Skip to main content

Table 2 The maximum displacement of the 4 models

From: Finite element analysis of different locking plate fixation methods for the treatment of ulnar head fracture

  Dorsal-side plate
(2 screws)
Dorsal-side plate
(3 screws)
Ulnar-side plate
(2 screws)
Ulnar-side plate
(3 screws)
20N axial compression 0.403 mm 0.386 mm 0.301 mm 0.248 mm
50N axial compression 1.007 mm 0.966 mm 0.837 mm 0.431 mm
1 N∙m torsion moments 0.176 mm 0.137 mm 0.156 mm 0.043 mm
1 N∙m lateral bending moments 0.400 mm 0.355 mm 0.800 mm 0.687 mm
1 N∙m extension bending moments 0.799 mm 0.676 mm 0.391 mm 0.248 mm