Skip to main content

Table 9 Stiffness of coronal shear fracture of the distal humerus models under different testing conditions

From: Biomechanical properties of a novel fixation system for intra-articular distal humerus fractures: a finite element analysis

Configuration

Axial (N/mm)

Bending (N/mm)

Rotational torsion (N mm/°)

Novel plating

194.17

38.46

2929.69

Orthogonal plating

153.85

33.33

2650.18

Three screws

115.61

28.30

2180.23