Skip to main content

Table 4 Maximum (mean) values of stress on the plate and J-bone in the three models under two levels of stress loading

From: J-bone graft with double locking plate: a symphony of mechanics and biology for atrophic distal femoral non-union with bone defect

Adduction load (N)DP+J-boneLLP+J-boneDP-only
Plate (MPa)J-bone (MPa)Plate (MPa)J-bone (MPa)Plate (MPa)
7501924121061264
1800715134907213970
  1. DP double-plate, LLP lateral locking plate