Table 6 Maximal von Mises stresses of the femur, sleeve, and stem in single-legged stance (MPa)

From: Influence of a metaphyseal sleeve on the stress-strain state of a bone-tumor implant system in the distal femur: an experimental and finite element analysis

 

Intact femur

Implanted femur

Sleeve

Stem

Maximal stress

29.6

113.8a

27.29a

221a

43.41b

24.66b

18.09b

  1. aData are obtained in the immediate postoperation condition
  2. bData are obtained in the osseointegration condition