Comparison of the accelerometer-based navigation system with conventional instruments for total knee arthroplasty: a propensity score-matched analysis

Abstract

Background

The accelerometer-based navigation (ABN) system is an emerging navigation system for total knee arthroplasty (TKA). This study aimed to determine whether the ABN system could improve the accuracy of mechanical alignment, component positioning, and short-term clinical outcomes for TKA when compared to conventional instruments (CON).

Methods

A total of 204 patients were selected and divided into two groups (CON: 135, ABN: 69) after applying the inclusion and exclusion criteria. Then, 1:1 propensity score matching was performed for age, gender, body mass index, hip-knee-ankle angle (HKA), Knee Society Score (KSS), Western Ontario and McMaster Universities (WOMAC) score, and follow-up times. A total of 82 consecutive patients (82 knees) underwent total knee arthroplasty using ABN (n = 41) or CON (n = 41) were enrolled in this study. The postoperative HKA, frontal femoral component (FFC) angle, frontal tibial component (FTC) angle, lateral femoral flexion (LFF) angle, and lateral tibial component (LTC) angle were compared between the two groups to evaluate mechanical alignment and component positioning. Additional clinical parameters, including haemoglobin reduction, the KSS, and the WOMAC score, were assessed at the final follow-up (the mean follow-up period was 20.9 months in the CON group and 21.2 months in the ABN group).

Results

The ABN group had a significantly improved mean absolute deviation of HKA (P = 0.033), FFC (P = 0.004), FTC (P = 0.017), LFF (P = 0.023), and LTC (P = 0.031) compared to those of the CON group. The numbers of FFCs and LTCs within 3° were significantly different (P = 0.021, P = 0.023, respectively) between the two groups. However, no differences in the numbers of FTCs within 3° (P = 0.166) and LFF within 3° (P = 0.556) were found. The ABN group had a significantly higher KS function score (P = 0.032), and the pain and stiffness scores were significantly different (P = 0.034, P = 0.020, respectively) between the two groups. Moreover, the ABN system could reduce hidden blood loss postoperatively. However, no difference was found in the KS knee score and the total WOMAC score between the two groups.

Conclusion

This study demonstrates that ABN system improved TKA mechanical alignment and component positioning and decreased the hidden blood loss postoperatively compared to conventional instruments. However, no significant differences were found in short-term clinical outcomes between ABN and conventional instruments at the final follow-up. However, whether this system contributes to revision rates and long-term clinical outcomes requires further study.

Background

Total knee arthroplasty (TKA) is an effective treatment to reduce pain and restore normal function and alignment in cases with advanced osteoarthritis of the knee. Ideal alignment is widely viewed as a critical factor for the long-term success of TKA [1, 2]. Nevertheless, in Parratte’s study [3], a postoperative mechanical axis of neutral ± 3° did not improve the implant survival rate after a 15-year follow-up. Although the ideal range of mechanical alignment for successful TKA is controversial, most surgeons favour mechanical alignment within 3° of neutral. Berend et al. followed 3124 TKA patients for 2–14 years and found that the failure rate increased by 17.2-fold in tibias with greater than 3° of varus and by 168-fold in tibias with greater than 3° of varus and in patients with a body mass index (BMI) > 33.7 [4]. Favourable alignment has a significant impact on function [2, 5, 6], pain [7], and quality of life [8, 9]. Ritter et al. [10] documented that among 6070 knees, only 60% achieved optimal tibial alignment (90° of valgus), femoral alignment (< 8.0° of valgus), and overall alignment (2.5–7.4° of valgus) simultaneously. Patients with severe deformities, a high BMI, and acute bowing can also limit the use of conventional instruments. Consequently, new and advanced techniques have been developed and combined with conventional instruments to assist surgeons in improving the precision of the mechanical axis and component alignment. One such innovation is the Computer Assist System (CAS). In Mason’s meta-analysis [11] of 29 studies, the percentage of femoral varus/valgus alignments within 2° perpendicular to the femoral mechanical axis was 90.4% in the CAS group (versus 65.9% in the conventional group), and the percentage of tibial varus/valgus alignment within 2° perpendicular to the tibial mechanical axis was 95.2% in the CAS group (versus 79.7% in the conventional group). However, some studies [12, 13] have noted that CAS has not been widely accepted due to inevitable problems, such as large initial start-up costs, extra pin sites, a substantial learning curve, and large consoles. Kim et al. [14] found that using CAS during TKA could lead to femoral notching and pin-site fracture. Moreover, extensive bone resection, intramedullary positioning, and soft-tissue release in TKA often contribute to significant blood loss [15]; thus, avoiding extra injury is necessary when using a navigation system.

Recently, the novel navigation system accelerometer-based navigation (ABN) has been used to avoid such negative outcomes. ABN does not require a large computer console, extra pin sites, navigation arrays, and intramedullary mechanical devices. Since the surgical procedure of ABN is similar to that of a conventional instrument, it has a shorter learning curve compared to CAS. However, there exists uncertainty whether this kind of accelerometer-based navigation can improve mechanical and component alignment or not [16, 17].

As accelerometer-based navigation systems are a relatively new technique, literatures evaluating its accuracy and clinical effectiveness in TKA are sparse. The purpose of this study is to compare the postoperative outcomes, including hip-knee-ankle angle (HKA), frontal femoral component (FFC), frontal tibial component (FTC), lateral femoral flexion (LFF), lateral tibial component (LTC), blood loss volume, Knee Society Score (KSS), and Western Ontario and McMaster Universities (WOMAC) score, in patients who underwent TKA surgery with conventional instruments and accelerometer-based navigation system. We hypothesised that using accelerometer-based navigation system during TKA would significantly improve the accuracy of mechanical alignment, component positioning, and short-term (mean follow-up period was 20.9 months in the CON group and 21.2 months in the ABN group) clinical outcomes postoperatively when compared to conventional instruments in TKA.

Materials and methods

Patient selection

This study was a retrospective, case-control study of patients who underwent TKA surgery with either conventional instruments or the ABN (iAssist, Zimmer, Inc., Warsaw, IN) in The First Hospital of China Medical University. The inclusion criteria were as follows: (1) a diagnosis of primary tricompartmental osteoarthritis and (2) a history of unilateral TKA after conservative treatment failure. The exclusion criteria were as follows: (1) a knee-related operative or injury history, (2) a severe knee deformity (defined as a hip-knee-ankle (HKA) angle valgus > 30° or varus > 20°), (3) rheumatoid arthritis or another medical disease involving the knee, (4) haematological disorders, and (5) hip pathology that severely limited the range of motion. A total of 204 patients were enrolled in this study after applying the inclusion and exclusion criteria. From April 2016 to November 2017, 69 patients received a TKA using the iAssist to perform distal femoral and proximal tibial resection. During the same period, 135 patients received a TKA using conventional instruments. All procedures were performed by one experienced surgeon being familiar with the iAssist navigation system. This retrospective study was approved by the Ethics Committee of The First Hospital of China Medical University.

Propensity-matched analysis

Propensity score matching analysis are statistical methods aimed to reduce confounding attributable to measured covariates in observational studies [18]. In observational studies, there is often bias derived from significant differences between the characteristic subjects of the treatment group and the no-treatment group. We attempted to limit such bias with propensity score-matched analysis to account for other possible confounding factors. A propensity score is generally defined as a patient’s conditional probability of being assigned a treatment based on patient’s pre-treatment characteristics by logistic regression using the “MatchIt” R package. Then, the ABN and CON groups were propensity score matched in a 1:1 ratio based on age, gender, BMI, HKA angel, Knee Society Score (KSS), and Western Ontario and McMaster Universities (WOMAC) score.

Surgical technique for the ABN group

The iAssist system is a novel, personalised guide system that is designed to guide both proximal tibial and distal femoral resections in TKA. The system uses surgical instruments and positioning sensors to determine the axes in relation to anatomical landmarks. The surgical workflow is similar to the conventional method, with independent resection of the femur and tibia along their respective mechanical axes. The only difference of surgical procedure between the two groups is the tool used to guide both proximal tibial and distal femoral resections.

Starting with the femur, a small spike was inserted into the distal femur using a pod clipped to the spike. Femur registration was performed using 13 stable positions by accelerating and stopping the leg to create a star-shaped or circular pattern and obtain the mechanical axis of the femur. Next, the femoral adjustment mechanism was installed with another pod clipped to it onto the anterior side of the femoral reference. Using gold and green screws, a light-emitting diode (LED) followed the pod to adjust the flexion/extension and varus/valgus independently (Fig. 1a). Then, the distal femur was resected using the distal femur resection instrument. A validation tool can be used to check the result or adjust the angle if necessary. Finally, a 4-in-1 cutting guide aligned in rotation on Whiteside’s line was used to finish the anterior and posterior femoral cuts. An EM tibial guide with a pod clipped to it was employed to obtain the mechanical axis of the tibia. The proximal tibial guide has two spikes. The longer spike was inserted into the proximal surface of the tibia between the two spines, and the claw of the distal tibial guide was positioned on the malleoli. The claw was confirmed to be at the centre of the ankle joint prior to the next step. Next, the tibia registration was completed through 3 stable positions: the abduction, adduction, and neutral positions. The tibial adjustment mechanism was used to adjust the flexion/extension and varus/valgus through the gold and green screws, respectively (Fig. 1b). The depth of the cut was determined using a classic stylus. A validation tool can be used to check the results or adjust the angle if necessary. In this study, there is one patient required additional resection of proximal tibia following validation. The soft-tissue release was carried out in 26 patients in this group.

Fig. 1
figure 1

Intraoperative view of the iAssist instruments. a View of the femoral cutting block with the pods (white arrow) for distal femoral cut. b View of the extramedullary tibial cutting block with the pods (white arrow) for tibial cut. A LED light on pods is used as an indicator for proper cutting angle in a real-time manner

Surgical technique for the conventional group

The surgical procedure for the conventional group followed the standard TKA method. Starting with the femur, an intramedullary guide was employed to acquire the reference femoral mechanical axis. Next, an appropriate 4-in-1 cutting jig aligned in rotation on Whiteside’s line was used to finish the anterior and posterior femoral cuts. For the proximal tibia, an extramedullary (EM) cutting guide was used to acquire the reference of the tibial mechanical axis. The soft-tissue release was carried out in 25 patients in this group. The trial components were placed to examine the tension of the knee and the overall lower limb alignment. Then, the corresponding prosthesis was implanted.

The objective in both groups was to obtain femoral and tibial implants at 90° to the mechanical alignment and femoral rotation aligned with the transepicondylar axis and checked using Whiteside’s line. All surgeries in both groups were performed using the medial para-patellar approach with patellar eversion. The depth of the resectable distal femur was 9 mm, and the femoral valgus angle was set at 6°. The depth of the resectable proximal tibia was determined using a classic stylus depending on the size of the prosthesis (generally, the depth of the resectable bone outside of the proximal tibia is 9 mm). Depending on the actual condition, soft-tissue balance techniques were used to release the medial collateral ligament, lateral collateral ligament, or posterior cruciate ligament at different degrees to reach a balanced state (before releasing the soft tissue, redundant osteophytes must be eliminated). All patients underwent the same closure of wounds in a standard method and the same postoperative rehabilitation protocol. Anti-coagulants were stopped 1 week before the operation. The drainage tube was activated 4 h after surgery. Oral anti-coagulants were introduced to all patients on the second postoperative day.

Radiological evaluation

A standardised radiographic assessment was performed to confirm the absence of malrotation in the radiographs (antero-posterior view, both the proximal tibia and fibula can be seen and are not overlying significantly; lateral view, the condyles of the femoral implant are overlying one another). X-rays were taken 1 week after surgery and included a total lower limb X-ray in a weight-bearing position, an antero-posterior view, and a lateral view. The assessor measured the HKA, frontal femoral component angle (FFC), frontal tibial component angle (FTC), lateral femoral flexion angle (LFF), and lateral tibial component angle (LTC) (Fig. 2). Calculations were completed to determine the absolute deviations of the HKA, FFC, FTC, LFF, and LTC from the ideal parameters (HKA, 180°; FFC, 90°; FTC, 90°; LFF, 87°; and LTC, 83°). The objectives used in this study were as follows: [1] restoration of an HKA value of 180° ± 3°, [2] satisfactory FFC and FTC defined as 90° ± 3°, [3] satisfactory LFF defined as 87° ± 3°, and [4] satisfactory LTC defined as 83° ± 3°. The radiographs were assessed twice by one experienced orthopaedic surgeon (CZH) who was not participated in surgery and blinded to the group to which the patient belonged using the IMPAX 6 (Agfa-Gevaert N.V., Mortsel, Belgium). To test intra-observer reliability, each set of measurements was repeated three times on 30 randomly selected patients. The intra-observer reliability based on ICC ranged from 0.878 to 0.901. All abovementioned parameters were compared between the two groups. The primary outcome of this evaluation was the proportion of radiological angle within 3° for HKA, FFC, FTC, LFF, and LTC, while the secondary outcomes include the absolute deviations of the HKA, FFC, FTC, LFF, and LTC.

Fig. 2
figure 2

Antero/posterior and sagittal radiographs showing the standard graphic patterns of HKA, FFC, FTC, LFF, and LTC. HKA, hip-knee-ankle angle; FFC, frontal femoral component angle; FTC, frontal tibial component angle; LFF, lateral femoral flexion angle; LTC, lateral tibial component angle

Short-term clinical evaluation

The gender, age, and BMI of the patients were recorded. Two experienced doctors handled the preoperative and postoperative clinical evaluations and were blinded to each other’s results. The Knee Society Score (KSS) and Western Ontario and McMaster Universities (WOMAC) scores were recorded at the final follow-up. The KSS involves 100 points each for a function (KSFS) and Knee Score (KSKS). The WOMAC contains 3 subscales (pain, stiffness, and physical function) consisting of 24 questions. Scores can range from 0 to 96 and higher scores indicate more severe disease. These two clinical outcomes, KSS and WOMAC scores, were assessed at the latest follow-up (mean follow-up period was 20.9 months in the CON group and 21.2 months in the ABN group). Haemoglobin (HGB) was recorded for all patients preoperatively and at postoperative days 1 and 3. The drop in HGB between the postoperative and preoperative days was calculated and compared between the two groups.

Statistical methods

For all analyses, the independent variable was the patient group, and the dependent variables were the abovementioned parameters. A power analysis for HKA absolute deviation revealed that a sample size of 39 knees in each group was required to provide appropriate power (beta = 0.80, effect size = 0.5714286) based on previous literature [18]. To detect a minimal clinically important difference (MCID) in the KSS of 10 points from a baseline mean score of 80 with a standard deviation of 13, a sample size of at least 28 patients in each group would be required to achieve a power of 0.80 (effect size = 0.7692308) [18]. To detect an MCID of 11.5 points in WOMAC from a baseline score of 35 with a standard deviation of 10, a sample size of at least 18 patients in each group would be required to achieve a power of 0.95 (effect size = 1.15) [19]. There is currently no MCID specifically to measure HKA, FFC, FTC, LFF, LTC, and blood loss volume. These calculations were done for a one-sided test with a type I error of 0.05. Continuous data (age, BMI, HKA, FFC, FTC, LFF, LTC, blood loss volume, KSS, and WOMAC scores) are presented as the means ± standard deviations and compared using Student’s t test or the Mann-Whitney U test (non-normally distributed data). Categorical data (gender, the number of radiological angle within 3°) are presented as counts and percentages. Categorical data were compared using Fisher’s exact test. The postoperative HKA, FFC, FTC, LFF, LTC, blood loss volume, KSS, and WOMAC scores are the outcomes compared between the CON group and ABN group. All statistical analyses were performed by an independent investigator using SPSS (version 20.0; IBM SPSS Statistics, Chicago, IL, USA). Statistical significance was set at P < 0.05.

Results

Patient characteristics

A total of 204 patients were selected and divided into two groups (CON: 135, ABN: 69) after applying the inclusion and exclusion criteria. Then, the ABN and CON groups were propensity score matched in a 1:1 ratio based on age, gender, BMI, HKA angel, KSS, and WOMAC score. After propensity-matched analysis, 82 patients (82 knees) were finally included in this study who underwent TKA using conventional instruments (n = 41) or the ABN system (n = 41). Clinical results before and after propensity matching are summarised in Table 1. No significant differences were found in gender, age, BMI, or the preoperative HKA, KSS, or WOMAC score between the two groups after matching.

Table 1 Demographic characteristic of patients undergoing TKA with or without the use of ABN

Radiographic outcomes

There was a significant difference in the number of TKAs within 3° of HKA between the two groups (95.1% in the ABN group vs. 80.5% in the CON group, P = 0.043). In addition, the ABN group had significantly improved mean absolute deviations in the HKA (P = 0.033), along with the above-described trend towards increased accuracy. The numbers of FFC and LTC within 3° were significantly different (P = 0.021, P = 0.023, respectively) between the two groups. The ABN group also had significantly improved mean absolute deviations in the FFC (P = 0.004) and LTC (P = 0.031) compared to those of the CON group. There was no difference in the number of TKAs within 3° of FTC (P = 0.166) and LFF (P = 0.556), even significant differences in favour of iAsisst were found for mean absolute deviations in FTC (P = 0.017) and LFF (P = 0.023). These results suggest that a trend exists towards significant improvement with the use of the ABN system for mechanical alignment and component positioning (Table 2).

Table 2 Radiographic outcomes at final follow-up

Short-term clinical outcomes

At the final postoperative follow-up, the ABN group had a significantly higher KS function score (P = 0.032) (Table 3). However, no difference was found in the KS knee score (P = 0.767). At the final postoperative follow-up, the total WOMAC score was not significantly different between the two groups, but the pain and stiffness scores were significantly different (P = 0.034, P = 0.020, respectively). The mean reduction in HGB from preoperative to both postoperative days 1 and 3 was significantly lower in the ABN group than in the CON group.

Table 3 Clinical outcomes at final follow-up

Discussion

Although the requirements of a neutral axis and optimal implant position have been challenged [19,20,21], a suboptimal implant position and alignment remain among the main reasons for TKA failure [10, 22, 23]. The Australian Orthopaedic Association National Joint Registry reported a reduced revision rate for navigated TKAs for loosening/lysis in patients aged < 65 years [24]. A set of studies have demonstrated that CAS has been shown to have greater accuracy than conventional instruments [25]. However, the limitations of CAS include large initial start-up costs, a substantial learning curve, an unstable workflow, femoral notching, large consoles, optical tracking, and line-of-sight problems. Moreover, the extra pin site may contribute to pain, infection, and pin-site fracture [14]. These complications will adversely affect TKA surgery, which is already a challenging invasive surgery. The aim of this study was to assess the accuracy of mechanical alignment, component positioning, and short-term clinical outcomes of TKA using a novel accelerometer-based navigation system to perform the proximal tibial and distal femoral resections when compared to conventional instruments.

Our study compared the accuracy of mechanical alignment and component positioning between CON and ABN groups. The results showed that there was a significant difference in the number of TKAs within 3° of HKA, FFC, and LTC between the two groups. However, we found no significant difference in the number of TKAs within 3° of FTC and LFF. In addition, the ABN group had significantly improved mean absolute deviations in HKA, FFC, FTC, LFF, and LTC when compared to the CON group. Our results confirm our initial hypothesis that compared with the use of conventional instruments, the use of the ABN for mechanical alignment and component positioning showed a trend towards significant improvement. Two recent meta-analysis studies found inconsistent results regarding the accuracy of ABN. Shigemura et al. reported that there were significantly fewer outliers for mechanical axis and coronal femoral component angle using ABN compare to CON group; however, no significant difference was observed for coronal tibial component angle outliers [16]. Sun et al. determined in their study that ABN was significantly superior to CON in reducing tibial component alignment out of ± 3°, femoral coronal angle out of ± 3°, and overall mechanical alignment out of ± 3°; and the two groups were comparable in tibial component posterior slope out of ± 3°, femoral sagittal angle out of ± 3° [17]. These results are partly consistent with our study.

In addition to the accuracy of mechanical alignment and component positioning, we also compared short-term clinical outcomes in terms of KSS, WOMAC, and blood loss between two groups. In this study, the ABN group had a better pain and stiffness score in WOMAC at the final follow-up. Our study also found that the KS function score was significantly better in the ABN group when compared to the CON group. Research shows that patients with increased symptoms of stiffness after TKA have a worse functional outcome and a lower rate of patient satisfaction [26]. Pain and stiffness are important factors related to the early functional rehabilitation and patient satisfaction. Postoperative pain after TKA plays an important role in delayed rehabilitation [27]. ABN system can reduce intraoperative injuries (with no need for an intramedullary location and less blood loss), which is a contributor to release pain after surgery and improve patient compliance during early rehabilitation. This may be one of the reasons of the better stiffness score in ABN group. However, our study found no significant difference in KS Knee Score and total WOMAC score between the two groups. This result is partly consistent with another study which is the only one assessed clinical outcomes of iAssist at 2-year follow-up and demonstrated no difference in KSS following TKA between ABN and CON groups [28]. No previous study has reported the WOMAC score in TKA using ABN. However, a high-quality meta-analysis study concluded that there is limited evidence which indicates that CAS improves functional outcomes at 5- to 8-year follow-up as measured by WOMAC and Knee Society Score-function scores [29].

Finally, we assessed blood loss using the reduction in HGB and found that the ABN system reduced hidden blood loss postoperatively, possibly because the ABN system did not need an intramedullary location or an extra pin site. Certainly, the concept fits with the research conducted by Ikawa et al. [30]. Diamond et al. found that higher HGB at baseline was associated with rehabilitation outcomes after TKA [31]. Although TKA is an effective, approved treatment for advanced arthritis of the knee, extensive bone resection, intramedullary positioning, and soft-tissue release during surgery often contribute to significant blood loss [15]. Despite recent advances in blood conservation, some patients require a transfusion after the operation, which can increase the risk of transfusion reactions and wound complications [32]. Reza et al. documented that compared with that of the conventional group, the CAS group can reduce transfusion risk [33].

Although the navigation system has been shown to reduce the risk of mal-alignment, whether the navigation system can improve the survival rate of the prosthesis and clinical outcomes for long-term follow-up is a question at issue. Several studies have documented a greater survival rate in a navigated group than in a conventional group 10 years after surgery [34, 35]. However, in some studies, no differences were found in long-term survival and the clinical outcomes between navigated TKA and conventional TKA [14, 20, 36, 37]. These discrepant reports may raise a concern that whether improved component and overall limb alignment accuracy guarantee improved clinical and functional outcomes for long-term follow-up. However, as summarised by Jones et al., by reviewing the long-term follow-up studies, the improved alignment accuracy achieved with navigation system may eventually translate to lower revision rates over time [38]. Based on clinical experience, precise mechanical alignment and implant positioning may significantly contribute to surgical success in patients with a high BMI, acute bowing, extra-articular deformity, or complex bone deformities in the long-term. Ritter et al. reported that compared with the failure rate of TKA in patients with a BMI of 23 to 26 kg/m2, the failure rate in patients with a BMI ≥ 41 kg/m2 increased from 0.7 to 2.6% (P = 0.0046) in well-aligned knees, from 1.6 to 2.9% (P = 0.0180) in varus knees, and from 1.0 to 7.1% (P = 0.0260) in valgus knees [10]. During the preparation of this study, one study reported that ABN system is accurate in achieving neutral mechanical alignment and optimal implant position after TKA in patients with extra-articular deformity [39].

There are several advantages of accelerometer-based navigation that should be noted here. The iAssist system has 4 pods that are clipped onto the cutting guide during surgery within the surgical field to facilitate direct observation of the cutting angle by surgeons. The surgical workflow of the iAssist system in TKA follows that of conventional instruments, thereby shortening the operation time and reducing the learning curve compared with those of CAS. The iAssist system exchanges information through a secure wireless local area network (LAN) to obtain a stable workflow. Compared with conventional navigation systems, the iAssist system is smaller and is not dependent on bone landmarks for location, which makes the operation easier [40]. Identifying the centre of the femoral head in TKA is difficult. The iAssist system provides a direct tracing method for the centre of the femoral head and allows accurate resection of the distal femur. Moreover, there are two literatures identified that iAssist system is as accurate as CAS [41, 42]. Moreover, the duration of surgery was significantly longer in the CAS group [42]. Due to the ABN system’s not needing of an intramedullary location for distal femur resection, it could be an appropriate tool for patients with femoral shaft disease or femoral implants [43].

Our study has several strengths. Propensity-based scoring helps reduce the bias inherent to observational studies, which can minimise possible confounding factors (age, gender, BMI, preoperative HKA, and clinical scores) [18]. All procedures in this study were performed by a single surgeon who was highly experienced with TKA and familiar with ABN technique prior to the commencement of this study (performed 30 navigated TKAs using ABN system before the first enrolled patient in this study), which can limit the risk of learning curve-induced bias. All patients were treated with standard postoperative rehabilitation protocols. The only difference was the way to guide bone resections between two groups. Although ABN system has been clinically applied for the past few years, the data published previously are still remaining few. The results from the present study could help to further understand the application of ABN system in TKA.

There were several limitations to this study. This study was not a prospective, randomised, and controlled comparison between ABN and CON. Due to the relatively new nature of ABN, this study could not determine the ABN system greatly contributed to long-term clinical outcomes and revision rates. These patients will need to be followed up annually to assess whether there are differences in clinical outcomes and revision rates between ABN and CON groups over the long term. Like most studies of alignment regarding TKA, the present study did not evaluate the rotational alignment. The cutting error between the intraoperative cut results and the postoperative radiographic alignment was not reported in this study. Relatively few cases of ABN were included in this study, though we controlled for this using PSM. Although this study was performed in a blinded manner, surgeon bias may have played a role. Furthermore, we did not compare the differences between the ABN system and other navigation systems.

Conclusion

In summary, this study demonstrates that ABN system improved TKA mechanical alignment, component positioning, and decreased the hidden blood loss postoperatively compared to conventional instruments. However, no significant differences were found in clinical outcomes between ABN system and conventional instruments at the final follow-up (20.9 months in the CON group versus 21.2 months in the ABN group). Satisfying follow-up times are needed to confirm that the ABN system can contribute to TKA regarding the quality of life and prosthesis survival rate. ABN system should be treated reasoningly until the day that it shows definite improvements in clinical outcomes.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

References

  1. Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468:57–63.

    Article  PubMed  Google Scholar 

  2. Rebal BA, Babatunde OM, Lee JH, Geller JA, Patrick DA Jr, Macaulay W. Imageless computer navigation in total knee arthroplasty provides superior short term functional outcomes: a meta-analysis. J Arthroplast. 2014;29:938–44.

    Article  Google Scholar 

  3. Parratte S, Pagnano MW, Trousdale RT, Berry DJ. Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am. 2010;92:2143–9.

    Article  PubMed  Google Scholar 

  4. Berend ME, Ritter MA, Meding JB, et al. Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res. 2004;428:26–34.

    Article  Google Scholar 

  5. Hohler SE. Total knee arthroplasty: past successes and current improvements. AORN J. 2008;87:143–58 quiz 159–162.

    Article  PubMed  Google Scholar 

  6. Kim YH, Park JW, Kim JS, Park SD. The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthop. 2014;38:379–85.

    Article  PubMed  Google Scholar 

  7. Nicoll D, Rowley DI. Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg Br. 2010;92:1238–44.

    Article  CAS  PubMed  Google Scholar 

  8. Gothesen O, Espehaug B, Havelin LI, et al. Functional outcome and alignment in computer-assisted and conventionally operated total knee replacements: a multicentre parallel-group randomised controlled trial. Bone Joint J. 2014;96-B:609–18.

    Article  CAS  PubMed  Google Scholar 

  9. Serbest S, Tosun HB. Dislocation of a revision total knee arthroplasty: rare but serious complication. Pan Afr Med J. 2015;20:193.

    PubMed  PubMed Central  Google Scholar 

  10. Ritter MA, Davis KE, Meding JB, Pierson JL, Berend ME, Malinzak RA. The effect of alignment and BMI on failure of total knee replacement. J Bone Joint Surg Am. 2011;93:1588–96.

    Article  PubMed  Google Scholar 

  11. Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplast. 2007;22:1097–106.

    Article  Google Scholar 

  12. Barrett WP, Mason JB, Moskal JT, Dalury DF, Oliashirazi A, Fisher DA. Comparison of radiographic alignment of imageless computer-assisted surgery vs conventional instrumentation in primary total knee arthroplasty. J Arthroplasty. 2011;26:1273–84 e1271.

    Article  PubMed  Google Scholar 

  13. Mizu-uchi H, Matsuda S, Miura H, Okazaki K, Akasaki Y, Iwamoto Y. The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Joint Surg Br. 2008;90:1025–31.

    Article  CAS  PubMed  Google Scholar 

  14. Kim YH, Park JW, Kim JS. The clinical outcome of computer-navigated compared with conventional knee arthroplasty in the same patients: a prospective, randomized, double-blind. Long-Term Study J Bone Joint Surg Am. 2017;99:989–96.

    Article  PubMed  Google Scholar 

  15. Hogan CA, Golightly LK, Phong S, Dayton MR, Lyda C, Barber GR. Perioperative blood loss in total hip and knee arthroplasty: outcomes associated with intravenous tranexamic acid use in an academic medical center. SAGE Open Med. 2016;4:2050312116637024.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Shigemura T, Murata Y, Yamamoto Y, Mizuki N, Toki Y, Wada Y. No definite advantage of a portable accelerometer-based navigation system over conventional technique in total knee arthroplasty: a systematic review and meta-analysis. Orthop Traumatol Surg Res. 2019;105:619–26.

    Google Scholar 

  17. Sun H, Li S, Wang K, Wu G, Zhou J, Sun X. Efficacy of portable accelerometer-based navigation devices versus conventional guides in total knee arthroplasty: a meta-analysis. J Knee Surg. 2019.

  18. Inacio MC, Chen Y, Paxton EW, Namba RS, Kurtz SM, Cafri G. Statistics in brief: an introduction to the use of propensity scores. Clin Orthop Relat Res. 2015;473:2722–6.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Bonner TJ, Eardley WG, Patterson P, Gregg PJ. The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br. 2011;93:1217–22.

    Article  CAS  PubMed  Google Scholar 

  20. Cip J, Obwegeser F, Benesch T, Bach C, Ruckenstuhl P, Martin A. Twelve-year follow-up of navigated computer-assisted versus conventional total knee arthroplasty: a prospective randomized comparative trial. J Arthroplast. 2018;33:1404–11.

    Article  Google Scholar 

  21. Matziolis G, Adam J, Perka C. Varus malalignment has no influence on clinical outcome in midterm follow-up after total knee replacement. Arch Orthop Trauma Surg. 2010;130:1487–91.

    Article  PubMed  Google Scholar 

  22. Abdel MP, Oussedik S, Parratte S, Lustig S, Haddad FS. Coronal alignment in total knee replacement: historical review, contemporary analysis, and future direction. Bone Joint J. 2014;96-B:857–62.

    Article  CAS  PubMed  Google Scholar 

  23. Tiftikci U, Serbest S, Burulday V. Can Achilles tendon be used as a new distal landmark for coronal tibial component alignment in total knee replacement surgery? An observational MRI study. Ther Clin Risk Manage. 2017;13:81–6.

    Article  Google Scholar 

  24. Graves SE, Davidson D, Ingerson L, et al. The Australian Orthopaedic Association National Joint Replacement Registry. Med J Aust. 2004;180:S31–4.

    PubMed  Google Scholar 

  25. Keyes BJ, Markel DC, Meneghini RM. Evaluation of limb alignment, component positioning, and function in primary total knee arthroplasty using a pinless navigation technique compared with conventional methods. J Knee Surg. 2013;26:127–32.

    PubMed  Google Scholar 

  26. Clement ND, Bardgett M, Weir D, Holland J, Deehan DJ. Increased symptoms of stiffness 1 year after total knee arthroplasty are associated with a worse functional outcome and lower rate of patient satisfaction. Knee Surg Sports Traumatol Arthrosc. 2018;27:1196–203.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Barrington JW, Halaszynski TM, Sinatra RS, Expert Working Group On Anesthesia And Orthopaedics Critical Issues In Hip And Knee Replacement Arthroplasty FT. Perioperative pain management in hip and knee replacement surgery. Am J Orthopedics (Belle Mead, NJ). 2014;43:S1–s16.

    Google Scholar 

  28. Goh GS, Liow MHL, Tay DK, Lo NN, Yeo SJ, Tan MH. Accelerometer-based and computer-assisted navigation in total knee arthroplasty: a reduction in mechanical axis outliers does not lead to improvement in functional outcomes or quality of life when compared to conventional total knee arthroplasty. J Arthroplast. 2018;33:379–85.

    Article  Google Scholar 

  29. Panjwani TR, Mullaji A, Doshi K, Thakur H. Comparison of functional outcomes of computer-assisted vs conventional total knee arthroplasty: a systematic review and meta-analysis of high-quality, prospective studies. J Arthroplasty. 2019;34:586–93.

    Article  PubMed  Google Scholar 

  30. Ikawa T, Takemura S, Kim M, Takaoka K, Minoda Y, Kadoya Y. Usefulness of an accelerometer-based portable navigation system in total knee arthroplasty. Bone Joint J. 2017;99-B:1047–52.

    Article  CAS  PubMed  Google Scholar 

  31. Diamond PT, Conaway MR, Mody SH, Bhirangi K. Influence of hemoglobin levels on inpatient rehabilitation outcomes after total knee arthroplasty. J Arthroplast. 2006;21:636–41.

    Article  Google Scholar 

  32. Abdel MP, Chalmers BP, Taunton MJ, et al. Intravenous versus topical tranexamic acid in total knee arthroplasty: both effective in a randomized clinical trial of 640 patients. J Bone Joint Surg Am. 2018;100:1023–9.

    Article  PubMed  Google Scholar 

  33. Tabatabaee RM, Rasouli MR, Maltenfort MG, Fuino R, Restrepo C, Oliashirazi A. Computer-assisted total knee arthroplasty: is there a difference between image-based and imageless techniques? J Arthroplast. 2018;33:1076–81.

    Article  Google Scholar 

  34. Baumbach JA, Willburger R, Haaker R, Dittrich M, Kohler S. 10-year survival of navigated versus conventional TKAs: a retrospective study. Orthopedics. 2016;39:S72–6.

    Article  PubMed  Google Scholar 

  35. Baier C, Wolfsteiner J, Otto F, et al. Clinical, radiological and survivorship results after ten years comparing navigated and conventional total knee arthroplasty: a matched-pair analysis. Int Orthop. 2017;41:2037–44.

    Article  PubMed  Google Scholar 

  36. Kim YH, Park JW, Kim JS. Computer-navigated versus conventional total knee arthroplasty a prospective randomized trial. J Bone Joint Surg Am. 2012;94:2017–24.

    Article  PubMed  Google Scholar 

  37. Ollivier M, Parratte S, Lino L, Flecher X, Pesenti S, Argenson JN. No benefit of computer-assisted TKA: 10-year results of a prospective randomized study. Clin Orthop Relat Res. 2018;476:126–34.

    Article  PubMed  Google Scholar 

  38. Jones CW, Jerabek SA. Current role of computer navigation in total knee arthroplasty. J Arthroplast. 2018;33:1989–93.

    Article  Google Scholar 

  39. Matassi F, Cozzi Lepri A, Innocenti M, Zanna L, Civinini R, Innocenti M. Total knee arthroplasty in patients with extra-articular deformity: restoration of mechanical alignment using accelerometer-based navigation system. J Arthroplast. 2019;34:676–81.

    Article  PubMed  Google Scholar 

  40. Niehaus R, Schilter D, Fornaciari P, et al. Experience of total knee arthroplasty using a novel navigation system within the surgical field. Knee. 2017;24:518–24.

    Article  PubMed  Google Scholar 

  41. Goh GS, Liow MH, Lim WS, Tay DK, Yeo SJ, Tan MH. Accelerometer-based navigation is as accurate as optical computer navigation in restoring the joint line and mechanical axis after total knee arthroplasty: a prospective matched study. J Arthroplast. 2016;31:92–7.

    Article  Google Scholar 

  42. Desseaux A, Graf P, Dubrana F, Marino R, Clave A. Radiographic outcomes in the coronal plane with iASSIST versus optical navigation for total knee arthroplasty: a preliminary case-control study. Orthop Traumatol Surg Res. 2016;102:363–8.

    Article  CAS  PubMed  Google Scholar 

  43. Tiftikci U, Serbest S. Periprosthetic proximal medial femoral cortical destruction caused by a femoral arterial pseudoaneurysm. Clin Interv Aging. 2015;10:1967–70.

    PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Funded by the Natural Science Foundation of Liaoning Province (CN) (2015020473).

Author information

Authors and Affiliations

Authors

Contributions

XL conceived and designed the study. XG performed data analyses and wrote the manuscript. YS contributed to the manuscript preparation. ZC, TD, and QL are responsible for the data collection. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Xu Li.

Ethics declarations

Ethics approval and consent to participate

In the original article of the datasets, the trials were approved by the local institutional review boards of all participating centres and informed consent was obtained from all patients.

Consent for publication

Patients provided written informed consent.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gao, X., Sun, Y., Chen, ZH. et al. Comparison of the accelerometer-based navigation system with conventional instruments for total knee arthroplasty: a propensity score-matched analysis. J Orthop Surg Res 14, 223 (2019). https://doi.org/10.1186/s13018-019-1258-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-019-1258-y

Keywords