Skip to main content

Table 2 Radiological evaluations just after TKA

From: Pressurized carbon dioxide lavage reduces the incidence of a radiolucent line around the tibial component two years after total knee arthroplasty

 

CO2− group (47 knees)

CO2+ group (53 knees)

p value

Implant alignment (°)

 α

96.9 ± 2.3

96.0 ± 1.9

n.s.

 β

88.6 ± 1.3

88.6 ± 1.4

n.s.

 γ

3.0 ± 2.0

2.5 ± 1.7

n.s.

 Φ

83.2 ± 2.5

84.0 ± 2.5

n.s.

Cement penetration of baseplate zones (mm)

 Medial

1.7 ± 0.8

3.5 ± 1.5

< 0.001

 Lateral

1.7 ± 0.9

3.2 ± 1.6

< 0.001

 Anterior

2.0 ± 1.0

3.0 ± 1.0

< 0.001

 Posterior

1.7 ± 0.7

3.2 ± 1.3

< 0.001

  1. n.s. not significant