Table 1 Material properties assigned to the FE model

From: Interlaminar stabilization offers greater biomechanical advantage compared to interspinous stabilization after lumbar decompression: a finite element analysis

Component

Element type

Young modulus (MPa)

Poisson ratio

Cross-sectional area (mm2)

Bone

    

 Cortical bone

C3D4

12,000

0.3

 

 Cancellous bone

C3D4

100

0.2

 

 Cartilage endplate

C3D8

24

0.4

 

 Intervertebral disc

    

 Nucleus pulpous

C3D8

1

0.49

 

 Annulus ground substance

C3D8H

Hyperelastic (C10 = 0.18, C01 = 0.045)

  

 Annulus fibers

T3D2

Hypoelastic (360–550 MPa)

  

Ligaments

    

 Anterior longitudinal

T3D2

15.6 (< 12%), 20 (> 12%)

 

63.7

 Posterior longitudinal

T3D2

10 (< 11%), 20 (> 11%)

 

20

 Ligamentum flavum

T3D2

15 (< 6.2%), 19.5 (> 6.2%)

 

40

 Capsular

T3D2

7.5 (< 25%), 32.9 (> 25%)

 

30

 Interspinous

T3D2

10 (< 20%), 11.6 (> 20%)

 

40

 Supraspinous

T3D2

8.0 (< 20%), 15 (> 20%)

 

30

 Intertransverse

T3D2

10 (< 18%), 58.7 (> 18%)

 

2

Implants

    

 Coflex device, screws, and rods (titanium)

C3D4

110,000

0.3

Â