Skip to main content

Table 2 Material parameters of cervical components and implants

From: Biomechanical effects of hybrid constructions in the treatment of noncontinuous cervical spondylopathy: a finite element analysis

Material

Elastic modulus (MPa)

Poisson ratio

Element type

Cross-sectional area (mm2)

Vertebra

Cortical bone

12,000

0.3

C3D4

–

Cancellous bone

100

0.2

C3D4

–

Endplate

500

0.4

S3

–

Cartilage

10

0.4

C3D4

–

Posterior structure

3500

0.25

C3D4

–

Intervertebral disc

Annulus fibre

450

0.45

T3D2

–

Annulus ground Substance

Hyperelasitc Mooney-Rivlin

 

C3D6

–

 

C10 = 0.56

   
 

C01 = 0.14

   

Nucleus pulposus

Hyperelasitc Mooney-Rivlin

 

C3D6

–

 

C10 = 0.56

   
 

C01 = 0.14

   

Ligaments

ALL

30

0.4

T3D2

6.1

PLL

20

0.4

T3D2

5.4

LF

10

0.4

T3D2

50.1

IL

10

0.4

T3D2

13.1

SL

10

0.4

T3D2

13.1

CL

10

0.4

T3D2

46.6

NuVasive® Helix ACP

Ti6Al4V

114.000

0.35

C3D4

–

CoRoent® Contour

PEEK

3400

0.4

C3D4

–

Synthes® Prodisc-C

End plates

(cobalt chromium)

210,000

0.3

C3D4

–

Core

(polyethylene)

800

0.3

C3D4

–

  1. ALL anterior longitudinal ligament, PLL posterior longitudinal ligament, LF ligament flavum, IL interspinous ligament, SL supraspinous ligament, CL capsular ligament, PEEK polyether-ether-ketone, C3D4 tetrahedron, S3 shell, T3D2 truss, tension only, C3D6 triangular prism