Skip to main content

Table 4 Anteversion of different measurements and correlations with postoperative CT-measured stem anteversion

From: Variety of femoral anteversion and its measurement in cementless total hip arthroplasty: Does robotic technology improve accuracy?

N = 64

Anteversion (Mean ± SD, range)

Postoperative CT-measured stem anteversion

r* (P value)

Intraclass correlation coefficient (P value)

Preoperative Mako-measured native femoral anteversion

14.64° ± 15.782°

(− 32° to 50°)

18.37° ± 8.20°

(3° to 38°)

0.459

(< 0.001)

0.546

(0.001)

Intraoperative surgeon estimated femoral stem anteversion

14.69° ± 6.41°

(0° to 30°)

0.281

(0.025)

0.428

(0.014)

Intraoperative goniometer-measured femoral anteversion**

20.03° ± 7.96°

(5° to 42°)

0.435

(< 0.001)

0.606

(< 0.001)

Intraoperative Mako-measured femoral stem anteversion

14.31° ± 7.87°

(− 8° to 35°)

0.806

(< 0.001)

0.892

(< 0.001)