Temperature Beneath Cast (at time in (m) minutes)

Test #

Cast Material

Layers of cast

Cotton Stockin-ette Present?

Cast Padding Type

Layers of Cast pad

H20 Dip Temp (°C)

1 m

5 m

10 m

15 m

20 m

25 m

30 m

35 m

40 m

Peak

Repeated Tests for Plaster and Fiberglass

1

SFPR

12

Yes

Webril

1

32

27

27.5

28.5

34

35

34

X

X

X

35

2

SFPR

12

Yes

Webril

1

32

28

28

28.5

31

34

34

X

X

X

34.5

3

SFPR

12

Yes

Webril

1

32

29

29

29

32

34

33.5

X

X

X

34

10

DLFR

7

Yes

Webril

1

32

31.5

33

33.5

31

X

X

X

X

X

33.5

11

DLFR

7

Yes

Webril

1

32

31.5

33

34.5

31.5

X

X

X

X

X

34.5

12

DLFR

7

Yes

Webril

1

32

30

31.5

32

30

X

X

X

X

X

33.5

Effect of Layers of Cast Material on Temperature Underneath

4

SFPS

16

Yes

Webril

1

32

27.5

29

28.5

29

32

33.5

X

X

X

33.5

5

SFPS

12

Yes

Webril

1

32

29.5

29.5

29

29.5

31

33.5

X

X

X

33.5

6

SFPS

20

Yes

Webril

1

32

30

30

30

30.5

32.5

33.5

X

X

X

35.5

7

SEFPR

12

Yes

Webril

1

32

28

30

35

41

38

X

X

X

X

41

8

SEFPR

16

Yes

Webril

1

32

30

32.5

40.5

41

36.5

X

X

X

X

41

9

SEFPR

20

Yes

Webril

1

32

33

37

46.5

45

40

X

X

X

X

46.5

14

DLFR

10

Yes

Webril

1

32

32

33.5

36

32.5

X

X

X

X

X

36

15

DLFR

10

Yes

Webril

1

32

32.5

35

38

36

X

X

X

X

X

38

16

DLFR

12

Yes

Webril

1

32

31

33

39

38

X

X

X

X

X

39

Effect of Cast Padding Thickness and Types

21

DLFR

7

Yes

Webril

3

32

31.5

34

34.5

33

X

X

X

X

X

34.5

22

DLFR

7

Yes

Webril

5

32

31.5

34

33.5

32

X

X

X

X

X

34

23

3MSFR

7

No

Procel

1

32

30

34.5

37

33

X

X

X

X

X

37

24

3MSFR

7

No

Procel

3

32

28

30

34

32

X

X

X

X

X

34

25

SFPR

20

Yes

Webril

1

39

29.5

30

32.5

37.5

38

35

X

X

X

38

26

SFPR

20

Yes

Webril

3

39

31.5

32

35.5

42.5

44

41

X

X

X

44

27

SFPR

20

Yes

Webril

5

39

33

33

36

42.5

44

41

X

X

X

44

Pillow Effect

28

SFPR

20

Yes

Webril

1

39

32

32.5

36.5

46.5

53

54

53

50.5

48

54

29

SFPR

12

Yes

Webril

1

32

29

29

31.5

32.5

43

45.5

45.5

X

X

45.5

Effect of Dip Water, Temp, Splints, etc.

13

3MSFR

7

Yes

Webril

1

32

32

36

36.5

32

X

X

X

X

X

36.5

17

DLFR

7

Yes

Webril

1

39

34

36.5

36.5

33

X

X

X

X

X

36.5

18

3MSFR

7

Yes

Webril

1

39

35

39.5

37.5

33

X

X

X

X

X

39.5

19

3MSFR

7

No

Prepadded

1

39

34

32

31.5

30

X

X

X

X

X

34

20

3MSFR

7

No

Prepadded

1

32

27

30

35

33

X

X

X

X

X

35

30

SFPR

20

Yes

Webril

1

39

33

34

39

43

42

39

X

X

X

43

31

SFPR

12

Yes

Webril

1

39

29

30

31

34

36

35.5

X

X

X

36

  1. Presents the data from all tests performed. The purpose of the test is described in the text and grouped by numbers (i.e., tests 1–3 and 10–12 are reproducibility of measures tests for plaster and fiberglass respectively. Data organized in columns represents controlled and variable factors including: plaster type, material thickness, padding thickness, dip water temperature, dip water purity, ambient humidity and temperature, as well as the outcomes measure of temperature over time.