Table 3 Range of motion with three models at four operated segments at a pure moment of 7.5 Nm

From: Finite element analysis comparing a PEEK posterior fixation device versus pedicle screws for lumbar fusion

Motion

Lumbar level

Implant location

ROM (°)

Difference in ROM (°)

Intact spine

CPPF

PSS

PSS versus intact spine

CPPF versus intact spine

CPPF versus PSS

Flexion

L4–L5

L4–L5

9.0

0.2

0.2

 − 8.8

 − 8.8

0.0

L5–S1

L5–S1

9.2

0.8

0.7

 − 8.5

 − 8.4

0.1

L4–S1

L4–L5

9.0

0.3

0.2

 − 8.8

 − 8.7

0.1

L4–S1

L5–S1

9.2

0.9

0.4

 − 8.8

 − 8.3

0.5

Extension

L4–L5

L4–L5

 − 3.5

 − 0.4

 − 0.3

3.2

3.1

 − 0.1

L5–S1

L5–S1

 − 6.0

 − 0.7

 − 0.6

5.4

5.3

 − 0.1

L4–S1

L4–L5

 − 3.5

 − 0.2

 − 0.2

3.3

3.3

0.0

L4–S1

L5–S1

 − 6.0

 − 0.4

 − 0.4

5.6

5.6

0.0

Lateral bending

L4–L5

L4–L5

4.5

0.9

0.8

 − 3.7

 − 3.6

0.1

L5–S1

L5–S1

3.2

0.8

0.6

 − 2.6

 − 2.4

0.2

L4–S1

L4–L5

4.5

0.9

0.6

 − 3.9

 − 3.6

0.3

L4–S1

L5–S1

3.2

0.7

0.1

 − 3.1

 − 2.5

0.6

Torsion

L4–L5

L4–L5

2.1

0.4

0.9

 − 1.2

 − 1.7

 − 0.5

L5–S1

L5–S1

2.6

0.6

0.9

 − 1.7

 − 2.0

 − 0.3

L4–S1

L4–L5

2.1

0.3

0.6

 − 1.5

 − 1.8

 − 0.3

L4–S1

L5–S1

2.6

0.5

0.4

 − 2.2

 − 2.1

0.1

  1. CPPF—cortico-pedicular fixation, PSS—pedicle screw system, ROM—range of motion