Table 4 Removal depth [μm] CoCrMo specimen pairs after RoD test

From: Study of the tribological properties of surface structures using ultrashort laser pulses to reduce wear in endoprosthetics

 StructureMeanStDevMinQ1MedianQ3Max
Depth removal [μm]Reference3.6911.6811.3442.3542.9505.3686.435
Grid lines1.6430.8630.2731.0461.3372.2473.746
Dimples1.9960.9600.8351.1691.7222.5524.125
Offset lines1,6620,8330,5451,0191,5651,9383,995
  1. StDev standard deviation, Min minimum, Q1 first quartile, Q3 third quartile, Max maximum