Skip to main content
  • Technical note
  • Open access
  • Published:

“90/90” Plating of proximal humerus fracture—a technical note

Abstract

Introduction

While locking plates have markedly improved fixation of proximal humerus fractures, a cohort of fractures remains difficult to treat. This cohort has been identified as fractures with marked medial comminution and varus deformity. Loss of reduction and fixation failure are the most frequently reported complications for this cohort. We report the use of an orthogonal 1/3 tubular plate to augment the proximal humerus locking plate.

Methods

The subject underwent osteosynthesis for a four-part proximal humerus fracture with medial comminution. Fixation was performed within 24 h of injury. Standard deltopectoral approach exposed the fracture. Sutures were sited to control the tuberosities and cuff. Initial reduction was held with a K-wire and augmented with a three-hole 1/3 tubular plate. Proximal humerus locking plate was sited in standard fashion including locked medial support screws. Reduction was confirmed both clinically and with intra-operative radiography.

Results

The technique provided satisfactory results. At 6 months, the fracture had fully united with no loss of reduction. At 1 year, the patient had excellent range of motion.

Conclusion

The use of a 1/3 tubular plate to augment fixation of proximal humerus fractures with medial comminution may provide a simple, reproducible, and cost-effective method to decrease loss of reduction and subsequent malunion.

Background

Locked plating systems allow surgeons greater control of both the fracture fragments and the rotator cuff for proximal humerus fractures. However, certain fracture configurations remain challenging, particularly with respect to maintaining reduction and avoiding malunion [1, 2]. While specifically designed locking plates provide excellent biomechanical control in the coronal plane, they have decreased ability to resist varus and extension forces [3]. This vulnerability is exacerbated intra-operatively, as extension deformities are more difficult to recognize [1, 2]. We present a novel use of an anterior 1/3 tubular plate paced orthogonally to a locking plate, with a 1-year radiological and functional follow-up.

Case and surgical technique

A 60-year-old female required osteosynthesis of a proximal humerus fracture following a fall. Figure 1 displays plain x-rays and selected computed tomography slices at presentation. This is an anteromedial impaction fracture as classified by Furoria et al. This fracture configuration is associated comminution of the medial calcar with varus, flexion, and anteversion of the head as seen in Fig. 1 [4].

Fig. 1
figure 1

Imaging on presentation

The patient was positioned beach chair with the affected arm controlled using a pneumatic limb positioner. Access was obtained via deltopectoral approach with biceps tenotomy (as is standard practice for the senior author). Sutures were placed in the tuberosities for both cuff control and fracture fragment manipulation. The varus impacted, apex-anterior extension deformity was reduced manually. This reduction was maintained using a K-wire and confirmed using an image intensifier. A three-hole 1/3 tubular plate was sited at the apex of correct the extension deformity. This was fixed with two uni-cortical non-locking screws. During fixation of the 1/3 tubular plate, the drill bit impacted the in situ K-wire and broke. The fragment was buried in the bone and considered unlikely to have a negative clinical impact. It was therefore left in situ.

Control of the fracture in the sagittal plane facilitated optimal positioning of a 3-hole proximal humerus locking plate without loss of reduction. A non-locking metaphyseal screw in the oblong hole allowed minor adjustments to final plate positioning prior to the insertion of seven proximal locking screws and two more (locking) metaphyseal screws. The tuberosities/cuff was reduced using the previously sited polyethylene sutures via the suture holes in the plate. These sutures were passed through the suture holes prior to locking screw insertion and tied after final plate positioning. Intra-operative x-rays are provided in Fig. 2. The tenotomized biceps tendon was sutured to the pectoral insertion completing a tenodesis, which is standard practice for the senior author.

Fig. 2
figure 2

Intra-operative radiographs

Rehabilitation, follow-up, and outcome

Following review by physiotherapy, the patient was discharged day 1 post-op. Range of motion was permitted immediately at the hand/wrist/elbow. Gentle active/active-assisted range of motion of the shoulder was commenced day 1 post-operatively. Subsequent reviews in the outpatients were at 2 weeks, 6 weeks, 2 months, 6 months and 1 year. The fracture was fully united at 6 months. X-rays at 1 year are provided in Fig. 3.

Fig. 3
figure 3

Radiographs at 1 year post-operatively

Active range of motion was assessed independently by the senior and second author at 1 year. Results for each assessor along with mean values are provided in Table 1. No cuff weakness was noted by either assessor.

Table 1 Active range of motion of shoulder at 1 year post-operatively+

Discussion

Current fixation options and potential sources of improvement

Locking plates have become the standard of care in osteosythesis of proximal humerus fractures; however, high rates of complications and re-operation persist [5, 6]. A 2011 systematic review identified collapse into varus malunion as the most common complication following osteosythesis [5]. Reduction loss is a leading cause of revision surgery and ultimately results in poor outcomes [7].

Numerous variations of fixation methods have been described in and attempt to control these fractures, depending on the fracture configuration. These have included fixation with K-wires, use of bone cement, humeral nails, medialization with impaction of humeral shaft, use of allografts, and medial support screws [8,9,10,11,12,13].

Continued technical improvements and identification of risk factors for failure have been proposed as the two key elements for improved outcomes in varus impacted cohort [5]. Attempts to improve locking plate fixation in these patients have used three main strategies: (1) medial support screws, (2) cement augmentation, and (3) bone grafts [6]. While there is a paucity of biomechanical data to support the use of medial support screws, clinical studies suggest improved rates of reduction loss [14]. Calcium cement augmentation has been proposed to improve biomechanical properties and successful clinical results have been reported [9, 15]. Despite this however, it has failed to become established practice. Finally, the addition of intra-medullary fibular allograft has been shown to increase biomechanical robustness, but superior clinical results have not been demonstrated [6].

To date, low bone mineral density, age, and initial displacement have been associated with reduction loss. With specific respect to more technical aspects of fixation, medial comminution and insufficient medial support have also been associated with failure [7, 16]. Without medial support, i.e., bony contact, the stiffness of the construct is dramatically decreased and up to 43% of these fractures can be expected to subside into varus [17]. It is postulated this is because locking plates for proximal humerus fractures have decreased ability to resist varus and extension deforming forces [3].

The importance of the sagittal plane

While the importance of coronal malalignment has long been acknowledged by the literature, it is only in more recent years that sagittal alignment has been incorporated into the decision-making process and thus featured in classification systems [18]. It has been suggested that failure to control sagittal malalignment may affect periosteal blood supply and thus fracture healing [18]. This is consistent with the observation that varus impacted fractures with extension deformity are associated with progressive displacement, particularly when managed non-operatively [19]. Robinson et al. recognized these fractures to be technically challenging and associated with high rates of fixation failure with the use of a single plate [20]. Robinson et al. went on to describe the use of sculpted femoral head allograft to maintain bony continuity and prevent apex anterior deformity.

Role of the “90/90” 1/3 tubular technique

In this technical note, we suggest the use of a single three-hole 1/3 tubular plate to control apex anterior deformity. This avoids both the increased costs associated with bone cement and allograft. Furthermore, it negates the potential risks associated with using allograft material. The placement of the 1/3 tubular plate at 90° has the further advantage of the fixation construct now directly resisting the two most important deforming forces—varus (locking plate) and extension (1/3 tubular plate). Thus, the “90/90” technique’s main role will be in the fractures with apex anterior sagittal deformities. One potential disadvantage of this method is restricting internal rotation. In this case the 1/3 tubular plate was placed over the tendinous portion of subscaularis. However, the results in this patient are comparable with the limited reports of internal rotation outcomes post-osteosythesis available in the literature [21, 22].

Rationale for the proposed technique

By examining a cohort of 67 consecutive patients, Krappinger et al. identified age, local bone mineral density, initial displacement/angulation, and successful reduction as key factors for successful fracture healing [16]. However, half of these patients were managed by percutaneous fixation rather than open reduction with internal fixation. In 2015, Jung et al. reported bone density, varus displacement, medial comminution, and insufficient medial support to be independent risk factors for loss of reduction [7]. In our case, the use of the 1/3 tubular plate allowed better control of the bony fragments in the context of decreased bone mineral density as well as anatomical reduction. This is consistent with the biomechanical study performed by Schliemann et al. whereby anteriorly directed head screws augmented with bone cement were demonstrated to significantly decrease movement at the bone-implant interface [23].

The study by Schliemann et al. was the first to recognize the potential importance of the rigidity of the link between the plate and the humeral head [23]. The concept of using two orthogonal plates to create a more rigid construct is well recognized in elbows; however, it has not been examined in proximal humerus fractures [24]. Previous authors have successfully used two orthogonal 1/3 tubular plates prior to the introduction of locking plates, but this technique was overshadowed by the success of the proximal humerus locking plate [25]. Given a cohort with a high failure rate with locking plates has been identified, the authors propose this option should be re-examined but with a difference—partnering locking plates with an orthogonal 1/3 tubular plate. While there are previous biomechanical studies comparing locking plates with 1/3 tubular plates, we could not find any biomechanical or clinical studies examining their use in combination [26].

Other described methods of augmenting locking plate fixation

Neviaser et al. described using an endosteal implant to aid fixation [27]. In four cases, a 1/3 tubular plate was used instead of fibular graft. This plate was placed medially with the intramedullary canal and locking screws were passed through both the locking plate and the 1/3 tubular plate. Thus, this technique differs in two ways to what we present. Firstly, the implant is intramedullary, and secondly the plates are in line rather than orthogonal [27]. While elegant, potential disadvantages such as increased difficult performing arthroplasty have been identified by other authors [28]. One potential advantage of the technique we describe is similar augmentation of the locking plate with less challenging access to both plates in the event of a revision.

Limitations of the 90/90 technique

Potential limitations of the 90/90 technique include compromising range of motion (secondary to metal impingement on cuff) and threatening the blood supply to the humeral head, specifically the anterior circumflex artery [29]. However, these complications were not noted with the previous two plate constructs using 1/3 tubular plates [25]. With respect to range of motion, perhaps counterintuitively, it is external rotation which has recovered most poorly.

Little is known about external rotation outcomes post fixation, an observation confirmed by the 2015 Cochrane on operative intervention for proximal humerus fractures. However, the authors noted no clinically significant differences with for external rotation with any forms of intervention; although, it is more widely reported for arthroplasty in fractures. One possible cause of restricted internal rotation may be tethering of subscapularis beneath the plate. However, given the 1/3 tubular plate is sited at the tendinous insertion, this is unlikely. The lack of vascular consequence is likely explained by the fact that the anterior circumflex artery is disrupted in approximately 80% of fractures which is markedly higher than clinical rates of osteonecrosis [28, 30]. Finally, it may be difficult to site the anterior plate without performing a tenotomy of the long head of the biceps; however, in the fracture cohort, the cosmetic impact of this is unlikely to affect patient satisfaction.

Conclusion

This technical note outlines a potentially cost-effective method of maintaining reduction in a challenging cohort of proximal humeral fracture patients (apex anterior deformity). No additional surgical exposure is required, and the only additional equipment needed is the widely available, cost-effective 1/3 tubular plate. The technique is therefore easily reproducible. By increasing rigidity at the fracture site, the 90/90 plate technique potentially mitigates a proposed source of failure.

References

  1. Beredjiklian PK, Iannotti JP, Norris TR, Williams GR. Operative treatment of Malunion of a fracture of the proximal aspect of the humerus*. J Bone Joint Surg Am. 1998;80:1484.

    Article  CAS  Google Scholar 

  2. Shrader MW, Sanchez-Sotelo J, Sperling JW, Rowland CM, Cofield RH. Understanding proximal humerus fractures: image analysis, classification, and treatment. J Shoulder Elbow Surg. 2005;14:497–505.

    Article  Google Scholar 

  3. Sanders BS, Bullington AB, McGillivary GR, Hutton WC. Biomechanical evaluation of locked plating in proximal humeral fractures. J Shoulder Elbow Surg. 2007;16:229–34.

    Article  Google Scholar 

  4. Foruria AM, de Gracia MM, Larson DR, Munuera L, Sanchez-Sotelo J. The pattern of the fracture and displacement of the fragments predict the outcome in proximal humeral fractures. J Bone Joint Surg Br. 2011;93:378–86.

    Article  CAS  Google Scholar 

  5. Sproul RC, Iyengar JJ, Devcic Z, Feeley BT. A systematic review of locking plate fixation of proximal humerus fractures. Injury. 2011;42:408–13.

    Article  Google Scholar 

  6. Schliemann B, Wähnert D, Theisen C, Herbort M, Kösters C, Raschke MJ, et al. How to enhance the stability of locking plate fixation of proximal humerus fractures? An overview of current biomechanical and clinical data. Injury. 2015;46:1207–14.

    Article  Google Scholar 

  7. Jung S-W, Shim S-B, Kim H-M, Lee J-H, Lim H-S. Factors that influence reduction loss in proximal humerus fracture surgery. J Orthop Trauma. 2015;29:276–82.

    Article  Google Scholar 

  8. Blonna D, Rossi R, Fantino G, Maiello A, Assom M, Castoldi F. The impacted varus (A2.2) proximal humeral fracture in elderly patients: is minimal fixation justified? A case control study. J Shoulder Elbow Surg. 2009;18:545–52.

    Article  Google Scholar 

  9. Robinson CM, Page RS. Severely impacted valgus proximal humeral fractures. Results of operative treatment. J Bone Joint Surg Am. 2003;85-A:1647–55.

    Article  Google Scholar 

  10. Dilisio MF, Nowinski RJ, Hatzidakis AM, Fehringer EV. Intramedullary nailing of the proximal humerus: evolution, technique, and results. J Shoulder Elb Surg. 2016;25:e130–8.

    Article  Google Scholar 

  11. Cha H, Park K-B, Oh S, Jeong J. Treatment of comminuted proximal humeral fractures using locking plate with strut allograft. J Shoulder Elb Surg. 2017;26:781–5.

    Article  Google Scholar 

  12. Gardner MJ, Weil Y, Barker JU, Kelly BT, Helfet DL, Lorich DG. The importance of medial support in locked plating of proximal humerus fractures. J Orthop Trauma. 2007;21:185–91.

    Article  Google Scholar 

  13. Neviaser AS, Hettrich CM, Beamer BS, Dines JS, Lorich DG. Endosteal strut augment reduces complications associated with proximal humeral locking plates. Clin Orthop Relat Res. 2011;469:3300–6.

    Article  Google Scholar 

  14. Zhang L, Zheng J, Wang W, Lin G, Huang Y, Zheng J, et al. The clinical benefit of medial support screws in locking plating of proximal humerus fractures: a prospective randomized study. Int Orthop. 2011;35:1655–61.

    Article  Google Scholar 

  15. Kwon BK, Goertzen DJ, O'Brien PJ, Broekhuyse HM, Oxland TR. Biomechanical evaluation of proximal humeral fracture fixation supplemented with calcium phosphate cement. J Bone Joint Surg Am. 2002;84-A:951–61.

    Article  Google Scholar 

  16. Krappinger D, Bizzotto N, Riedmann S, Kammerlander C, Hengg C, Kralinger FS. Predicting failure after surgical fixation of proximal humerus fractures. Injury. 2011;42:1283–8.

    Article  Google Scholar 

  17. Lescheid J, Zdero R, Shah S, Kuzyk PRT, Schemitsch EH. The biomechanics of locked plating for repairing proximal humerus fractures with or without medial cortical support. J Trauma Acute Care Surg. 2010;69:1235–42.

    Article  Google Scholar 

  18. Resch H, Tauber M, Neviaser RJ, Neviaser AS, Majed A, Halsey T, et al. Classification of proximal humeral fractures based on a pathomorphologic analysis. J Shoulder Elb Surg. 2016;25:455–62.

    Article  Google Scholar 

  19. Court-Brown CM, McQueen MM. The impacted varus (A2.2) proximal humeral fracture: prediction of outcome and results of nonoperative treatment in 99 patients. Acta Orthop Scand. 2004;75:736–40.

    Article  Google Scholar 

  20. Robinson CM, Wylie JR, Ray AG, Dempster NJ, Olabi B, Seah KTM, et al. Proximal humeral fractures with a severe varus deformity treated by fixation with a locking plate. J Bone Joint Surg Br. The British editorial Society of Bone and Joint Surgery; 2010;92:672–678.

  21. Voigt C, Geisler A, Hepp P, Schulz AP, Lill H. Are polyaxially locked screws advantageous in the plate osteosynthesis of proximal humeral fractures in the elderly? A prospective randomized clinical observational study. J Orthop Trauma. 2011;25:596–602.

    Article  Google Scholar 

  22. Zhu Y, Lu Y, Shen J, Zhang J, Jiang C. Locking intramedullary nails and locking plates in the treatment of two-part proximal humeral surgical neck fractures: a prospective randomized trial with a minimum of three years of follow-up. J Bone Joint Surg Am. 2011;93:159–68.

    Article  Google Scholar 

  23. Schliemann B, Seifert R, Rosslenbroich SB, Theisen C, Wähnert D, Raschke MJ, et al. Screw augmentation reduces motion at the bone-implant interface: a biomechanical study of locking plate fixation of proximal humeral fractures. J Shoulder Elb Surg. 2015;24:1968–73.

    Article  Google Scholar 

  24. Got C, Shuck J, Biercevicz A, Paller D, Mulcahey M, Zimmermann M, et al. Biomechanical comparison of parallel versus 90-90 plating of bicolumn distal humerus fractures with intra-articular comminution. J Hand Surg. 2012;37:2512–8.

    Article  Google Scholar 

  25. Wanner GA, Wanner-Schmid E, Romero J, Hersche O, Smekal v A, Trentz O, et al. Internal fixation of displaced proximal humeral fractures with two one-third tubular plates. J Trauma Injury Infect Crit Care. 2003;54:536–44.

    Article  Google Scholar 

  26. Hessmann MH, Korner J, Hofmann A, Sternstein W, Rommens PM. Angle-fixed plate fixation or double-plate osteosynthesis in fractures of the proximal humerus: a biomechanical study. Biomed Tech (Berl). 2008;53:130–7.

    Article  Google Scholar 

  27. Neviaser AS, Hettrich CM, Dines JS, Lorich DG. Rate of avascular necrosis following proximal humerus fractures treated with a lateral locking plate and endosteal implant. Arch Orthop Trauma Surg. 2011;131:1617–22.

    Article  Google Scholar 

  28. Laux CJ, Grubhofer F, Werner CML, Simmen H-P, Osterhoff G. Current concepts in locking plate fixation of proximal humerus fractures. J Orthop Surg Res. 2017;12:137.

    Article  Google Scholar 

  29. Gerber C, Schneeberger AG, Vinh TS. The arterial vascularization of the humeral head. An anatomical study. J Bone Joint Surg Am. 1990;72:1486–94.

    Article  CAS  Google Scholar 

  30. Duparc F, Muller JM, Fréger P. Arterial blood supply of the proximal humeral epiphysis. Surg Radiol Anat. 2001;23:185–90.

    Article  CAS  Google Scholar 

Download references

Acknowledgments

The authors would like to acknowledge and thank the individual patient involved who was agreeable to publication of this article.

Funding

No funding was provided for this study.

Availability of data and materials

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.

Author information

Authors and Affiliations

Authors

Contributions

JTC was responsible for the study design, data collation, and manuscript generation. EC was responsible for being an independent assessor of range of motion at 1 year and contributed to data collection. DM was the senior author who performed the surgery and contributed to study design and manuscript editing. All authors read and approved final manuscript.

Corresponding author

Correspondence to John Tristan Cassidy.

Ethics declarations

Ethics approval and consent to participate

This is a retrospective, de-identified technical note and thus ethics approval is not required. Nevertheless, the patient was contacted and provided permission for the de-identified images as well as details of the case to be used for publication.

Consent for publication

As stated above, consent for publication was obtained from the patient in the outpatients’ department of our institution.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cassidy, J.T., Coveney, E. & Molony, D. “90/90” Plating of proximal humerus fracture—a technical note. J Orthop Surg Res 14, 41 (2019). https://doi.org/10.1186/s13018-019-1083-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-019-1083-3

Keywords