Table 4 The stiffness of models under different testing conditions

From: Biomechanical analysis of a novel plating for intra-articular distal humerus fractures: combined anteromedial and anterolateral plating

Configuration Axial compression
(N/mm)
Rotation torsion
(Nmm/deg)
Bending torsion
(Nmm/deg)
Valgus torsion
(Nmm/deg)
Novel plating 149.4 5549.2 5972.9 12803.3
Orthogonal plating 125.1 2491.5 4779.3 8303.5