From: The risk for complications and reoperations with the use of mega prostheses in bone reconstructions
Ā | Total number of implants | Soft tissue failure | Aseptic loosening | Mechanical failure | Infection | Tumour relapse/progression | Total (%) |
---|---|---|---|---|---|---|---|
Site | |||||||
Ā Proximal humerus | 15 | 1 | 1 | 1 | 3 | 1 | 7 (47) |
Ā Total humerus | 1 | 0 | 0 | 0 | 0 | 0 | 0 (0) |
Ā Diaphyseal humerus | 1 | 0 | 1 | 0 | 0 | 0 | 1 (100) |
Ā Distal humerus | 3 | 0 | 0 | 1 | 1 | 0 | 2 (67) |
Ā Pelvis | 5 | 0 | 0 | 0 | 3 | 1 | 4 (80) |
Ā Proximal femur | 30 | 4 | 0 | 3 | 4 | 2 | 13 (43) |
Ā Total femur | 3 | 0 | 0 | 1 | 0 | 1 | 2 (67) |
Ā Arthrodesis implant | 2 | 0 | 0 | 0 | 0 | 0 | 0 (0) |
Ā Diaphyseal femur | 4 | 0 | 1 | 0 | 0 | 0 | 1 (25) |
Ā Distal femur | 33 | 4 | 2 | 5 | 8 | 2 | 21 (61) |
Ā Proximal tibia | 19 | 6 | 0 | 1 | 6 | 1 | 14 (68) |
Total | 116 | 15 | 5 | 12 | 25 | 8 | 65 |