Skip to main content

Table 3 Native femoral anteversion and intraoperative stem anteversion measured by robotic system for DDH and non-DDH patients

From: Variety of femoral anteversion and its measurement in cementless total hip arthroplasty: Does robotic technology improve accuracy?

 

Native femoral anteversion

Intraoperative stem anteversion measured by robotic system (with anteversion adjusted stem)

Intraoperative stem anteversion measured by robotic system (without anteversion adjusted stem)

 

 < 5° or > 25°

5°–25°

 < 5° or > 25°

5°–25°

 < 5° or > 25°

5°–25°

DDH (n = 63)

33 (52.4%)

30 (47.6%)

11 (17.7%)

51 (82.3%)**

10 (18.2%)

45 (81.8%)

Non-DDH (n = 54)

17 (32.1%)

36 (67.9%)*

11 (20.4%)

43 (79.6%)

10 (19.0%)

40 (81.0%)

P Value

0.028

0.719

0.813

  1. *One case of traumatic arthritis was difficult to measure preoperatively
  2. **One case of DDH lost Intraoperative Mako-measured femoral stem anteversion