Skip to main content

Table 2 Data for stress shielding (SS) expressed as percent of initial load not reflected at fusion surface.

From: Biomechanical investigation of a novel ratcheting arthrodesis nail

 

SS 3 mm

SS 6 mm

Solid 1

33.0

32.0

Solid 2

41.0

28.6

Solid 3

30.4

25.4

Solid Avg.

34.8

28.7

Solid S.D.

4.9

3.0

Threaded 1

54.5

48.6

Threaded 2

29.1

33.7

Threaded 3

37.2

40.5

Threaded Avg.

40.3

40.9

Threaded S.D.

11.6

6.7

Ratcheting 1

29.0

11.5

Ratcheting 2

22.6

12.2

Ratcheting 3

3.8

10.7

Ratcheting Avg

18.5

11.5

Ratcheting S.D.

11.7

11.5