Skip to main content

Table 3 Univariate analysis results for augmented screws with extensive epidural cement leakage

From: Anatomic distribution of basivertebral foramen with a magistral form in vertebral bodies of T10~L5 and its clinical significance for extensive epidural cement leakage in cement-augmented pedicle screw fixation: a multicenter case–control study

 

Extensive epidural Leakage (n = 32)

Non-extensive epidural Leakage (n = 1866)

T/χ2/Z

P values

The type of basivertebral foramen (magistral/dispersed)

26/6

775/1091

20.346

 < 0.001*

Volume of cement injected (ml)a

2.56 ± 0.83

2.16 ± 0.78

− 2.876

0.004*

The distance between screw tip and the midline of vertebral body (mm)a

4.27 ± 1.57

5.49 ± 2.39

2.894

0.004*

The distance between screw tip and the posterior wall of the vertebral body (mm)a

21.56 ± 3.58

24.15 ± 3.22

4.491

 < 0.001*

Cement extension into pedicle area (yes/no)

5/27

105/1761

5.760

0.016*

Screw design (solid/fenestrated)

10/22

418/1448

1.411

0.235

  1. Bold values indiacte statistically significant difference
  2. aQuantitative variables are expressed as mean ± SD