Table 1 Reproducibility results showing the difference in calculated component angular alignment across two cases performed by three engineers at two different time points. The maximum average difference for each case and a 95% confidence interval are shown for all three axes of rotation for the femoral and tibial components

From: Accurate determination of post-operative 3D component positioning in total knee arthroplasty: the AURORA protocol

Case

Operator

Run

Femoral component alignment

Tibial component alignment

F/E

V/V

IE

F/E

V/V

IE

Case 1

Sim Eng A

Run 1

1.1

− 0.6

− 0.8

7.1

0.6

5.9

Case 1

Sim Eng A

Run 2

1.0

− 0.5

− 1.2

7.0

0.4

6.2

Case 1

Sim Eng B

Run 1

1.0

− 0.8

− 1.5

7.0

0.5

4.5

Case 1

Sim Eng B

Run 2

0.8

− 0.7

− 0.8

7.4

0.4

4.7

Case 1

Sim Eng C

Run 1

1.1

− 0.6

− 1.8

6.9

0.8

5.5

Case 1

Sim Eng C

Run 2

1.1

− 0.4

− 1.0

7.2

0.6

5.9

Maximum difference from average (°) ± 95% CI

0.2 ± 0.1

0.2 ± 0.1

0.5 ± 0.3

0.3 ± 0.1

0.2 ± 0.1

0.9 ± 0.6

Case 2

Sim Eng A

Run 1

1.9

− 0.9

0.1

8.2

1.5

3.0

Case 2

Sim Eng A

Run 2

2.3

− 0.8

− 0.1

7.7

1.8

2.7

Case 2

Sim Eng B

Run 1

2.3

− 0.7

− 0.2

7.9

1.7

1.7

Case 2

Sim Eng B

Run 2

2.1

− 0.8

0.4

8.4

1.4

2.8

Case 2

Sim Eng C

Run 1

2.0

− 0.7

− 0.2

7.9

1.0

2.3

Case 2

Sim Eng C

Run 2

1.8

− 0.8

0.0

8.4

1.5

2.8

Maximum difference from average (°) ± 95% CI

0.3 ± 0.2

0.1 ± 0.1

0.3 ± 0.2

0.3 ± 0.2

0.4 ± 0.2

0.7 ± 0.4