Skip to main content

Table 2 Loading information of included studies with cadaver specimens

From: Comparisons of the surface micromotions of cementless femoral prosthesis in the horizontal and vertical levels: a network analysis of biomechanical studies

Author, year

Material testing machine

Testing jig

Loading information

Action simulated

Anatomical features simulated

Load force

Femur orientation

Test cycles

Gotze 2002 [20]

Zwick, Typ 1454, Einsingen, Germany

6 LVDTs

NR

NR

Vertical force 200–2000 N, 0.5 Hz

8° valgus, 6° flexion

1

Klestil 2006 [21]

Bionics 851.2, MTS

NR

NR

Acetabulum

50–2100 N; frequency (2 Hz)

10° flexion, 10° adduction

50000

Abdul-Kadir 2008 [17]

Instron 5565, Instron Corp., Canton, MA, USA

2 LVDTs

NR

NR

5000 N load cell with at a rate of 1 KN/min; maximum load of 2000 N

Long axis coaxial to load direction

50

Pettersen 2009 [23]

MTS 858 Minibionix II, MTS Systems Corporation, Eden Prairie, MN, USA

6 LVDTs

SLS, SC

Abductor muscles, acetabulum

Vertical load 600 N (74 Kg BW); torsional moment 13.5 Nm (1.82%BW)

12° valgus tilt

50

Østbyhaug 2010 [22]

MTS 858 Minibionix II, MTS Systems Corporation, Eden Prairie, MN, USA

6 LVDTs

SLS, SC

Abductor muscles, acetabulum

Vertical load 600 N (74 Kg BW); torsional moment 15 Nm

12° valgus tilt

50

Wik 2011 [24]

MTS 858 MiniBionix II, MTS Systems Corporation, Eden Prairie, MN, USA

3 LVDTs

SLS, SC

Abductor muscles, acetabulum

Vertical force 600 N; torsional moment 13.8 Nm; for the strain measurement, torsional moment 10.0 Nm

12° valgus tilt

10S

Bieger 2012 [18]

Servo hydraulic machine, instron, Typ 8871, Pfungstadt, Germany

6 LVDTs

SLS

Acetabulum

1 Hz for first 4000 cycles; 0–1000 cycles 100–400 N; 1000–2000 cycles 100–800 N; 2000–3000 cycles 100–1200 N; 3000–4000 cycles 100–1600 N (250%BW); 2 Hz for 96,000 cycles with 100–1600 N

8° valgus, 6° flexion

250; 500

Bieger 2013 [19]

Instron, Typ 8871, Pfungstadt, Germany

6 LVDTs

SLS

NR

Vertical load 100–1600 N; frequency (2 Hz)

8° valgus, 6° flexion

500

Østbyhaug 2013 [15]

MTS 858 Minibionix II, MTS Systems Corporation, Eden Prairie, MN, USA

6 LVDTs

SLS, SC

Abductor muscles, acetabulum

Vertical load 600 N (73 Kg BW); torsional moment 15 Nm

12° valgus tilt

500

Bieger 2016 [13]

Instron, Typ 8871, Pfungstadt, Germany

6 LVDTs

SLS

NR

Vertical load 100–1600 N (250%BW); frequency (2 Hz)

8° valgus, 6° flexion

40 K, 100 K

  1. SLS single-leg stance, SC stair climbing, BW body weight, NR not reported