193). Though the position of COR was corrected to be nearly as the same as contralateral site, no statistical significance was found in our study.

Table 2 Anteversion, abduction, position of COR before and after revision THA

The correlation between planned and postoperative cup orientation is shown in Fig. 4. The mean planned cup anteversion value did not differ from the postoperative value (− 1.39 ± 4.1; p = 0.168), and a strong correlation was found (r = 0.894; p < 0.001). There was deviation between the mean planned abduction and the postoperative value (2.24 ± 3.02; p = 0.006), but a strong correlation between these two values was found (r = 0.921, p < 0.001).

Fig. 4
figure 4

a, b There was a significant correlation between the planned and the final postoperative values of cup anteversion and abduction

Number and size of 3D planned and used TM augments are shown in Table 3. Thus, the number of planned and used augments was same in all the cases. In 15 cases (83.33%), the size of planned and used TM augments was the same. In other cases, the difference was not over 4 mm.

Table 3 Comparison of number and size of planned and used augments

Clinical outcome

Follow-up data were available for all the patients included with a mean length of follow-up of 27.72 ± 12.18 months. The mean HHS was significantly improved from 27.50 ± 6.54 preoperatively to 80.94 ± 5.19 at final follow-up (p < 0.001). To sum up, the final scores were excellent in 0 hip, good in 13 hips, modest in 4 hips, and poor in 1 hip. The patient with lowest score was mainly influenced by the non-surgical site, which was also aseptic loosening but without surgery.

Complications

Two cases had intraoperative periprosthetic femoral fracture, which were both fixed by locking plates and proceeded to bony union at final follow-up. There was no dislocation or revision required in any cases. There was no migration of acetabular component or loosening of the TM augment.

Discussion

For complex revision THA with large acetabular defect, loss of normal anatomic landmarks increases the difficulty of implant positioning. An optimal implant position can prevent many complications, such as dislocation, acetabular component migration, wear, and osteolysis which affect the overall clinical outcome [15].

The type of implant used for acetabular defect reconstruction was variable in literature, such as reconstruction rings with structural allograft [16], cup cage reconstruction [17], and trabecular metal cups and augments [18]. The use of trabecular cup and TM augments maximized the contact with the host bone regardless of the size and shape of the bone defect, which was proved to have satisfied clinical outcome in previous studies [18, 19]. This study proved that using TM cup and augments had encouraging short- to mid-term outcomes. There was no sign of loosening or implant failure at final follow-up. Furthermore, there was high accuracy in predicting the number and size of used augments with 3D technique, which would greatly decrease the preoperative preparing time and provide valuable plan for the surgeons.

Restoring the COR to an anatomic position is important to hip biomechanics and implant stability [20]. Using cup and TM augments allows the surgeon to achieve a reduction of the migration of hip COR. In our study, we compared preoperative and postoperative ratio of position of COR in surgical site/contralateral site, which we wished the postoperative ratio to be more closed to 1. Both the vertical and horizontal positions of COR were corrected to be more similar with contralateral site, though there was no statistical significance, which may be due to the sample size.

Though the accuracy of implant position deeply influences the clinical outcome, the technique aiming to improve acetabular component position accuracy in revision THA was limited. For accuracy of freehand cup position in primary THA, Bosker et al. reported accuracy for cup placement in the Lewinnek safe zone were 85.2% and 82.7% in radiographic abduction and radiographic anteversion respectively [21]. Another study conducted by Minoda et al. also reported the accuracy of freehand cup position was over 80% [22]. However, freehand cup positioning has lower accuracy in revision THA. In a series case study involving 34 patients with Paprosky type III defects, only 19 (56%) were freehand positioned within the safe zone of Lewinnek postoperatively [23]. The ability of a surgeon to determine the difference between 15° and 30° was difficult and unreliable without the use of advanced technology [24]. Therefore, a supportive tool was needed to assist implant positioning in revision THA.

This study showed the 61.1% of revised cups were optimally positioned based on Lewinnek safe zone, which demonstrated good accuracy. In addition, 3D technology was also proved to be effective for primary THA in patients with complex acetabular deformity. Recently, Coral et al. reported a case of using 3D technology to assist primary THA in patient diagnosed with hip coxarthrosis due to previous untreated acetabular fracture (Paprosky type III). They found the center of rotation of the hip and length discrepancy was accurately recovered as preoperative plan without signs of loosening, subsidence, or osteolysis at final follow-up [25]. Compared with CT-based navigation, the 3D preoperative simulation technology and 3D model had lower accuracy [26,27,28]. Additionally, previous study reported there was a high accuracy by using imageless navigation-assisted cup positioning [29]. However, in contrast to CT-based navigation technique, the 3D planning was easy to achieve without increasing operative time and extra cost. Consequently, 3D planning was a good compromise method between accuracy on the one hand and extra cost and operative time on the other hand.

With the 3D simulation and pelvis model, preoperative 3D acetabula anatomy was assessed clearly, and the plan of cup and TM augments were known beforehand. This study found the strong correlation between final cup anteversion and abduction angles with the 3D planned value. There was deviation between the postoperative and 3D planned abduction value. This may be due to a poor exposure of the proximal part of the acetabulum, which influences measuring the distance from the cup to the superior edge. Another reason may be related to the change of patient position during surgery. All the cases in our study were in lateral supine position; the change of position may affect the pelvic tilt which influences the surgeon to assess the abduction angle.

There are several limitations of this study. Firstly, being a retrospective study design, it is never as ideal as randomized controlled trial for comparison of this technique with other technique of reconstruction. Secondly, this was a study with small but reasonable sample size, because the trabecular metal augments have limited indication in revision surgery. In the future, we plan to compare revision groups with or without 3D simulation and models.

Conclusion

There was a statistically significant correlation between 3D planned and postoperative value. Preoperative 3D simulation and model were considered the useful method to assist implant positioning in complex revision THA, with moderate to high accuracy, and with satisfied clinical outcome and lower complication rate. Moreover, it had high accuracy in predicting number and size of TM augments used intraoperatively.

Availability of data and materials

The data and materials are available from the medical records department of Guangdong Provincial People’s Hospital

Abbreviations

3D:

Three-dimensional

COR:

Center of rotation

HHS:

Harris hip score

ICCs:

Intraclass correlation coefficients

PACS:

Picture Archiving and Communication System

THA:

Total hip arthroplasty

TM:

Trabecular metal

References

  1. Kurtz S, Ong K, Lau E, et al. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. The Journal of bone and joint surgery American. 2007;89(4):780–5 https://doi.org/10.2106/jbjs.f.00222.

    Article  Google Scholar 

  2. Kinkel S, Kaefer W, Reissig W, et al. Revision total hip arthroplasty: the influence of gender and age on the perioperative complication rate. Acta chirurgiae orthopaedicae et traumatologiae Cechoslovaca. 2003;70(5):269–73.

    CAS  PubMed  Google Scholar 

  3. Meek RM, Garbuz DS, Masri BA, et al. Intraoperative fracture of the femur in revision total hip arthroplasty with a diaphyseal fitting stem. The Journal of bone and joint surgery American. 2004;86-a(3):480–5.

    Article  Google Scholar 

  4. Sheth NP, Nelson CL, Springer BD, et al. Acetabular bone loss in revision total hip arthroplasty: evaluation and management. J Am Acad Orthop Surg. 2013;21(3):128–39 https://doi.org/10.5435/jaaos-21-03-128.

    Article  Google Scholar 

  5. Flecher X, Sporer S, Paprosky W. Management of severe bone loss in acetabular revision using a trabecular metal shell. The Journal of arthroplasty. 2008;23(7):949–55 https://doi.org/10.1016/j.arth.2007.08.019.

    Article  Google Scholar 

  6. Gallart X, Fernandez-Valencia JA, Riba J, et al. Trabecular TitaniumTM cups and augments in revision total hip arthroplasty: clinical results, radiology and survival outcomes. Hip international: the journal of clinical and experimental research on hip pathology and therapy. 2016;26(5):486–91 https://doi.org/10.5301/hipint.5000378.

    Article  Google Scholar 

  7. Meneghini RM, Meyer C, Buckley CA, et al. Mechanical stability of novel highly porous metal acetabular components in revision total hip arthroplasty. The Journal of arthroplasty. 2010;25(3):337–41 https://doi.org/10.1016/j.arth.2009.03.003.

    Article  Google Scholar 

  8. Johnston RC, Brand RA, Crowninshield RD. Reconstruction of the hip. A mathematical approach to determine optimum geometric relationships. The Journal of bone and joint surgery American volume. 1979;61(5):639–52.

    Article  CAS  Google Scholar 

  9. Kosashvili Y, Backstein D, Safir O, et al. Dislocation and infection after revision total hip arthroplasty: comparison between the first and multiply revised total hip arthroplasty. The Journal of arthroplasty. 2011;26(8):1170–5 https://doi.org/10.1016/j.arth.2011.04.022.

    Article  Google Scholar 

  10. Sariali E, Boukhelifa N, Catonne Y, et al. Comparison of Three-Dimensional Planning-Assisted and Conventional Acetabular Cup Positioning in Total Hip Arthroplasty: A Randomized Controlled Trial. The Journal of bone and joint surgery American. 2016;98(2):108–16 https://doi.org/10.2106/jbjs.n.00753.

    Article  Google Scholar 

  11. Kavalerskiy GM, Murylev VY, Rukin YA, et al. Three-Dimensional Models in Planning of Revision Hip Arthroplasty with Complex Acetabular Defects. Indian journal of orthopaedics. 2018;52(6):625–30 https://doi.org/10.4103/ortho.IJOrtho_556_16.

    Article  Google Scholar 

  12. Zerr J, Chatzinoff Y, Chopra R, et al. Three-dimensional printing for preoperative planning of total hip arthroplasty revision: case report. Skeletal radiology. 2016;45(10):1431–5 https://doi.org/10.1007/s00256-016-2444-1.

    Article  Google Scholar 

  13. Paprosky WG, Perona PG, Lawrence JM. Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. The Journal of arthroplasty. 1994;9(1):33–44.

    Article  CAS  Google Scholar 

  14. Lewinnek GE, Lewis JL, Tarr R, et al. Dislocations after total hip-replacement arthroplasties. The Journal of bone and joint surgery American volume. 1978;60(2):217–20.

    Article  CAS  Google Scholar 

  15. Moskal JT, Capps SG. Acetabular component positioning in total hip arthroplasty: an evidence-based analysis. The Journal of arthroplasty. 2011;26(8):1432–7 https://doi.org/10.1016/j.arth.2010.11.011.

    Article  Google Scholar 

  16. Saleh KJ, Jaroszynski G, Woodgate I, et al. Revision total hip arthroplasty with the use of structural acetabular allograft and reconstruction ring: a case series with a 10-year average follow-up. The Journal of arthroplasty. 2000;15(8):951–8.

    Article  CAS  Google Scholar 

  17. Hipfl C, Janz V, Lochel J, et al. Cup-cage reconstruction for severe acetabular bone loss and pelvic discontinuity: Mid-term Results of a Consecutive Series of 35 Cases. The bone & joint journal. 2018;100-b(11):1442–8 https://doi.org/10.1302/0301-620x.100b11.bjj-2018-0481.r1.

    Article  CAS  Google Scholar 

  18. Abolghasemian M, Tangsataporn S, Sternheim A, et al. Combined trabecular metal acetabular shell and augment for acetabular revision with substantial bone loss: a mid-term review. The bone & joint journal. 2013;95-b(2):166–72 https://doi.org/10.1302/0301-620x.95b2.30608.

    Article  CAS  Google Scholar 

  19. Grappiolo G, Loppini M, Longo UG, et al. Trabecular Metal Augments for the Management of Paprosky Type III Defects Without Pelvic Discontinuity. The Journal of arthroplasty. 2015;30(6):1024–9 https://doi.org/10.1016/j.arth.2015.01.001.

    Article  Google Scholar 

  20. Delp SL, Wixson RL, Komattu AV, et al. How superior placement of the joint center in hip arthroplasty affects the abductor muscles. Clinical orthopaedics and related research. 1996;328:137–46 https://doi.org/10.1097/00003086-199607000-00022.

    Article  Google Scholar 

  21. Bosker BH, Verheyen CC, Horstmann WG, et al. Poor accuracy of freehand cup positioning during total hip arthroplasty. Archives of orthopaedic and trauma surgery. 2007;127(5):375–9 https://doi.org/10.1007/s00402-007-0294-y.

    Article  CAS  Google Scholar 

  22. Minoda Y, Kadowaki T, Kim M. Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clinical orthopaedics and related research. 2006;445:186–91 https://doi.org/10.1097/01.blo.0000201165.82690.f8.

    PubMed  Google Scholar 

  23. Choi HR, Anderson D, Foster S, et al. Acetabular cup positioning in revision total hip arthroplasty with Paprosky type III acetabular defects: Martell radiographic analysis. International orthopaedics. 2013;37(10):1905–10 https://doi.org/10.1007/s00264-013-2008-0.

    Article  Google Scholar 

  24. Dorr LD, Wan Z, Malik A, et al. A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. The Journal of bone and joint surgery American. 2009;91(11):2598–604 https://doi.org/10.2106/jbjs.h.01225.

    Article  Google Scholar 

  25. Sanchez-Perez C, Rodriguez-Lozano G, Rojo-Manaute J, et al. 3D surgical printing for preoperative planning of trabecular augments in acetabular fracture sequel. Injury, 2018,49 Suppl 2: p. S36-s43. https://doi.org/10.1016/j.injury.2018.07.014

    Article  Google Scholar 

  26. Kubota Y, Kaku N, Tabata T, et al. Efficacy of Computed Tomography-Based Navigation for Cup Placement in Revision Total Hip Arthroplasty. Clinics in orthopedic surgery. 2019;11(1):43–51 https://doi.org/10.4055/cios.2019.11.1.43.

    Article  Google Scholar 

  27. Nakamura N, Nishii T, Kitada M, et al. Application of computed tomography-based navigation for revision total hip arthroplasty. The Journal of arthroplasty. 2013;28(10):1806–10 https://doi.org/10.1016/j.arth.2012.11.015.

    Article  Google Scholar 

  28. Yun HH, Yoon JR, Yu JJ, et al. Navigated Acetabular Cup Fixation for Acetabular Deformity or Revision Total Hip Arthroplasty. Hip & pelvis. 2014;26(3):150–6 https://doi.org/10.5371/hp.2014.26.3.150.

    Article  Google Scholar 

  29. Chang JD, Kim IS, Prabhakar S, et al. Revision Total Hip Arthroplasty Using Imageless Navigation With the Concept of Combined Anteversion. The Journal of arthroplasty. 2017;32(5):1576–80 https://doi.org/10.1016/j.arth.2016.12.030.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

The study was supported by Major Science and Technology Project of Guangdong Province (2015B020225007), Natural Science Foundation of Guangdong Province (2018A0303130206), Guangdong Medical Science and Technology Research Fund (A2019150).

Author information

Authors and Affiliations

Authors

Contributions

QTL, XPC, and QJZ designed the study and modified and approved the final version of the manuscript. XPC and QTL wrote the first draft of the manuscript. XPC and QTL collected the data. YCM and JXL analyzed and interpreted the data. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Qiujian Zheng.

Ethics declarations

Ethics approval and consent to participate

The review board of the Guangdong Provincial People's Hospital approved the whole research plan.

Consent for publication

All authors reviewed the final version of the manuscript and approved it for submission.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Q., Chen, X., Lin, B. et al. Three-dimensional technology assisted trabecular metal cup and augments positioning in revision total hip arthroplasty with complex acetabular defects. J Orthop Surg Res 14, 431 (2019). https://doi.org/10.1186/s13018-019-1478-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-019-1478-1

Keywords