Biomimetic design and clinical application of Ti-6Al-4V lattice hemipelvis prosthesis for pelvic reconstruction

Abstract

Objective

This study aims to biomimetic design a new 3D-printed lattice hemipelvis prosthesis and evaluate its clinical efficiency for pelvic reconstruction following tumor resection, focusing on feasibility, osseointegration, and patient outcomes.

Methods

From May 2020 to October 2021, twelve patients with pelvic tumors underwent tumor resection and subsequently received 3D-printed lattice hemipelvis prostheses for pelvic reconstruction. The prosthesis was strategically incorporated with lattice structures and solid to optimize mechanical performance and osseointegration. The pore size and porosity were analyzed. Patient outcomes were assessed through a combination of clinical and radiological evaluations.

Results

Multiple pore sizes were observed in irregular porous structures, with a wide distribution range (approximately 300–900 μm). The average follow-up of 34.7 months, ranging 26 from to 43 months. One patient with Ewing sarcoma died of pulmonary metastasis 33 months after surgery while others were alive at the last follow-up. Postoperative radiographs showed that the prosthesis’s position was consistent with the preoperative planning. T-SMART images showed that the host bone was in close and tight contact with the prosthesis with no gaps at the interface. The average MSTS score was 21 at the last follow-up, ranging from 18 to 24. There was no complication requiring revision surgery or removal of the 3D-printed hemipelvis prosthesis, such as infection, screw breakage, and prosthesis loosening.

Conclusion

The newly designed 3D-printed lattice hemipelvis prosthesis created multiple pore sizes with a wide distribution range and resulted in good osteointegration and favorable limb function.

Introduction

The pelvis is a common anatomical site for primary malignant bone tumors as well as metastatic disease [1]. With the application of adjuvant chemotherapy and improvement in radiologic technology and surgical techniques, limb salvage surgery has become the standard treatment for pelvic tumors [2,3,4]. The utilization of metallic prostheses is the currently preferred reconstruction method following tumor resection. However, the complex anatomy of the pelvis as well as its load-bearing function pose a great challenge for surgeons. Until now, several prostheses have been developed for pelvic reconstruction, such as saddle prosthesis [5, 6], ice cream cone prosthesis [7, 8], and modular hemipelvis prosthesis [1, 9, 10]. However, when the ilium and acetabulum are completely removed, the saddle prosthesis and ice cream cone prosthesis cannot be fitted and fixed, which requires sufficient bone stock of the residual ilium for implant anchorage [11, 12]. In this situation, modular hemipelvis prosthesis has gained popularity for pelvic reconstruction due to its flexible features in surgery [13]. Nevertheless, implant integration with host bone is still unsatisfying, and the prosthesis relies on simple screw fixation. Therefore, the potential for complications in long-term follow-up remains high, such as screw breakage and prosthesis loosening [14].

In recent years, the development of 3D printing technology has enabled the fabrication of personalized prostheses and patient-specific instruments [15, 16]. Due to the unique advantage of design freedom for individual bone defects, 3D-printed hemipelvis prosthesis presents a compelling opportunity for pelvic reconstruction following tumor resection [2, 17,18,19,20,21]. Additionally, 3D printing allows for the creation of porous structures incorporated within the implant, which effectively reduces the elastic modulus and provides space for bone in-growth, thereby increasing osteointegration [22, 23]. Lattice structure refers to a 3D framework characterized by a network of interconnected struts or beams arranged in a repeating pattern. In medical devices and implants, lattice porous structures can provide lightweight, excellent interconnectivity, and good mechanical strength [24]. However, it is noted that the natural bone trabecular network is a complex and irregular porous structure [25]. Recent studies have shown that irregular lattice structures facilitate osteointegration due to the biomimetic property [26, 27]. However, regular lattice structure remains the currently used design in clinical prostheses. To our knowledge, there is no literature on the use of irregular lattice design in hemipelvis prostheses.

Therefore, this study aims to biomimetic design a new 3D-printed lattice hemipelvis prosthesis with irregular porous structures and evaluate its clinical efficiency for pelvic reconstruction following tumor resection, focusing on feasibility, osseointegration, and patient outcomes.

Methods

Patients

From January 2020 to October 2021, twelve patients with pelvic tumors underwent tumor resection and subsequently received 3D-printed lattice hemipelvis prostheses for reconstruction. Inclusion criteria were as follows: (1) pathological diagnosis of primary sarcoma, chondrosarcoma, or solitary metastatic lesion (2) lesions invade the acetabulum as well as most of the ilium, requiring complete removal of the ilium and acetabulum; (3) lesions do not extend to the sacroiliac joint; (4) follow-up period of more than 2 years. Exclusion criteria were: (1) life expectancy less than six months; (2) involvement of important neurovascular structures; (3) lack of complete follow-up information. Patient demographic and clinical characteristics, including gender, age, and pathological diagnosis were collected and shown in Table 1. Preoperatively, patients underwent detailed radiographic examinations of the pelvis, including X-ray, computed tomography (CT), and magnetic resonance imaging (MRI) (Fig. 1). The workflow for designing 3D-printed lattice hemipelvis implants was carried out after obtaining written informed consent from each patient.

Table 1 Demographics, clinical characteristics, and follow-up outcomes of twelve patients
Fig. 1
figure 1

Preoperative X-ray (A), CT (B), and MRI (C) scans of a 50-year-old patient with osteosarcoma involving the acetabulum as well as most of the ilium

Personalized hemipelvis prosthesis design

The collected CT data (DICOM format) were imported into Mimics software (Materialise, Leuven, Belgium) to segment anatomical 3D models of the pelvis and tumor (Fig. 2). According to the MRI examination results, the tumor resection boundary was set as 10 mm for chondrosarcoma, 20 mm for metastasis lesions, and 30 mm for high-grade sarcomas. Following the determination of the resection location, the osteotomy plane of the pubic and ischial were set as perpendicular as possible to the axis of the bone. Then, hemipelvis tumor resection was simulated based on the anatomical 3D model. The preliminary shape of the prosthesis was designed by mirroring the 3D model of the contralateral hemipelvis. Then, the prosthesis shape was simplified, including minimizing the ilium wing and removing the posterior iliac spines. For all cases, the pubis superior ramus of the prosthesis model was preserved to reconstruct the pelvic ring. Then, specific features were added to the prosthesis model, including plate, screw holes, and suture holes.

Fig. 2
figure 2

Profile of designing the personalized hemipelvis prosthesis for the individual bone defect. (A) determination of the osteotomy plane; (B) simulation of hemipelvis tumor resection based on the anatomical 3D model; (C, D) preliminary shape of the prosthesis with specific features, including plate, screw holes, and suture holes

Hemipelvis prosthesis was strategically incorporated with lattice structures and solid to optimize mechanical performance and osseointegration (Fig. 3). 3-Matic (Materialise, Leuven, Belgium) software was employed to create the lattice structure, and the dodecahedron cell was selected. Unit cell parameters were adjusted for the irregular porous lattice part and regular large porous lattice part, respectively (Table 2). The pore size (Figs. 4 and 5) and porosity (Supplementary 1) were analyzed. The finalized prosthesis design file was exported in the STL format.

Fig. 3
figure 3

Profile of designing lattice structure for the hemipelvis prosthesis. (A) three parts were strategically incorporated to optimize mechanical performance and osseointegration; (B) dodecahedron cell was selected; (C) a regular large porous lattice part for reducing weight; (D, E) an irregular porous lattice part with the property to facilitate osteointegration

Table 2 Details of the personalized hemipelvis prosthesis and lattice structure design
Fig. 4
figure 4

Pore size analysis of personalized lattice hemipelvis prosthesis for one patient. Multiple pore sizes were observed in the irregular porous lattice part with a wide distribution range (350–860 μm)

Fig. 5
figure 5

Pore size analysis of personalized lattice hemipelvis prosthesis for one patient. Multiple pore sizes were observed in the irregular porous lattice part with a wide distribution range (300–850 μm)

Ti-6Al-4V prosthesis manufacture

The lattice hemipelvis prostheses were manufactured using the electron beam melting (EBM) machine (ARCAM Q10plus, Mölndal, Sweden), utilizing the Ti-6Al-4V alloy powder. Then, the lattice hemipelvis prosthesis was polished, processed, and cleaned (Fig. 6). Prosthesis trial mold and patient-specific bone-cutting guides were fabricated by selective laser sintering with nylon powder. In addition, resin anatomical 3D models were prepared. Preoperatively, reconstruction of the pelvic defect with the 3D-printed lattice hemipelvis prosthesis was re-simulated was performed.

Fig. 6
figure 6

Final 3D model (A) and photos (B, C) of the lattice hemipelvis prosthesis for one patient

Surgical procedure

All the surgeries were performed by the same senior surgeon (CQ T). Patients adopted a lateral or floating position, and an ilioinguinal incision was selected to expose the lesion and a pubic incision was added if necessary. After visualization of the surgical site, osteotomies were performed with the assistance of patient-specific bone-cutting guides. Then, the nylon trial prosthesis was used to check the resection accuracy. After satisfactory implantation of the prosthesis trial mold, the sacroiliac articular cartilage was removed using bone rongeurs, exposing the sacral trabecula. The true hemipelvis prosthesis was carefully inserted into the prepared bone defect. To improve the accuracy of screw insertion, a drilling guide was used to drill bony screw paths (Fig. 7). Then, screws were inserted to secure the prosthesis in place. Then, the constrained acetabular pad was fixed in the acetabular cup of the prosthesis, and the proximal femur prosthesis was implanted. The remaining muscles were sutured to the corresponding parts of the prosthesis through suture holes. Finally, the incision was sutured layer by layer, and a drainage tube for continuous negative pressure drainage was placed.

Fig. 7
figure 7

Intraoperative photo of using a drilling guide to improve the accuracy of screw insertion

Follow-up and outcomes assessment

A third-generation cephalosporin combined with vancomycin was given for ten to 14 days, followed by oral antibiotics for four weeks. The patient was followed monthly during the first 3 months, and every 3 months thereafter. Patient outcomes were assessed through a combination of clinical and radiological evaluations. Radiographic imaging, such as X-rays or CT scans, was used to monitor implant place, stability, and alignment. To evaluate osseointegration, tomosynthesis-Shimadzu metal artifact reduction technology (T-SMART) was utilized. Functional outcomes were assessed by the Musculoskeletal Tumor Society (MSTS) 93 score.

Results

Lattice hemipelvis prosthesis design

This new type of lattice hemipelvis prosthesis incorporated by three parts: an irregular porous part with the biomimetic property to facilitate osteointegration, a regular large porous part for reducing weight, and a solid load-bearing part. In total, 12 prostheses were created based on individual bone defects (Table 2). Through porous analysis in 3-Matic, multiple pore sizes were observed in the irregular porous lattice part (Figs. 4 and 5) with a wide distribution range (approximately 300–900 μm). The porosity analysis results showed that the average porosity of irregular porous structures was around 75.02% in the 12 prostheses, ranging from 74.7 to 75.6%.

Clinical and radiologic outcomes

The average follow-up of 34.7 months, ranging 26 from to 43 months. No patient was lost to follow-up. One patient with Ewing sarcoma died of pulmonary metastasis 33 months after surgery while others were alive at the last follow-up. Postoperative radiographs showed that the prosthesis’s position was consistent with the preoperative planning. T-SMART images showed that the host bone was in close and tight contact with the prosthesis with no gaps at the interface, as shown in Fig. 8. Additionally, favorable bone density around the prosthesis was observed, without bone resorption or osteolysis. The average MSTS score was 21 at the last follow-up, ranging from 18 to 24. During the follow-up, three out of twelve patients experienced complications. Delayed wound healing seems to be the most frequent complication, which occurred in two patients. Both patients were cured after wound debridement. Dislocation occurred in one patient, requiring a closed reduction. In addition, there was no other complication requiring revision surgery or removal of the 3D-printed hemipelvis prosthesis, such as infection, screw breakage, and prosthesis loosening.

Fig. 8
figure 8

Osseointegration assessment according to T-SMART images of two patients, respectively

Discussion

3D printing technology has the potential to revolutionize the field of pelvic reconstruction, allowing design freedom of the prosthesis for individual bone defects and fabrication of complex porous structures [2]. In the present study, we biomimetic designed a new 3D-printed lattice hemipelvis prosthesis with irregular porous structures inspired by the natural bone trabecular network. The preliminary results revealed multiple pore sizes with a wide distribution range in prostheses and subsequent good osteointegration in patients.

The prosthetic replacement has become a promising pelvic reconstruction treatment following tumor resection due to its flexible features, acceptable functional outcomes, early weight-bearing, as well as the possibility of rapid recovery [17]. Many studies focused on surgical technique and prosthesis design in periacetabular defect reconstruction and reported reasonable clinical and function outcomes [28,29,30,31,32]. However, for tumor resection without ilium preservation, fixing the prosthesis to the sacrum is surgical difficult because the flat surface of the sacroiliac joint is vulnerable to shear loading [11]. Despite the utilization of multiple screw fixation could achieve primary stability, long-term non-integration between the implant and host bone may result in a high incidence of reconstruction failure [17]. Under this situation, osseointegration is essential for the long-term stability of the hemipelvis prosthesis.

The porous structure acts as an osteoconductive scaffold for bone ingrowth [33] and therefore often suggested on the hemipelvis prosthesis to facilitate osseointegration and long-term stabilization [14]. It is noted that the porous structure design plays a significant role in osseointegration. Previous studies have reported that porous structures with pore sizes of 300 to 800 μm and high porosity (75%) are the most effective in promoting bone ingrowth [34,35,36,37]. In addition, a large number of studies have shown that the optimal pore size for improving osseointegration is determined by a specific range rather than a single value [38]. Therefore, irregular porous structures with different pore sizes are specially developed, and the property of facilitating osteointegration has been demonstrated in many vitro experiments [26, 27]. However, current hemipelvis prosthesis porous structures were often designed with regular lattice structures, which were composed of a single-element superposition and had a fixed pore size and porosity (Table 3). In the present study, to create the irregular lattice structure, dodecahedral units were set as randomized filling. Pore analysis results showed that multiple pore sizes with a wide distribution range (approximately 300–900 μm) in the irregular porous part (Figs. 4 and 5). In theory, this irregular porous structure had the advantage of promoting osteointegration compared to regular porous structures, which was significant for long-term stabilization. In our cases, osseointegration assessments according to T-SMART images indicated a high degree of bone-to-implant contact, along with favorable bone density around the prosthesis. During the follow-up period, good prosthesis stability was achieved without screw breakage or prosthesis loosening.

Table 3 Review of previous studies on the porous structure design of 3D-printed hemipelvis prosthesis

In the present study, 3D-printed lattice hemipelvis prosthetic replacement following tumor resection allowed patients to restore favorable functional outcomes. The average MSTS score was 21 (range, 18–24) at the last follow-up, comparable with recent studies regarding using 3D-printed personalized hemipelvis prosthesis for pelvic reconstruction [2, 18, 20, 21, 39]. 3D-printed personalized prostheses were designed according to the mirror model of the contralateral hemipelvis. This procedure enabled the prosthesis to match individual bone defects completely while keeping a reasonable rotation center. In the present study, minimizing the ilium wing attributed to a much smaller prosthesis volume and made it easier for flap coverage. Additionally, the resected muscles around the hip were carefully repaired with suture holes on the prosthesis wing. Delayed wound healing was the most common complication in the present case series, occurring in two patients. In addition, dislocation occurred in one patient although muscles around the hip were reconstructed. In this study, no infection was detected during the follow-up. Besides strict aseptic procedures and rational use of antibiotics, osteotomies assisted with patient-specific bone-cutting guides shortened surgical time, which was believed to reduce the risk of infection. Additionally, the remaining muscles were sutured carefully to the corresponding parts of the prosthesis through suture holes. Reducing dead space around the prosthesis was another crucial strategy for preventing infection. It is noteworthy that the majority of previous studies have reported high rates of complications following similar surgeries [2, 40]. In light of the low incidence of complications observed in our study, it is imperative to acknowledge the limitations inherent in our sample size and follow-up duration. While our results are promising, further confirmation through longer follow-up periods and broader patient cohorts is warranted.

Conclusion

In conclusion, the newly designed 3D-printed lattice hemipelvis prosthesis created multiple pore sizes with a wide distribution range and resulted in good osteointegration and favorable limb function.

Data availability

No datasets were generated or analysed during the current study.

References

  1. Guo W, Li D, Tang X, Yang Y, Ji T. Reconstruction with modular hemipelvic prostheses for periacetabular tumor. Clin Orthop Relat Research®. 2007;461:180–8.

    Article  Google Scholar 

  2. Liang H, Ji T, Zhang Y, Wang Y, Guo W. Reconstruction with 3D-printed pelvic endoprostheses after resection of a pelvic tumour. bone Joint J. 2017;99(2):267–75.

    Article  PubMed  Google Scholar 

  3. Wang J, Min L, Lu M, Zhang Y, Wang Y, Luo Y, Zhou Y, Duan H, Tu C. Three-dimensional-printed custom-made hemipelvic endoprosthesis for primary malignancies involving acetabulum: the design solution and surgical techniques. J Orthop Surg Res. 2019;14:1–12.

    Article  Google Scholar 

  4. O’connor M, Sim F. Salvage of the limb in the treatment of malignant pelvic tumors. JBJS. 1989;71(4):481–94.

    Article  Google Scholar 

  5. Jansen J, Van de Sande M, Dijkstra P. Poor long-term clinical results of saddle prosthesis after resection of periacetabular tumors. Clin Orthop Relat Research®. 2013;471:324–31.

    Article  CAS  Google Scholar 

  6. Danışman M, Mermerkaya MU, Bekmez Ş, Ayvaz M, Atilla B, Tokgözoğlu AM. Reconstruction of periacetabular tumours with saddle prosthesis or custom-made prosthesis, functional results and complications. Hip Int. 2016;26(2):e14–8.

    Article  PubMed  Google Scholar 

  7. Issa S-P, Biau D, Babinet A, Dumaine V, Le Hanneur M, Anract P. Pelvic reconstructions following peri-acetabular bone tumour resections using a cementless ice-cream cone prosthesis with dual mobility cup. Int Orthop. 2018;42:1987–97.

    Article  PubMed  Google Scholar 

  8. Fisher N, Patton J, Grimer R, Porter D, Jeys L, Tillman R, Abudu A, Carter S. Ice-cream cone reconstruction of the pelvis: a new type of pelvic replacement: early results. J Bone Joint Surg Br Volume. 2011;93(5):684–8.

    Article  CAS  Google Scholar 

  9. Wang B, Xie X, Yin J, Zou C, Wang J, Huang G, Wang Y, Shen J. Reconstruction with modular hemipelvic endoprosthesis after pelvic tumor resection: a report of 50 consecutive cases. PLoS ONE. 2015;10(5):e0127263.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Zang J, Guo W, Yang Y, Xie L. Reconstruction of the hemipelvis with a modular prosthesis after resection of a primary malignant peri-acetabular tumour involving the sacroiliac joint. bone Joint J. 2014;96(3):399–405.

    Article  PubMed  Google Scholar 

  11. Guo Z, Peng Y, Shen Q, Li J, He P, Yuan P, Liu Y, Que Y, Guo W, Hu Y. Reconstruction with 3D-printed prostheses after type I + II + III internal hemipelvectomy: Finite element analysis and preliminary outcomes. Front Bioeng Biotechnol. 2023;10:1036882.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Enneking WF, Dunham W. Resection and reconstruction for primary neoplasms involving the innominate bone. JBJS. 1978;60(6):731–46.

    Article  CAS  Google Scholar 

  13. Zhou Y, Min L, Liu Y, Shi R, Zhang W, Zhang H, Duan H, Tu C. Finite element analysis of the pelvis after modular hemipelvic endoprosthesis reconstruction. Int Orthop. 2013;37:653–8.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Zhang Y, Tang X, Ji T, Yan T, Yang R, Yang Y, Wei R, Liang H, Guo W. Is a modular pedicle-hemipelvic endoprosthesis durable at short term in patients undergoing enneking type I + II tumor resections with or without Sacroiliac involvement? Clin Orthop Relat Research® 2018, 476(9).

  15. Javaid M, Haleem A, Singh RP, Suman R. 3D printing applications for healthcare research and development. Global Health J 2022.

  16. Javaid M, Haleem A. Additive manufacturing applications in orthopaedics: a review. J Clin Orthop Trauma. 2018;9(3):202–6.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Wang B, Hao Y, Pu F, Jiang W, Shao Z. Computer-aided designed, three dimensional-printed hemipelvic prosthesis for peri-acetabular malignant bone tumour. Int Orthop. 2018;42:687–94.

    Article  PubMed  Google Scholar 

  18. Zoccali C, Baldi J, Attala D, Scotto di Uccio A, Cannavò L, Scotto G, Luzzati A. 3D-printed titanium custom-made prostheses in reconstruction after pelvic tumor resection: indications and results in a series of 14 patients at 42 months of average follow-up. J Clin Med. 2021;10(16):3539.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Peng W, Zheng R, Wang H, Huang X. Reconstruction of bony defects after tumor resection with 3D-printed anatomically conforming pelvic prostheses through a novel treatment strategy. BioMed Research International 2020, 2020.

  20. Zhu D, Fu J, Wang L, Guo Z, Wang Z, Fan H. Reconstruction with customized, 3D-printed prosthesis after resection of periacetabular ewing’s sarcoma in children using triradiate cartilage-based surgical strategy: a technical note. J Orthop Translation. 2021;28:108–17.

    Article  Google Scholar 

  21. Li Z, Lu M, Min L, Luo Y, Tu C. Treatment of pelvic giant cell tumor by wide resection with patient-specific bone-cutting guide and reconstruction with 3D-printed personalized implant. J Orthop Surg Res. 2023;18(1):648.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Ma L, Wang X, Zhao N, Zhu Y, Qiu Z, Li Q, Zhou Y, Lin Z, Li X, Zeng X. Integrating 3D printing and biomimetic mineralization for personalized enhanced osteogenesis, angiogenesis, and osteointegration. ACS Appl Mater Interfaces. 2018;10(49):42146–54.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Wang H, Su K, Su L, Liang P, Ji P, Wang C. The effect of 3D-printed Ti6Al4V scaffolds with various macropore structures on osteointegration and osteogenesis: a biomechanical evaluation. J Mech Behav Biomed Mater. 2018;88:488–96.

    Article  CAS  PubMed  Google Scholar 

  24. Pei X, Wu L, Zhou C, Fan H, Gou M, Li Z, Zhang B, Lei H, Sun H, Liang J. 3D printed titanium scaffolds with homogeneous diamond-like structures mimicking that of the osteocyte microenvironment and its bone regeneration study. Biofabrication. 2020;13(1):015008.

    Article  Google Scholar 

  25. Chao L, Jiao C, Liang H, Xie D, Shen L, Liu Z. Analysis of mechanical properties and permeability of trabecular-like porous scaffold by additive manufacturing. Front Bioeng Biotechnol. 2021;9:779854.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Pei X, Wang L, Zhou C, Wu L, Lei H, Fan S, Zeng Z, Deng Z, Kong Q, Jiang Q. Ti6Al4V orthopedic implant with biomimetic heterogeneous structure via 3D printing for improving osteogenesis. Mater Design. 2022;221:110964.

    Article  CAS  Google Scholar 

  27. Pei X, Wu L, Lei H, Zhou C, Fan H, Li Z, Zhang B, Sun H, Gui X, Jiang Q. Fabrication of customized Ti6AI4V heterogeneous scaffolds with selective laser melting: optimization of the architecture for orthopedic implant applications. Acta Biomater. 2021;126:485–95.

    Article  CAS  PubMed  Google Scholar 

  28. Bus MPA, Boerhout EJ, Bramer JAM, Dijkstra PDS. Clinical outcome of pedestal cup endoprosthetic reconstruction after resection of a peri-acetabular tumour. Bone Joint J. 2014;96–B(12):1706–12.

    Article  PubMed  Google Scholar 

  29. Falkinstein Y, Ahlmann ER, Menendez LR. Reconstruction of type II pelvic resection with a new peri-acetabular reconstruction endoprosthesis. J Bone Joint Surg Br Volume. 2008;90–B(3):371–6.

    Article  Google Scholar 

  30. Ji T, Guo W, Yang RL, Tang XD, Wang YF. Modular hemipelvic endoprosthesis reconstruction–experience in 100 patients with mid-term follow-up results. Eur J Surg Oncol (EJSO). 2013;39(1):53–60.

    Article  CAS  PubMed  Google Scholar 

  31. Wang J, Min L, Lu M, Zhang Y, Wang Y, Luo Y, Zhou Y, Duan H, Tu C. What are the complications of three-dimensionally printed, custom-made, integrative hemipelvic endoprostheses in patients with primary malignancies involving the acetabulum, and what is the function of these patients? Clin Orthop Relat Res. 2020;478(11):2487.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Wong K, Kumta S, Geel N, Demol J. One-step reconstruction with a 3D-printed, biomechanically evaluated custom implant after complex pelvic tumor resection. Comput Aided Surg. 2015;20(1):14–23.

    Article  CAS  PubMed  Google Scholar 

  33. Kim HD, Amirthalingam S, Kim SL, Lee SS, Rangasamy J, Hwang NS. Biomimetic materials and fabrication approaches for bone tissue engineering. Adv Healthc Mater. 2017;6(23):1700612.

    Article  Google Scholar 

  34. Palmquist A, Snis A, Emanuelsson L, Browne M, Thomsen P. Long-term biocompatibility and osseointegration of electron beam melted, free-form–fabricated solid and porous titanium alloy: experimental studies in sheep. J Biomater Appl. 2013;27(8):1003–16.

    Article  CAS  PubMed  Google Scholar 

  35. Shah FA, Omar O, Suska F, Snis A, Matic A, Emanuelsson L, Norlindh B, Lausmaa J, Thomsen P, Palmquist A. Long-term osseointegration of 3D printed CoCr constructs with an interconnected open-pore architecture prepared by electron beam melting. Acta Biomater. 2016;36:296–309.

    Article  CAS  PubMed  Google Scholar 

  36. Taniguchi N, Fujibayashi S, Takemoto M, Sasaki K, Otsuki B, Nakamura T, Matsushita T, Kokubo T, Matsuda S. Effect of pore size on bone ingrowth into porous titanium implants fabricated by additive manufacturing: an in vivo experiment. Mater Sci Engineering: C. 2016;59:690–701.

    Article  CAS  Google Scholar 

  37. Hara D, Nakashima Y, Sato T, Hirata M, Kanazawa M, Kohno Y, Yoshimoto K, Yoshihara Y, Nakamura A, Nakao Y. Bone bonding strength of diamond-structured porous titanium-alloy implants manufactured using the electron beam-melting technique. Mater Sci Engineering: C. 2016;59:1047–52.

    Article  CAS  Google Scholar 

  38. Kapat K, Srivas PK, Rameshbabu AP, Maity PP, Jana S, Dutta J, Majumdar P, Chakrabarti D, Dhara S. Influence of porosity and pore-size distribution in Ti6Al4 V foam on physicomechanical properties, osteogenesis, and quantitative validation of bone ingrowth by micro-computed tomography. ACS Appl Mater Interfaces. 2017;9(45):39235–48.

    Article  CAS  PubMed  Google Scholar 

  39. Xu L, Qin H, Tan J, Cheng Z, Luo X, Tan H, Huang W. Clinical study of 3D printed personalized prosthesis in the treatment of bone defect after pelvic tumor resection. J Orthop Translation. 2021;29:163–9.

    Article  Google Scholar 

  40. Dai K-R, Yan M-N, Zhu Z-A, Sun Y-H. Computer-aided custom-made hemipelvic prosthesis used in extensive pelvic lesions. J Arthroplast. 2007;22(7):981–6.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported by Qingdao research institutes of Sichuan University, Research of biomedical materials and 3D printing related products (20GZ30301), the Science and Technology Research Program of Sichuan Province (2023YFH0099), and the Natural Science Foundation of Sichuan Province (2023NSFSC1749).

Author information

Authors and Affiliations

Authors

Contributions

ZZ L, Y L, L M, and CQ T were involved with the concept and design of this manuscript. TJ G, XH H, JJJ L, and X H were involved with the acquisition of the subject and data. ZZ L, MX L, and CQ T were involved in the design of the prosthesis. YT W, Y L, Y Z, L M, and CQ T were involved in the postsurgical evaluation of the patient. All authors contributed to the article and approved the submitted version.

Corresponding authors

Correspondence to Li Min or Chongqi Tu.

Ethics declarations

Ethics approval and consent to participate

This study was performed in accordance with the Declaration of Helsinki as revised in 2008 and was approved by the Ethics Committee of West China Hospital. The patients signed the informed consent form before surgery.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Z., Luo, Y., Lu, M. et al. Biomimetic design and clinical application of Ti-6Al-4V lattice hemipelvis prosthesis for pelvic reconstruction. J Orthop Surg Res 19, 210 (2024). https://doi.org/10.1186/s13018-024-04672-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-024-04672-5

Keywords