Advantages of customization of osseointegrated implants in transfemoral amputees: a comparative analysis of surgical planning

Abstract

Background

Commercially available osseointegrated devices for transfemoral amputees are limited in size and thus fail to meet the significant anatomical variability in the femoral medullary canal. This study aimed to develop a customized osseointegrated stem to better accommodate a variety of femoral anatomies in transfemoral amputees than off-the-shelf stems. Customization is expected to enhance cortical bone preservation and increase the stem-bone contact area, which are critical for the long-term stability and success of implants.

Methods

A customized stem (OsteoCustom) was designed based on the statistical shape variability of the medullary canal. The implantability of the OsteoCustom stem was tested via 70 computed tomography (CT) images of human femurs and compared to that of a commercial device (OFI-C) for two different resection levels. The evaluations included the volume of cortical bone removed and the percentage of stem-bone contact area for both resection levels. Statistical significance was analyzed using paired and unpaired t tests.

Results

The OsteoCustom stem could be virtually implanted in all 70 femurs, while the OFI-C was unsuitable in 19 cases due to insufficient cortical thickness after implantation, further emphasizing its adaptability to varying anatomical conditions. The OsteoCustom stem preserved a greater volume of cortical bone than did the OFI-C. In fact, 42% less bone was removed at the proximal resection level (3.15 cm³ vs. 5.42 cm³, p ≤ 0.0001), and 33% less at the distal resection level (2.25 cm³ vs. 3.39 cm³, p = 0.003). The stem-bone contact area was also greater for the OsteoCustom stem, particularly at the distal resection level, showing a 20% increase in contact area (52.3% vs. 32.2%, p = 0.002) compared to that of the OFI-C.

Conclusions

The OsteoCustom stem performed better than the commercial stem by preserving more cortical bone and achieving a greater stem–bone contact area, especially at distal resection levels where the shape of the medullary canal exhibits more inter-subject variability. Optimal fit in the distal region is of paramount importance for ensuring the stability of osseointegrated implants. This study highlights the potential benefits of customized osseointegrated stems in accommodating a broader range of femoral anatomies, with enhanced fit in the medullary canal.

Background

The use of osseointegrated prostheses for the treatment of lower limb amputations has been practiced since 1990 [1,2,3]. This solution provides a viable option for patients whose needs cannot be adequately addressed through traditional suspension sockets [4, 5], which are associated with issues such as skin irritation, discomfort, volume fluctuations, and a high rate of abandonment [6,7,8]. Osseointegrated prostheses establish a direct structural link between the external prostheses and the remaining living bone, showcasing related benefits, such as improved somatosensory feedback, an increase in hip joint range of motion, and a decrease in metabolic energy expenditure while ambulating [5, 9, 10].

The widespread adoption of osseointegrated transfemoral prostheses, however, has been hindered by concerns among clinicians about several possible complications, which include septic and aseptic causes of implant failure. The septic failure rates are a significant concern in clinical practice [11]. Nevertheless, there exist several aseptic complications, which affect a concerningly high fraction of patients. Aseptic complications include postoperative periprosthetic fractures and aseptic loosening [12, 13]. Periprosthetic fractures are reported to occur with an incidence between 4% and 44% [14,15,16,17]. This mode of failure is mainly related to the removal of cortical bone that occurs during the surgical procedure [18]. Excessive removal during surgery results in weakening of the bone structure, compromising the overall stability and integrity of the implant [13, 15, 19,20,21]. No general indication is available for osseointegrated femoral implants. However, the guidelines for one of the most common stems on the market (e.g. OPRA, [22]) recommend preserving at least 2 mm of cortical bone thickness. Therefore, careful consideration of cortical bone removal is essential to ensure optimal implant longevity and performance. It should also be considered that osseointegration implants are often placed after wearing the socket for many years, which has acted as an external brace and thus protected the weakened bone after surgery [17]. Concerning the second problem, aseptic loosening refers to the gradual separation of an implant from surrounding bone tissue without the presence of infections and is reported in 3–29% of cases [23]. This phenomenon occurs when there is insufficient osseointegration between the bone and the prosthesis, which depends on excessive relative movements at the bone-implant interface and inadequate contact area [24, 25]. This issue may arise both in the short and long-term following the stem implantation [26,27,28,29,30,31]. To reduce the problems related to adverse bone remodeling caused by stress shielding, osseointegration should mainly occur near the level of bone resection (i.e., proximal to the femoral component of total hip replacement, distal to the stem for transfemoral amputees) to maintain physiological load transfer from the implant to the host bone as much as possible [32, 33]. Therefore, to reduce the risk of aseptic loosening, surgeons aim to achieve as much contact area as possible, especially at the implant-bone interface near the resection site [34,35,36].

To avoid possible fractures and mitigate aseptic loosening, it is crucial that the prosthesis design respects the anatomy of the femoral medullary canal while also considering its anatomical variability. Indeed, it has been demonstrated that there is significant inter-subject variability in various anatomical parameters (e.g., diameter and ellipticity of the canal, radius of curvature of the canal, and conicity) and that these variations also depend on the level of resection and on the residual canal segment [37, 38].

To date, however, osseointegrated off-the-shelf prostheses are manufactured in a limited number of sizes, where only the diameter and length of the implant vary, thus failing to consider the variability of the other anatomical parameters of the medullary canal [22, 39, 40]. Consequently, several inclusion criteria are in place, within which not all patients qualify. For instance, commercial osseointegrated prostheses such as OPRA, OPL, ILP, and BADAL X are typically available with a diameter ranging from 14 mm to 24 mm [22, 39, 40]. In addition, some of these prostheses (e.g. OPL, ILP, BADAL X) are manufactured with a fixed radius of curvature. These sizes do not cover the anatomical variability of the entire population [38], as the femoral canal may exhibit smaller/larger values and/or have different radii of curvature compared to those proposed by the current prostheses off the shelf. In addition, osseointegrated implantation is not advisable where canal reaming would result in excessively thin cortical bone remaining around the implant. Hence, it is imperative to evaluate whether customization would enhance the implant-bone fit-and-fill for this type of stem.

The aim of this study was to design a customized osseointegrated stem for transfemoral amputees capable of the following:

  • Being suitable for a broader range of patient anatomies than the currently available commercial osseointegrated stems.

  • Preserving a greater amount of cortical bone.

  • Increasing the stem-bone contact area.

Materials and methods

Design of the customized stem

The customized stems (hereafter referred to as OsteoCustom, Fig. 1) consisted of two main parts:

  • A proximal portion, with customized diameter and radius of curvature of the canal. This part was designed to have a number of ribs as a function of the diameter. The tip – at the proximal end of this portion – consisted of a hemisphere with a diameter equal to the inner diameter of the stem, excluding the ribs. This portion was designed similarly to other commercially available models.

  • A distal portion, customized to fit the shape of the medullary canal more precisely by adjusting the main anatomical parameters (e.g., diameter, conicity, ellipticity). The intra-extra rotation angle between the planes of curvature of the proximal portion was also customized. This portion terminated distally with the connector for the external components of the prosthesis (e.g., the abutment).

Fig. 1
figure 1

Example of an OsteoCustom stem. (a) Overall view of the stem; (b) zoom on the proximal part, with ribs; (c) zoom on the distal part, to be connected with the abutment. (d) Example of a 3D-printed stem, with the external connection

To customize the OsteoCustom stem for each femur examined in the present study (see below, in Sect. 2.3), the anatomical parameters were extracted by following an existing procedure, which was based on the reconstruction of the model at cross sections spaced by 10 mm [38]. The anatomical parameters were defined as follows:

  • Each cross-section was first best fitted with a circumference, thus estimating the diameters at each cross section. The diameter of the proximal portion of the stem was calculated as the mean diameter of the medullary canal in the relevant proximal cross sections.

  • Each cross-section was then best fitted with ellipses, thus identifying the centroid and estimating the major and minor axes at each cross section. The radius of curvature of the canal was computed by reconstructing the arc of the circle passing through the centroids of the canal along the entire length of the stem. It must be specified that the radius of curvature used in this study was the radius of curvature of the segment of medullary canal where the stem would be inserted, rather than the radius of curvature of the entire femur.

  • The conicity was computed for the distal portion of the stem as the difference between the diameter in the most distal section and the minimum diameter within the segment considered (this often but not always corresponded to the diameter in the most proximal cross section).

  • The ellipticity was computed for the distal portion of the stem as the difference between the major axis and the minor axis at the most distal cross section of the canal.

The range of variability for these anatomical parameters to fit 95% of the population was derived from the statistical shape model (SSM) of the distal medullary canal of a large cohort of femurs [38] (Table 1). To reduce the risk of mechanical failure of the implant, the smallest diameter considered for the most distal section was 15 mm since smaller diameters were found to be a significant predictor of mechanical failure [41]. This corresponds to the smallest size available for the OFI-C stem (see below).

The customized stems were designed using Autocad Inventor (vers. 23, Autodesk, San Francisco, CA, USA).

Table 1 Range of variability and average values for the main anatomical parameters. The diameter is expressed as the mean diameter of the medullary canal. The radius of curvature of the canal was computed by reconstructing the arc of the circle passing through the centroids of the canal. The conicity is expressed as the difference between the diameter in the most distal section and the minimum diameter within the segment considered. The ellipticity is calculated as the difference between the major axis and the minor axis at the most distal section of the canal. More information is provided in [38]

Commercial stem selected for comparison

To compare the customized stem with a gold standard, a commercial stem was used (OFI-C, Badal X, OTN Implants, Fig. 2). Such stem was selected for comparison as it is one of the most commonly utilized designs, among those employing a press-fit fixation, other than being the most similar to OsteoCustom in terms of design [42].

Fig. 2
figure 2

Example of an OFI-C stem. (a) CAD of the stem without the distal connection to the abutment; (b) OFI-C stem on the market

Collection of CT scans, canal segmentation, and extraction of anatomical parameters.

To detect a significant difference between the two types of stems, a power analysis (\(\:\alpha\:=0.05,\:\:\beta\:=80\%\)) was performed on the results while the procedure was carried out, including an increasing number of cases as needed [43]. This analysis indicated that 70 samples provided adequate statistical power both for bone volume removal and for the contact area. The CT scans of the lower limb were collected by combining data from two different databases:

  • Twenty-two femur scans were obtained from the repository of ex vivo specimens from the Laboratory of Biomechanics at the University of Bologna. All of these previous studies received approval from the bioethics committee of the University of Bologna.

  • Forty-eight scans from individuals chosen from the HipOp registry (Rizzoli Orthopedic Institute) were used, for which the data are partially available at [44]. Informed consent was obtained from all subjects participating in the study.

All CT scans had a pixel size within the range of 0.41 to 0.78 mm, and the slice thickness varied from 0.5 to 2 mm. The dataset was selected to satisfy the guidelines of osseointegrated prosthesis implantation (e.g., age < 65 years old, BMI < 30) [22, 40] and, more in general, press-fit implantation (e.g. cementless hip prostheses) [45,46,47]. A similar mean age was chosen among follow-up studies on the literature [14, 26, 29, 42]. While these ranges of age and BMI are generally recommended and found in the literature, also subjects with higher BMI or older than 65 years are sometimes recruited for osseointegrated implantation. Additionally, a balanced representation of both men and women, encompassing a broad spectrum of age, height, and weight, was maintained throughout the data collection process (Table 2).

Table 2 Details about the study, with ‘Age’ referring to the subjects’ age at the time of the CT scan

As each femur was positioned within its respective CT scan reference system, all femurs underwent initial realignment along their longitudinal axis using Mimics software (version 24.0, Materialise NV, Leuven, Belgium). This axis was defined as the line connecting the lateral edge of the piriformis fossa to the intercondylar notch [48]. Threshold-based segmentation was performed to isolate the entire femur [226–2999 HU] and the cortical bone [662–2999 HU] in the diaphyseal area. The shape of the medullary canal was then obtained from the Boolean subtraction of the previous masks.

Due to the specific length of the stem, at least 140 mm of residual canal is needed to provide the necessary space for implantation. Patients with a shorter femoral stump are not eligible for standard osseointegrated implantation [40]. In addition, to leave enough space for the external knee prosthesis, at least 150 mm are needed from the osteotomy to the intercondylar notch. Therefore, to include only the region of interest where an osseointegrated stem can be implanted, two extreme scenarios were simulated. Each canal was indeed cropped to simulate the two cases (Fig. 3), describing the range specified in [22, 40]:

  • Resection level A, representing the most proximal possible osteotomy: 140 mm distal to the lesser trochanter.

  • Resection level B, representing the most distal possible osteotomy: 150 mm proximal to the intercondylar notch.

The masks of the canals were then exported as triangulated surfaces. A total of 140 canal segments (70 for level A and 70 for level B) was therefore analyzed.

Fig. 3
figure 3

(a) Sagittal view of a typical femur for reference; (b) Sagittal view of the canal considering the most proximal resection (level A, on the left, in yellow); (c) 3D view of the mask of the canal; (d) Sagittal view of the canal considering the most distal resection (level B, on the right, in yellow); and (e) 3D view of the mask of the canal for the same femur

For both resection levels, the cross-section of the canal was then reconstructed every 10 mm and best-fitted with an ellipse, based on a validated procedure (Fig. 4) [38]. By measuring the major/minor axes and centroids, the anatomical parameters already cited in Sect. 2.1 were calculated for each mask (diameter, radius of curvature of the canal, conicity, and ellipticity).

Fig. 4
figure 4

(a) Sagittal view of a typical femur for reference; (b) Sagittal view of the canal; (c) Visual representation of the canal (in blue) reconstructed every 10 mm and best-fit with an ellipse (in red), with the major axes in pink and the minor axes in green

Stem selection and preoperative planning

The customized OsteoCustom stems were designed based on the aforementioned anatomical parameters for both resection levels A and B for each of the 70 femurs examined. In parallel, the size of the commercial stem (OFI-C, Badal X, OTN Implants) was provisionally chosen based on the mean diameter of the medullary canal. Preliminarily, for both levels A and B, the two stems were virtually aligned on the same femur following the centroids of the canal. The actual preoperative planning was then carried out by two experienced surgeons, who assessed whether the size chosen through the procedure for both prostheses (OsteoCustom and OFI-C) was appropriate. If the planned stems were considered too small/large, adjustments were made accordingly to correct the size. Adjustments of 1 mm were needed in 17 cases.

To evaluate whether a customized stem would allow the inclusion of more cases, the canals were categorized into two groups:

  • Group 1: canals in which a commercially available size of the OFI-C stem preserved at least 2 mm of cortical bone around the stem (as per the guidelines for osseointegrated prostheses [22]).

  • Group 2: canals where the insertion of the commercial prosthesis did not meet these criteria. In these canals, implantation of the customized stem was tested.

Measurement of bone volume removal

The volume of cortical bone removed was calculated as the difference between the volume before and after the insertion (Fig. 5) using a script in Python (version 3.11, Python Software Foundation). This volume was calculated for both stems and both resection levels (\(\:{Vol}_{A}\) and \(\:{Vol}_{B}\)) for all femurs.

Fig. 5
figure 5

Evaluation of bone volume removal in the canal. (a) Sagittal view of the femur, for reference; (b) Sagittal view of an example of an OsteoCustom stem virtually inserted in a femoral canal; (c) Axial view of proximal (top) and distal slices (bottom) with the volume of the cortical shell before (yellow + blue) and after (blue) the stem insertion (green)

Measurement of the stem-bone contact area

To measure the contact area (\(\:Area\)) between the stem and the canal, their surfaces were reconstructed as a series of points every 5 mm (Fig. 6). The intersection between the stem and the canal was quantified as the percentage of points where the canal was inside the stem volume over the total number of points of the canal at each cross section. The contact area was calculated for the proximal and distal portions of the stem separately. These values were then compared between the two stems for both resection levels (\(\:{Area\:}_{prox,\:\:A}\), \(\:{\:Area\:}_{dist,A}\), \(\:\:{Area}_{\:prox,\:B}\), \(\:{Area\:}_{dist,B}\)) for all femurs.

Fig. 6
figure 6

Evaluation of the stem-bone contact area in the canal. (a) Sagittal view of the femur, for reference; (b) 3D representation of the contact area (in red) between the stem (in green) and the canal (in blue); (c) An example of the contact area calculated for the proximal (top) and distal (bottom) portions of the stem

Statistical analysis

Continuous variables were verified for normality using the Shapiro‒Wilk test. Since all the variables were found to be normal, parametric statistical tests were chosen. Paired T-tests were performed to compare:

  • The cortical bone removed to implant the OsteoCustom and OFI-C stems for both resection levels A (\(\:{Vol}_{\:A}\)) and B (\(\:{Vol}_{\:B}\)). These evaluations were performed for both Groups 1 and 2.

  • The contact areas of the OsteoCustom and OFI-C stems in the proximal and distal portions of the stem for resection levels A (\(\:{Area\:}_{prox,\:\:A}\) and \(\:{Area\:}_{dist,\:\:A}\), respectively) and B (\(\:{Area\:}_{prox,\:\:B}\) and \(\:{Area\:}_{dist,\:\:B}\), respectively). These evaluations were performed for both Groups 1 and 2, but the results for Group 2 are reported in the Supplementary Materials for brevity.

To assess whether similar results would be achieved by implanting the OsteoCustom stem in medullary canals that are not eligible for implantation with the OFI-C, the contact area for the OsteoCustom stem was compared between Group 1 and Group 2 with an unpaired t test.

A two-tailed p value ≤ 0.05 was considered to be statistically significant. All the data were analyzed using GraphPad Prism (version 10, GraphPad Software, La Jolla, CA, USA).

Results

Canals assigned to Group 1 and Group 2

Among the 70 medullary canals analyzed, 51 retained at least 2 mm of cortical bone thickness after the virtual insertion of an OFI-C stem, both for resection levels A and B (and were thus assigned to Group 1), while 19 did not meet this criterion (thus categorized as Group 2). For most of the cases who fell under Group 2, the diameter of the medullary canal was less than 12 mm (\(\:\sim70\%\) of the cases). For each of Group 1 and Group 2 cases, it was possible to design an OsteoCustom stem capable of meeting the criteria outlined by osseointegrated prosthesis guidelines. In case of canals with a diameter smaller than 15 mm in the isthmus, the proximal part of the OsteoCustom stem was designed with a diameter from 12 mm to 15 mm. By adjusting the parameters of conicity and ellipticity, a diameter of at least 15 mm was achieved in the distal section.

Comparison of bone volume removal for the OsteoCustom and OFI-C

From a visual comparison of the bone volume removed, it was already clear that implantation with the OsteoCustom stem spared more cortical bone tissue than implantation with the OFI-C in the same canal (Fig. 7). For Group 1, a significantly lower volume of cortical bone needed to be removed to implant the OsteoCustom stem (3.15 cm3 on average) than for the OFI-C (5.42 cm3) for resection level A (\(\:p\le\:0.0001\)). Similarly, less cortical bone needed to be removed to implant the OsteoCustom stem (2.25 cm3) than to implant the OFI-C (3.39 cm3) for resection level B (\(\:p=0.003\), Fig. 8).

Although, in principle, the OFI-C was not recommended for the medullary canals in Group 2 due to excessively thin remaining cortical bone, a comparison of the theoretical results that would be obtained is reported to highlight the different volumes of bone to be removed. In this group, significantly less cortical bone needed to be removed for the OsteoCustom group than for the OFI-C group for both resection level A (2.67 cm3 vs. 7.37 cm3, \(\:p\le\:0.0001\)) and level B (2.11 cm3 vs. 7.28 cm3, \(\:p\le\:0.0001\), Fig. 9).

Fig. 7
figure 7

Example of visual comparison of the cortical bone removed by OsteoCustom (red) and OFI-C (green) in the same medullary canal

Fig. 8
figure 8

Group 1: volume of cortical bone removed by OsteoCustom (OC, in red) and OFI-C (in green) for (a) resection level A (\(\:{Vol}_{\:A}\)) and (b) level B (\(\:{Vol}_{\:B}\))

Fig. 9
figure 9

Group 2: volume of cortical bone removed by the OsteoCustom (OC, in red) and OFI-C (in green) stems for (a) resection level A (\(\:{Vol}_{\:A}\)) and (b) level B (\(\:{Vol\:}_{B}\))

Comparison of the stem-bone contact areas of the OsteoCustom and OFI-C stems

For Group 1 (which met the indications for both implantation with OsteoCustom and OFI-C), a slightly greater percentage of the stem-bone contact area was observed for the OsteoCustom stem than for the OFI-C stem for resection level A, both in the proximal portion of the stem (OsteoCustom: 8.7% vs. OFI-C: 6.7%, \(\:p=0.05\)) and in the distal portion (OsteoCustom: 77.0% vs. OFI-C: 72.7%, \(\:p=0.03\), Fig. 10). A significantly greater percentage of contact area was observed for resection level B, both in the proximal portion (OsteoCustom: 27.3% vs. OFI-C: 15.8%, \(\:p=0.001\)) and in the distal portion (OsteoCustom: 52.3% vs. OFI-C: 32.2%, \(\:p=0.002\)).

The results for Group 2 (where according to the guidelines [22], the OFI-C should not be implanted due to excessively thin remaining cortical bone) are not detailed here for brevity but are reported in the Supplementary Material.

For resection level A, the contact area achieved with the OsteoCustom stem for Group 2 was comparable to that for Group 1, both in the proximal portion of the stem (Group 1: 8.7% vs. Group 2: 9.1%, \(\:p=0.7\)) and in the distal portion (Group 1: 77.0% vs. Group 2: 77.1%, \(\:p=0.7\), Fig. 11). For resection level B, a smaller contact area was achieved for Group 2 in the proximal portion of the stem (Group 1: 27.3% vs. Group 2: 12.1%, \(\:p=0.6\)). Conversely, for the distal stem portion of resection level B, a significantly greater contact area was achieved for Group 2 (Group 1: 52.3% vs. Group 2: 67.2%, \(\:p=0.008\)).

Fig. 10
figure 10

Group 1: The stem-bone contact area of the OsteoCustom (OC, in red) and OFI-C (in green, crossed lines) stems in the proximal and distal portions for resection levels A and B

Fig. 11
figure 11

The stem-bone contact area of the OsteoCustom stem for Group 1 (in red) and Group 2 (in red, crossed lines) in the proximal and distal portions for resection levels A and B

Discussion

This study focused on the development of a customized stem, OsteoCustom, designed to meet a wider spectrum of anatomies than existing commercially available stems while maintaining greater cortical bone volume preservation and a greater stem–bone contact area. To achieve this goal, a customized stem was designed based on the anatomical parameters of the femoral medullary canal. The parameters on which the design of the OsteoCustom stem was based had been evaluated in a previous study, where the mean and variability range were calculated [38]. To the best of authors’ knowledge, no previous studies computed the conicity and ellipticity of the femoral medullary canal. Concerning the canal diameter, the results were in line with the literature. Indeed, a minimum diameter of 9–10 mm was found for the isthmus [49, 50], and an average value of 13–14 mm was reported for the canal diameter [51, 52]. Concerning the radius of curvature, the mean value and the range in [38] were smaller than those reported for commercially available osseointegrated stems (e.g., OPL, ILP and BADAL-X [53]). These numbers were also smaller when compared to studies that calculated the radius of curvature of the femoral canal as the femoral bow in the sagittal plane (689 – 1885 mm) [54, 55]. However, such values are comparable to the radius of curvature reported in other studies where a similar methodology was applied, and in particular the one considering only the proximal femur [56, 57].

This study focused on 70 CT scans in which a customized stem was designed for two resection levels and was compared to a commercial stem (OFI-C).

Out of the 70 femurs examined, 19 were not suitable for implantation of the commercial stem OFI-C (Group 2), since less than 2 mm of cortical bone thickness remained after virtual implantation both in level A and B. This was due to several reasons, such as the combination of very narrow femoral canals with the constant radius of curvature of the OFI-C stem. This issue arose for both the most proximal and distal resection levels. Therefore, shortening the femur above the isthmus to bypass the narrower canal sections would still not enable to implant the OFI-C stem.

In contrast, it was possible to implant the OsteoCustom stem in these femurs while preserving at least 2 mm of cortical bone thickness. The ability to adjust the conicity and ellipticity in the distal portion of the stem allowed designing stems thinner than 15 mm in the proximal portion (e.g., between 12 mm and 15 mm). Indeed, this approach ensured that the distal portion always maintained a diameter of at least 15 mm to prevent the risk of stem breakage [58].

This outcome highlighted the potential benefits of the OsteoCustom prosthesis, which is capable of accommodating a wider range of femoral anatomies than the current standard.

In comparison with a commercially available stem (OFI-C), qualitative and quantitative differences were observed both for the volume of cortical bone removed and the stem-bone contact area between the stem and the canal.

When both the OFI-C and the OsteoCustom stem could be implanted (Group 1), the OsteoCustom stem implantation reduced the amount of cortical bone removed by 42% compared to that of the commercial stem at the more proximal resection level (3.15 cm3 vs. 5.42 cm3). Similarly, compared with the OFI-C, the OsteoCustom stem spared 33% of the cortical bone at the distal resection level (2.25 cm3 vs. 3.39 cm3, Fig. 8). These values represent the absolute amounts of bone removed for the two prostheses. Even if these amounts of bone removed are scaled, by normalizing with respect to the stem length, there would still be roughly one third bone sparing using OsteoCustom compared to OFI-C for the proximal resection level. Similarly, for the distal resection level the OsteoCustom would spare nearly one fourth of the bone that would be removed for implanting the OFI-C.

Regarding the contact area, the OsteoCustom stem achieved 8.7% and 77.0% of the contact area in the proximal and distal portions, respectively, of the most proximal resection level (Level A). These numbers were slightly greater than those found for the OFI-C stem. A more significant increase was instead observed at the most distal resection (level B). Particularly, in the distal portion of the stem, a 20% increase in the contact area was noted between the OsteoCustom and OFI-C stems (52.3% vs. 32.2%). These results highlighted the improved fit-and-fill granted by a customized stem compared to a commercial stem, especially at the most distal resection level. This outcome was not unexpected, considering that the statistical shape model of the femoral medullary canal demonstrated that certain anatomical parameters (e.g., conicity) were more pronounced in the distal part of the medullary canal, whereas the proximal part exhibited a more regular geometry [38]. Therefore, the customization of the stem that accounts for these anatomical parameters (e.g., conicity, ellipticity, and others, as described in Sect. 2.1) enabled better matching with the subject-specific anatomy, especially at the distal resection level. Unfortunately, a direct comparison with literature data was not possible, as to the best of the authors’ knowledge, this is the first study evaluating the benefits of customizing osseointegrated stems for transfemoral amputees. In the field of hip replacement, numerical studies have been performed to estimate the contact area needed to achieve long-term secondary stability of cementless hip stems: a contact area of 72 \(\:\pm\:\) 8% was needed for osseointegration [59]. These numbers are comparable to those achieved for the most proximal resection level in this study (level A) but greater than those found at the most distal resection level (level B). However, the present findings are in agreement with in vivo and ex vivo observations in cementless hip implants, where generally 50–55% of the contact area was found in stable implants (approximately 75% under the most favorable conditions) [60,61,62]. The findings for the OsteoCustom stem seem very promising, as a greater contact area facilitates better osseointegration, leading to improved implant stability [63]. Indeed, an increase in the stem-bone contact area of 23% already showed several benefits in cementless hip stems [64, 65].

For completeness, the implantation of the OFI-C stem was also simulated in the subset of patients for whom implantation of the OFI-C stem was not advisable (Group 2). In this group, the OsteoCustom stem achieved less contact area than did the OFI-C stem (Fig. S1 – Supplementary Materials). However, in such group, the potential benefit of the OsteoCustom stem with respect to the OFI-C was even more relevant regarding the volume of cortical bone removed, with almost 5 cm3 of spared cortical bone for both resection levels. The large amount of bone removed in part can explain why the OFI-C stem achieved a larger contact area than did the OsteoCustom stem in Group 2. Such massive cortical bone removal, particularly considering that - for Group 2 - there would not have been even 2 mm of cortical bone remaining around the stem, would have strongly discouraged surgeons from implanting the OFI-C stem. In a study on the effect of cortical thickness on stress and strain on osseointegrated prostheses, Thesleff et al. reported that an increase in cortical thickness from 2 mm to 5 mm reduced the stress in the host bone from 100 MPa to 43 MPa, with a significant decrease in the risk of periprosthetic fractures [20]. Hence, in scenarios where 2 mm would not be preserved if a commercial stem was implanted, the customization of the OsteoCustom stem would facilitate the avoidance of extensive cortical bone removal, thus allowing for the inclusion of these 19 cases from Group 2 as viable candidates for osseointegrated implants. This suggests that while commercial implants would be unsuitable for approximately 30% of the anatomies, implant customization would make nearly all patients eligible from an anatomical point of view for osseointegrated implantation.

To assess whether the customized stem would yield similar results in the anatomies that were not eligible for the commercial stem (Group 2) as in the anatomies that were eligible for the commercial stem (Group 1), the contact areas achieved with the OsteoCustom stem were compared between the two groups. A similar contact area was found for the most proximal resection level, both for the proximal and for the distal portions of the OsteoCustom stem. For the most distal resection level, however, a greater contact area was found in Group 1 than in Group 2 in the most proximal portion of the stem. This is explained by the fact that more ribs are present in the stems designed for the femurs of Group 1, as they have a larger canal diameter compared to Group 2. Conversely, in the most distal portion, a significantly larger contact area was achieved in Group 2 than in Group 1. As the stem diameter was at least 15 mm to prevent the risk of stem breakage [41, 58], in Group 2, which had a narrower canal, more bone was removed, resulting in an increased contact area.

The results of this study could have a significant impact on the design of osseointegrated stems, particularly on how customization is related to the optimization of bone removal and on improving fit-and-fill. Indeed, preserving more bone stock can be beneficial if an osseointegrated implant needs revision, as it helps prevent further weakening of the bone [66].

This study presents some limitations. First, the study focused on a limited number of subjects (n = 70). This sample size was comparable to that in a study on the statistical shape modeling of the distal femoral medullary canal [38]. Indeed, the power analysis confirmed that this number of subjects was adequate to detect statistically significant differences between the two stems. Another limitation relies on the study population included for this study. First of all, patients older than 65 years were excluded since aging generally leads to a deterioration in bone quality [67] and in the majority of the studies reported an average patient age below 50 years [26, 29, 68]. However, in some follow-up studies, patients older than 70 years old were included [69]. Similarly, patients with a BMI>30 were excluded following the guidelines for press-fit implants [40]. Furthermore, to the best of authors’ knowledge, most follow-up studies on osseointegrated prostheses have focused on patients with a BMI<30 [70,71,72]. However, other studies have found that BMI is not associated with an increased risk of aseptic or overall failure [69].

Furthermore, this work focused primarily on the immediate outcomes of stem design customization, such as cortical bone preservation and stem-bone contact area. However, the long-term performance and durability of the customized stem, including factors such as implant stability, were not directly assessed here. However, the indicators measured in the present study are good predictors of the fit-and-fill of the implant, which is a key factor for short- and long-term success. The promising results obtained from the methodology developed in this article led to further preclinical validation of the OsteoCustom concept, for which biomechanical tests are currently underway to assess implant stability following a validated procedure [23].

Another limitation relies on how the comparative analysis is performed. Indeed, while the study compared the customized stem (OsteoCustom) with a commercially available stem (OFI-C), it is essential to acknowledge that other prosthetic designs exist on the market, not only press-fit stems such as the one developed in the study (e.g., OPL stem [39]) but also threaded implants, which are screwed inside the medullary canal (e.g., OPRA [73]). Among these stems, the one with the most similar geometry to the OsteoCustom was chosen for comparison. However, commercial designs appear to be more or less similar in terms of geometry: these stems indeed have a constant cross-section and curvature, with a diameter in the range of 14–25 mm [53]. No significant changes are anticipated if the design of other stems is included in the study.

Throughout this study, it was possible to explore whether customization could have a potential benefit in terms of bone volume removal and stem-bone contact area. Given the good results obtained, one of the next steps will be a health technology assessment to evaluate several aspects, as the potential costs associated with possible implant mobilization, the expected reduction in the number of mobilized implants with OsteoCustom, and the potential increase in the number of patients treated with OsteoCustom. Throughout such analysis, it will be possible to understand if the customization process of osseointegrated stem would be advantageous in economic terms, therefore whether it is worthwhile to commercialize customized prostheses.

It must also be acknowledged that this study evaluated the advantages of a customized stem in terms of anatomical fit-and-fill. More pre-clinical work is needed to confirm if this would translate to an actual advantage in terms of implant stability or load transfer. In parallel, our group has developed a biomechanical test on ex vivo specimens to measure implant stability and load transfer [23].

Conclusions

This study aimed to develop a customized osseointegrated prosthesis stem (OsteoCustom) to address the main limitations of the existing commercial designs. According to anatomical data from femoral medullary canals, the OsteoCustom stem demonstrated superior cortical bone preservation and increased contact area compared to a commercial stem (OFI-C). Notably, the use of an OsteoCustom stem offered a viable solution for patients for whom the OFI-C failed to meet the cortical bone thickness criteria. The promising results from this study represent a crucial step in exploring the potential benefits of customized osseointegrated prostheses in terms of fit-and-fill.

Data availability

The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

CT:

Computed Tomography

HU:

Hounsfield Unit

BMI:

Body Mass Index

OC:

OsteoCustom

SSM:

Statistical Shape Modeling

References

  1. Windrich M, Grimmer M, Christ O, Rinderknecht S, Beckerle P. Active lower limb prosthetics: a systematic review of design issues and solutions. Biomed Eng OnLine. 2016;15(S3):140.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Brånemark R, Berlin Ö, Hagberg K, Bergh P, Gunterberg B, Rydevik B. A novel osseointegrated percutaneous prosthetic system for the treatment of patients with transfemoral amputation: a prospective study of 51 patients. Bone Jt J. 2014;96–B(1):106–13.

    Article  Google Scholar 

  3. Brånemark R, Brånemark PI, Rydevik B, Myers RR. Osseointegration in skeletal reconstruction and rehabilitation: a review. J Rehabil Res Dev. 2001;38(2):175–81.

    PubMed  Google Scholar 

  4. Dunne S, Coffey L, Gallagher P, Desmond D. If I can do it I will do it, if I can’t, I can’t: a study of adaptive self-regulatory strategies following lower limb amputation. Disabil Rehabil. 2014;36(23):1990–7.

    Article  PubMed  Google Scholar 

  5. Overmann AL, Forsberg JA. The state of the art of osseointegration for limb prosthesis. Biomed Eng Lett. 2019;10(1):5–16.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Anderst W, Fiedler G, Onishi K, McKernan G, Gale T, Paulus P. Within-subject effects of standardized prosthetic socket modifications on physical function and patient-reported outcomes. Trials. 2022;23(1):299.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Lee KH, Heo HS, Kim J, Cho JH, Kim KT, Hur JY, et al. A pneumatically controlled prosthetic socket for Transfemoral amputees. Sensors. 2024;24(1):133.

    Article  Google Scholar 

  8. Galteri G, Cristofolini L. In vitro and in silico methods for the biomechanical assessment of osseointegrated transfemoral prostheses: a systematic review. Front Bioeng Biotechnol. 2023;11:1237919.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Van de Meent H, Hopman MT, Frölke JP. Walking ability and quality of life in subjects with transfemoral amputation: a comparison of osseointegration with socket prostheses. Arch Phys Med Rehabil. 2013;94(11):2174–8.

    Article  PubMed  Google Scholar 

  10. Leijendekkers RA, van Hinte G, Nijhuis-van der Sanden MW, Staal JB. Gait rehabilitation for a patient with an osseointegrated prosthesis following transfemoral amputation. Physiother Theory Pract. 2017;33(2):147–61.

    Article  PubMed  Google Scholar 

  11. Reif TJ, Jacobs D, Fragomen AT, Rozbruch SR. Osseointegration Amputation Reconstruction. Curr Phys Med Rehabil Rep. 2022;10(2):61–70.

    Article  Google Scholar 

  12. Lee WCC, Frossard LA, Hagberg K, Haggstrom E, Gow DL, Gray S, et al. Magnitude and variability of loading on the osseointegrated implant of transfemoral amputees during walking. Med Eng Phys. 2007;30(7):825–33.

    Article  PubMed  Google Scholar 

  13. Hoellwarth JS, Tetsworth K, Kendrew J, Kang NV, van Waes O, Al-Maawi Q, et al. Periprosthetic osseointegration fractures are infrequent and management is familiar. Bone Jt J. 2020;102–B(2):162–9.

    Article  Google Scholar 

  14. Juhnke DL, Beck JP, Jeyapalina S, Aschoff HH. Fifteen years of experience with Integral-Leg-Prosthesis: Cohort study of artificial limb attachment system. J Rehabil Res Dev. 2015;52(4):407–20.

    Article  PubMed  Google Scholar 

  15. Örgel M, Petri M, Ranker A, Wirries N, Graulich T, Krettek C, et al. Management, outcome, and novel classification system of periprosthetic fractures in patients with transcutaneous osseointegrated prosthetic systems (TOPS)—a retrospective cohort analysis. Arch Orthop Trauma Surg. 2022;142(7):1499–509.

    Article  PubMed  Google Scholar 

  16. Leijendekkers RA, van Hinte G, Frölke JP, van de Meent H, Nijhuis-van der Sanden MWG, Staal JB. Comparison of bone-anchored prostheses and socket prostheses for patients with a lower extremity amputation: a systematic review. Disabil Rehabil. 2017;39(11):1045–58.

    Article  PubMed  Google Scholar 

  17. Kunutsor SK, Gillatt D, Blom AW. Systematic review of the safety and efficacy of osseointegration prosthesis after limb amputation. Br J Surg. 2018;105(13):1731–41.

    Article  CAS  PubMed  Google Scholar 

  18. Karadsheh M. Orthobullets. [cited 2024 May 29]. THA Periprosthetic Fracture - Recon - Orthobullets. https://www.orthobullets.com/recon/5013/tha-periprosthetic-fracture.

  19. Tanaka K, Sailer I, Iwama R, Yamauchi K, Nogami S, Yoda N, et al. Relationship between cortical bone thickness and implant stability at the time of surgery and secondary stability after osseointegration measured using resonance frequency analysis. J Periodontal Implant Sci. 2018;48(6):360–72.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Thesleff A, Ortiz-Catalan M, Brånemark R. The effect of cortical thickness and thread profile dimensions on stress and strain in bone-anchored implants for amputation prostheses. J Mech Behav Biomed Mater. 2022;129:105148.

    Article  CAS  PubMed  Google Scholar 

  21. Örgel M, Liodakis E, Jaratjitwilai P, Harb A, Wirries N, Omar M, et al. Three-year follow-up of changes of cortical bone thickness after implantation of endo-exo-prosthesis (EEP) for transfemoral amputees. J Orthop Surg. 2020;15(1):164.

    Article  Google Scholar 

  22. OPRA Implant System. Instructions for Use [Internet]. [cited 2024 Apr 23]. https://www.accessdata.fda.gov/cdrh_docs/pdf8/H080004S002D.pdf.

  23. Galteri G, Palanca M, Alesi D, Zaffagnini S, Morellato K, Gruppioni E et al. Reliable in vitro method for the evaluation of the primary stability and load transfer of transfemoral prostheses for osseointegrated implantation. Front Bioeng Biotechnol [Internet]. 2024 Mar 21 [cited 2024 Apr 24];12. https://www.frontiersin.org/articles/https://doi.org/10.3389/fbioe.2024.1360208.

  24. Engh CA, Bobyn JD, Glassman AH. Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br. 1987;69(1):45–55.

    Article  CAS  PubMed  Google Scholar 

  25. Lerch M, Kurtz A, Stukenborg-Colsman C, Nolte I, Weigel N, Bouguecha A, et al. Bone remodeling after total hip arthroplasty with a short stemmed metaphyseal loading implant: finite element analysis validated by a prospective DEXA investigation. J Orthop Res off Publ Orthop Res Soc. 2012;30(11):1822–9.

    Article  Google Scholar 

  26. Hagberg K, Ghassemi Jahani SA, Kulbacka-Ortiz K, Thomsen P, Malchau H, Reinholdt C. A 15-year follow-up of transfemoral amputees with bone-anchored transcutaneous prostheses. Bone Jt J. 2020;102–B(1):55–63.

    Article  Google Scholar 

  27. Roberts P, Chan D, Grimer RJ, Sneath RS, Scales JT. Prosthetic replacement of the distal femur for primary bone tumours. J Bone Joint Surg Br. 1991;73–B(5):762–9.

    Article  Google Scholar 

  28. Aschoff HH, Juhnke DL. [Endo-exo prostheses: Osseointegrated percutaneously channeled implants for rehabilitation after limb amputation]. Unfallchirurg. 2016;119(5):421–7.

    Article  PubMed  Google Scholar 

  29. Muderis MA, Tetsworth K, Khemka A, Wilmot S, Bosley B, Lord SJ, et al. The Osseointegration Group of Australia Accelerated Protocol (OGAAP-1) for two-stage osseointegrated reconstruction of amputated limbs. Bone Jt J. 2016;98–B(7):952–60.

    Article  Google Scholar 

  30. Kagan R, Adams J, Schulman C, Laursen R, Espana K, Yoo J, et al. What factors are Associated with failure of compressive osseointegration fixation? Clin Orthop. 2017;475(3):698–704.

    Article  PubMed  Google Scholar 

  31. Atallah R, Leijendekkers RA, Hoogeboom TJ, Frölke JP. Complications of bone-anchored prostheses for individuals with an extremity amputation: A systematic review. PLOS ONE. 2018 ago;13(8):e0201821.

  32. Khemka A, FarajAllah CI, Lord SJ, Bosley B, Al Muderis M. Osseointegrated total hip replacement connected to a lower limb prosthesis: a proof-of-concept study with three cases. J Orthop Surg. 2016;11(1):13.

    Article  Google Scholar 

  33. Schwarze M, Hurschler C, Seehaus F, Correa T, Welke B. Influence of transfemoral amputation length on resulting loads at the osseointegrated prosthesis fixation during walking and falling. Clin Biomech. 2014;29(3):272–6.

    Article  Google Scholar 

  34. Kang N, Al-Ajam Y, Keen P, Woollard A, Steinitz H, Farrant J, et al. Radiological evaluation before and after treatment with an osseointegrated bone-anchor following major limb amputation—a guide for radiologists. Skeletal Radiol. 2024;53(6):1033–43.

    Article  PubMed  Google Scholar 

  35. Apostu D, Piciu D, Oltean-Dan D, Cosma D, Lucaciu O, Popa C, et al. How to prevent aseptic loosening in Cementless Arthroplasty: a review. Appl Sci. 2022;12(3):1571.

    Article  CAS  Google Scholar 

  36. Apostu D, Lucaciu O, Berce C, Lucaciu D, Cosma D. Current methods of preventing aseptic loosening and improving osseointegration of titanium implants in cementless total hip arthroplasty: a review. J Int Med Res. 2018;46(6):2104–19.

    Article  CAS  PubMed  Google Scholar 

  37. Chantarapanich N, Sitthiseripratip K, Mahaisavariya B, Wongcumchang M, Siribodhi P. 3D geometrical assessment of femoral curvature: a reverse engineering technique. J Med Assoc Thai. 2008;91(9):1377–81.

    PubMed  Google Scholar 

  38. Betti V, Aldieri A, Cristofolini L. A statistical shape analysis for the assessment of the main geometrical features of the distal femoral medullary canal. Front Bioeng Biotechnol [Internet]. 2024 Apr 10 [cited 2024 Apr 23];12. https://www.frontiersin.org/articles/https://doi.org/10.3389/fbioe.2024.1250095.

  39. Osseointegration Group. Osseointegration International launches the Type-D range of OPL [Internet]. 2023 [cited 2024 Apr 25]. https://osseointegration.org/2023/02/23/osseointegration-international-launches-type-d/.

  40. OTN Implants. OTN Implants - BADAL X In- and Exclusion criteria. [cited 2024 Apr 23]. OTN Implants - BADAL X In- and Exclusion criteria. https://www.otnimplants.nl/inexclusion.

  41. Reif TJ, Jacobs D, Fragomen AT, Rozbruch SR. Osseointegration Amputation Reconstruction. Curr Phys Med Rehabil Rep. 2022;10(2):61–70.

  42. Atallah R, van de Meent H, Verhamme L, Frölke JP, Leijendekkers RA. A Eshraghi editor 2020 Safety, prosthesis wearing time and health-related quality of life of lower extremity bone-anchored prostheses using a press-fit titanium osseointegration implant: a prospective one-year follow-up cohort study. PLoS ONE 15 3 e0230027.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  43. Singh ND. and M. Sample Size Calculator for Comparing Two Independent Means [Internet]. [cited 2024 Apr 23]. https://statulator.com/SampleSize/ss2M.html.

  44. Aldieri A, La Mattina AA, Szyszko JA, Baruffaldi F, Viceconti M. HFValid collection: Hip-Fracture validation collection [Internet]. University of Bologna; 2023 [cited 2024 Apr 23]. https://amsacta.unibo.it/id/eprint/7277/.

  45. Cemented Total Hip Replacement. through the ABMS Approach | SpringerLink [Internet]. [cited 2024 Jul 4]. https://link.springer.com/chapter/10.1007/978-3-031-02059-9_8.

  46. Overview |. Hip fracture: management | Guidance | NICE [Internet]. NICE; 2011 [cited 2024 Jul 4]. https://www.nice.org.uk/guidance/cg124.

  47. Zhang C, Yan CH, Zhang W. Cemented or cementless fixation for primary hip arthroplasty—evidence from The International Joint Replacement Registries. Ann Jt [Internet]. 2017 Oct 16 [cited 2024 Jul 4];2(10). https://aoj.amegroups.org/article/view/3848.

  48. Wu CC. Is clinical measurement of anatomic axis of the femur adequate? Acta Orthop. 2017;88(4):407–10.

    Article  PubMed  PubMed Central  Google Scholar 

  49. Jung IJ, Choi EJ, Lee BG, Kim JW. Population-based, three-dimensional analysis of age- and sex-related femur shaft geometry differences. Osteoporos Int. 2021;32(8):1631–8.

    Article  CAS  PubMed  Google Scholar 

  50. Su Xyun, Zhao Jxin, Zhao Z, Zhang L, cheng, Li C, Li J et al. tao,. Three-Dimensional Analysis of the Characteristics of the Femoral Canal Isthmus: An Anatomical Study. BioMed Res Int. 2015;2015:459612.

  51. Milligan DJ, O’Brien S, Bennett D, Hill JC, Beverland DE. The effects of age and gender on the diameter of the femoral canal in patients who undergo total hip replacement. Bone Jt J. 2013;95–B(3):339–42.

    Article  Google Scholar 

  52. Uzel AP, Deloumeaux J, Rouvillain JL, Laflamme GY, Durandeau A, Caix P. Comparative study of femoral diaphyseal morphometry in two male populations, in France and a French West Indies island: an example of clinical relevance of comparative anatomy for orthopedic practice. Surg Radiol Anat SRA. 2011;33(3):235–40.

    Article  PubMed  Google Scholar 

  53. Atallah R. Safety of bone-anchored prostheses in lower extremity amputation [PhD Thesis]. Radboud Institute for Health Sciences; 2023.

  54. Egol KA, Chang EY, Cvitkovic J, Kummer FJ, Koval KJ. Mismatch of current intramedullary nails with the anterior bow of the femur. J Orthop Trauma. 2004;18(7):410–5

  55. Harper MC, Carson WL. Curvature of the femur and the proximal entry point for an intramedullary rod. Clin Orthop. 1987;(220):155–61

  56. Thiesen DM, Prange F, Berger-Groch J, Ntalos D, Petersik A, Hofstätter B, et al. Femoral antecurvation-A 3D CT analysis of 1232 adult femurs. PLoS ONE. 2018;13(10):e0204961.

    Article  PubMed  PubMed Central  Google Scholar 

  57. Liaw CK, Chen YP, Wu TY, Fuh CS, Chang RF. New Computerized Method in measuring the Sagittal bowing of Femur from Plain Radiograph—A validation study. J Clin Med. 2019;8(10):1598.

    Article  PubMed  PubMed Central  Google Scholar 

  58. Wyatt M, Hooper G, Frampton C, Rothwell A. Survival outcomes of cemented compared to uncemented stems in primary total hip replacement. World J Orthop. 2014;5(5):591–6.

    Article  PubMed  PubMed Central  Google Scholar 

  59. Banducci E, Al Muderis M, Lu W, Bested SR. The safety of one-stage versus two-stage approach to osseointegrated prosthesis for limb amputation. Bone Jt Open. 2023;4(7):539–50.

    Article  PubMed  PubMed Central  Google Scholar 

  60. Kagan R, Adams J, Schulman C, Laursen R, Espana K, Yoo J, et al. What factors are Associated with failure of compressive osseointegration fixation? Clin Orthop. 2017;475(3):698–704.

  61. Mohamed J, Reetz D, van de Meent H, Schreuder H, Frölke JP, Leijendekkers R. What are the risk factors for mechanical failure and loosening of a Transfemoral Osseointegrated Implant System in patients with a lower-limb amputation? Clin Orthop Relat Res. 2022;480(4):722.

    Article  PubMed  Google Scholar 

  62. Viceconti M, Pancanti A, Dotti M, Traina F, Cristofolini L. Effect of the initial implant fitting on the predicted secondary stability of a cementless stem. Med Biol Eng Comput. 2004;42(2):222–9.

    Article  CAS  PubMed  Google Scholar 

  63. Hardy DC, Frayssinet P, Guilhem A, Lafontaine MA, Delince PE. Bonding of hydroxyapatite-coated femoral prostheses. Histopathology of specimens from four cases. J Bone Joint Surg Br. 1991;73(5):732–40.

    Article  CAS  PubMed  Google Scholar 

  64. Tonino AJ, Thèrin M, Doyle C. Hydroxyapatite-coated femoral stems. Histology and histomorphometry around five components retrieved at post mortem. J Bone Joint Surg Br. 1999;81(1):148–54.

    Article  CAS  PubMed  Google Scholar 

  65. Søballe K, Gotfredsen K, Brockstedt-Rasmussen H, Nielsen P, Rechnagel K. Histologic analysis of a retrieved hydroxyapatite-coated femoral prosthesis. Clin Orthop. 1991;272:255–8.

    Article  Google Scholar 

  66. Staedt H, Kämmerer PW, Goetze E, Thiem DGE, Al-Nawas B, Heimes D. Implant primary stability depending on protocol and insertion mode — an ex vivo study. Int J Implant Dent. 2020;6(1):49.

    Article  PubMed  PubMed Central  Google Scholar 

  67. Glismann K, Konow T, Lampe F, Ondruschka B, Huber G, Morlock MM. Small design modifications can improve the primary stability of a fully coated tapered wedge hip stem. PLoS ONE. 2024;19(4):e0300956.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  68. Walsh WR, Pelletier MH, Bertollo N, Lovric V, Wang T, Morberg P, et al. Bone ongrowth and mechanical fixation of implants in cortical and cancellous bone. J Orthop Surg. 2020;15(1):177.

    Article  Google Scholar 

  69. Ong G, Hoellwarth JS, Testworth K, Al Muderis M. Techniques to remove Press-Fit Osseointegration implants. JBJS Essent Surg Tech. 2024;14(1):e2300017.

    Article  Google Scholar 

  70. Hansen RL, Langdahl BL, Jørgensen PH, Petersen KK, Søballe K, Stilling M. Does migration of osseointegrated implants for transfemoral amputees predict later revision? A prospective 2-year radiostereometric analysis with 5-years clinical follow-up. Orthop Traumatol Surg Res OTSR. 2019 ;105(5):1013–20.

  71. Reetz D, Atallah R, Mohamed J, van de Meent H, Frölke JPM, Leijendekkers R. Safety and performance of bone-anchored prostheses in persons with a transfemoral amputation: a 5-year follow-up study. J Bone Joint Surg Am. 2020;102(15):1329–35.

  72. Hagberg K, Ghasemi Jahani SA, Omar O, Thomsen P. Osseointegrated prostheses for the rehabilitation of patients with transfemoral amputations: a prospective ten-year cohort study of patient-reported outcomes and complications. J Orthop Transl. 2023;38:56–64.

  73. Integrum Implants. OPRAâ„¢ Implant System [Internet]. [cited 2024 Apr 30]. https://integrum.se/what-we-do/our-products-future-solutions/opra-implant-system/.

Download references

Acknowledgements

The authors wish to acknowledge Alessandra Aldieri, Federico Stacchiotti, Agostino Igor Mirulla and Edoardo Ognisanto for the stimulating discussions, and Antonino Amedeo La Mattina and Giorgio Davico for their advice on image processing.

Funding

This study was funded by INAIL (PR19-CR-P5-OsteoCustom, CUP: E59C20000730005).

Author information

Authors and Affiliations

Authors

Contributions

First authorship is equally shared between authors V.B. and G.G. V.B. contributed to the study design, performed the study, interpreted the results and wrote the main manuscript text. G.G., K.M., M.P. and E.G. contributed to the study design, and manuscript conceptualization. S.Z. and D.A. contributed to the interpretation of the results. L.C. designed the study design, contributed to manuscript conceptualization, acquired the funding and supervised the project. All authors reviewed and approved the manuscript.

Corresponding author

Correspondence to Luca Cristofolini.

Ethics declarations

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary Material 1

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it.The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions