Table 3 Accuracy of imageless and CT-based navigation systems reported in the literature

From: A method combining the use of a mobile application and a dedicated pelvic positioner for acetabular cup insertion

 

Inclination (°)

Absolute value error

Anteversion (°)

Absolute value error

Type

Navigation system

Kalteis

(30 hips)

3.0 ± 2.6

3.3 ± 2.3

CT-based

The Vector Vision hip 3.0

Iwana

(117 hips)

1.8 ± 1.6

1.2 ± 1.1

CT-based

Stryker CT-Hip System V1.0–29

Nakahara

(49 hips)

1.2 ± 1.3

1.0 ± 0.5

CT-based

Stryker Navigation System2

Tetsunaga

(30 hips)

3.2 ± 2.4

3.0 ± 2.5

CT-based

The Vector Vision Hip CT-based version 3.5.2

Kalteis

(30 hips)

2.9 ± 2.2

4.2 ± 3.3

Image-free

The Vector Vision hip 3.0

Ybinger

(37 hips)

3.5 ± 4.4

6.5 ± 7.3

Image-free

The PiGalileo THR, Plus

Lass

(62 hips)

3.2 ± 2.4

6.5 ± 3.7

Image-free

The Navitrack

Takeda

(118 hips)

3.6 ± 2.6

6.7 ± 3.6

Image-free

The Orthopilot THA Pro

Current study

(30 hips)

2.4 ± 2.1

2.8 ± 2.6