Skip to main content

Table 3 Diagnostic screening tests and ROC curve results according to USG screw penetration

From: Ultrasonography or direct radiography? A comparison of two techniques to detect dorsal screw penetration after volar plate fixation

 

Diagnostic scan

Kappa agreement

ROC curve

p

Sensitivity

Specificity

Positive predictive value

Negative predictive value

Area

95% Confidence interval

Lateral radiograph

41.67

100.0

100.0

83.33

0.515

0.708

0.511–0.905

0.033*

45° supination view

50.00

100.0

100.0

85.37

0.598

0.750

0.560–0.940

0.010*

45° pronation view

16.67

100.0

100.0

77.78

0.230

0.583

0.382–0.784

0.393

Dorsal tangential view

58.33

100.0

100.0

87.50

0.676

0.792

0.611–0.972

0.003**

  1. *p < 0.05; **p < 0.01