Skip to main content

Table 2 Comparison of mechanical femorotibial angle (x ± s)

From: A novel 3D-printed patient-specific instrument based on “H-point” for medial opening wedge high tibial osteotomy: a cadaver study

 

PSI group

(n = 15)

Control group

(n = 11)

P-value

Preoperative (°)

178.13 ± 1.44

179.23 ± 2.25

0.143

Postoperative (°)

186.25 ± 1.47

187.53 ± 2.33

0.100

Degree of planing correction (°)

8

8

 

Degree of actual correction (°)

8.11 ± 0.84

8.3 ± 1.42

0.685

Absolute correction error (°)

0.62 ± 0.56

1.20 ± 0.73

0.032*

Absolute correction error within 1° (%)

86.7%

63.6%

 

Absolute correction error within 2° (%)

100%

81.8%

 
  1. *Statistically significant difference (P < 0.05)