Skip to main content

Table 4 KSS Score preoperatively, at 6 month and 5-year follow-up

From: Early and mid-term results with the ATTUNE total knee replacement system compared to PFC Sigma: a prospective comparative study

KSS score (max 200)

N

Mean

SD

Median

Min–max

p value

Preoperatively

ATTUNE

PFC

30

30

105.6

96.5

17.5

26.3

103

100

63–139

44–145

0.249

6 month FU

ATTUNE

PFC

29

30

168.9

158.6

22.9

34.0

176

174

104–192

23–192

0.160

5-year FU

ATTUNE

PFC

20

24

178.2

170.2

26.6

22.2

189

175

110–200

120–200

0.106