Skip to main content

Table 1 Material properties for each part of the assembly

From: Biomechanical comparison of all-polyethylene total knee replacement and its metal-backed equivalent on periprosthetic tibia using the finite element method

Part

Material

E [MPa]

µ [−]

References

All-poly components

UHMWPE

670

0.46

[19]

Metal part of the MBT component

Tivanium® alloy

110,000

0.30

[9, 20]

Femoral component

Zimaloy®

210,000

0.29

[9, 21]

Bone cement

Palacos R

2891

0.40

[22]

‘A’—155 mm

Cortical bone

15,000

0.30

[6, 18]

‘B’—20 mm

Cortical bone

10,500

0.30

[6, 18]

‘C’—25 mm

Cortical bone

7500

0.30

[6, 18]

Cancellous bone

Cancellous bone

Heterogeneous

0.30

[6, 18]