Skip to main content

Table 1 Test results of classical and novel locking plates

From: A novel model of locking plate and locking spongious screw: a biomechanical in vitro comparison study with classical locking plate

 

Axial Pull-out Strength (N)

Specimen

Classic Locking Plate

Novel Locking Plate

1

729,01

1098,54

2

426,79

1100,22

3

917,99

943,86

4

883,31

1167,94

5

632,40

1162,91

6

426,88

1203,83

7

231,33

1155,01

Average

606,82

1118,91

Standart deviation

256,46

85,88