Ultrasonic cartilage thickness measurement is accurate, reproducible, and reliable—validation study using contrast-enhanced micro-CT

Abstract

Background

Ultrasonography is a fast and patient-friendly modality to assess cartilage thickness. However, inconsistent results regarding accuracy have been reported. Therefore, we asked what are (1) the accuracy, (2) reproducibility, and (3) reliability of ultrasonographic cartilage thickness measurement using contrast-enhanced micro-CT for validation?

Methods

A series of 50 cartilage–bone plugs were harvested from fresh bovine and porcine joints. Ultrasonic cartilage thickness was determined using an A-mode, 20-MHz hand-held ultrasonic probe with native (1580 m/s) and adjusted speed of sound (1696 m/s). All measurements were performed by two observers at two different occasions. Angle of insonation was controlled by tilting the device and recording minimal thickness. Retrieval of exact location for measurement was facilitated by aligning the circular design of both cartilage–bone plug and ultrasonic device. There was no soft tissue interference between cartilage surface and ultrasonic probe. Ground truth measurement was performed using micro-CT with iodine contrast agent and a voxel size of 16 μm. The mean cartilage thickness was 1.383 ± 0.402 mm (range, 0.588–2.460 mm).

Results

Mean accuracy was 0.074 ± 0.061 mm (0.002–0.256 mm) for native and 0.093 ± 0.098 mm (0.000–0.401 mm) for adjusted speed of sound. Bland–Altman analysis showed no systematic error. High correlation was found for native and adjusted speed of sound with contrast-enhanced micro-CT (both r = 0.973; p < 0.001). A perfect agreement for reproducibility (intraclass correlation coefficient [ICC] 0.992 and 0.994) and reliability (ICC 0.993, 95% confidence interval 0.990–0.995) was found.

Conclusions

Ultrasonic cartilage thickness measurement could be shown to be highly accurate, reliable, and reproducible. The A-mode ultrasonic cartilage thickness measurement is a fast and patient-friendly modality which can detect early joint degeneration and facilitate decision making in joint preserving surgery.

Background

Early cartilage degeneration is characterized by softening and thinning. In the advanced stage, the cartilage deteriorates with partial and full thickness defects [1]. Measurement of cartilage thickness allows early and objective evaluation of joints in an early arthritic stage. In addition, outcome following disease modification, e.g., joint-preserving surgery, can be quantified by monitoring cartilage thickness. While conventional radiography or computed tomography (CT) imaging of the joint could be shown to be relatively insensitive for early arthritic changes, magnetic resonance (MR) imaging is the today’s gold standard for cartilage evaluation [2]. However, MR imaging is a costly and time-consuming imaging modality often necessitating intraarticular contrast agent for accurate evaluation of cartilage. Ultrasonographic imaging is an inexpensive, fast, and patient-friendly alternative for cartilage thickness evaluation. It offers the possibility of real-time cartilage evaluation and could potentially be applied intraoperatively during joint surgery.

There are, however, inconsistent results regarding the accuracy of ultrasonic cartilage thickness measurement in literature [3]. While some studies have shown a good accuracy for this method of cartilage thickness measurement [4,5,6], others did not recommend it as valid [7,8,9,10]. Inaccurate results could potentially be due to unconsidered acoustic principals of ultrasonography [3], the use of an inaccurate ground truth measurement [8, 10, 11], or soft tissue interference between the measurement probe and cartilage surface.

The current study evaluated a hand-held device for A-mode ultrasonic cartilage thickness measurement. For validation of ultrasonic cartilage thickness measurement, the results were compared to the thickness measurements using contrast-enhanced micro-CT, a method with very high spatial resolution in the micrometer range. The ex vivo evaluation was performed using fresh bovine and porcine cartilage samples. The purpose of the study was to determine if using contrast-enhanced micro-CT as the validation tool would demonstrate that A-mode ultrasound is an (1) accurate (2), reproducible, and (3) reliable method of measuring cartilage thickness. In addition, the goal was to summarize and compare the results in literature about A-mode cartilage thickness measurement and to show potential errors resulting in decreased accuracy of ultrasonic measurement.

Methods

Specimens

A total of 50 cylindrical cartilage–bone plugs were harvested from bovine and porcine joints. This included six bovine hips (15 samples from femoral heads), six bovine knees (11 samples from femoral condyles, four from tibial plateaus, three from patellae), five porcine shoulders (15 samples from humeral heads), and one porcine elbow (two samples from the trochlea). All joints had macroscopically intact cartilage and a closed joint capsule until cartilage–bone plugs were harvested using a 10-mm cylindrical punch. All plugs had a minimum thickness of 5 mm subchondral bone. Ultrasonic cartilage thickness measurement was performed immediately following the harvest of the plugs. Until micro-CT measurement was performed, the cartilage–bone plugs were stored in phosphate-buffered saline solution at 4 °C. All measurements were performed within maximum 18 h from slaughter.

Ultrasonographic cartilage thickness measurement

Ultrasonic cartilage thickness was determined with an ultrasonic hand-held probe (DUB micro®, Rev. 2.14a, Taberna pro medicum, Lüneburg, Germany). Thickness was measured at the center of the cartilage plug, and the ultrasonic probe was manually aligned. Both the cartilage probe and the ultrasonic device had a circular design (Fig. 1), which facilitated alignment. The gap between the ultrasonic probe and the cartilage surface was filled with ultrasound transmission gel (Aquasonic® 100, Parker Laboratories, Fairfield NJ, USA). An average layer of transmission gel of 0.745 ± 0.018 mm (range, 0.704 to 0.792 mm) was used. The speed of sound of the ultrasonic probe was 1580 m/s. The 20-MHz echoes were plotted on a screen as a function of depth and time (A-mode sonography). Cartilage thickness was measured between the first and second reflection (Fig. 1). The band-shaped interference following the first reflection, also known as the “leading interface” [3], is a part of the cartilage and was included in the thickness measurement (Fig. 1). This “leading interface” occurs due to the considerable difference in acoustic impedance of the transmission gel and cartilage [3]. The angle of insonation has a significant influence on cartilage thickness measurement. Deviation from true orthogonal insonation will result in increased cartilage thickness measurement. Therefore, the ultrasonic probe was tilted manually during continuous measurement, and the smallest plotted cartilage thickness was recorded (Fig. 1). Cartilage thickness was measured with native speed of sound of the ultrasonic probe with 1580 m/s. An optimal speed of sound of 1696 m/s for cartilage thickness measurement has been recommended in literature [3, 9, 12]. Therefore, cartilage thickness evaluated with native speed of sound was multiplied by the factor 1.07 (1696/1580) for calculation of thickness with optimal speed of sound. The thickness measurements were performed by two observers and twice for each cartilage–bone plug by each observer. The average cartilage thickness of all 200 ultrasonic measurements ranged from 0.595 to 2.464 mm (mean 1.372 ± 0.412 mm).

Fig. 1
figure 1

a A pen-like ultrasonic transducer was used (DUB micro®, Rev. 2.14a, Taberna pro medicum, Lüneburg, Germany). b The ultrasonic signal fades while traveling through cartilage. At the cartilage–bone border, the signal is partially reflected (dashed lines). Cartilage thickness is calculated based on the time the reflected signal needs to travel through cartilage. Therefore, thickness directly relates to speed of sound in cartilage. c A-mode ultrasonic image: upper cartilage border (arrow); “leading interface” (LI), an interference pattern which occurs due to the great difference in acoustic impedance of cartilage and gel; cartilage–bone border (asterisk)

Contrast-enhanced micro-CT

Ground truth measurement of cartilage thickness was done using contrast-enhanced micro-CT (Scanco Medical μCT 40, version 6.1, Brüttisellen, Switzerland). The scans were performed with the following parameters: voxel size of 16 μm, maximum voltage of 70 kVp, and electric current of 114 μA. Each cartilage–bone sample was embedded in a radiolucent sampling tube with 15-mm diameter (Fig. 2). To restrict movements of the cartilage–bone plugs during micro-CT scan, a sponge was inserted into the tube. While the bone–cartilage interface is clearly identifiable using native CT, the cartilage–air interface of the cartilage is difficult to be distinguished. Therefore, the cartilage–bone plugs were embedded in a solution with a 1:1 ratio of iodine contrast agent (Iopamidol, 300 mg/ml, Bracco Suisse SA, Manno, Switzerland) and phosphate-buffered saline (Fig. 2). Scan volume was determined on the scout view with a constant 15-mm diameter and individual height ranging from 5.9 to 15.5 mm. Cartilage thickness was measured using a DICOM viewer (Osirix, version 5.8, Geneva, Switzerland). Multiplanar reconstruction was used to obtain a cross-section perpendicular to the cartilage surface, and the cartilage thickness was measured at the center of the cartilage–bone plug (Fig. 2). The mean cartilage thickness measured using micro-CT was 1.383 ± 0.402 mm (range, 0.588–2.460 mm; see Additional file 1 named “Data validation ultrasonic cartilage thickness measurement.xls”).

Fig. 2
figure 2

For validation of ultrasonic cartilage thickness measurement we used the micro-CT with iodine contrast agent (a) for comparison. For micro-CT measurements the cartilage-bone samples were harvested with a punch biopsy and embedded in a radiolucent tube. While the lower limit of the cartilage (c) adjacent to the bone (b) was clearly visible in the micro-CT, detection of the upper limit of the cartilage was enhanced by iodine contrast agent (a). This allowed exact determination of cartilage thickness

Validation

Accuracy of ultrasonic cartilage thickness measurement was evaluated by calculating the difference between the ultrasonic and the micro-CT measurements of cartilage thickness. In addition, the error was calculated as the quotient of the difference of cartilage thickness using the two modalities and the thickness evaluated using micro-CT. The linear relationship between the ultrasonic and micro-CT techniques for cartilage thickness was calculated. All measurements were made for both native and adjusted speed of sound in cartilage.

Reproducibility was evaluated by comparing the ultrasonic measurements performed at two occasions. Reliability was evaluated by comparing the measurements of the two observers. Reproducibility and reliability measurements were performed with native speed of sound in cartilage.

Statistical analysis

Accuracy was calculated as the absolute difference of the ultrasonic and micro-CT-based cartilage thickness measurements. Differences in accuracy or error between ultrasonic measurement with native and adjusted speed of sound were detected using the independent t test. To detect a systematical error of the ultrasonic measurements, the Bland–Altman analysis [13] was calculated by plotting the difference between the two measurement techniques against their average. Correlation between the two measurements techniques was evaluated with the Pearson correlation coefficient. The Pearson correlation coefficient was graded as r < 0.2 for very weak, 0.20–0.39 for weak, 0.40–0.59 for moderate, 0.60–0.79 for strong, and ≥ 0.8 for very strong correlation [14]. The ICC was used for calculation of reproducibility and reliability and was graded as ICC < 0.20 for slight agreement, 0.21–0.40 for fair, 0.41–0.60 for moderate, 0.61–0.80 for substantial, and > 0.80 for almost perfect agreement [15]. The level of significance was set at 0.05.

Results

For the ultrasonic measurement with native speed of sound, the mean accuracy was 0.074 ± 0.061 mm (0.002–0.256 mm) with a corresponding mean error of 6% ± 5% (0–31%). For adjusted speed of sound, the mean accuracy was 0.093 ± 0.098 mm (0.000–0.401 mm) with a corresponding mean error of 7% ± 8% (0–40%). No difference existed for accuracy or error between the measurements using native or adjusted speed of sound (p = 0.237 and p = 0.289, respectively). The Bland–Altman analysis showed that the mean of the measurement pairs was spread evenly and randomly with no evidence for a systematic error for both the ultrasonic measurements with native or adjusted speed of sound (Fig. 3). A very strong linear correlation was found between the contrast-enhanced micro-CT measurement and the ultrasonic measurement using native or adjusted speed of sound (p < 0.001; r = 0.973 for both; Fig. 4).

Fig. 3
figure 3

The Bland–Altman analysis [13] was performed by plotting the difference between the two measurement techniques against their average. The analysis showed that the means of the measurement pairs were spread evenly and randomly, and therefore, no systematic error existed for a native (1580 m/s) and b adjusted speed of sound (1696 m/s) in cartilage. CT = computed tomography, US = ultrasonography, CI = confidence interval

Fig. 4
figure 4

Correlation of micro-CT measurement of cartilage thickness with ultrasonic measurement a with native (1580 m/s) or b adjusted speed of sound (1696 m/s). A strong correlation between ultrasonic and micro-CT measurement was found for both native and adjusted speed of sound (both p < 0.001 and r = 0.973). CT = computed tomography, US = ultrasonograph

An almost perfect agreement for both reproducibility (ICC of 0.992 and 0.994) and reliability (ICC 0.993, 95% confidence interval of 0.990–0.995) was found (Table 1).

Table 1 Results of reproducibility and reliability

Discussion

Ultrasonic measurement of cartilage thickness is an inexpensive, radiation-free, and patient-friendly alternative to MRI. It offers the possibility of real-time imaging which could also be applied intraoperatively, e.g., for topographical cartilage thickness assessment in joint-preserving surgery. Inconsistent results regarding accuracy for ultrasonic thickness measurement have been reported (Table 2). Several factors need to be controlled to minimize error for validation of ultrasonic cartilage thickness measurement including the angle of insonation, retrieval of location of measurement, speed of sound in cartilage, and handling of cartilage samples. In addition, ground truth measurement needs to be precise and error-free. We validated a pen-like ultrasonic device for cartilage thickness measurement. Contrast-enhanced micro-CT measurement of cartilage thickness was used as ground truth measurement for validation with an iodine contrast agent and multiplanar reconstruction. We could show that ultrasonic assessment of cartilage thickness is very accurate (mean accuracy of 0.074 mm [6%]) and has no systematical error (Fig. 3). We found a very strong correlation with micro-CT measurement of cartilage thickness (Fig. 4), and we could show an almost perfect agreement for both reproducibility and reliability (Table 1). No improvement was found for the adjusted speed of sound in cartilage (Figs. 3 and 4).

Table 2 Selected literature on accuracy of ultrasonic cartilage thickness measurement

The high resolution of ultrasonography makes it an optimal imaging modality for cartilage. However, several basic acoustic principals of ultrasonography must be taken into account for correct cartilage thickness measurement [3]. First, the borders of cartilage have to be detected correctly. At the upper cartilage border, a typical interference pattern occurs due to the great difference in acoustic impedance of cartilage and the gel (Fig. 1). This interference pattern, also called the “leading interface,” is part of the cartilage and must be included for accurate cartilage thickness measurement [3]. Second, the angle of insonation needs to be controlled. The true thickness is measured with cartilage insonated orthogonally. With a 10° and 20° error, the measured cartilage thickness is increased by 1.5% and 6.4%, respectively. The angle of insonation was controlled by manually tilting the probe and recording the thinnest cartilage thickness. Third, for evaluation of accuracy, it is mandatory to find the same location of thickness measurement with both measurement techniques (ultrasonography and micro-CT). A circular design of the cartilage samples was chosen, and the center was defined as the location of measurement. The pen-like ultrasonic probe also had a circular design which simplified manual alignment with the cartilage probe. For the micro-CT measurements, the center was defined with a digital ruler after multiplanar reconstruction to achieve orthogonal slices. Fourth, cartilage thickness directly relates on the speed of sound (Fig. 1). Most studies recommend an increased speed of sound for cartilage evaluation (average of 1696 m/s) and reported an underestimated thickness using native speed of sound [3, 9, 12]. We calculated cartilage thickness using both native and adjusted speed of sound. However, some uncertainty exists in literature regarding the correct speed of sound in cartilage, and a wide range of optimal speed from 1419 to 2428 m/s has been reported [9, 12, 16]. These discrepancies in speed were related to the different anatomical sites, cartilage degeneration, and the inhomogeneous structural components of cartilage [9, 12, 16].

Besides the acoustic principals of ultrasonography, the handling of the cartilage samples can affect accuracy. We performed both ultrasonic and micro-CT measurements within a maximum of 18 h from slaughter, and the joints were not opened before measurement. No fixation of cartilage samples was used since fixation can affect cartilage thickness by swelling or shrinking. The ground truth measurement should be precise and error-free. We used contrast-enhanced micro-CT measurement as the ground truth for validation. The micro-CT offers a very high spatial resolution up to 16 μm. The iodine contrast agent enhanced the detection of the cartilage border. The multiplanar reconstruction allowed reconstruction of orthogonal slices to measure true cartilage thickness. All these factors were controlled as good as possible to reduce potential sources of error. All cartilage samples showed macroscopically intact cartilage, and therefore, no statement can be made about the accuracy of ultrasonographic thickness measurement of cartilage with degenerative changes. In vivo measurements of cartilage thickness usually include soft tissue between the ultrasonic probe and the cartilage surface potentially resulting in increased error of cartilage thickness measurement. The setup in the current study did not include the validation with soft tissue interference. Therefore, the results of the current study do not allow to draw a conclusion on transcutaneous cartilage thickness measurement.

A high mean accuracy of ultrasonic cartilage thickness measurement of 0.074 mm (6% mean error) and 0.093 mm (7% mean error) was revealed in the current study for native and adjusted speed of sound, respectively. In addition, an almost perfect agreement between the ultrasonic and micro-CT measurements with a correlation coefficient of r = 0.973 was found (Fig. 4). In literature, two different methods of ground truth measurement were used including imaging (MRI or CT) or visual measurement (microscope, needle probe). For both methods of ground truth measurement, different results for accuracy have been found previously (Table 2). Some studies using imaging [6, 17, 18] or visual measurements [4, 5, 19, 20] for ground truth measurement in knee joints found an almost perfect correlation (correlation coefficient ranging from 0.82 to 0.98). In contrast, some in vivo MRI-based studies did not find satisfactory agreement with ultrasonic measurement (correlation coefficient ranging from 0.38 to 0.71) [8, 10, 11]. This might be due to inadequate spatial resolution with a slice thickness ranging from 1.5 to 4.5 mm [8, 10], no or an insufficient guidance to retrieve the exact same location of cartilage thickness measurement [8, 10, 11], and no control of the angle of insonation [8, 10]. The moderate correlation coefficient of r = 0.509 in another study might be due to the use of coarse ground truth measurement with calibrated photos of cross-sections of metacarpal cartilage [7]. The mean difference between ultrasonic cartilage thickness measurement and ground truth measurement published in literature ranged from 0.01 to 0.33 mm (Table 2). In the studies reporting a mean difference exceeding 0.1 mm, the inferior result of accuracy might be due to the use of a simple ruler as ground truth [21, 22], historic ultrasonic techniques [21, 22], a MRI slice thickness of 1.56 mm [18], or transcutaneous measurements with soft-tissue interference resulting in increased error of measurements [11, 22].

We found an almost perfect agreement for reproducibility and reliability with a mean ICC exceeding 0.99 (Table 1). In literature, the mean ICC for reproducibility and reliability ranged from 0.70 to 0.99 and 0.62 to 0.99, respectively [7, 10, 23, 24]. Direct insonation of the cartilage without interfering soft tissue might have decreased potential sources of error in the current study. In addition, the circular design of both the cartilage–bone plugs and the ultrasonic probe simplified retrieval of the same location for cartilage thickness measurement. The angle of insonation was controlled by manually tilting the probe and recording the thinnest cartilage thickness. These factors might also have improved the reproducibility and reliability in the current study. In literature, results for reproducibility and reliability of ultrasonic thickness measurement were based on transcutaneous measurements of cartilage in the knee and metacarpal joints with soft-tissue interference [7, 10, 23, 24]. However, this may hinder exact retrieval of the same location of cartilage thickness measurement and complicate orthogonal insonation of the cartilage.

Despite the efforts to control factors negatively affecting accuracy, several possible sources for errors in accuracy exist. First, the angle of insonation and retrieval of location of measurement were controlled manually only. Second, a maximum of 18 h existed between harvesting of cartilage sample and micro-CT measurement. During this time, the cartilage was stored in phosphate-buffered saline solution at 4 °C. This could potentially have resulted in dehydration or swelling of the cartilage affecting thickness measurement. However, since the samples were only stored for a few hours in this isotonic solution, this should not have jeopardized our results significantly. Third, adjustment of speed of sound did not improve accuracy. In contrast, with native speed of sound (1580 m/s), a slightly decreased mean error of 6% was found compared to 7% using adjusted speed of 1696 m/s. This might be due to the use of an average layer off transmission gel of 0.745 mm (54% of total thickness) mainly consisting of water (Fig. 1). Water has a lower speed of sound of 1480 m/s [3]. Thus, the use of transmission gel may have decreased the average speed of sound and affecting accuracy measurement.

Conclusion

We could show that cartilage thickness can be assessed very accurately, reproducibly, and reliably using A-mode ultrasonography. We tried to control factors adversely affecting measurement of accuracy. This included the use of fresh cartilage samples, ultrasonic and micro-CT measurements within hours of slaughter, control of the angle of insonation, and retrieval of exact location of cartilage thickness measurement. In addition, we used contrast-enhanced micro-CT measurements with multiplanar reconstruction for the true cartilage thickness measurement. No difference in accuracy was found for adjustment of higher speed of sound in cartilage. We compared the literature on ultrasonic cartilage thickness measurement (Table 2) and compared potential factors resulting in decreased accuracy for the reported measurements. Ultrasonic measurement offers the advantage of a fast, patient-friendly, and relatively inexpensive cartilage thickness assessment. It can also be applied intraoperatively for topographical cartilage thickness evaluation. Due to the lack of radiation exposure, this method seems suitable for consecutive monitoring of cartilage thickness following disease modification, e.g., joint-preserving surgery.

Abbreviations

CT:

Computer tomography

DICOM:

Digital imaging and communications in medicine

MR:

Magnetic resonance

References

  1. Outerbridge RE. The etiology of chondromalacia patellae. 1961. Clin Orthop Relat Res. 2001;389:5–8.

  2. Hayashi D, Guermazi A, Crema MD, Roemer FW. Imaging in osteoarthritis: what have we learned and where are we going. Minerva Med. 2011;102:15–32.

    CAS  PubMed  Google Scholar 

  3. Torp-Pedersen S, Bartels EM, Wilhjelm J, Bliddal H. Articular cartilage thickness measured with US is not as easy as it appears: a systematic review of measurement techniques and image interpretation. Ultraschall Med. 2011;32:54–61.

    Article  CAS  Google Scholar 

  4. Jurvelin JS, Rasanen T, Kolmonen P, Lyyra T. Comparison of optical, needle probe and ultrasonic techniques for the measurement of articular cartilage thickness. J Biomech. 1995;28:231–5.

    Article  CAS  Google Scholar 

  5. Ohashi S, Ohnishi I, Matsumoto T, et al. Evaluation of the accuracy of articular cartilage thickness measurement by B-mode ultrasonography with conventional imaging and real-time spatial compound ultrasonography imaging. Ultrasound Med Biol. 2012;38:324–34.

    Article  Google Scholar 

  6. Ohashi S, Ohnishi I, Matsumoto T, et al. Measurement of articular cartilage thickness using a three-dimensional image reconstructed from B-mode ultrasonography mechanical scans feasibility study by comparison with MRI-derived data. Ultrasound Med Biol. 2012;38:402–11.

    Article  Google Scholar 

  7. Mandl P, Supp G, Baksa G, et al. Relationship between radiographic joint space narrowing, sonographic cartilage thickness and anatomy in rheumatoid arthritis and control joints. Ann Rheum Dis. 2015;74:2022–7.

    Article  CAS  Google Scholar 

  8. Tarhan S, Unlu Z. Magnetic resonance imaging and ultrasonographic evaluation of the patients with knee osteoarthritis: a comparative study. Clin Rheumatol. 2003;22:181–8.

    Article  CAS  Google Scholar 

  9. Yao JQ, Seedhom BB. Ultrasonic measurement of the thickness of human articular cartilage in situ. Rheumatology (Oxford). 1999;38:1269–71.

    Article  CAS  Google Scholar 

  10. Yoon CH, Kim HS, Ju JH, Jee WH, Park SH, Kim HY. Validity of the sonographic longitudinal sagittal image for assessment of the cartilage thickness in the knee osteoarthritis. Clin Rheumatol. 2008;27:1507–16.

    Article  Google Scholar 

  11. Pradsgaard DØ, Fiirgaard B, Spannow AH, Heuck C, Herlin T. Cartilage thickness of the knee joint in juvenile idiopathic arthritis: comparative assessment by ultrasonography and magnetic resonance imaging. J Rheumatol. 2015;42:534–40.

    Article  Google Scholar 

  12. Patil SG, Zheng YP, Wu JY, Shi J. Measurement of depth-dependence and anisotropy of ultrasound speed of bovine articular cartilage in vitro. Ultrasound Med Biol. 2004;30:953–63.

    Article  CAS  Google Scholar 

  13. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.

    Article  CAS  Google Scholar 

  14. Evans JD. Straightforward statistics for the behavioral sciences. Pacific Grove: Brooks/Cole; 1996.

  15. Montgomery AA, Graham A, Evans PH, Fahey T. Inter-rater agreement in the scoring of abstracts submitted to a primary care research conference. BMC Health Serv Res. 2002;2:8–11.

    Article  Google Scholar 

  16. Patil SG, Zheng YP, Chen X. Site dependence of thickness and speed of sound in articular cartilage of bovine patella. Ultrasound Med Biol. 2010;36:1345–52.

    Article  CAS  Google Scholar 

  17. Aula AS, Töyräs J, Tiitu V, Jurvelin JS. Simultaneous ultrasound measurement of articular cartilage and subchondral bone. Osteoarthr Cartil. 2010;18:1570–6.

    Article  CAS  Google Scholar 

  18. Ostergaard M, Stoltenberg M, Gideon P, et al. Effect of intraarticular osmic acid on synovial membrane volume and inflammation, determined by magnetic resonance imaging. Scand J Rheumatol. 1995;24:5–12.

    Article  CAS  Google Scholar 

  19. Chérin E, Saïed A, Pellaumail B, et al. Assessment of rat articular cartilage maturation using 50-MHz quantitative ultrasonography. Osteoarthr Cartil. 2001;9:178–86.

    Article  Google Scholar 

  20. Töyräs J, Nieminen HJ, Laasanen MS, et al. Ultrasonic characterization of articular cartilage. Biorheology. 2002;39:161–9.

    PubMed  Google Scholar 

  21. Aisen AM, McCune WJ, MacGuire A, et al. Sonographic evaluation of the cartilage of the knee. Radiology. 1984;153:781–4.

    Article  CAS  Google Scholar 

  22. Martino F, Ettorre GC, Angelelli G, et al. Validity of echographic evaluation of cartilage in gonarthrosis. Preliminary report. Clin Rheumatol. 1993;12:178–83.

    Article  CAS  Google Scholar 

  23. Bevers K, Zweers MC, van den Ende CH, et al. Ultrasonographic analysis in knee osteoarthritis: evaluation of inter-observer reliability. Clin Exp Rheumatol. 2012;30:673–8.

    PubMed  Google Scholar 

  24. Naredo E, Acebes C, Moller I, et al. Ultrasound validity in the measurement of knee cartilage thickness. Ann Rheum Dis. 2009;68:1322–7.

    Article  CAS  Google Scholar 

  25. Modest VE, Murphy MC, Mann RW. Optical verification of a technique for in situ ultrasonic measurement of articular cartilage thickness. J Biomech. 1989;22:171–6.

    Article  CAS  Google Scholar 

  26. Pellaumail B, Watrin A, Loeuille D, et al. Effect of articular cartilage proteoglycan depletion on high frequency ultrasound backscatter. Osteoarthr Cartil. 2002;10:535–41.

    Article  CAS  Google Scholar 

  27. Mathiesen O, Konradsen L, Torp-Pedersen S, Jørgensen U. Ultrasonography and articular cartilage defects in the knee: an in vitro evaluation of the accuracy of cartilage thickness and defect size assessment. Knee Surg Sports Traumatol Arthrosc. 2004;12:440–3.

    Article  Google Scholar 

  28. Spannow AH, Stenboeg E, Pfeiffer-Jensen M, et al. Ultrasound and MRI measurements of joint cartilage in healthy children: a validation study. Ultraschall Med. 2011;32(Suppl 1):S110.

    PubMed  Google Scholar 

Download references

Acknowledgements

We thank Prof. Wilhelm Hofstetter and Mr. Mark Sigrist (Group of Bone Biology and Orthopaedic Research, Department Clinical Research, University of Bern, Bern, Switzerland) for their help and expertise in harvesting and storing the cartilage-bone plugs. In addition, we thank them for giving us the possibility to use the micro-CT machine and for their assistance with the micro-CT measurements.

Funding

One or more of the authors have received funding from (SDS) the Deutsche Arthrose-Hilfe e.V. (Project number: P186-A163-Siebenrock- EP2-step1-hüfte-op-II-66 k-2008-10) and (MT) from the Swiss National Science Foundation (SNSF) (Project number: PP00P3_144856). Both funding sources had no role in study design, data collection, analysis, interpretation, writing or submission of the manuscript.

Availability of data and materials

All data generated or analyzed during this study are included in this published article and its supplementary information files.

Author information

Authors and Affiliations

Authors

Contributions

SDS contributed to the concept and design, analysis and interpretation of data, and manuscript editing. MSH contributed to the data collection, initial draft, and creation of figures. CAZ contributed to the data collection, creation of figures, and manuscript editing. PCH contributed to the data collection, writing of the initial draft, and manuscript editing. FMK contributed to the concept and design, data collection, manuscript editing, and critical review. MT contributed to the concept and design, manuscript editing, and critical review, and is responsible for ensuring integrity of work. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Simon Damian Steppacher.

Ethics declarations

Ethics approval and consent to participate

Not applicable. The study does not report on human or live animal data. The specimens were obtained from the slaughterhouse. Therefore, no ethical approval was needed.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional file

Additional file 1:

The raw data of the current study for the ultrasonic and CT measurements of cartilage thickness were included as an additional file in the file. Data validation ultrasonic cartilage thickness measurement. (XLSX 10 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions