Table 3 Rotation excursions during squatting activity

From: Mobile-bearing insert used with total knee arthroplasty does not rotate on the tibial tray during a squatting activity: a cross-sectional study

 

10 weeks postoperatively

1 year postoperatively

p values

Knee extension angle (degrees)

− 2.0 [− 5.5, 1.6]

− 3.2 [− 10.9, 4.4]

0.709

Knee flexion angle (degrees)

55.3 [44.6, 66.0]

57.1 [46.7, 67.6]

0.760

FEM/TIB (degrees)

5.7 [4.2, 7.3]

6.3 [4.3, 7.8]

0.702

FEM/INS (degrees)

5.9 [4.5, 7.2] ※

5.5 [3.8, 7.2] †

0.547

INS/TIB (degrees)

1.8 [1.4, 2.2] ※

1.6 [1.1, 2.2] †

0.517

p value (FEM/INS with INS/TIB)

< 0.001

< 0.001

 
  1. Mean [95% CI]
  2. ※FEM/INS was significantly greater than that of INS/TIB at 10 weeks postoperatively
  3. † FEM/INS was significantly greater than that of INS/TIB at 1 year postoperatively
  4. FEM/TIB rotation of the femoral component relative to the tibial component, FEM/INS rotation of the femoral component relative to the polyethylene insert, INS/TIB: rotation of the polyethylene insert relative to the tibial component, p value probability value, 95% CI 95% confidence interval